Service Manual
IMPORTANT
Servicing and Repairs
In order to ensure the full operational life of this
absorber, servicing by a Penlon-trained engineer
should be undertaken periodically.
We recommend that the absorber should be
serviced on the following schedule:
(a)
(b)
(c)
(i)
FOREWORD
This manual has been produced to provide
authorised personnel with information on the
function, routine, performance and maintenance
checks applicable to the A100 Absorber.
(ii)
CONTENTS
Page No.
USER RESPONSIBILITY
1.
2.
PURPOSE
3.
DESCRIPTION
3.1
Canisters
3.2
3.3
3.4
3.5
3.6
Manometer
3.7
3.8
3.9
10
3.10
10
4.
SPECIFICATION
4.1
13
4.2
13
4.2.1
Expiratory Resistance
13
4.2.2
Inspiratory Resistance
13
4.3
14
4.4
15
4.5
15
4.5.1
Canister Capacity
15
4.5.2
Canister Resistance
15
4.6
Non-return valves
16
5.
5.1
5.2
System Connections
19
5.2.1
19
5.2.2
Fresh Gas
19
5.2.3
19
5.2.4
Oxygen Monitor
19
5.3
20
5.4
Optional Accessories
21
5.4.1
Manometer
21
5.4.2
PEEP Valve
21
17
(iii)
CONTENTS
Page No.
6.
PRE-USE CHECKS
6.1
Pre-use Checklist
22
6.2
Leak Test
23
6.3
23
6.4
24
6.5
24
6.6
25
6.7
25
7.
USER MAINTENANCE
7.1
26
7.2
26
7.3
Condensate Drainage
26
7.3.1
Polemount Models
26
7.3.2
27
7.4
Non-return Valves
29
7.5
Manometer
29
7.6
29
7.7
APL Valve
29
8.
STERILISATION
8.1
Bacterial Filters
8.2
Sterilisation Policy
30
8.3
30
8.4
Absorber Assembly
31
8.5
32
9.
SERVICING
33
10.
PARTS LIST
38
11.
APPENDIX
30
41
User Notes
44
(iv)
USER RESPONSIBILITY
Statements in this manual preceded by the
following words are of special significance:-
WARNING
means there is a
possibility of injury to
yourself or others.
CAUTION
NOTE
1.
7.
8.
9.
10.
After
servicing
and
cleaning
procedures, verify positive action of the
bag/ventilator and absorber On/Off
(bypass) selector switches (if fitted)
before the unit is used clinically.
WARNINGS
General Information
1.
2.
3.
4.
12.
The
use
of
patient
Y-pieces
containing non-return valves in
connection with the Penlon Circle
System Absorber is hazardous,
because two sets of non-return valves
may
easily
be
connected
in
opposition, by error.
6.
14.
CAUTIONS
15.
1.
2.
3.
4.
5.
2.
PURPOSE
b)
with an anaesthesia
Ventilator is selected.
ventilator
when
1
13
14
15
2
3
12
11
10
5
8.
Upper canister
2.
9.
3.
10.
4.
11.
5.
Reservoir bag
12.
6.
13.
Manometer (optional)
7.
Lower canister
14.
15.
Manifold block
3.
DESCRIPTION
3.1
Canisters
Further Information
Mounting the absorber section 5.1
3.2
Capacity
Each canister is designed to hold 1.3 kg of loose
absorbent, or to take prepacks.
The maximum fill level is marked on the canister
equivalent to 1500 ml of absorbent.
Seals
Canisters each have a rubber seal attached at
their lower end. The top seal attached to the
metalwork of the frame is identical to the canister
seals.
Further Information
b)
DESCRIPTION
3.3
Adjustable Pressure
Limiting (APL) Valve
3.4
A
Ventilator mode
In Ventilator
mode the
reservoir bag is
closed off from
the breathing
system and the
ventilator
connection port
(A) at the rear of
the manifold
block, is in circuit.
WARNING The APL valve is out of circuit
when the system is in Ventilator mode. The
ventilator must be equipped with a pressure
relief valve.
60
Pressure
(cmH2O)
1.9
0
0.5
1.0
1.5
Bag/Ventilator Switch
(Optional)
2.0
Turns Clockwise
Bag mode
The breathing bag acts as an additional overpressure protection device, preventing pressure
exceeding 60 cmH2O.
