Академический Документы
Профессиональный Документы
Культура Документы
Biological and Chemical Research, Volume 2015, 111-122 | Science Signpost Publishing
Received: January 08, 2014 / Accepted: February 06, 2015 / Published: March 25, 2015
Abstract: Investigation of total petroleum hydrocarbons (THC) and polycyclic aromatic hydrocarbons (PAHs) in sea sediment
samples that were collected from the areas of the Caspian Sea. Samples are extracted with methylene chloride and methanol, after the
extracts are cleaned on silica-gel columns and then injected into GC/FID for determination of THC and GC/MS operating in the
selected-ion-monitoring (SIM) mode for determination of the PAHs and their alkylated.
Keywords: THC, PAHs, GC/MS, SIM, RRF, Sediment
1. Introduction
Most methods for measuring the semivolatile compounds of petroleum in environmental samples first extract
them with an organic solvent, fractionate with chromatography, and then analyze the fractions by infrared (IR)
spectroscopy, ultraviolet fluorescence (UVF) spectroscopy, or gas chromatography (GC) [1]. Each of these
methods targets a particular characteristic of the extracted material and gives an operationally-defined
concentration of total hydrocarbons (THC). Although IR and UVF provide some useful information and a
measure of THC values, they lack the specificity and sensitivity of GC methods.
Polycyclic aromatic hydrocarbons (PAHs) are recognized as potent carcinogens and numerous studies have
shown that they are ubiquitous contaminants in a wide variety of matrices such as air, food, fly ash, soil,
sediments, water, crude oil and petrochemicals [25].
There are a wide variety of solvent extraction techniques commonly used for extracting hydrocarbons from
soils and sediments. Traditional extraction procedures include Soxhlet [69] ultrasonication [7,10,11], mechanical
shaking [12,13]. Modern techniques include supercritical fluid extraction, pressur-ized liquid extraction,
accelerated solvent extraction, microwave assisted extraction and focused microwave assisted. Each technique has
its own merits and the choice of extraction depends on several factors including capital cost, operating cost,
simplicity of operation, amount of organic solvent required, sample throughput and the availability of a
standardized method.
In this work used ultrasonic procedure for extracting hydrocarbons and PAHs in sea sediments.
Polynuclear aromatic hydrocarbons (PAHs) are important environmental pollutants because many of them are
known or suspect carcinogens. In environmental studies, the extent of PAH contamination as a group is often
Corresponding author:
Navai A. IBADOV, Institute of Radiation Problems of Azerbaijan NAS, 121 H.Javid aven., AZ1143,Baku, Azerbaijan. E-mail:
navai@azecolab.com.
112
quantified by the concentrations of 16 of the representative PAH species, which are included in the list of priority
pollutants as defined by USEPA. Many analytical techniques have been developed in the past years for the
determination of these 16 PAH species, and among them two of the most widely practiced ones are
HPLC-Fluorescence [14] and GC/MS [15]. HPLC with fluorescence detection (HPLC-FLD) cannot be used for
the detailed analysis of individual alkylated PAHs. Gas chromatography with mass spectrometry (GC-MS) is
presently the preferred analytical technique for the analysis of both parent and alkylated PAHs [17]. As is often
the case in environmental analysis, effective sample preparation and cleanup holds the key of success for the
analysis of trace PAHs.
Aromatic ring structures in petroleum products range from one- to five-ring combinations. Two or more fiveor
six-member carbon rings are fused together to form polyaromatic hydrocarbons (PAHs). These petroleum PAHs
have abundant alkyl group substitution on their ring structures. The alkyl groups generally have one to four
saturated carbon atoms, and thus can produce many different structural isomers and homologs for each aromatic
hydrocarbon family. The most abundant aromatic hydrocarbon families have two or three fused rings with one to
four carbon atom alkyl group substitutions (denoted C1-, C2-, C3-, and C4- by GC/MS/SIM expanded scans). It is
important to point out that crude oils contain primarily the alkyl homologs of aromatic compounds and relatively
small quantities of the unsubstituted "parent" aromatic structures. C1-, C2-, C3-, or C4- followed by a PAH name
(for example, C1-naphthalene) is a naming convention for reporting the total of all detected C1-, C2-, C3-, or
C4-alkyl homologs of the noted PAH. For example, C1-naphthalene reported concentrations represent the total
concentration of all C1 naphthalenes. C1-compounds differ from C2-compounds in that there is one rather than
two carbon groups attached. Groups of alkyl homologs are often analyzed by a GC/MS/SIM expanded scan for
polyaromatic hydrocarbons (PAHs) and alkyl PAHs.
