DIGITAL WEIGHT INDICATOR

Users guide
The Digital Weight Indicator is designed for general weighing, counting and check weighing applications.
The main features of the indicators are:
- Full front panel setup and calibration
- Easy to read large LED display
- Parts counting
- Check weighing
- RS232 data Interface
- Supplies enough current for up to 4-350 ohm Load Cells
- Powered by AC/DC Adaptor or Rechargeable Battery
GENERAL
Unpack the Indicator and keep carton for storing and/or future transportation. In the carton you should
find:Indicator with mounting bracket and this instruction manual. Connection of loadcell :

FRONT PANEL

ON/OFF

Key

0 /

Key

T/

Key

Turn the indicator ON/OFF;


Re-zero the indictor ; Shift the digit when setting;
Tare; Select the value of parameter;

MODE/ESC Key

Quit;Enter/Cancel the counting mode from weighing mode;

F/SET

Function key, confirm and save the seletion; Accumulate;

Key

POWER OPTIONS
AC/DC Adapter 220V 50HZ AC / 9V DC 800mA. LED in the front panel is ON when powered by
Adapter.
Rechargeable battery 6V 4Ah: Charging cycle is about 8 hours with the indicator turned off, providing
about 20 hours of service. LED is RED when battery is being charged and turns GREEN when charge is
completed. Display will flash when the battery is low.
When the black switch on the back cover of the indicator is at the O state, the power is off.
When I state, the indicator is powered by AC/DC Adapter 220V 50HZ AC / 9V DC 800mA.
When II state, the indicator is powered by battery.

SETUP MENU CHART


Four front panel keys F, MODE, T and 0 are navigation keys in the setup menu. MODE key is used to
enter the menu and to save the selection. Turn the indicator ON While pressing the 0 Key. The indicator
will display the first menu F1 KEY. And then release the 0 key. To quit the setup press T key.
DISPLAY

DESCRIPTION

F1 KEY

Counting/Accumulation/Holding

VALUE
the

weight/Pre-set

1,2,3,4

Battery saving mode. The indicator will power off if

SLP no

Tare Value
F2 SLP

No Sleep

no weight is loaded or no key is pressed in the set time.

F3 CHK

F4 CAL

SLP 5

after 5 min

SLP 10

after 10 min

SLP 15

after 15 min

Check weighing

beep 1 NO beep

After setting beep 1, 2 or 3 , set LO and/or HI targets.

beep 2 beep if within range

See Check weighing operation

beep 3 beep if outside the range

13 sub-menus(described as follow)

P1-P13

In the F4 CAL menu, there are 13 sub-menus.


DISPLAY

DESCRIPTION

VALUE

P1 INP

Internal value

P2 AUT

Power-on Zero

0, 2, 5, 10, 20% of the capacity

P3 RAN

Manual zero band range

2, 4, 10, 20, 50, 100% of capacity

P4 AZ

Auto zero tracking band

0.5d; 1d; 2d; 4d;

P5 DEC

Decimal point position

0; 0.0; 0.00; 0.000; 0.0000

P6 CAP

Capacity

Enter the target capacity via keypad

P7 INC

Display division

1; 2; 5; 10; 20; 50

P8 CAL

Calibration.

P9 INT

Initiate the indicator back to the original setting

P10 SP

A to D conversion speed

1, 2, 3, 4, 5

P11 RS

RS232 transmission options, selects when

Aut - after each weighing

data will be sent out via serial port

Cot - continuous

Press F to enter.

See Calibration procedure

1- fastest 5 - slowest

Prt - use with serial printer


P12 FL

To setup the field level calibration value.

P13 BD

To charge the battery simply plug into the

mains power.

RS 232 SERIAL OUTPUT


Baud Rate: 9600
Start Bit: 1
Data Bits: 8
Stop Bit: 1

Parity:

None

CALIBRATION PROCEDURE
Enter P8 CAL of the F4 CAL menu
1.

noLoad appears in the display, unload the scale and press F key when STABLE LED is on.

2.

Load appears in the display, load the scale with the calibration weight and press F key when
STABLE LED is on.

3.

When the display shows 000000,input the calibration weight. Press the F key.

4.

The indicator returns to weighing mode.

COUNTING OPERATION
F1 KEY should be set to be value 1 or 2
Testing step:
1.

Place a container on the scale and zero the indicator.

2.

Press MODE key and display will show number of samples P 10.

3.

Use T key to change number of samples.

4.

Place the required number of samples into the container and press F key.

5.

Wait till the display is stable and it will show number of pieces.

Counting step:
1.

Place the samples on the scale and it will show the number.

2.

Press F key circularly, it will show the total weight, unit weight and number of the samples.

3.

Press MODE key to shift between weighing mode and counting mode.

CHECKWEIGHING OPERATIO
1. Enter the F3 CHK.
2. Select Beep 1,Beep 2 or Beep 3 and then it display 000000.
3. Use T and 0 Key to set the parameter HI, press F key to confirm.
4. Set the parameter LO as step 3.
5. If LO or/and HI values have been set to others, not zero, the checkweighing is ON.
6. LED Display segments status: HI & LO segments ON - load within the target range
HI

ON - load accedes the higher target

LO

ON - load below the lower target

7. To have just one target LO or HI, the other value has to be 0.

8. If both HI and LO values are set to be 0 , the checkweighing is OFF.

ACCUMULATION OPERATION

F1 KEY should be set to be value 2 and the indicator is not connected with an available printer.
1. When the weight is stable, press F key. The display shows the No. and the value of weight is stored in
the memory.
2. Empty the scale by removing the weights. Load the next weight and press F key to store the weight.
3. The No. and total weight could be viewed by pressing the F key when the weight on scale is zero.
4.

The No. and total weight could be deleted by pressing the 0 key when viewing the No. and total

weight.

PRINT OPERATION
1.

The wires of printer should be connected to the 3 rd and 5th connector of the male socket sent as

accessory.
2. Make sure F1 KEY is set to be value 2 .
3. When the weight is stable, press the F key to print.
Print Format
1.

When P11 RS is set to be prt, and prt is set to be lab 1:


No.

2.

Weight

For example:

001

0.233 kg

When P11 RS is set to be prt, and prt is set to be lab 2:


No.
NT(or GS)

For example:

No.

001

Weight

NT

0.233 kg

THE MEANING OF ERROR MESSAGES

ERROR MESSAGES

DESCRIPTION

SOLUTION

-----

Overload

Remove excessive load


If there is still a problem contact your dealer

Err 4

Zero setting error

Zero setting range is less then the load.


Remove the load and try again

If there is still a problem contact your dealer


Err 6

A/D out of range

Remove load from the scale


Load cell or electronics may be at- fault
If there is still a problem contact your dealer

SPECIFICATION*
Display Digits
Internal Resolution

Easy to read, 6 Digit LED


Up to 500,000 counts

Display Resolution

Up to 20,000 counts, selectable

Display Division

1,2,5,10,20,50 selectable

Decimal Point

0 to 4 places, selectable

Load Cell Excitation

5 VDC, 4 x 350 ohm Load Cells

Analog Input Range

from 0.5 mV/V to 3.0 mV/V

Analog Sensitivity

0.5uV/grad min., 1uV/grad recommended

Conversion Rate

20 meas/sec, nominal

Overload Limit

102% of selected Display Resolution

Tare Range
Operating Temperature
Power

ON/OFF, 0, T, MODE, F
-10o to +40oC (14o to 104oF)
AC/DC Adapter 9 - 12VDC, 300 500mA
or Rechargeable Battery