Optical Biometry & Topography System

ALADDIN

WE LET YOU SEE THE COMPLETE PICTURE


Today accurate IOL-power calculation is no longer enough.
The final optical result and patients satisfaction are paramount.
By incorporating full corneal topography and wave front analysis
of the cornea the ALADDIN addresses the limitations of conventional
biometers.
Topcon provides the complete picture with the
ALADDIN Biometer with Corneal Topography.
The ALADDIN supports the surgeon not only in his/her choice
of the spherical power of the IOL, but also assists in the
choice of the right premium IOL for each individual eye.
Posterior & Anterior interferometry

Biometry results are complemented with anterior


topography, Zernike analysis and pupillometry in one fast,
accurate and easy acquisition. The Interferometer of the
ALADDIN also provides anterior measurements such as
the Central Corneal Thickness (CCT), Anterior Chamber
Depth (ACD) and Lens Thickness (LT).
The ALADDIN provides the complete picture for
cataract surgeries. Whether during standard
cataract surgery or premium IOL implantation,
the ALADDIN allows the surgeon to screen
for corneal aberrations, Keratoconus and
previous corneal refractive surgery procedures
all at once. ALADDIN only requires one acquisition.

9 Functions; 1 Instrument
Axial Length
Keratometry
Anterior Chamber Depth
Central Corneal Thickness
Lens Thickness
Corneal Topography
Corneal Wave-front Analysis
Pupillometry
White to White

ALADDIN features

Fully integrated patient database


- Patient search function
- Input post op data
Easy acquisition 9 in 1
- Pre op input of lens and vitreous body
Conventional IOL calculation formulae
- SRK II, SRK/T , Hoffer Q, Holladay 1, Haigis
- Multiple surgeon pre-settings
- ULIB database compatible
- Database customizable
Post refractive IOL calculation formulae
- Camellin-Calossi, Shammas (no history)
Generic toric IOL calculation
- Toric IOL rotation simulator
Topography
- Full featured mapping
- Accurate corneal radii
- Keratoconus probability index
Corneal wavefront (Zernike) analysis
- Maps (pupil size 2,5 7.0 mm)
- Simulation graphs
Interferometer graphs
- Axial length
- Central corneal thickness
- Anterior chamber depth
- Lens thickness
Pupillometry
- Dynamic, Photopic, Mesopic
- Decentralization and Latency graph
White to white measurement
Reports
- To USB, shared folder and printer
- Topography report
- IOL report
- Biometry report (AL, K, ACD, LT, CCT WTW)
- Pupillometry

The ALADDIN was developed with three key points in mind:


SPEED
Point and shoot acquisition. All
measurements are taken in
under 5 seconds. Anterior
Chamber Depth (ACD), Axial
Length (AL), Lens Thickness (LT),
Central Corneal Thickness,
Corneal Topography, as well as
full pupillometry are easily
obtained.

ACCURACY
Proven interferometry
accuracy and the advanced
technology for keratometry
in combination with the use
of interferometry provide
accurate Axial Length and
cornea radii information for
precise calculation of IOL
spherical and toric powers.

Keratometry / Topography
Full corneal topography provides substantially more
information than conventional central keratometry.
Corneal topography data is especially useful in the
selection of toric IOLs to quickly differentiate regular and
irregular astigmatism as well as corneal aberrations. The
ALADDIN provides accurate corneal topography obtained
from the reflection of a set of 24 Placido rings in
combination with a low coherence interferometer.

EASE OF USE
The operator is only 3 clicks
away from printing the
ALADDIN report. The 10.1
inch full color touchscreen
monitor with its wide angle
of view is responsive and
comfortable to use. The
friendly user interface guides
you through the main
functionalities with ease.

Keratoconus screening
The Keratoconus Screening function screens the
corneal surface for keratoconus probability. This
information provides the surgeon detailed corneal
indices to assist in making the correct toric IOL
selection. The Keratoconus Probability Index is shown in
percentage as well as in color codes.
Green : Not compatible with keratoconus
Yellow: Suspected keratoconus
Red : Compatible with keratoconus

Aberrometry analysis (Zernike)


Zernike analysis of the topographic data
provides the Optical Path Difference (OPD)
and information on astigmatism, spherical
aberrations, higher order aberrations and
Coma for pupil sizes of 2.5mm to 7.0mm
When using the actual Spherical Aberration
provided by Zernike analysis, you can select
the appropriate aspherical IOL with
standardized spherical aberration correction
according to the patients individual required
spherical aberration.

Axial Length / Anterior biometry


Using a low-coherence interferometry
system with a super luminescent
diode of 850 nm and signal processing
ALADDIN achieves Axial Length
measurement with high signal-tonoise ratio and is able to penetrate
even high grade dense cataracts. Axial
length measurements can be
performed on normal eyes as well as
on aphakic, pseudophakic and silicone
oil-filled eyes.

Anterior biometry with the ALADDIN


makes it possible to measure Central
Corneal Thickness, Anterior Chamber
Depth and the Crystalline Lens
Thickness. Pachymetry is a key
measurement for all cataract surgery
procedures. ACD is measured through
interferometry providing high precision
and reproducibility. All interferometry
measurements are shown in a single
graph quick reference.

