Академический Документы
Профессиональный Документы
Культура Документы
2. We switched to Program, then we set the flow of how the TGA would run. The entire
program can also be seen exactly the same under View Program. The program of our very
first run is shown below. We based this from the one provided by the Perkin Elmer
technician/operator.
3. Once everything is set, the Method can now be saved: File Save Method.
C. Running the Method/Program
Once the Method is defined, the user can now run the program using the following procedure:
1. Load/place the pan in TGA. In the Control Panel located at the right side of the window,
click the Zero Weight button. See Figure 3, for the important buttons in the Control
Panel.
Start/Stop
Go to Temp
Cooling Air
Sample Weight
Zero Weight
2. Ensure the Sample Temp 50 C and the Status is At Temp. This can be viewed as
shown in Figure 4. If the Sample Temp 50 C then click the Go to Temp button in the
Control Panel.
3. Place sample in the pan and then cover. The TGA weighs the sample and the reading is
indicated under Weight in Figure 4. Then, in the Control Panel, click the Sample
Weight button. The value of the Weight that is indicated in Figure 4 should then replace
the value of Weight in the Sample Info section of the Method Editor in Figure 1 and
the Weight % should become 100%.
COMMENT: In Figure 4, once the sample is placed in the pan, the Weight value should
ideally reflect the actual mass of the sample. However, during the actual operation, that
Weight does not give the expected value (only about 5% during the first run) or worse does
not give any value at all (0.000 mg in the second and run) as can be seen in Figure 4. Perhaps,
there could be some problem in the precision balance of the TGA or with the calibration.
4. If the TGA is still reading, wait for the Weight to be equal to the samples weight and the
Weight % to be equal to 100%.
COMMENT: During the actual run, the TGA reading stabilizes only to ~5.66% of the
expected value. This is clearly a big problem that a TGA cannot give an accurate value!
According to Perkin Elmers Technician, the Weight under the Sample Info in the
Method Editor dialog box can be manually inputted. By doing so, the Weight in Figure 4
does not really change (still does not give the accurate reading) but the Weight %. The only
way to change its value and for the Weight% to be equal to 100% is to manually adjust the
value of the Zero under the Sample Info in the Method Editor dialog box. This is what
the Perkin Elmer technician did. As a matter of fact, in the most common concept of
weighing, this is very wrong! The Zero there reflects the weight of the pan (or any other
else in the TGA furnace during calibration) which is set to be the reference value.
5. In the control panel, click the Start button.
6. After a while, a TGA curve will appear on the screen like in Figure 5. The Program Temp
value will change depending on the Program of your Method. The Sample Temp,
Weight and Weight% also change during the run. The TGA curves of the other methods
and runs are in the Appendix.
7. The TGA curve can be exported as an ASCII file: File Export As ASCII. Save
the ASCII file to your own folder.
D. Running for the next sample
Before running for the next sample, make sure that the Sample Temp value drops to 50.00 C.
Cooling down may take longer but for a faster cooldown, click the Cooling Air button in the
Control Panel.
E. Shutting down of the equipments
To turn off the TGA, the procedure are followed:
1. Make sure the Program Temp is 50.00 C. Dont turn off at temperature above 50 C.
2. Click Start Pyris button Close All. Wait for all the windows to close.
3. Shutdown the computer.
4. Turn off the TGA.
5. Turn off the chiller starting from the face switch button then to the back switch button.
6. Turn off the compressor.
7. Close the Nitrogen tank.
APPENDIX
TGA DRY RUN METHODS AND RESULTS
A. Method and TGA curve for Dry Run 1