Академический Документы
Профессиональный Документы
Культура Документы
analysis. The study was conducted entirely by Ohmicron according to AOAC requirements for collaborative studies (6).
Collaborative Study
A. Principle
Water sample is mixed with horseradish peroxidase-labeled
atrazine conjugate, and then paramagnetic particles with attached triazine-specific antibodies are added to the sample mixture. Atrazine present in sample competes with enzyme-labeled
atrazine for binding sites on antibodies. After 30 min incubation, a magnetic field is applied to hold paramagnetic particles
in tube while excess reagents and other unbound material are
washed away. After separation of antibody-bound atrazine and
atrazineenzyme conjugate from free components, particles are
washed with wash solution. Atrazine bound to paramagnetic
particles is identified by adding substrate and a chromogen (hydrogen peroxide and 3,3,5,5-tetramethylbenzidine [TMB]
chromogen solution) that generates blue color. After another
(20 min) incubation period, reaction is stopped. Intensity of
color is proportional to enzyme-labeled atrazine and inversely
proportional to initial concentration of atrazine in test sample.
Absorbance is measured at 450 nm. Atrazine concentration is
determined from standard curve.
B. Apparatus
(a) Test tubes.12 75 mm, polystyrene.
(b) Micropipets.Capable of accurately delivering
200 L; used for test samples, atrazine standard solutions, and
diluent/zero standard solution.
(c) Repeater pipet.Capable of accurately delivering 250,
500, and 1000 L; used for enzyme conjugate, antibody particles, wash, and stop solutions.
(d) Spectrophotometer.Capable of measuring absorbance at 450 nm through 12 75 mm polystyrene test tubes
containing 1 mL sample.
(e) Magnetic separation rack.2-piece plastic unit consists of test tube holder that fits over magnetic separation rack
containing rare earth magnets (available from Ohmicron Environmental Diagnostics).
(f) Vortex mixer.
(g) Absorbent pad.About 5 layers of paper towels.
(h) Filter paper.0.2 m.
(i) Glass tubes.Borosilicate, holding 50 mL.
C. Reagents
Items C(a)(j) are available as Atrazine RaPID Assay Kit
(Ohmicron Environmental Diagnostics, Newtown, PA; Part
No. A00002, 30 tests/kit, or Part No. A00071, 100 tests/kit).
(a) Atrazineantibody-coupled magnetic particles.Rabbit antitriazine antibodies covalently bound to paramagnetic
particles and suspended in buffered saline containing antimicrobial preservatives and protein stabilizers.
(b) Atrazineenzyme conjugate.Horseradish peroxidase-labeled atrazine analog diluted in buffered saline containing antimicrobial preservatives and protein stabilizers.
(c) Atrazine standard solutions.0.1, 1.0, and 5.0 g/L;
prepared in buffered saline containing antimicrobial preservatives and protein stabilizers.
(d) Quality control (QC) solution.Approximately 3 g
atrazine/L in buffered saline containing antimicrobial preservatives and protein stabilizers.
(e) Diluent/zero standard solution.Buffered saline with
antimicrobial preservatives and protein stabilizers, containing
no detectable atrazine.
(f) Hydrogen peroxide.0.02% H2O2 in citric acid buffer.
(g) TMB chromogen solution.3,3,5,5-Tetramethylbenzidine, 0.4 g/L, in N,N,-dimethylformamide.
(h) Stop solution.2M H2SO4. Provided in ready-to-use
form in 30-test kit. When using 100-test kit, prepare stop solution before analysis by pouring contents of stop solution bottle
into stop solution diluent and mixing. Some warming of solution will occur.
(i) Wash solution.Preserved deionized H2O.
(j) Loglogit graph paper.
D. General Instructions
Reagents are stable until the expiration date when stored at
4C. Do not use reagents after expiration date.
Bring all reagents to room temperature (2025C) before
analysis. Mix thoroughly antibody-coupled paramagnetic particles before use. Do not transfer antibodies to glass containers.
Treat each test tube in identical manner to obtain optimal
performance and precision.
Add reagents directly to the bottom of test tube; avoid contact between reagents and pipet tip to ensure consistent quantities of reagent in test mixture.
