Академический Документы
Профессиональный Документы
Культура Документы
The half cell is a stable reference electrode used to measure the potential of another electrode. There are several types of reference
electrodes, but the most common types are saturated Copper/Copper Sulfate (CSE) electrode, Silver/Silver Chloride (SSC) electrode, and
Saturated Calomel Electrode (SCE).
a) Half-Cell Selection
The selection of reference electrode depends on the environment of application:
b) Half-Cell Calibration
This can be done by measuring the voltage difference between the used reference and another standard (unused) reference, as shown
below.
If the voltage difference between the two reference electrodes is more than 5 mv, maintenance is required for the field reference
electrode.
c) Half-Cell Maintenance
The copper rod may be cleaned with a 10% nitric acid solution and the rod should be immersed for several minutes to remove
all surface products and contaminants. Sandpaper may be used for cleaning the copper rod.
The Copper Sulfate solution should be saturated, with loose crystals visible, and the end cap placed over the plug when not in
use (to prevent leakage and contamination).
Field Connections
This is the potential of the buried pipeline to be measured with respect to a stable reference electrode. For accurate measurements, the
reference electrode needs to be placed directly over the center line of the pipe.
The pipeline needs to be connected to the voltmeter negative terminal and the reference electrode to the positive terminal. By
this connection, the current will flow from the voltmeter positive terminal to the negative terminal.
The displayed value is positive, indicating that the reference electrode is more positive than the pipeline. The electrical continuity of the
pipeline needs to be checked before the measurement.
The input impedance of the voltmeter needs to be higher than the total measurement circuit resistance. Input impedance of 10
Ohm is sufficient for accurate measurement, while lower values may be accepted if they're higher than the total circuit
resistance.
Note: The difference between errors can be due to the design accuracy of the voltmeter and due to voltage drop errors in the measuring
circuit.
The contact resistance of the reference electrode needs to be as low as possible. This can be done by wetting the contact area,
especially for frozen, concrete or asphalt layers.
Test leads needs to be checked for any broken points, bare areas or bad connections.
For accurate measurements, it must be taken in the upper two-thirds of the selected range for a particular instrument.
In order to eliminate voltage drop errors due to the passage of current in the electrolyte, one of these needs to be considered:
The reference electrode should be placed very close to the buried pipeline, maybe at a distance twice the reference electrode
diameter.
Interruption of CP current sources instantaneously in order to measure the polarized potential of the pipeline. CP current
sources includes impressed currents, galvanic anodes, stray currents and electrical bonds.