Laboratory

Calitron Calibration Laboratory, U1: 207, Kohinoor Arcade, Tilak Chowk,


Nigdi, Pune, Maharashtra

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

01.08.2014

Certificate Number

C-0433

Valid Until

31.07.2016

Last Amended on

Page

1 of 7

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

SOURCE
1.

DC Voltage#

1 mV to 200 mV
200 mV to 1000 V

1.15% to 0.01%
0.01%

Using MFC (Transmille


3050) by Direct Method

2.

DC Current#

100 A to 200 mA
200 mA to 20 A

0.04%
0.04% to 0.13%

Using MFC (Transmille


3050) by Direct Method

20A to 1000 A

0.4%

With Current Coil

10 to 1 k
1 k to 100 k

0.3% to 0.01%
0.01%

Using MFC (Transmille


3050) by Direct Method

(2 Wire)

100 k to 10 M

0.02% to 0.07%

Using MFC (Transmille


3050) by Direct Method

(2 Wire)

10 m to 100 m
100 m to 1

2.5%
2.3% to 0.57%

By Using Decade Box


(Vaishashika 7400) by
Direct Method

50 Hz to 1 kHz
25A to 200 mA
200 mA to 20 A

1.47% to 0.27%
0.27% to 0.32%

50 Hz
20 A to 1000 A

0.6% to 2%

50 Hz to 1kHz
20 mV to 2 V
2 V to 1000 V

0.3% to 0.10%
0.10%

3.

4.

5.

DC Resistance#
(4 Wire)

AC Current#

AC Voltage#

Avijit Das
Program Manager

Using MFC (Transmille


3050) by Direct Method

With Current Coil


Using MFC (Transmille
3050) by Direct Method

Naveen Jangra
Convenor

Laboratory

Calitron Calibration Laboratory, U1: 207, Kohinoor Arcade, Tilak Chowk,


Nigdi, Pune, Maharashtra

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

01.08.2014

Certificate Number

C-0433

Valid Until

31.07.2016

Last Amended on

Page

2 of 7

Quantity Measured/
Instrument

6.

Range / Frequency

Capacitance$

Capacitance

7.

Inductance#

8.

Temperature Simulation
(Indicator/ Controller/ Recorder)

a.

Thermocouple Type#
K Type
J Type
R Type
S Type
T Type

Avijit Das
Program Manager

*Calibration Measurement
Capability ()

Remarks

1 kHz
1 nF to 10 uF

3.90% to 2.28%

Using Decade
Capacitance Box
(Agronic CDB6) by
Direct Method

1 kHz
1 nF to 10 uF

2.26%

Using Decade
Capacitance Box
(Agronic CDB6) by
Direct Method

1 kHz
100 H to 10 H

3.9% to 2.28%

Using Decade
Inductance Box
(Agronic LDB6) by
Direct Method

1C to 1200 C
1C to 750 C
1C to 1700 C
1C to 1700 C
1C to 400 C

0.43 C
0.43 C
0.43 C
0.43 C
0.43 C

Using MFC (Transmille


3050 & Temperature
Simulator EA001) by
Direct Method

Naveen Jangra
Convenor

Laboratory

Calitron Calibration Laboratory, U1: 207, Kohinoor Arcade, Tilak Chowk,


Nigdi, Pune, Maharashtra

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

01.08.2014

Certificate Number

C-0433

Valid Until

31.07.2016

Last Amended on

Page

3 of 7

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

b.

PRT Mode$

-200C to 600 C

0.03 C to 0.47 C

Using Micro Ohm Box


(Vaishashika 7400) by
Direct Method

9.

Frequency#

100 Hz to 1 MHz

0.11% to 0.007%

Using MFC (Transmille


3050) by Direct Method

DC Voltage$

100 V to 100 mV
100 mV to 1000 V

0.5% to 0.002%
0.002% to 0.001%

Using 8 digit DMM


(Transmille 8081) by
Direct Method

DC Voltage

1 mV to 1 V
1 V to 1000 V

1.30% to 0.02 %
0.02% to 0.01%

Using 6 digit DMM


Agilent 34401A by
Direct / Comparison
Method

DC Current$

10 nA to 100 mA
100 mA to 1 A
1 A to 20 A

1.4% to 0.006%
0.006% to 0.03%
0.03% to 0.1%

Using 8 digit DMM


(Transmille 8081) by
Direct Method

DC Current

100 uA to 1 mA
1 mA to 10 A
10A to 20 A

1.74% to 5.70%
5.70% to 1.60%
0.35%

Using DMM Kusum


Meco KM 857 Agilent 6
+Shunt by Direct /
Comparison Method

AC Voltage$

50 Hz to 1 kHz
10 mV to 10 V
10 V to 1000 V

0.18% to 0.29 %
0.29% to 0.07%

Using 8 digit DMM


(Transmille 8081) by
Direct Method

2 kHz
100 mV to 1000 V

0.05% to 0.07%

Using 8 digit DMM


by Direct Method

MEASURE
10.

11.

12.

Avijit Das
Program Manager

Naveen Jangra
Convenor

Laboratory

Calitron Calibration Laboratory, U1: 207, Kohinoor Arcade, Tilak Chowk,


Nigdi, Pune, Maharashtra

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

01.08.2014

Certificate Number

C-0433

Valid Until

31.07.2016

Last Amended on

Page

4 of 7

Quantity Measured/
Instrument

AC Voltage

13.