WARNING If no ventilator is connected to
the absorber, care must be
taken to ensure that the bag/
ventilator switch is kept in the
Bag position, to avoid gross
loss of gas from the breathing
system and to maintain the
reservoir bag in the system.
The ventilator port must be left
uncapped when no ventilator
is connected.
AGSS connector
On UK specification models, the APL valve is
fitted with 30 mm taper connector.
US specification models have a 19 mm taper
connector.
Further Information
Pre-use check section 6.3
Sterilisation section 8.5
Further Information
Gas flow schematic section 3.10
Pre-use check section 6
DESCRIPTION
3.5
3.7
This facilitates:
WARNING
3.6
b)
Manometer (optional)
NOTE:
a)
Manometer scale:
10 to +100 cmH2O
Manometer accuracy:
5%
(within range
+10 to 80 cmH2O)
CAUTION
Further Information
CO2 monitoring section 3.8
3.8
Further Information
Installation section 5.4
DESCRIPTION
3.9
3.10
1.
2.
3.
4.
Further Information
Gas flow diagrams section 10
Component layout
In all models the gas flow through the canisters is
from bottom to top.
Note that the bag/ventilator connection is between
the absorber and the patient. Bag squeezing or
the use of mechanical ventilation does not result
in the transport of dust toward the patient, but
tends to drive dust back into the absorber.
10
DESCRIPTION
Circle System Absorber Basic Model
Flow
through
canister
Patient
7.
Breathing bag
2.
Manometer (optional)
8.
3.
9.
4.
5.
Absorber canister
6.
link (optional)
Flow
through
canister
Flow
through
bypass
11
DESCRIPTION
Flow
through
canister
Flow
through
bypass
Flow
through
canister
Patient
Manometer (optional)
Expiratory NRV (non-return valve)
Note: not fitted if PEEP valve is used)
APL (adjustable pressure limiting) valve
Absorber canister
Fresh gas inlet
Bag/ventilator control / mechanical link
Breathing bag
Ventilator
Inspiratory NRV (Non-return valve)
On/Off (bypass) control / mechanical link
PEEP valve (optional)
Note: fitted in place of expiratory NRV
Flow
through
bypass
12
4.
SPECIFICATION
4.1
General Dimensions
Overall height
Width
186 mm
Depth
240 mm
Weight (empty)
Mounting system
4.2
25 mm (1 inch) diameter
polemount
(B)
Absorber only.
Both canisters filled to the MAX level with fresh absorbent, and the APL
valve fully open.
4.2.1
Expiratory Resistance
Tested with a flow of 6 L/min of air through the fresh gas inlet and an
induced flow of 60 L/min through the breathing system.
(A)
(B)
(B)
Bacterial Filter:
Bacterial filters may be used in this breathing system provided they do
not raise the resistance values of the whole system to above 0.6 kPa
(6 cmH2O).
13
SPECIFICATION
4.2.2
Inspiratory Resistance
Tested with a flow of 6 L/min of air through the fresh gas inlet
and an induced flow of 60 L/min through the breathing system.
(A)
(B)
(B)
Bacterial Filter:
Bacterial filters may be used in this breathing system provided they do
not raise the resistance values of the whole system to above 0.6 kPa
(6 cmH2O).
4.3
(B)
Absorber only.
Note that the reservoir bag is not fitted and the bag mount blocked.
(A)
(B)
14
SPECIFICATION
4.4
(B)
Absorber only.
The patient connection port is sealed and the APL valve fully closed.
(A)
(B)
4.5
4.5.1
Canister Capacity
When filled to the MAX level mark, each canister holds 1.3 kg (2.87 lb)
of absorbent (1500 ml).
Recommended absorbent Soda lime or barium lime, with a colour
indicator, 4-8 mesh, supplied in bulk.
Alternatively, pre-packs may be used.
Note
i)
ii)
4.5.2
Canister Resistance
The resistance of a freshly filled canister is less than 0.2 kPa (2 cmH2O)
at 60 L/min.
15
SPECIFICATION
4.6
60
Non-return Valves
50
30
Flow
(L/min)
40
20
10
Pressure
(kPa)
0.02
16
0.04
0.06
0.08
0.01
5.