To avoid cross-contaminations and carryover of reagents,
use new pipet tips for each addition of test sample. Avoid contact of pipet tips with droplets of reagent on walls of test tubes.
Magnetic separation rack consists of (1) upper rack that securely holds test tubes and (2) lower separator containing magnets used to attract antibody-coupled paramagnetic particles.
Before incubation, upper rack is removed from lower separator
to allow paramagnetic particles to remain suspended during incubation. For separation steps, rack and separator are combined
to pull paramagnetic particles to sides of test tubes.
Mix antibody-coupled paramagnetic particles just before
pipetting and add mixture to all test tubes within 2 min to ensure consistent results.
E. Preparation of Sample
Filter through filter paper, B(h), water samples containing
gross particulate matter. If acid was added to environmental
water sample to preserve analyte, neutralize sample with strong
base (e.g., 6N NaOH) before analysis.
G. Determination
(1) Before beginning assay, label polystyrene test tubes for
standards, QC solution, and test samples. See Table 995.08C
for labeling pattern of test tubes.
(2) Separately place 200 L atrazine standard solutions,
C(c), QC solution, or test sample into labeled tubes.
(3) Add 250 L atrazineenzyme conjugate, C(b), to each
tube.
(4) Mix thoroughly atrazine antibody-coupled paramagnetic particles, C(a). Add 500 L to each tube.
(5) Mix contents of test tubes for 12 s on Vortex mixer.
Minimize foaming by gently pressing test tube to Vortex mixer
preset at low speed.
(6) Incubate test tubes 15 min at room temperature (20
25C).
(7) Allow antibody-coupled paramagnetic particles to
separate from test mixture for 2 min in magnetic separation
rack, B(e). Note: To obtain optimum precision of assay, perform separation steps carefully and consistently.
(8) Decant excess reagent from tubes. Gently blot test tubes
in a consistent manner, as follows: Invert rack away from analyst by using smooth turning action, so liquid flows consistently along only one side of test tubes. While rack is still inverted, blot all tubes briefly on absorbent pad. Lift rack and
place it gently onto absorbent pad several times to remove liquid completely from rim of test tubes.
(9) Add 1 mL wash solution, C(i), to each tube and allow
tubes to remain in magnetic separation unit for 2 min.
(10) Decant wash solution from test tubes, and then gently
and briefly blot tubes on absorbent pad in consistent manner.
(11) Repeat steps (9) and (10).
(12) Remove upper rack from separator and add 500 L
color reagent from F to each tube.
(13) Mix contents of test tubes for 12 s on Vortex mixer.
Minimize foaming as in (5).
(14) Incubate tubes 20 min at room temperature.
(15) Add 500 L stop solution, C(h), to each test tube.
Note: No mixing on Vortex mixer is required.
(16) Read absorbance, A, for each test tube at 450 nm
wavelength 15 min after adding stop solution.
H. Manual Calculations
(1) Calculate mean A for each atrazine standard solution.
(2) Calculate A/ADZ ratio (%) for each atrazine standard solution, where ADZ = mean absorbance for diluent/zero standard
solution.
(3) Construct standard curve on loglogit graph paper, by
plotting data as y-axis = A/ADZ ratio for each atrazine standard
solution, %; and x-axis = corresponding atrazine concentration, g/L.
(4) Calculate A/ADZ ratios for QC solution and test samples
using corresponding A values. Read atrazine concentration in
QC solution and test samples using standard curve from (3).
If atrazine concentration in test sample is >5 g/L, dilute
sample at least 10-fold (e.g., 50 + 450 L) with diluent/zero
standard solution, C(e), and repeat analysis, beginning at G. It
is not recommended to use <50 L sample for dilution. Mix
solution thoroughly before analysis. Calculate atrazine concentration by multiplying obtained value by dilution factor (e.g., 10).
J. Quality Control
Results of analysis are valid when the following QC requirements are met:
(1) Absorbance of diluent/zero standard solution measured
at 450 nm wavelength must be >0.8.
(2) Correlation coefficient (r) of standard curve must be
>0.990.