A gas chromatograph/mass spectrometer (GC/MS) in selected ion mode (SIM) coupled to a capillary column is
used to resolve, detect and quantitate polycyclic aromatic hydrocarbons (PAH) in solids at parts per billion levels.
Samples are injected into a temperature-programmed GC/MS, operated in splitless mode. The capillary column is
a DB-5MS (30 m x 0.25 mm ID and 0.25 mm film thickness). The mass spectrometer is capable of scanning from
50 to 500 AMU every second or less and uses 70 electron volts energy in electron impact ionization mode. The
data acquisition system continuously acquires and stores all data analyses [15-18].
113
Calibration Solution
For GC/FID calibration mix used C7-C40 Saturated Alkanes (Sigma Aldrich, Cat#49452-U). Calibrations
solutions (16EPA PAHs and their 34 individual Alkyl PAHs components, AccuStandard, Inc. or Sigma-Aldrich))
are prepared at 7 concentrations ranging from approximately 0.01 to 1 g/mL, by diluting commercially available
certified solutions containing analytes of interest.
Initial Calibration
Prepare calibration standards at a minimum of seven concentration levels for each parameter (Naphthalene,
2-Methylnaphthalene, 1,3-Dimethylnaphthalene, Acenaphthylene, Acenaphthene, Fluorene, Dibenzothiophene,
Phenanthrene, Anthracene, 4-Methyldibenzothiophene, 1-Methylphenanthrene, 3,6-Dimethylphenanthrene,
Fluoranthene,
Pyrene,
2-Methylfluoranthene,
1-Methylpyrene,
Benz[a]Anthracene,
Chrysene,
9-Methylbenz[a]Anthracene, 6,8-Dimethylbenz[a]Anthracene, Benzo[b]fluoranthene, Benzo[k]fluoranthene,
Benzo[a]Pyrene, Perylene, 9-Methylbenzo[a]pyrene, 7,10-Dimethylbenzo[a]pyrene, Indeno(1,2,3-cd)pyrene,
Benz[g,h,i]perylene, Dibenz[a,h]Anthracene) of interest by adding volumes of one or more stock standards to a
volumetric flask and diluting to volume with dichloromethane. The following calibration levels are used: 2, 5, 10,
25, 50, 100 and 200 mg/L for the individual hydrocarbons and 0.010, 0.020, 0.050, 0.100, 0.200, 0.500 and 1.000
g/L for the PAH components. A 7-point relative response factor (RRF) calibration curve is established for
analytes of interest prior to the analysis of samples and quality control samples.
A RRF is determined, for each analyte, for each calibration level using the following equation 1:
RRF =
Where:
Astd Amount(IS)
Ais Amount(S)
(1)
114
The response factors determined for each calibration level are averaged to produce a mean relative response
factor (RRF) for each analyte. The percent relative standard deviation (%RSD) for the 7 response factors must be
less than or equal to 25%, for each analyte.
GC/FID ANALYSIS
The hydrocarbons are analyzed on a gas chromatograph with a flame ionization detector (GC-FID), and the
THCs are usually estimated by integrating the areas of the resolved and unresolved components [19].
Chromatographic separation of THC was accomplished on a CP-Sil 5CB (100% dimethylpolysiloxane) capillary
column (15 m0.25 mm I.D., 0.25 mm film thickness). Nitrogen was the carrier gas and a pressure 15 psi
(constant) was used for column elution. Sample injection was carried out in the splite/splitless mode with an
injection volume of 1 L. Split ON at 0.75min, Split Ratio-20. Autosampler CP8410 (Varian). The GC oven
temperature was programmed first from 40C (hold 2 min), 40C to 320C at a rate of 10C/min (hold 2.0 min).
Fid temperature were 340C (Make-up- 30 ml/min Hydrogen-30 ml/min, Air 300 ml/min ). The temperatures of
the injection port were set at 300C. Galaxie software acquisition. Capable of continuous acquisition and storage
of all data during analysis.
GC/MS-SIM ANALYSIS
The initial calibration of the GC/MS must meet the previously described criteria prior to sample analysis.