5.1
CAUTION
If the absorber has to be lifted or carried by hand,
always support the weight of the unit under the
base.
Do not lift the absorber by gripping any of the
components attached to the manifold blocks at the
top of the absorber the manometer, APL valve,
breathing circuit connectors, etc.
Polemount bracket
A 25 mm (1 inch) diameter pole clamp is
provided on the rear of the absorber.
Suitable pole mount brackets are provided on the
Penlon Prima anaesthetic machine, as illustrated.
WARNING
17
EXHAUST TO
SCAVENGE
SYSTEM
PRESSURE MONITOR
LINE
VENTILATOR
DRIVE GAS
SENSOR LINE
FOR O2 MONITOR
(AV900 OPTION)
F
G
FRESH
GAS
SUPPLY
E
CIRCLE
SYSTEM
ABSORBER
DRIVE
GAS
PATIENT
Bacterial filter
CGO block
APL valve
Ventilator connection
Spirometer
exchanger
L
Note
1)
To protect the expiratory limb of the breathing circuit, and the spirometer (H), use a
breathing circuit bacterial filter (J), or a heat and moisture exchanger (K) at the
patient tee-piece.
2)
Follow the instructions given in the relevant user manuals for connections to analysers
and monitors.
3) Prima SP Inboard Mounting System
Refer to the Prima SP user instruction manual for installation and connection details.
18
System Connection
B
5.2.1
5.2.2
The fresh gas hose assembly (E) supplied with the unit
has a Penlon connector at the absorber inlet and a 22
mm
Safelock
taper
at
the
other
end.
This should be connected to the common gas
outlet (F) of the anaesthetic machine.
5.2.3
WARNING
5.2.4
Oxygen Monitor
19
Inspiratory
valve
Sensor
WARNING
If the absorbent is to be
changed during clinical use,
the bypass switch must be in
the Absorber OFF position,
and adequate fresh gas
flow must be maintained to
prevent excessive build up
of CO2.
WARNING
WARNING
2.
(a)
(b)
1.
(a)
(b)
2.
CAUTION
1.
2.
20
Optional Accessories
5.4.1
Manometer
CAUTION
5.4.2
PEEP Valve
1.
2.
3.
Function test
clinical use.
the
absorber
before
21
6.
PRE-USE CHECKS
6.1
Pre-use Checklist
1.
2.
3.
4.
5.
Check the inspiratory and expiratory nonreturn valves for correct operation see
section 6.4.
6.
7.
8.
9.
22
PRE-USE CHECKS
The procedures detailed in sections 6.2 to
6.6 must be carried out in the order listed.
The absorber must be attached to an
anaesthetic machine, which must be leak
tested before the checks are carried out.
If a manometer is fitted, check that it is zeroed
before use.
6.2
C
B
Leak Test
2.
3.
6.3
3.
4.
5.
6.
23
PRE-USE CHECKS
6.4
1.
2.
3.
4.
Turn off the gas flow and check that the bag
does not empty by reverse flow through the
expiratory valve (B).
5.
6.
7.
8.
6.5
2.
3.
4.
5.
6.
7.
8.
9.
24
1.
PRE-USE CHECKS
6.6
1.
2.
3.
4.
5.
6.
7.
6.7
A
C
1.
2.
3.
4.
5.
6.
Turn the control knob fully anticlockwise and check that the pressure
falls.
7.
25
7.
MAINTENANCE
7.1
Maintenance and
Service Policy
7.3
Condensate Drainage
WARNING
Service Frequency
Servicing and repairs must only be carried
out by Penlon-trained technicians and
engineers.
(a)
(b)
7.2
7.3.1
Polemount Models
Weekly:
1.
2.
3.
26
MAINTENANCE
7.3.2
Daily Procedure:
1.
Weekly Procedure:
1.
2.
B
C
27
3.
4.
MAINTENANCE
Prima SP
In-board A100
Absorber and
Ventilator Bellows
(Ventilator control unit
mounted on side bracket
or shelf)
Prima SP
In-board A100
Absorber
(Ventilator mounted
on side bracket or
shelf)
28
MAINTENANCE
7.4
Inspiratory and
Expiratory Non-return
Valves
A
7.5
CAUTION
7.6
7.7
APL Valve
an
automatic
29
8.
