Вы находитесь на странице: 1из 2

Model 551

Small Parts Phantom

Vertical Measurement Calibration


Horizontal Measurement Calibration
Axial Resolution
Lateral Resolution
Sensitivity
Focal Zone
Functional Resolution
Image Uniformity
Gray Scale
Dead Zone

Product Description

!Model 551 provides a comprehensive means of evaluating the performance of a diagnostic ultrasound system
at frequencies ranging from 7.5 to 15 MHz. It is designed with a combination of monofilament line targets and
tissue mimicking cylindrical targets of varying sizes and contrasts. The monofilament line targets have a
diameter of 0.05 mm, to optimize the displayed image at frequencies of 7.5 MHz or greater. Four groups of
line targets are provided to evaluate the vertical and horizontal calibration measurements, dead zone and
axial-lateral resolution. The Model 551 is constructed of our rubber-base tissue mimicking material to provide
a phantom which is accurate, durable and maintenance free. Lifetime warranty.

!
Webpage: www.atslaboratories-phantoms.com

!
Email: atslaboratories@yahoo.com

Tel:1- 203 579 2700

SPECIFICATIONS
General
Operating Frequency Range
Weight
Scan Surface Dimensions
Line Targets
Diameter
Vertical Group
Number of Targets
Depth
Dead Zone Group
Number of Targets
Interval Spacing
Horizontal Group
Number of Targets
Interval Spacing

7.5 - 15.0 MHz


3.1 Kg*
25 x 8 cm

Overall Dimensions
Scan Surfaces

28 x 11 x 11 cm*
1

10
5 - 50 mm

Interval Spacing

5 mm

5
1.0 mm

Lateral Displacement
Depth

5 mm
1.0 - 5.0 mm

9
5 mm

Number of Groups
Depth

2
Linear Array: 20 & 40 mm
Sector Array: 25 & 45 mm

Lateral Displacement
Depth

NA
20 & 45 mm

.05 mm

Axial-Lateral Resolution
Number of Targets
10
Interval Spacing (edge to edge) 3, 2, 1, .5, .25 mm
Anechoic Target Structures
Targets per group
6
Group 1, 2, 3, 4

Gray Scale Target Structures

Diameter
1, 2, 3, 4 mm
Interval Spacing
10 mm (Center-to-Center)
Diameter
Number of Targets
6 mm
6
Contrast relative to the background material (dB)
+15, +6, +3, -3, -6, -15

Webpage: www.atslaboratories-phantoms.com

Email: atslaboratories@yahoo.com

Depth
10 - 60 mm

Depth
20 mm
Interval Spacing
10 mm (Center-to-Center)

Tel:1- 203 579 2700