Samples are analyzed in analytical sets that consist of standards, samples and quality control samples. Quality
control samples are method blanks, laboratory duplicates, blank spikes, matrix spikes and standard reference
materials. The type and number of quality control samples depend upon client requests and material availability.
Thermo-Electron GC/MS Trace DSQ System equipped with a CombiPal Auto Sampler was used for analysis.
Chromatographic separation of 2-6 ring PAHs was accomplished on a DB-5MS capillary column (30 m0.32 mm
I.D., 0.25 mm film thickness). Helium was the carrier gas and a flow-rate of 1.2 mL/min was used for column
elution. Sample injection was carried out in the splitless mode with an injection volume of 1 L. The GC oven
temperature was programmed first from 40C (hold 1 min), 40C to 120C at a rate of 15C/min, then to 256C at
a rate of 6/min (hold 5.0 min), and then to 300C at a rate of 6/min finally held constant for 5 min. The
temperatures of the injection port and the interface to the MS system were set at 300C and 300C, respectively.
Peak quantification was carried out in Selected Ion Monitoring (SIM) mode. MS Ion source temperature at 250C.
Software - ThermoElectron-Finnigan Technologies Xcalibur 1.4SR1, capable of continuous acquisition and
115
storage of all data during analysis. For identifications of components used 4 segments (in the each segments
present 12 ions), for each segments scan time equal to 1.39s and dwell time 0.1 sec per ion.
Analyte
Quantitation
Ion, m/z
Reference to
Internal Standard
RRF
and Surrogate
Naphthalene
128
Naphthalene-d8
1.22
C1-Naphthalenes
142
Acenaphthene-d10
1.07
C2-Naphthalenes
156
Acenaphthene-d10
1.04
C3-Naphthalenes
170
Acenaphthene-d10
1.04
C4-Naphthalenes
184
Acenaphthene-d10
1.04
Acenaphthylene
152
Acenaphthene-d10
1.50
Acenaphthene
153
Acenaphthene-d10
1.25
Fluorene
166
Acenaphthene-d10
1.31
Dibenzothiophene
184
Phenanthrene-d10
1.18
C1-Dibenzothiophenes
198
Phenanthrene-d10
0.86
C2-Dibenzothiophenes
212
Phenanthrene-d10
0.86
C3-Dibenzothiophenes
226
Phenanthrene-d10
0.86
Phenanthrene
178
Phenanthrene-d10
1.21
Anthracene
178
Phenanthrene-d10
1.01
C1-178 PAHs
192
Phenanthrene-d10
1.07
C2-178 PAHs
206
Phenanthrene-d10
0.95
C3-178 PAHs
220
Phenanthrene-d10
0.95
116
Fluoranthene
202
Pyrene-d10
1.38
Pyrene
202
Pyrene-d10
1.31
C1-202 PAHs
216
Pyrene-d10
1.14
C2-202 PAHs
230
Pyrene-d10
1.14
C3-202 PAHs
244
Pyrene-d10
1.14
Benz[a]anthracene
228
Chrysene-d12
0.86
Chrysene
228
Chrysene-d12
1.19
C1-228 PAHs
242
Chrysene-d12
0.68
C2-228 PAHs
256
Chrysene-d12
0.53
Benzo[b]fluoranthene
252
Perylene-d12
1.78
Benzo[k]fluoranthene
252
Perylene-d12
1.88
Benzo[a]Pyrene
252
Perylene-d12
1.13
Perylene
252
Perylene-d12
1.08
C1-252 PAHs
266
Perylene-d12
0.51
C2-252 PAHs
280
Perylene-d12
0.39
Indeno(1,2,3-cd)pyrene
276
Perylene-d12
1.00
Benz[g,h,i]perylene
276
Perylene-d12
1.32
C1-276 PAHs
290
Perylene-d12
1.00
C2-276 PAHs
304
Perylene-d12
1.00
Dibenz[a,h]Anthracene
278
Perylene-d12
1.07
Data not meeting the criteria established in this section are appropriately qualified or re-analyzed.
Quantitation Calculations
Sample analyte concentrations are calculated based on the concentration and response of the internal standard
compounds. The equations 1 used to calculate the RRF of each analyte relative to the concentration and area of
the internal standard in the initial calibration. Response factors for same alkyl homologues are presumed equal to
the average response factor of the respective unsubstituted (parent) compound.
