Академический Документы
Профессиональный Документы
Культура Документы
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, VOL. 9, NO. 3, JUNE 2015
it may be difcult to maintain adequate medical care. Therefore, health monitoring technology will enter the personal
environment of humans. Among the available technologies,
bioimpedance is a potential candidate, as it facilitates assessment of body composition and cardiac status in an easy-to-use,
inexpensive and non-invasive way. Once integrated into textiles, the monitoring of patients 24/7 becomes possible. The
integration of bioimpedance devices into textiles is challenging
since textile electrodes imply a high load for current sources
and often the impedance may be larger than the value given in
the specications.
In this work, a simulator was developed for validating bioimpedance instruments which aim to assess the
bioimpedance on the human thorax. Since the simulator produces a time-variant complex impedance, both bioimpedance
spectroscopy (BIS) devices and impedance cardiography
(ICG) devices (or combinations thereof) can be tested. An
electrocardiographic signal (ECG) is produced in parallel
to the impedance for ICG measurements since ICG devices
calculate certain parameters from the combination of ECG and
impedance measurements. In addition, electrode-skin contact
impedances caused by textile electrodes can be altered to test
current sources of different devices. Previous studies concentrated on resistive simulations, thereby neglecting all capacitive
effects and focusing only on standard ICG devices [2][4].
Below we describe the measurement methods, the technical
implementation of the simulator, and the measurement results.
I. INTRODUCTION
1932-4545 2014 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
413
with its
(2)
with the following real and imaginary part:
(3)
(4)
For low and high frequencies, the impedance becomes real.
The above-mentioned (2) would represent a semicircle in a
complex frequency locus plot. Since real measurements show
a depressed semi-circle, a heuristic factor as a tissue-specic
dispersion parameter was introduced. In that way, the electrical
behavior of biological tissue can be described completely by the
Cole equation [7]
(5)
C. Electrodes
Electrodes are transducers between electron-based charge
transports in conductors and ionic charge transports in the body.
Attaching electrodes to the body, a half cell is formed with the
414
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, VOL. 9, NO. 3, JUNE 2015
(8)
In comparison to standard metal electrodes, textile electrodes
at the interface due to
show a strong capacitive behavior
the lack of hydrogel which is often used as electrolyte layer.
Sweat and skin humidity result in an additional conductive
path
in parallel to the capacitance. Thus, for textile
electrodes an equivalent circuit can be assumed (Fig. 3) and the
total impedance for the equivalent circuit can be calculated as
follows [12], [13]
(9)
In practice, two different measurement methods are used
when measuring bioimpedance by attaching electrodes on the
skin: bipolar and tetrapolar measurements. For bipolar measurements, the same electrode pair is used for current injection
as well as voltage measurement. That way,
for standard
electrodes [see (8)] or
[see (9)] will contribute twice to the measurement result since the current ow
through the electrodes causes a voltage drop over the interface
impedance. To avoid this, since only the tissue under test is
intended to be measured, tetrapolar measurements are preferred
using a second electrode pair for voltage measurements only
(Fig. 4).
Fig. 7. Schematic of
415
schematic of
composed of two double-throw switches and
implemented resistances.
in combination with the Cole circuit represent the frequency dependent complex dynamic impedance of the thorax.
Note that changes in
are much larger than changes in the
impedance during the heartbeat
.
Fourth, in parallel to this body impedance, an ECG has to be
created since ICG devices need this signal to calculate certain
parameters such as the pre-ejection period (PEP). This was implemented using a look-up table created from a measured ECG
in combination with the microcontroller's 12 bit digital/analog
output. The ECG generation circuit is shown in Fig. 8 without
the impedance circuit.
Since the microcontroller produces voltages up to 3.3 V
, and maximum ECG values are between 1 mV and
2 mV, a voltage divider was implemented by
and
. To
avoid a current ow through the voltage divider caused by the
bioimpedance device, since this resistance path is not considered to be part of the overall bioimpedance circuit, operational
ampliers were inserted between voltage divider and thoracic
impedance. In addition, the ampliers are serving as a buffer
to provide a precise ECG voltage level. The temporal shift
between ECG and ICG signal was physiologically matched
according to the cardiac phase.