*) ALADDIN LT model

Pupillometry

During Placido evaluation, pupillary response is observed to assess a pseudo Photopic and pseudo
Mesopic pupil size, indicating response and normal range of the pupil. Full pupillometry screening assists
to evaluate eyes for multifocal IOL implantation or refractive surgery. For any refractive procedure it is
important to carefully evaluate the pupil size in different light conditions to address cases of extremely
small or decentered pupils. The ALADDIN can perform pupillometry in three different modes:

Dynamic
Photopic
Mesopic

White to white
The ALADDIN automatically measures
white to white length which can be
manually edited. Reliable white to
white measurement is used for
anterior chamber intraocular lens and
sulcus fixated posterior chamber
intraocular lens calculation in highly
myopic eyes.

IOL & Toric IOL calculation


The ALADDIN guides you through the
choice of the appropriate IOL for each
patient. A combination of IOL brand, type
and formulae can be viewed and
compared to various chosen
combinations, in order to obtain the
most effective post-operative Visual
Acuity result for the patient.
A pre-defined IOL selection can be
programmed for each surgeon.
When implanting a toric IOL, specific toric
calculation software assists the surgeon
in calculating the best choice of toric IOL.
This integrated toric IOL calculator saves
you time and avoid unnecessary mistakes
when manually entering data online.
IOL toric rotation simulation software
calculates the induced spherical and
cylindrical power for every 5 degrees that
the toric IOL rotates.

Post refractive IOL


In eyes that have previously undergone
refractive surgery such as RK, PRK, Lasik,
Lasek, LK and PTK, spherical aberrations are
often outside the standard values. In these
cases the ALADDIN provide the Camellin
Calossi formula. If there is no patient history
the Shammas No-history formula can be
used for the correct IOL calculation.

Customizable IOL database


The ALADDIN provides a full ULIB
database which can be easily
upgraded and customized. The
surgeon can manually upgrade the Aconstant for each individual IOL to
obtain even a higher accuracy every
time a cataract surgery is performed.
Favorite IOLs can be selected and
programmed for each surgeon,
making IOL selection simple and
personalized.

Topcons Cataract Workstation


Visual acuity (VA) is the most common clinical measure of the quality of cataract surgery. It is
how the success of surgery is measured and it is therefore, critical that it is measured correctly.
Measurement of VA must be standardized and systematic. Topcons KR-800S Auto KeratoRefractometer with subjective VA check will do exactly that. With the KR-800S the VA can be subjectively
tested before and after cataract surgery. With the unique features of the KR-800S such as Glare test
and Contrast test, the progression and differentiation of nuclear cataract from cortical cataract can be
evaluated.

KR-800S
Premium IOL simulation
The KR-800S offers a Spherical Equivalent mode, which can simulate
the benefit of a premium (toric) IOL , to encourage the patient to reach
an even higher post-operative VA. The subjective VA test for near vision will
demonstrate to the patient the benefits of a Multifocal IOL.

1.
2.
3.
4.
5.
6.

VA check far vision


VA check near vision
VA check glare condition
VA check contrast condition
Grid test (AMD screening)
Simulate benefit Premium IOL

Reports Measurement Summary

Reports IOL Calculation

Reports Generic Toric IOL Calculation*

Reports Biometry

Reports Pupillometry

Specifications
Measurement Range for IOL
Axial Length (Interferometry)
Corneal Radii
Anterior Chamber Depth Measurement
White-to-White Measurement
Pupillometry
Lens Thickness (Interferometry)
Central Corneal Thickness (Interferometry)
On-board Calculation Formulae
IOL Formulae
Post Refractive IOL Formulae
Placido Topography Specifications
Keratoscopic Cone
Points Analyzed
Points Measured
Cornea Coverage
Guided Focus System
Keratoconus Screening
Apical Curvature
Apical Gradient of Curvature
Symetric Index
KpI (Keratoconus Probability Index)
Software Features
Toric IOL Calculator
Zernike Analysis
Print To
Instrument Specifications
Display
Storage
Operating System
Processor
Internal Memory
Power Input
Dimensions
Weight
Connections
Supports
Markings
Reports
Aladdin Report
Measurement Overview
Pupillometry
IOL
Generic Toric IOL

Super luminescent diode 850Nm, 15mm 38mm


3.3mm 37.5mm / 9.0D 101.5D
Interferometer 1.5mm 5.5mm
6.0mm 18.0mm
Dynamic, Photopic and Mesopic 0.5mm 10mm
0.300mm 0.800mm
0.300mm 0.800mm
Haigis, Hoffer Q, Holladay 1, SRKII, SRKT
Camellin Calossi and Shammas No History
24 rings on a 43D sphere, working distance 80mm
Over 100,000
Over 6,200
Up to 9.8mm (on a 8mm sphere) 42.2D with N=1.3375
Yes
Yes
Yes
Yes
Yes
Generic Toric IOL
Pupil size from 2.5mm to 7.0mm
USB printer, network printer, PDF to shared network
folder, PDF to USB drive
10.1 touch screen
320 GB HDD + 32 GB SSD
Windows 8 Embedded OS
AMD G-T56N
2GB RAM
AC 100 240V 46-63 Hz
320 mm (W) x 490 mm (H) x 470 mm (L)
18 kg
1x LAN, 2x USB
USB barcode scanner, external USB keyboard / mouse
CE, ETL
Yes
Yes
Yes
Yes
Yes