The concentration of TPH or PAHs components in the original wet sample, Cw is calculated from the following
equation 2:
Cw =
As Amount(IS)
Ais RF Msamp
(2)
Where:
Cw - Concentration of compound of interest in the wet sample, ng/g for PAHs compounds;
As peak area of the compound of interest;
Ais peak area of the corresponding internal standard;
Amount(IS) the amount of internal standard added to the sample, blank, calibration satandard solution and QC samples
in 1 ml of extract,ng;
RF the response factor; and
Msamp mass of wet sample taken for extraction, g.
117
To convert the result to a dry matter basis, the following equation is used:
Cd = Cw
100
DM
(3)
Where:
Cd concentration of compound of interest in the dry sample (ng/gdm for PAHs compounds);
Cw - concentration of compound of interest in the wet sample (ng/g for PAHs compounds); and
DM the percent dried weight of the sample.
Quality Control
The initial calibration must pass established criteria before sample analysis can begin. All continuing
calibration checks must pass established criteria for analysis to continue. An acceptable method blank analysis
may not contain more than two target analytes at concentrations three times greater than the MDL. This criterion
does not apply if the analytes detected in the method blank are not detected in the associated samples or if the
sample analyte concentrations are 10 times greater than the blank analyte concentrations. If the method blank
exceeds these criteria then the analytical procedure is not in control. The source of the contamination must be
investigated, and corrective measures taken and documented before further sample analysis occurs.
All samples and quality control samples are spiked with deuterated PAH standards compounds prior to
extraction. The deuterated compounds evaluate sample matrix effects and analytical efficiencies associated with
sample preparation and analysis. The recovery of deuterated surrogate compounds is monitored in each sample
and quality control sample. The laboratory will take corrective action if the average surrogate recovery, with the
exception of perylene-d12, is less than 40% or greater than 140%. The following corrective action will be taken if
the above criteria are not met:
Calibration checks solution. If relative percent difference (RPD) of any analyte within a calibration check
standard varies from the predicted response by more than 25%, a new calibration curve must be prepared
for that analyte -per 10 samples
Laboratory Blank -per 20 samples. The criteria less than MDL (for TPH equal to 1 g/gdm and for PAHs
equal to 0.5 ng/gdm, RSD less than 20% for each components) or less than 30 % of minimal values of
sample.
Quality Control Samples (Certified Reference Material (CRM), Laboratory Fortified Blank (LFB) or
Matrix Spike (MS), depend from lab QC program)- at least per 20 samples. The recovery criteria equal to
40%-140 %.
Sample duplicates (laboratory or field, depend from project)- at least per 20 samples. The % RPD of duplicate
samples must not exceed 50%.
Sampling
Composite sediment samples (including the organic top layer) were taken between 10 july and 20 september
2013 from Apsheron Peninsula in Caspian Sea, South-East of the Baku of the Azerbaijan areas for 50 points (see
fig.1), collected in aluminum pans and transported to the laboratory under cool conditions. Before analysis, the
samples were air homogenized in ambient temperature. No losses of analyts should occur under these conditions
and potential contamination from laboratory air or other samples is considered insignificant.
118
Checking of lab glassware and materials preparatory to check status each type lab glassware and material
before starting of samples processing according of below chain: Solvent cleaning, Detergent washing,
Acetone treatment, DCM treatment, Drying in oven, Annealing in oven at 400 deg C, Aluminium foiling
and send to storage.
Sodium sulphate is treated as possible before usage and in amount required for batch samples as below:
Bake in a muffle furnace at 400 C for 2 hours. Allow to cool before using. Store sodium sulphate in a large
mouth bottle with cap Glass wool treatment procedure.
Glass wool: Bake in a muffle furnace at 400 C for 2 hours. Store in an oven at 100 C.
Silica Gel treatment procedure: Silica Gel is treated as possible before usage and in amount required for
batch samples as below: 400 deg C, 5 hours. Activating at 200 deg C 1 hours before use and Cooling before
use.
Sample Extraction
Weigh 40g of each sample in Erlenmeyer flasks. The sample was spiked of each of the individual internal
standards and the remaining sediment was added to the Erlenmeyer flask. Add 60 ml DCM and 50 ml methanol
into Erlenmeyer flask and stopper. Now samples are ready for ultrasonic extraction (US) for 30min. After
extraction remove the flask from the US bath and allow the sample to settle. Carefully decant the supernatant
solvent through a glass wool into a 1 liter separating funnel containing 100 ml of distillated water and stopper.