STERILISATION
8.1
Bacterial Filters
8.2
Sterilisation Policy
8.3
30
STERILISATION
8.4
Absorber Assembly
Cleaning Procedure Before
Sterilisation
F
WARNING
CAUTION
CAUTION
1.
2.
3.
4.
5.
Refitting
Note - Refit the discs - the face with the white marking
(F) must be uppermost.
31
STERILISATION
8.5
Soap water
Cidex Sonacid
(Note 1)
Steam
Autoclave
Maximum
Temperature
oF
oC
Breathing hoses
(Penlon)
yes
yes
yes
278
137
Safelock fittings
yes
yes
yes
278
137
Reservoir bag
(Penlon)
yes
yes
yes
278
137
Manifold block
(including nonreturn valves)
yes
yes
yes
278
137
Frame assembly
yes
yes
yes
278
137
Canisters
yes
yes
no
APL valve
yes
yes
yes
278
137
Pressure gauge
no
no
no
PEEP valve
yes
yes
yes
278
137
General Notes:
1.
2.
3.
Thorough rinsing in warm water and drying in air should follow chemical disinfection.
Do NOT clean any component in an automatic cleaning/washing machine.
Before clinical use, ALWAYS carry out the Pre-use Checks listed in section 6 of this manual.
32
9.
SERVICING
Service Schedule
1
1.1
1.2
1.3
1.3
2
Initial Checks
Check serial numbers to determine service required.
Check general condition of Absorber and fittings.
Drain condensate from absorber. : Caution Caustic Solution
Check toggle labels for security of attachment.
If necessary, replace with new toggle with labels already attached.
On / Off Valve Assembly
2.1
2.2
Replace;
O' Ring
O' Ring
O' Ring
O' Ring
O' Ring
O' Ring
O' Ring
041220
041216
041207
39897
041204
020414
041219
x2
x3
2.3
Degrease Shaft A
Lightly lubricate Shafts A, B, C with Fomblin.
Reassemble Valve
NB; Use Loctite 242 at thread 'D' when assembling.
2
2.4
3.2
3.3
Degrease Shaft A
3.4
3.5
Ring
Ring
Ring
Ring
Ring
041220
041212
041204
39896
041207
x3
x2
33
SERVICING
3.6
Reassemble Valve
NB: Use Loctite 242 at point 'D' when assembling.
3.7
3.8
4.1
4.2
4.3
Dismantle
4.4
Replace:
O' Ring
O' Ring
O' Ring
Washer M4 Nylon
4.5
0501
001
041218
025609
x2
x3
Reassemble
5.1
Dismantle
5.2
Replace:
O' Ring
Washer M4 Nylon
Cable Tie (Canister)
041218
025609
90252
x 4
5.3
5.4
Reassemble
6
6.1
6.2
6.3
SERVICE SCHEDULE
7
7.1
7.2
Remove valve discs and wash under running water and dry carefully.
7.3
7.4
8.1
Part No 041218
8.2
Part No 041231
9
9.1
Miscellaneous
Replace remaining O-rings as supplied in Service kit referring to relevant diagrams.
9.2
9.3
10
10.1
11
Set Up
Connect absorber to anaesthetic machine that has been checked for LEAKS.
Check that manometer reads zero.
If no manometer fitted tee in 0 - 10 kPa ( 0-100 cmH2O) test gauge between Bag and bag
port.
For all tests refer to this gauge where manometer gauge mentioned.
Leak Test
11.1
11.2
Ensure bag fitted to bag port and position lever to 'Bag' position.
11.3
11.4
11.5
11.6
Stop the gas flow when Manometer reads 3 kPa (30 cmH2O).
Ensure pressure does not fall to zero within one minute.
12
12.1
12.2
12.3
35
SERVICE SCHEDULE
12.4
13
Use O2 flush
Check APL valve relieves pressure at 6 kPa
(60cm H2O) 10 % at 40 l/min flow rate.
Inspiratory and Expiratory Non-return Valve Test
13.1
13.2
13.3
13.4
13.5
Remove bung from inspiratory port and fit a second re-breathing bag.
13.6
13.7
14
14.1
14.2
14.3
15
15.1
15.2
Pressurise the system to 3 kPa (30 cmH20) and turn off gas flow.
15.3
15.4
Select On position
Check that system pressure exhausts through canisters.