(3) Concentration of atrazine in QC solution must fall
within 3.0 0.6 g/L.
When routinely testing environmental water samples, QC
measurements might involve inclusion of replicate or multilevel kit control samples in the middle and at the end of test tube
sequence; use of blanks, spikes, or spike duplicates in each
batch; and tracking ratio or distribution of positive and negative
results in stable test sample population.
Before performing analyses, analyst should practice procedure. Video tapes with detailed instructions are available from
Ohmicron Environmental Diagnostics.
Ref.: J. AOAC Int. 79, 529(1996)
Results and Discussion
Treatment of Data
The data from 14 laboratories (Table 1) were reviewed for
completeness and acceptability according to the minimum QC
criteria described earlier. No data points were missing and all
reported runs met the acceptability criteria. All nondetectable
concentrations of atrazine (<0.05 g/L) were converted to zero
before data entry. Results were grouped by water type in sub-
Method Recovery
Summary statistics, after removal of outliers, are shown in
Table 995.08A. Mean recovery at each concentration level in
each water matrix was calculated as the percentage of added
atrazine. Waters with naturally incurred atrazine were tested in
the authors laboratory to estimate the initial atrazine content so
spiking concentrations would allow total measured atrazine to
remain within the assay calibration curve. Blank waters used to
prepare study samples were collected in sufficient quantity and
prescreened in the authors laboratory with the atrazine immunoassay to ensure that atrazine-free starting material was used.
In all cases, results were below the detection limit of 0.05 g/L
atrazine. This detection limit is recommended by Ohmicron to
its kit users on the basis of validation studies performed with
multiple lots of kit reagents with different operators over more
than 20 standard curve runs. The 0.05 g/L detection limit represents the analyte level predicted by a regression analysis of
an average standard curve at 10% displacement from the zero
standard response, i.e., 90% B/Bo. When the 4 blank samples
were analyzed by the collaborators, 9 of 112 determinations
(8%) were reported as detectable, ranging from 0.07 to
0.54 g/L. Seven of the 9 detectable results were from analysis
of municipal or surface water blanks. We have frequently observed 0.05 to 0.20 g/L levels of atrazine in randomly sampled environmental waters. It is possible that these blanks contained very low background concentrations of atrazine such
that they did not behave as a true zero in the system and were
therefore detectable in a small percentage of the results. Another explanation for the positive results on blank waters is
found by further analysis of data from laboratory 5. Four of the
9 detectable blank results occurred in a single run performed in
this laboratory. Comments from the operator suggest that some
delays in timing of reagent addition and reading of optical den-
sity may have occurred in this run. Delays of more than about
2 min in addition of antibody-coupled particles to sample tubes
compared with calibrator tubes will cause sample results to
drift higher. This occurs because the antibodyantigen reaction
proceeds to a greater extent in the calibrator tubes when the
reaction is stopped, and the sample tubes show relatively lower
optical density translated as higher analyte concentrations.
When other nonblank samples included in the same batch by
this laboratory were checked for this trend, results on most
samples were among the highest reported by all collaborators.
This observation supports a kinetic drift explanation for blank
samples. Data from this run were not excluded from the database, because the QC criteria were met and no data points qualified for removal as outliers. However, the result of 0.54 g/L
on the reagent water blank reported by laboratory 9 appears to
be a true outlier. When the rule of the huge error (10) is applied
to the B/Bo ratios of reagent water results from all laboratories,
the 0.54 g/L result is obviously eliminated.
Average recovery for all waters was 104%. Average recoveries for individual spiked matrixes ranged from 80 to 120%,
with one exception: 134% for natural water with background
atrazine at 0.47 g/L. The precision of measurement on this
sample was good, suggesting that the spiking procedure may
have produced the high bias in the results. Recoveries from
samples containing atrazine at 5 g/L are expected to fall below
100% as the loglogit fit of the calibration curve becomes
poorer when binding of unlabeled atrazine is in great excess of
binding of atrazineenzyme conjugate. The average calculated
concentration by regression for the 5.0 g/L calibrator was
4.54 g/L in 41 runs by 14 collaborators. Absorbance data in
this range can be corrected by use of an excess antigen tube
whose optical density is subtracted from each unknown absorbance, producing a more linear response in the calibration
curve at elevated concentrations. Such a reagent was provided
to the collaborators, and it was included in each run for this
purpose. When the naturally incurred waters spiked with
4.0 g/L were corrected for nonspecific antigen binding, recovery of the incurred low sample was not affected (81% recovery), but results on the high sample improved from 89 to 96%.