Shake the separating funnel for 1.0 0.1 minute, and allow the layers to separate. Run the lower dichloromethane
layer into a 500 ml round bottomed flask.
Add 50 ml of dichloromethane to the sediment and ultrasonicate for a further 15 minutes. Remove the flask
from the bath and allow the sample to settle. Carefully decant the supernatant solvent through the filter paper into
119
the separating funnel containing the residual aqueous methanol. Stopper and shake the separating funnel for 1.0
minute. Allow the layers to separate and combine the dichloromethane extracts by running the lower
dichloromethane layer into the 500 ml round bottomed flask and this process repeats once again and combine the
dichloromethane extracts. Add sufficient (typically 1g) sodium sulfate to the combined dichloromethane extracts
and dry the solvent. When dry, the solvent is filtered through a second (clean) filter paper .Approximately 5-10 ml
of dichloromethane are added to the flask and the flask rinsed. The solvent is then filtered and the extracts
combined.
Prepare the chromatographic column by plugging cotton wool into a long bodied Pasteur pipette. Prepare
slurry by adding 20ml DCM to 5gm silica gel. Swirl the contents and add the slurry into the column. Wait
until silica gel will sink. Add 1-2sm sodium sulfate. Open the column tap and discard the solvent. Do not
allow the meniscus of the solvent to fall below the surface of the column.
Transfer the sample extract from volumetric tube to the column. Place a round bottomed 500ml flask
under the column and pour the sample extract. Elute the column 20ml of a 2:1 pentane and DCM mixture.
Rinse the sample extract container with 2-3ml o a mixture and transfer the washings to the column. Elute
the column. Add a further 20ml mixture to the column and continue eluting.
Concentrate extracts by Kuderna Danish apparatus at 50-55C till 2 ml and transfer extract into
volumetric tube.
Rinse flask with a little portion of solvent and add the washings to extract to complete the quantitative
transfer.
After place the volumetric tube in nitrogen blow-down apparatus. Evaporate the extract till 1ml.
The solution is now ready for GC or GC/MS determinations.
120
chromatograms. Accuracy of the applied method was regularly verified by participation in the WEPALs quarterly
SETOC ring tests.
THC
UCM
2-6 PAHs
NPD
(g/g)
(g/g)
UCM
(ng/g)
(ng/g)
ACGR-01
2.4
1.2
52
88
57
64
17
ACGR-A
10.4
8.2
78
111
65
56
29
ACGR-04
3.4
56
158
95
61
40
ACGR-B
84
63
75
454
269
59
113
ACGR-22
221
160
72
524
257
49
132
ACGR-46
1.9
1.3
62
42
26
60
ACGR-62
18
13
75
62
35
57
12
ACGR-W6
6.8
4.8
63
93
62
64
17
ACGR-C
15
12
77
54
33
64
11
ACGR-26
2.9
1.3
45
91
58
64
17
ACGR-33
268
197
74
770
427
55
153
ACGR-D
23
18
77
177
102
58
33
Min
1.9
1.2
45
42
26
49
Max
268
197
78
770
427
64
153
Med
13
10
73
102
64
60
23
Mean
55
40
67
219
124
59
49
Points
%NPD
16 EPA
(ng/g)
There are a few different types of correlations, but the one we will use is the Pearson correlation. A correlation,
r, is a single number that represents the degree of relationship between two measures. The correlation coefficient
is a value such that between from -1 to 1.
A positive correlation indicates a relationship between x and y measures such that as the values of x increases,
the values of y will also increase. A negative correlation indicates the opposite as the values of x increase, the
values of y will decrease. The closer the correlation, r, is to -1 or 1, the stronger x and y are related. The negative
sign does not indicate anything about strength - it just signifies that correlation is negative. If r is close or equal to
0, it means there is a weak or no relationship between the measures.
The relationship between hydrocarbon concentrations was tested by a Pearsons r correlation analysis of
replicate values. Pearson correlation coefficient, also known as Pearson's r, a measure of the strength and direction
of the linear relationship between two variables that is defined as the (sample) covariance of the variables divided
by the product of their (sample) standard deviations.