Fifth, up to 8 complex electrode interface impedances [see
(9)] should be implemented since most ICG devices rely on 2
tetrapolar measurement leads.
and
were neglected and the interface impedances were calculated using one
bipolar
and one tetrapolar
measurement. The bipolar
measurement is necessary to measure the interface impedances
of both electrodes and the tetrapolar measurement assesses only
the impedance of the object of interest. Hence, subtracting both
measured impedances and dividing the resulting impedance by
2 calculates the interface impedance of one electrode
416
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, VOL. 9, NO. 3, JUNE 2015
TABLE I
TEXTILE ELECTRODE CHARACTERISTICS
TABLE II
ESTIMATED MODEL PARAMETERS FOR TEXTILE ELECTRODE-SKIN
CONTACT IMPEDANCES
. The measurements were made using a commercially available BIS device on a male human with three
kinds of textile electrodes from Clothing Plus, Finland.
As can be seen in Table I, all electrodes had the same size;
however, because they were made of different materials the material thickness varies resulting in different material weights.
Fig. 9 shows the measured and modeled contact impedance of
one exemplary textile.
Interestingly, these contact impedances does not reect model
assumptions found in literature. On the one hand, the contact impedances do not tend to zero for high frequencies which would
be the case for a textile contact impedance as shown in Fig. 3.
On the other hand, a standard RC element would represent a
semi-circle in the negative half-plane of the frequency locus
plot. Here, we have a deviant semi-circle which would correspond to the factor known from biological materials or tissues.
These ndings lead to a Cole-like impedance model for textile
electrode-skin contact impedances and (9) must be modied
(10)
Hence, this impedance is represented by a resistor
nected in parallel with a resistor
and a capacitor
estimated values for this model are listed in Table II.
con. All
417
Fig. 12. Complex impedance measured with the Xitron, SFB7 and BCM.
TABLE III
MEASUREMENT RESULTS: BIS
418
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, VOL. 9, NO. 3, JUNE 2015
Fig. 13. ECG, Impedance and ICG measured with the Sensatron.
Fig. 16. ECG, Impedance and ICG measured with the Niccomo.
average the signal to cancel thermal noise or respiratory and motion artifacts, since this is necessary to determine the characteristic points of the ICG curve. In commercially available devices,
this is done automatically before calculating the SV [18], [19].
Fig. 14 visualizes the very good match of the measured
unltered impedance and the original preset impedance since
the peak-to-peak impedances match perfectly, the RMS error
is 0.0064 , and the peak value of the cross-correlation is
, which shows that the morphology is nearly identical. The remaining ripple is an inherent error of the switches,
which have an on-state resistance between 0.26 and 0.3 ,
producing an error of 0.04 in the worst case. Here, this ripple
is negligible due to the subsequent ltering of the ICG curve
before characteristic point extraction. Thus, none of the characteristic points are shifted. The only difference is a decrease
of the ICG maximum (0.02 /s) leading to a SV difference of
TABLE IV
MEASUREMENT RESULTS: ICG
reect realistic values measured on a human thorax. These parameters are heart rate (HR), stroke volume (SV), pre-ejection
period (PEP), left ventricular ejection time (LVET) and cardiac
output (CO). For this work, we used the Bernstein-Sramek equation to calculate the SV since the Niccomo uses the same equation internally. The results are shown in Table IV.
The implemented impedances, as described in Section III, do
not affect the measurements, as stated in Section II.C, if the current source of the measurement device was designed correctly.
Hence, the same measurement results were obtained with all devices, independent from the type of electrode.
V. DISCUSSION
Only three works are known to the authors that deal with
the development of ICG simulators for testing and calibration
of ICG devices. All works have in common, that they show
no physiologic impedance curves produced by the simulator.
Only one paper describes a rectangular impedance change used
for the simulation of an electroglottographic signal [3]. However, a rectangular signal would be not sufcient for testing of
ICG devices since a rectangular impedance change would produce dirac-like pulses for rising and falling edges. This temporal derivative could not be analyzed to extract characteristic
points. Unfortunately, no minimal quantization is mentioned in
this paper so that no statement can be given if the circuit would
basically be capable of producing realistic impedance changes
for ICG. But since this circuit does not contain an ECG generator, it does not seem to be developed to serve as an ICG
simulator.
A second work claiming to present an ICG simulator rather
introduced a square impedance generator with an amplitude
of 0.2
[4]. Another work which describes a simulator
containing an ECG generator and aiming to develop an ICG
simulator seems to follow the same approach: the minimal
resolution of the simulated resistance is so small that only
steps of 0.0237
can be realized using the lowest possible
base impedance. The dynamic impedance was implemented as
[2]. These steps would also cause
an ICG device to produce only uninterpretable derivatives of
the measured impedance.