In the authors experience, if a matrix interference is present in
an environmental sample, high recoveries typically would be
observed, with the lowest spiking concentrations returning to
normal recoveries as more analyte is added. This was not the
case. The water type did not appear to influence the ability of
the immunoassay to recover added atrazine. Regression analysis of spiked atrazine versus measured atrazine in 15 samples
ranging from 0.15 to 4.47 g/L predicts X = 0.91C + 0.17 (r =
0.974).
Results for the QC sample are shown in Table 2. Mean concentration recovered was 3.13 g/L, or 104% of the true (spiked)
value for 14 collaborators in 41 immunoassay batch runs.
Precision
Precision parameters are reported in Table 3. The method
detection limit (MDL) is estimated at 0.05 g/L or 90% binding of the optical density of the zero concentration reaction tube
(maximum-binding tube). Precision of immunoassay results
typically is optimal between about 20 and 80% maximum binding. Because collaborative studies of chromatographic methods for pesticides have been evaluated for precision at concentrations with reference to the MDL, precision parameters for
this study were assessed in this way.
Samples with atrazine concentrations less then 5 times the
MDL (up to 0.25 g/L) exhibited an average single-analyst
relative standard deviation (RSDr) of 23.3 (range, 16.629.9)
and an average overall relative standard deviation (RSDR) of
39.4 (range, 35.843.4). In samples with atrazine concentrations above 5 times MDL (0.25 to 5.0 g/L), the RSDr averaged
10.1 (range, 7.7420.1) while RSDR averaged 15.7 (range,
9.0830.8). Water type did not appear to affect measurement
precision.
Conclusions
The magnetic-particle-based immunoassay is accurate and
precise in the detection range defined by the kit calibrators.
Overall and single-analyst relative standard deviations are
comparable with similar parameters for chromatographic
methods for pesticides (11, 12) approved by EPA for analysis
of drinking water and by AOAC at concentrations above
5 times MDL. The quality of the returned data was excellent,
with less than 2% of data points qualifying as outliers. The simplicity of the immunoassay method, which requires no laborious sample pretreatment steps such as solvent extraction or
analyte concentration, contributes to its robustness in a variety
of laboratory types. The majority of the collaborators were
regular users of immunoassay tests and were familiar with the
pipetting and decanting techniques and reaction timing issues
which can most substantially influence the precision of results.
Proper training in micropipetting techniques, monitoring of
quality control results, and understanding of assay kinetic properties must be established for routine achievement of good results.
This collaborative study format using relatively clean
atrazine-spiked waters has shown the atrazine immunoassay
performance comparable with traditional methods in accuracy
and precision. The method had previously been shown by the
developers to exhibit no interference with 250 mg/L of the ions
of calcium, copper, nickel, magnesium, sulfate, nitrate, and
thiosulfate. Test results were not affected by 100 mg/L of added
humic acid, iron, sulfide, and sulfite or by 0.65M sodium chloride. Reactivity of compounds that are either metabolites of
atrazine or similar in chemical structure and that might appear
in the sample need to be understood if the immunoassay is to
be properly applied. Similar interpretation would be necessary
when coelutants are potentially present in a chromatographic
procedure. This atrazine immunoassay was developed with an
antibody possessing broad specificity for triazine residues. The
analyst must incorporate this facet of immunochemical technology into the interpretation of results from unknown samples
and the design of a QC system. The most practical use for the
method may be as a screening test for evaluation of large numbers of water samples or for monitoring drinking water supply
systems with greater frequency for presence of atrazine at regu-
Recommendation
On the basis of the results of this study it is recommended
that the magnetic particle immunoassay for determination of
atrazine in water be adopted first action.