High correlation coefficients have been given between THC & UCM and between all PAH parameters given in
Table 2:
121
NPD
Total EPA
THC
UCM
(g/g)
(g/g)
THC (g/g)
1.00
UCM (g/g)
1.00
1.00
% UCM
0.35
0.36
1.00
0.95
0.96
0.33
1.00
NPD (ng/g)
0.93
0.93
0.32
1.00
1.00
% NPD
-0.73
-0.73
-0.56
-0.64
-0.59
1.00
0.94
0.94
0.34
0.99
0.98
-0.69
1.00
% UCM
ring PAH
(ng/g)
(ng/g)
% NPD
16PAHs
(ng/g)
4. Conclusion
THC concentration in the 50 points contained from 1.9 mg/kgdw -268 mg/kgdw and UCM present in samples
from 45%-78%. Total 2-6 rings PAHs were present in the sediments observation sites in concentrations from 42
g/kgdw to 770 g/kgdw, 16EPA PAHs concentrations from 9.0 g/kgdw to 153 g /kgdw and NPD from 49 %
to 64% (at 2-6 Total PAHs rings).
References
[1]. NRC (1985) Oil in the Sea: Inputs, Fates, and Effects. National Research Council, National Academy Press, Washington, DC.
[2]. Neff J.M., Polycyclic Aromatic Hydrocarbons in the Aquatic Environment, Applied Science Publishers, London, 1979.
[3]. Bjorseth A. (Ed.), Handbook of Polycyclic Aromatic Hydrocarbons, 1983, p. 727.
[4]. Blumer M.W., Sci. Am. 234 (1976) 34.
[5]. Blumer M., Youngblood W.W., Science 188 (1975) 53.
[6]. Luque de Castro M.D., L.E. Garca-Ayuso, Anal. Chim. Acta. 369 (1998) 1.
[7]. J.O. Grimalt, C. Marfil, Albaiges J., Int. J. Environ. Anal. Chem. 18 (1984) 183.
[8]. Lee H.-B., Dookhran G., Chau A.S.Y., Analyst 112 (1987) 31.
[9]. Wong M.K., Williams J.P.B., Mar. Chem. 9 (1980) 183.
[10]. Luque-Garca J.L., Luque de Castro M.D., Trends Anal. Chem. 22 (2003) 41.
[11]. Krahn M.M., Ylitalo G.M., Joss J., Chan S.-L., Mar. Environ. Res. 31 (1991) 175.
[12]. Berset J.D., Ejem M., Holzer R., Lischer P., Anal. Chim. Acta 383 (1999) 263.
[13]. MacLeod W.D., Brown D.W., Friedman A.J., Burrows D.G., Maynes O., Pearce R.W., Wigren C.A., Bogar R.G., Standard
Analytical Procedures of the National Oceanic Atmospheric Administration (NOAA), National Analytical Facility, 19851986:
Extractable Toxic Organic Compounds, NOAA Tech. Memo. NMFS F/NWC-92, 1985, p. 121.
[14]. Ibadov N.A., Huseynov V.I., Suleymanov B.A. (2004) Determination of Polynuclear Aromatic Hydrocarbons By High
Performance Liquid Chromatography, Journal of the Chemical Problems, 2, pp.40-48.
[15]. Reddy Christopher M. And Quinn James G. (1999), Marine Pollution Bulletin, Vol. 38, No. 2, pp. 126-135, 1999.
122
[16]. Hawthorne, S. B., Grabanski, C. B., and Miller, D. J., Measured Partitioning Coefficients for Parent and Alkyl Polycyclic
Aromatic Hydrocarbons in 114 Historically Contaminated Sediments: Part I, Koc Values, Environmental Toxicology and
Chemistry, 25, 2006, pp. 2901-2911.
[17]. Thomas J. McDonald, Bo Wang, Susanne J. McDonald and James M. Brooks. Quantitative Determination Of Aromatic
Hydrocarbons Using Selected Ion Monitoring Gas Chromatography/Mass Spectrometry TDI-Brooks International./B&B
Laboratories Inc. College Station, Texas 77845.
[18]. UK- A guide to practices, procedures and methodologies following oil spill contamination incidents (2004).
[19]. Douglas, G. S., McCarthy, K. J., Dahlen, D. T., Seavey, J. A., Steinhauer, W.G., Prince, R. C. and Elmendorf, D. L. (1992).
The use of hydrocarbon analyses for environmental assessment and remediation. Journal of Soil Contamination 1, 197-216.