Another commonality of all works is that they implemented
only resistors so that all capacitive effects occuring inside and
outside the body are neglected. One the one hand, this restricts
the simulator to be used for ICG devices only. Neither bioreactance devices nor BIS devices can be tested. On the other hand,
pure resistive changes can not reect the capacitive load of the
419
420
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, VOL. 9, NO. 3, JUNE 2015
[4] N. S. Manigandan, V. K. Pandey, and P. C. Pandey, Thoracic simulator for impedance cardiography, in Proc. Nat. Symp. Instrumentation, Pantnagar, India, Nov. 2003.
[5] International Electrotechnical Comission, IEC 60601-1-1: Medical
electrical equipment, 2000.
[6] H. Schwan, Electrical properties of tissue and cell suspensions, Adv.
Biol. Med. Phys., vol. 5, pp. 147209, 1957.
[7] S. Grimnes and O. G. Martinsen, Bioimpedance and Bioelectricity Basics, 2nd ed. Amsterdam, The Netherlands: Elsevier, 2008.
[8] T. Debski, Y. Zhang, J. R. Jennings, and T. W. Kamarck, Stability of
cardiac impedance measures: Aortic opening (b-point) detection and
scoring, Biol. Psych., vol. 36, pp. 6374, 1993.
[9] J. M. V. D. Water, T. W. Miller, R. L. Vogel, B. E. Mount, and M. L.
Dalton, Impedance cardiographyThe next vital sign technology?,
Chest, vol. 123, pp. 20282033, 2003.
[10] D. P. Bernstein and H. J. M. Lemmens, Stroke volume equation
for impedance cardiography, Med. Biol. Eng. Comput., vol. 43, pp.
443450, 2005.
[11] L. A. Geddes and L. E. Baker, Principles of Applied Biomedical Instrumentation. New York, NY, USA: Wiley, 1968.
[12] L. Beckmann, C. Neuhaus, G. Medrano, N. Jungbecker, M. Walter,
T. Gries, and S. Leonhardt, Characterization of textile electrodes and
conductors using standardized measurement setups, Phys. Meas., vol.
31, pp. 233274, 2010.
[13] T. InOh, S. Yoon, T. E. Kim, H. Wi, K. J. Kim, E. J. Woo, and R.
Sadleir, Nanober web textile dry electrodes for long-term biopotential recording, IEEE Trans. Biomed. Circuits Syst., vol. 7, no. 2, pp.
204211, Apr. 2013.
[14] P. Squara, Bioreactance: A new method for non-invasive cardiac
output monitoring, Intensive Care Med., pp. 619630, 2008.
[15] L. A. H. Critchley, Impedance cardiography: The impact of new technology, Anaesthesia, vol. 53, pp. 677684, 1998.
[16] L. Ward, B. H. Cornish, N. I. Paton, and B. J. Thomas, Multiple
frequency bioelectrical impedance analysis: A cross-validation study
of the inductor circuit and cole models, Phys. Meas., vol. 20, pp.
333347, 1999.
[17] J. Muehlsteff, P. Carvalho, J. Henriques, R. P. Paiva, and H. Reiter,
Cardiac status assessment with a multi-signal device for improved
home-based congestive heart failure management, in Proc. Annu. Int.
Conf. IEEE Engineering in Medicine and Biology Soc., Aug. 30-Sep. 3,
2011, pp. 876879.
[18] W. Hu, H. H. Sun, and X. Wang, A study on methods for impedance
cardiography, in Proc. 19th Annu. Int. IEEE Conf. Engineering in
Medicine and Biology Soc., 1997, vol. 5, pp. 20742077.
[19] O. Dromer, O. Alata, and O. Bernard, Impedance cardiography ltering using scale fourier linear combiner based on rls algorithm, in
Proc. Annu. Int. Conf IEEE Engineering in Medicine and Biology Soc.,
Sep. 2009, pp. 69306933.
[20] P. Carvalho, Robust characteristic points for ICG: Denition and comparative analysis, in Proc. Int. Conf. Bio-Inspired Systems and Signal
Processing, 2011.
[21] M. Min, QuadraImpedance Spectroscopy Device for Dynamic Measurements, 2012.
[22] M. Ulbrich, J. Mhlsteff, M. Walter, and S. Leonhardt, Simulation of
lung edema in impedance cardiography, Comput. Cardiol., vol. 39,
pp. 3336, 2012.
Daniel Teichmann (S'12) was born in Essen, Germany, in 1982. He received the Dipl.-Ing. degree in
electrical engineering from RWTH Aachen University, Aachen, Germany.
Currently, he is working toward the Dr.-Ing.
(Ph.D.) degree at the Chair of Medical Information
Technology, RWTH Aachen University, where he is
a Research Assistant. His research interests include
physiological measurement techniques and signal
processing.