AC Current$

AC Current

14.

DC Resistance $
(4 Wire)
(2 Wire)

DC Resistance
(4 Wire)

Avijit Das
Program Manager

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

20 kHz
100 mV to 100 V

0.16%

Using 8 digit DMM by


Direct Method

100 kHz
1 V to 10 V

0.27%

Using 8 digit DMM by


Direct Method

50 Hz to 1 kHz
100 mV to 750 V

0.12%

50 Hz to 1 kHz
25 A to 200 mA
200 mA to 2 A
2 A to 20 A

0.25% to 0.18%
0.18% to 0.36%
0.36% to 0.2%

Using 6 digit DMM


Agilent 34401A by
Direct / Comparison
Method
Using 8 digit DMM
(Transmille 8081) by
Direct Method

50 Hz
100 mA to 1A
1A to 10 A
1 KHz
10A to 20 A(1 KHz)
10 m to 1
1 to 100 k

0.40%
0.10% to 0.004%
0.004% to 0.002%

100 k to 100 M
100 M to 100 G

0.002% to 0.03%
0.03% to 3.0%

Using 8 digit DMM


(Transmille 8081) by
Direct Method

10 to 100 k

0.15% to 0.025%

Using 6 digit DMM


Agilent 34401A by
Direct / Comparison
Method

1.70%
1.70%

Using DMM Kusum


Meco KM 857 Agilent 6
+Shunt by Direct /
Comparison Method
Using 8 digit DMM
(Transmille 8081) by
Direct Method

Naveen Jangra
Convenor

Laboratory

Calitron Calibration Laboratory, U1: 207, Kohinoor Arcade, Tilak Chowk,


Nigdi, Pune, Maharashtra

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

01.08.2014

Certificate Number

C-0433

Valid Until

31.07.2016

Last Amended on

Page

5 of 7

Quantity Measured/
Instrument

(2 Wire)

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

100 K to 100 M

0.03% to 0.95%

Using 6 digit DMM


Agilent 34401A by
Direct / Comparison
Method

15.

Frequency & Period#

1 Hz to 1 GHz

0.08% to 0.0011%

Using High Resolution


Counter (HTC - FC3165)
by Direct / Comparison
Method

16.

DC High Voltage$

2 kV to 30 kV

3.9%

Using HV Probe and 5 4/5


digit DMM by Direct
Method

DC High Voltage

1 kV to 30 kV

3.9%

Using HV Probe and 5 4/5


digit DMM by Direct
Method

AC High Voltage$

50 Hz
2 kV to 10 kV

4.25%

Using HV Probe and 5 4/5


digit DMM by Direct
Method

50 Hz
1 kV to 10 kV

4.25%

Using HV Probe and 5 4/5


digit DMM by Direct
Method

17.

AC High Voltage

18.

Power Factor#

-1 to 1

0.0026 PF

Using Power Analyser


(Voltech PM 1000+) by
Direct / Comparison
Method

19.

Time#

1 min to 1 hr

0.06 s to 1.8 s

Using Stop Watch by

Avijit Das
Program Manager

Naveen Jangra
Convenor

Laboratory

Calitron Calibration Laboratory, U1: 207, Kohinoor Arcade, Tilak Chowk,


Nigdi, Pune, Maharashtra

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

01.08.2014

Certificate Number

C-0433

Valid Until

31.07.2016

Last Amended on

Page

6 of 7

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

Comparison Method
20.

AC Power#
@UPF

50 W to 6 kW
(50 V to 300 V, 1A to 20 A),
50 Hz

0.35%

Using Power Analyser


(Voltech PM 1000+) by
Direct / Comparison
Method

21.

DC Power#

10 W to 12 kW
(10 V to 600 V, 1A to 20 A)

0.75%

Using Power Analyser


(Voltech PM 1000+) by
Direct / Comparison
Method

22.

Capacitance#

1 kHz
20 pF to 1 mF

1.1%-0.75%

Using Agilent LCR meter


(U1732C) by Direct /
Comparison Method

23.

Inductance$

1 kHz
100 H to 10 H

1.3% to 0.6%

Using Agilent LCR meter


(U1732C) by Direct /
Comparison Method

24.

Temperature Measurement$

140C to 1340 C
180C to 750 C
50 C to 1700 C
50 C to 1700 C
1 C to 400 C

0.6 C
0.5 C
1.2 C
1.2 C
0.4 C

Using 8 digit DMM


(Transmille 8081) by
Direct / Comparison
Method

-100 to 400 C

0.09 C to 0.36 C

Using 8 digit DMM


(Transmille 8081) by
Direct / Comparison

Thermocouple Type
K Type
J Type
R Type
S Type
T Type
25.

PRT Measurement$

Avijit Das
Program Manager

Naveen Jangra
Convenor

Laboratory

Calitron Calibration Laboratory, U1: 207, Kohinoor Arcade, Tilak Chowk,


Nigdi, Pune, Maharashtra

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

01.08.2014

Certificate Number

C-0433

Valid Until

31.07.2016

Last Amended on

Page

7 of 7

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

Method
* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%
$

Only in Permanent Laboratory


Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual
environmental conditions and master equipment used.

Avijit Das
Program Manager

Naveen Jangra
Convenor