Acknowledgments
We express appreciation to the following collaborators who
participated in this study:
P. Barnes, Kansas State University, Manhattan, KS
C. Charan and Viorica Lopez-Avila, Midwest Research Institute, Mountain View, CA
J. Fleeker, North Dakota State University, Fargo, ND
H. Francis and T. Haezlit, Kentucky Geological Survey,
Lexington, KY
D.W. Litke and D. Goolsby, U.S. Geological Survey, Denver, CO
C. Hall, K. Sagan, and K. Fuchs, University of Guelph,
Guelph, Ontario, Canada
P. Luitweiler and P. Staudte, Philadelphia Suburban Water
Co., Bryn Mawr, PA
A. Lucas, DuPont Corp., Wilmington, DE
L. Marti, U.S. Department of Agriculture, Tifton, GA
C. Mihaliak, Dow Elanco, Indianapolis, IN
(1) Hammock, B.D., & Mumma, R.O. (1980) in Pesticide Identification at the Residue Level, ACS Symposium Series 136,
American Chemical Society, Washington, DC, pp. 321352
(2) Van Emon, J.M., & Lopez-Avila, V. (1992) Anal. Chem. 64,
79A88A
(3) Drinking Water Health Advisory: Pesticides (1989) U.S.
EPA, Office of Drinking Water Health Advisories, Lewis
Publishers, Chelsea, MI
(4) Meister, R.G. (Ed.) (1987) Farm Chemicals Handbook, 3rd
Ed., Meister Publishing Co., Willoughby, OH
(5) Rubio, F.M., Itak, J.I., Scutellaro, A.M., Selisker, M.Y., &
Herzog, D.P. (1991) Food Agric. Immunol. 3, 113125
(6) Youden W.Y., & Steiner, E.M. (1975) Statistical Manual of
the AOAC, AOAC, Arlington, VA
(7) Rodbard, D. (1974) Clin. Chem. 2/1, 12551270
(8) Baud, M., Mercier, M., & Chatelain, F. (1990) Scand. J.
Clin. Lab. Invest., Suppl. 205, 120
(9) Statistical software, AOAC INTERNATIONAL, Arlington,
VA, AOACBUBR.WK1, 7/20/93 version
(10) Taylor, J.K. (1987) Quality Assurance of Chemical Measurements, Lewis Publishers, Inc., Chelsea, MI, pp. 3435
(11) Edgell, K.W., Erb, E.J., Longbottom, J.E., & Lopez-Avila, V.
(1992) J. AOAC Int. 75, 858871
(12) Edgell, K.W., Jenkins, E.L., Lopez-Avila, V., & Longbottom,
J.E. (1991) J. Assoc. Off. Anal. Chem. 74, 295309
Table 995.08A. Method performance for determination of atrazine in water by magnetic particle immunoassay
Water
Reagent
Municipal
Well
Surface
Contaminated
municipal water
Contaminated well
water
a
b
c
d
e
Spike, g/L
Na
_
xb, g/L
sr, g/L
sR, g/L
rc, g/L
Rd, g/L
RSDr, %
RSDR, %
Recovery,
%
0.00
0.00
0.15
1.00
3.00
0.00
0.15
1.00
3.00
0.00
0.15
1.00
3.00
14
14
14
13
14
14
13
14
14
14
14
14
13
0.02
0.02
0.16
1.13
2.85
0.00
0.15
1.05
2.88
0.02
0.17
1.06
3.44
0.10
0.04
0.03
0.10
0.26
0.02
0.03
0.10
0.29
0.06
0.05
0.09
0.40
0.10
0.06
0.06
0.18
0.26
0.02
0.07
0.12
0.29
0.07
0.07
0.17
0.66
0.28
0.11
0.08
0.28
0.73
0.06
0.08
0.28
0.81
0.11
0.14
0.25
1.12
0.28
0.17
0.17
0.50
0.73
0.06
0.20
0.34
0.81
0.20
0.20
0.48
1.85
19.1
9.09
9.08
16.6
9.53
9.95
27.7
8.22
11.7
39.6
15.6
9.08
43.3
11.5
9.95
39.0
16.2
19.1
107
113
95
100
105
96
113
106
114
0.00
0.60
2.00
4.00
14
14
14
14
0.24
0.93
2.19
3.48
0.07
0.09
0.20
0.29
0.09
0.17
0.30
0.32
0.20
0.25
0.56
0.81
0.25
0.48
0.84
0.90
29.9
9.27
9.13
8.29
35.8
17.9
13.8
9.28
115
98
81
0.00
0.80
2.00
4.00
14
13
14
14
0.47
1.28
3.15
4.03
0.09
0.13
0.24
0.38
0.14
0.25
0.58
0.50
0.25
0.36
0.67
1.06
0.39
0.70
1.62
1.40
20.1
10.0
7.74
9.52
30.8
19.5
18.4
12.5
101
134
89
N = Number of laboratories.
