Cyclic Voltammetry
Cyclic Voltammetry (CV) is an electrochemical technique which measures the current that develops in an electrochemical cell under conditions where voltage is
in excess of that predicted by the Nernst equation. CV is performed by cycling the potential of a working electrode, and measuring the resulting current.
Introduction
The potential of the working electrode is measured against a reference electrode which maintains a constant potential, and the resulting applied potential
produces an excitation signal such as that of figure 1. In the forward scan of figure 1, the potential first scans negatively, starting from a greater potential (a) and
ending at a lower potential (d). The potential extrema (d) is call the switching potential, and is the point where the voltage is sufficient enough to have caused an
oxidation or reduction of an analyte. The reverse scan occurs from (d) to (g), and is where the potential scans positively. Figure 1 shows a typical reduction
occurring from (a) to (d) and an oxidation occurring from (d) to (g). It is important to note that some analytes undergo oxidation first, in which case the potential
would first scan positively. This cycle can be repeated, and the scan rate can be varied. The slope of the excitation signal gives the scan rate used.
24-05-2016 09:37
2 of 7
http://chemwiki.ucdavis.edu/Core/Analytical_Chemistry/Instrumental_Analysis/Cyclic_Voltammetry
M + + e M
24-05-2016 09:37
3 of 7
http://chemwiki.ucdavis.edu/Core/Analytical_Chemistry/Instrumental_Analysis/Cyclic_Voltammetry
E = Ei + vt
(1)
where
4 of 7
http://chemwiki.ucdavis.edu/Core/Analytical_Chemistry/Instrumental_Analysis/Cyclic_Voltammetry
When the direction of the potential sweep is switched, the equation becomes,
E = Es vt
(2)
Ep Ep/2 >
0.0565
n
(3)
where
Epa + Epc
2
.
cA
2 c A
= DA 2
t
x
(4)
During a reduction, current increases until it reaches a peak: when all M+ exposed to the surface of an electrode has been reduced to M. At this point additional
M+ to be reduced can travel by diffusion alone to the surface of the electrode, and as the concentration of M increases, the distance M+ has to travel also
increases. During this process the current which has peaked, begins to decline as smaller and smaller amounts of M+ approach the electrode. It is not practical to
obtain limiting currents Ipa, and Ipc in a system in which the electrode has not been stirred because the currents continually decrease with time.
In a stirred solution, a Nernst diffusion layer ~10-2 cm thick, lies adjacent to the electrode surface. Beyond this region is a laminar flow region, followed by a
turbulent flow region which contains the bulk solution. Because diffusion is limited to the narrow Nernst diffusion region, the reacting analytes cannot diffuse
24-05-2016 09:37
5 of 7
http://chemwiki.ucdavis.edu/Core/Analytical_Chemistry/Instrumental_Analysis/Cyclic_Voltammetry
into the bulk solution, and therefore Nernstian equilibrium is maintained and diffusion-controlled currents can be obtained. In this case, Ficks Law for mass
transfer diffusion can be simplified to give the peak current
1/2
ip = (2.69 x 105 ) n3/2 SD1/2
CA
A V
(5)
Here, (n) is equal to the number of electrons gained in the reduction, (S) is the surface area of the working electrode in cm, (DA) is the diffusion coefficient, (v)
is the sweep rate, and (CA) is the molar concentration of A in the bulk solution.
Instrumentation
A CV system consists of an electrolysis cell, a potentiostat, a current-to-voltage converter, and a data acquisition system. The electrolysis cell consists of a
working electrode, counter electrode, reference electrode, and electrolytic solution. The working electrodes potential is varied linearly with time, while the
reference electrode maintains a constant potential. The counter electrode conducts electricity from the signal source to the working electrode. The purpose of
the electrolytic solution is to provide ions to the electrodes during oxidation and reduction. A potentiostat is an electronic device which uses a dc power source
to produce a potential which can be maintained and accurately determined, while allowing small currents to be drawn into the system without changing the
voltage. The current-to-voltage converter measures the resulting current, and the data acquisition system produces the resulting voltammogram.
Applications
Cyclic Voltammetry can be used to study qualitative information about electrochemical processes under various conditions, such as the presence of
intermediates in oxidation-reduction reactions, the reversibility of a reaction. CV can also be used to determine the electron stoichiometry of a system, the
diffusion coefficient of an analyte, and the formal reduction potential, which can be used as an identification tool. In addition, because concentration is
proportional to current in a reversible, Nernstian system, concentration of an unknown solution can be determined by generating a calibration curve of current
vs. concentration.
References
1. Skoog, D.; Holler, F.; Crouch, S. Principles of Instrumental Analysis 2007
2. Kissinger, P. T., Heineman, W. R., Cyclic Voltammetry, Journal of Chemical Education, 60, 702 (1983).
Outside Links
Carriedo, Gabino A. "The use of cyclic voltammetry in the study of the chemistry of metal-carbonyls: An introductory experiment." J. Chem. Educ. 1988,
65, 1020.
http://www.jce.divched.org/Journal/I...E1983p0697.pdf
http://www.jce.divched.org/Journal/I...E1983p0702.pdf
24-05-2016 09:37
6 of 7
http://chemwiki.ucdavis.edu/Core/Analytical_Chemistry/Instrumental_Analysis/Cyclic_Voltammetry
http://www.jce.divched.org/Journal/I...E1984p0820.pdf
http://www.jce.divched.org/Journal/I...V78N01/p89.pdf
http://www.jce.divched.org/Journal/I...E1983p0702.pdf
Contributors
Amanda Quiroga (UCD)
Powered by MindTouch
Unless otherwise noted, content in the UC Davis ChemWiki is licensed under a Creative Commons
Attribution-Noncommercial-Share Alike 3.0 United States License. Permissions beyond the scope of this license may be available at copyright@ucdavis.edu. Questions and
24-05-2016 09:37
7 of 7
http://chemwiki.ucdavis.edu/Core/Analytical_Chemistry/Instrumental_Analysis/Cyclic_Voltammetry
concerns can be directed toward Prof. Delmar Larsen (dlarsen@ucdavis.edu), Founder and Director.
Terms of Use
24-05-2016 09:37