EN Installation and configuration instructions

VistaCam Digital
Only trained experts may position and install the appliance and start it
up for the first time.
Minimum system requirements
Imaging program

DBSWIN or VistaEasy by Drr Dental


Third-party software on request

Computer system

See "System requirements for computer systems for


imaging" (9000-618-148/01)

1 Installation

1
2.1 Setting up the configuration in DBSWIN

The USB Y adapter for the connection to the computer is included in the
scope of delivery. If the power supply provided with this adapter is insufficient, use a quick-connect adapter with a power unit (order no. 2106-170-51
(20 cm), 2106-170-52 (100 cm)).

Start DBSWIN.
Choose Options > Show configuration from the menu.
Click the "Modules"

button.

Only connect the camera to the computer when the installation


wizard prompts you to do so.
Close all programs.
Insert the supplied DBSWIN DVD (from Version 5.1) into the DVD drive.
The start screen appears.
If the DVD does not start to play automatically, execute the file "CD_Start.
exe".
Select the required language.
Open the "Driver" tab.

Double-click "Video".
The "Video Properties" window opens.
Select the required configuration.
Setting up image export
Go to the "Settings" tab.


Click "Driver Installation".


Confirm the message with "OK".
The "Drr Dental Driver Setup" installation wizard opens.
Follow the instructions of the installation wizard. When the wizard asks
you to connect or disconnect the camera to/from the computer, do so.

2 Configuration
Activate/deactivate "Autoexport images".
Each image is automatically copied to a defined path. The path, file format, and other settings can be made in the light table module.

The configuration differs depending on the imaging program used:


- DBSWIN: See "2.1 - Setting up the configuration in DBSWIN" (additional
settings can be made in VistaCamConfig)
- VistaEasy and third-party software (e.g. Sidexis): See "2.2 - Setting up the
configuration in VistaCamConfig

1 / 3

Setting the video source

Setting pressure sensitive release

Two cameras can be connected up.

Go to the "Pressure sensitive release" tab.

If you are only connecting up one camera, make the settings on the
"Video source 1" tab.
If you are connecting up two cameras, make the settings for the second
camera on the "Video source 2" tab.
If you are connecting up a VistaCam and a VistaProof, use the
"Video source 2" tab for the VistaProof.
Open the "Video source 1" or "Video source 2" tab.

You can choose from four profiles depending on the required pressure:
- Standard (default setting with default values)
- Low
- Medium
- High
You can modify the values for the low, medium, and high profiles:
Delay time

Time between the pressing of the pressure sensitive


release and the capturing of the image.
The delay time prevents blurred images as a result
of movement when actuating the pressure sensitive
release.

Sensitivity
(light touch)

Minimum pressure required to actuate the pressure


sensitive release

Threshold
(strong touch)

Pressure required to actuate the pressure sensitive


release if "Strong touch" is selected

Select the control method "VistaCam Digital/VistaProof".


In "Use half-frames", select "Both (full frame)".
Setting the WDM driver
Go to the "WDM" tab.
The WDM driver is selected automatically.
Click the "Update camera list"
button.
All connected cameras appear in the list.

Check the settings on the basis of the value and coloured indicator.

Select the relevant camera from the list.


Activate/deactivate noise reduction.
If noise reduction is active, the set number of images are captured one
after the other for each imaging operation. The system uses these images
to generate a new image that eliminates interference to the greatest extent
possible.


To save changed settings, click the "Save profile"

button.

Select the function for "Light touch" and "Strong touch".


Saving the configuration
Click "Apply".
Click "OK".
The window closes automatically.

2 / 3

2.2 Setting up the configuration in VistaCamConfig


Open VistaCamConfig by choosing Start > All Programs > Duerr Dental
> VistaCamConfig.
The camera is automatically detected and activated.

To change the configuration, click

Change the settings.


If necessary, check the settings for the pressure sensitive release on the
basis of the value and coloured indicator.

Changing settings
Go to the "Settings" tab.

To save the configuration, click

To cancel the changes to the configuration, click

Testing the camera and configuration


Go to the "Test" tab.

Display

The best possible values are automatically entered


here during the installation.

WDM driver

The best possible values are automatically entered


here during the installation.

Advanced settings
LED intensity
configuration

If LED intensity configuration is active, you can set


the intensity of the LEDs here.
An optimum value is preset.

Click "Take picture".


VistaEasyView starts up.

Pressure sensitive release

Capture an image with the camera.

Function

Two different functions can be selected depending


on the pressure.
You can check the pressure in the "Display received
signals" pane.

Close VistaEasyView.

Four profiles can be selected depending on the


required pressure:

Click "Close configuration".


The configuration is saved. The window closes.

Profile configuration

To check the image file, click "View image file".


Saving the configuration

- Standard (default setting with default values)


- Low

3 Electrical safety test

- Medium
- High
You can modify the values for the low, medium, and
high profiles.
Delay time

Time between the pressing of the pressure sensitive


release and the capturing of the image.
The delay time prevents blurred images as a result
of movement when actuating the pressure sensitive
release.

Sensitivity
(light touch)

Minimum pressure required to actuate the pressure


sensitive release

Threshold
(strong touch)

Pressure required to actuate the pressure sensitive


release if "Strong touch" is selected

Carry out an electrical safety test in accordance with national and local
regulations (e.g. measure the leakage current for the patient).
Record the results.

SW_Settings
CAM_Activate_
Delay_Time

Time between removal from handpiece holder and


activation of camera - the optimum value is preset.

9000-618-89/01

DRR DENTAL AG
Hpfigheimer Strasse 17
74321 Bietigheim-Bissingen, Germany

2011/02