Mean recovery.
r = 2.8 sr.
R = 2.8 sR.
= not calculated.
IC50a, g/L
0.72
Reactivityb, %
100
0.39
0.64
0.74
2.22
4.90
5.50
15.5
>10000
185
113
97.0
32.4
14.7
13.1
4.6
<0.1
3.21
217
148
>10000
22.4
0.3
0.5
<0.1
Tube content
Diluent/zero standard solution
Atrazine standard solution, 0.1 g/L
Atrazine standard solution, 1.0 g/L
Atrazine standard solution, 5.0 g/L
QC solution
Test sample 1
Test sample 2
Test sample 3
ID No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
Sample
Spike,
g/L
Laboratory No.
Sample
code
10
11
12
13
14
5
10
29
33
19
25
34
37
6
11
21
18
30
35
4
3
20
26
36
38
12
14
22
17
31
39
2
13
24
27
23
15
1
8
32
41
7
0.17
0.15
1.01
0.90
2.91
2.72
0.09
0.09
0.97
0.89
2.76
2.80
0.13
0.13
1.11
0.99
3.73
3.70
0.20
0.17
0.87
0.93
2.03
2.09
3.61
3.78
0.34
0.38
1.08
0.11
0.11
1.31
0.99
2.76
2.01
0.16
0.16
0.96
0.90
2.57
2.36
0.07
0.08
0.96
0.80
2.80
2.70
0.18
0.12
0.78
0.74
1.60
1.64
3.27
2.66
0.50
0.40
0.95
0.18
0.17
1.25
1.29
3.01
3.12
0.20
0.21
1.04
1.04
2.83
3.22
0.22
0.26
1.01
1.06
2.82
2.74
0.26
0.25
0.98
0.91
2.43
2.34
3.30
3.99
0.53
0.57
1.44
0.18
0.10
0.07
b
1.32b
b
2.67b
3.06
2.57
b
0.47b
b
0.28b
1.44
1.14
2.41
2.88
0.32
0.26
0.77
0.79
3.22
3.42
0.50
0.25
0.87
0.92
1.86
2.16
3.51
3.59
0.65
0.53
1.47
0.22
0.09
0.13
0.17
1.28
1.53
2.85
2.70
0.11
0.25
0.34
1.04
1.15
2.39
3.12
0.36
0.07
0.26
0.09
1.04
1.08
3.82
2.77
0.27
0.45
0.96
1.29
2.45
2.53
3.68
3.77
0.54
0.89
1.96
0.05
0.94
0.81
3.03
2.52
0.07
0.06
1.11
1.03
2.74
2.90
0.12
0.11
0.80
0.71
3.16b
5.11b
0.29
0.11
0.63
0.49
1.93
1.76
3.19
3.97
0.43
0.24
1.14
0.08
0.21
0.27
1.29
1.30
2.67
2.70
0.25
0.22
0.97
0.96
2.68
2.92
0.07
0.19
0.17
1.05
1.28
3.38
4.89
0.26
0.30
0.99
1.12
2.46
2.32
3.64
3.61
0.54
0.72
2.75b
0.20
0.11
0.82
0.94
2.76
2.85
0.09
0.12
1.00
0.92
3.24
2.66
0.13
0.13
1.09
1.14
4.03
4.36
0.17
0.22
0.77
0.77
1.72
2.03
3.26
3.84
0.30
0.36
1.21
0.54
0.18
0.24
1.23
1.19
2.85
2.93
0.19
0.19
1.09
0.99
3.02
3.12
0.18
0.20
1.08
1.23
3.71
3.58
0.21
0.22
1.11
1.06
2.43
2.31
3.41
3.50
0.53
0.52
1.11
0.19
0.24
1.03
1.18
2.95
2.81
0.13
0.18
1.12
1.06
3.39
2.90
0.22
0.18
1.22
1.15
2.63
2.76
0.29
0.35
0.93
1.03
2.35
2.35
3.51
4.05
0.42
0.35
1.31
0.22
0.23
1.12
1.02
3.07
2.87
0.18
0.15
1.08
1.02
2.64
3.19
0.19
0.21
1.19
1.15
4.27
4.90
0.25
0.27
1.00
0.92
2.33
2.33
3.45
3.52
0.51
0.53
1.34
0.18
0.15
1.19
1.32
2.89
2.84
0.14
0.13
0.91
1.02
3.23
2.94
0.12
0.10
1.17
1.38
3.01
3.29
0.14
0.19
0.97
1.07
2.43
1.96
3.34
3.41
0.44
0.41
1.38
0.10
0.14
1.09
1.15
2.90
3.08
0.12
0.06
1.01
1.10
2.82
3.36
0.26
0.11
0.98
1.00
3.05
3.29
0.24
0.22
0.85
0.89
2.85
2.06
2.93
3.10
0.25
0.31
1.07
0.19
0.20
1.15
1.05
3.52
2.73
0.12
0.13
0.99
1.33
2.78
2.79
0.12
0.15
1.21
1.29
3.38
3.08
0.17
0.23
1.16
1.01
2.37
2.28
3.05
3.44
0.36
0.47
1.09
Table 1. (continued)
ID No.
38
39
40
41
42
a
b
Sample
Duplicate
Incurred high plus 5x
Duplicate
Incurred high plus 10x
Duplicate
, not detected
Excluded as outlier pair.
Laboratory No.
Spike,
g/L
Sample
code
10
11
12
13
14
IH plus 0.80
IH plus 2.00
IH plus 2.00
IH plus 4.00
IH plus 4.00
9
40
42
16
28
1.23
2.54
3.26
3.55
3.57
0.88
3.98
3.62
3.15
3.61
1.51
2.73
2.70
3.83
3.63
1.31
4.56
4.06
4.20
4.07
1.77
2.54
2.64
3.55
3.96
1.29
3.41
3.84
4.37
3.69
1.64b
4.09
3.64
3.62
3.60
1.39
3.03
2.77
4.98
3.87
0.85
3.13
3.09
4.70
4.55
1.34
2.39
2.44
4.26
3.90
1.30
3.18
3.23
4.60
4.45
1.28
2.89
3.17
4.15
4.80
1.29
3.23
2.91
4.66
4.02
1.42
2.43
2.77
3.91
3.40
Run 1, g/L
Run 2, g/L
Run 3, g/L
2.94
2.66
3.26
3.19
2.76
3.09
3.19
2.90
2.86
3.30
3.03
3.41
3.18
3.10
2.98
2.89
3.46
3.06
3.04
2.99
3.35
3.01
3.41
3.08
3.30
3.30
2.85
3.40
3.04
2.53
3.22
3.52
3.12
2.64
3.56
3.37
3.54
3.08
3.55
3.01
3.00
Atrazine spike or
incurred load, g/L
Sample matrix
RSDr
RSDR
0.15
0.15
0.15
0.24
Spiked municipal
Spiked well
Spiked surface
Naturally incurred load
19.1
16.6
27.7
29.9
23.3
39.6
43.4
39.0
35.8
39.4
0.47
0.60
0.80
1.00
1.00
1.00
2.00
2.00
3.00
3.00
3.00
4.00
4.00
20.1
9.27
10.0
9.09
9.53
8.22
9.13
7.74
9.08
9.95
11.7
8.29
9.52
10.1
30.8
17.9
19.5
15.6
11.5
16.2
13.8
18.4
9.08
9.95
19.1
9.28
12.5
15.7
Average
High in detection range
(more than 5 MDL)
Average