15.5
Replace canisters
Ensure correct alignment of seals and tighten canister wheel clamp.
36
SERVICE SCHEDULE
16
Peep Valve
16.1
16.2
16.3
16.4
17
Paperwork
17.1
17.2
17.3
17.4
17.5
17.6
17.7
37
Tel:
Fax:
E-mail:
UK:
01235 547036
International:
+44 1235 547001
01235 547023
uksales@penlon.co.uk
export@penlon.co.uk
52590
58428
58436
58429
52584
52582
Note:
1.
Absorber detachables consists of:
3 litre breathing bag
1.05 m (42 inch) breathing tube
Female mount
Connector mount
Facemask elbow
Y piece
2.
All standard connectors are for use with 30 mm taper anti-pollution systems.
(Models compatible with 19 mm systems are available on request.)
38
A100 Absorber
Preventive Maintenance Kit
12 Month Service Kit
Kit Contents
Part No
041261
462631
103612
043
041216
041207
041204
041212
041219
39883
0518
39442
011160
011162
Qty.
1
1
2
4
1
2
3
1
1
2
2
2
1
1
Component Description
O-ring - water trap (A100 built-in to Prima SP)
Fresh Gas Hose - 1 m
Cable Tie - fresh gas hose
O-ring - Neoprene (Safelock Connectors)
O-ring - Viton
O-ring - Viton
O-ring - Viton
O-ring - Viton
O-ring - Viton
Seal - Dome
O-ring - Viton - Dome (early models)
Valve Disc
Corrugated Hose - Twin Canister
Corrugated Hose - Single Canister
WARNING
Function test the absorber before
clinical use.
See section 4 in the service manual.
041261
011160
(Twin canister)
011162
(Single canister)
462631
103612
39
PARTS LIST
041204
041219
041204
0501
043
041207
041216
043
043
0501
043
39442
39883
0518
041204
Note:
Use O-ring
0518 on early
models
041207
041212
40
PARTS LIST
Preventive Maintenance Kit
60 Month Service Kit
Qty.
1
4
1
2
3
1
Component Description
O-ring - water trap
O-ring - Neoprene
O-ring - Viton
O-ring - Viton
O-ring - Viton
O-ring - Viton
39883
0518
2
2
39442
011160
2
1
011162
Seal - Dome
O-ring - Viton - Dome
(early models)
Valve Disc
Corrugated Hose
(Twin Canister)
Corrugated Hose
(Single Canister)
WARNING
462631
103612
39897
39896
041220
041204
020414
01052
0501
001
041214
041231
090252
1
2
4
1
2
6
1
2
5
1
3
3
1
025609
041218
7
3
0432231
041218 (Manometer fitted)
0432231 (plug fitted)
0501
041222
025609
011160
011162
462631
103612
025609
001
41
041261
041204
020414
39897
041212
041204
Absorber On/Off
Switch Assembly
39897
041231
041219
041204
0501
39897
043
041216
041207
0501
043
041218
(ventilator connection
not fitted)
043
01052
041220
0501
Bag/Vent Switch
On absorbers built before end 1999,
check if latest type shaft is fitted.
Measure distance X.
Old type: X = 27.3 mm
New type: X = 24.8 mm.
Fit a new shaft (order Part No 39952)
if X = 27.3 mm.
041231
39442
39883
0518
041204
Note:
Use O-ring
0518 on
early models
39896
041204
043
041207
39897
041214
041212
01052
041212
041220
42
11. APPENDIX
Gas flow diagrams
The following pages contain gas flow diagrams as follows:
1.
Absorber ON / Bag ON
2.
3.
Absorber ON / ventilator ON
4.
NOTE
1.
2.
The APL valve is out of circuit when the system is in Ventilator mode.
3.
The absorbent canisters are sealed from the breathing circuit when the absorber is in
bypass mode (Absorber OFF).
41
Absorber ON / Bag ON
8
7
9
10
1.
2.
3.
4.
5.
6.
7.
Fresh gas in
8.
9.
11
3
6
2
1
circuit)
10.
11.
2.
3.
4.
5.
Fresh gas in
6.
7.
6
5
7
4
8
3
9
circuit)
8.
9.
Absorber ON / Ventilator ON
9
1.
2.
3.
4.
5.
6.
7.
Fresh gas in
8.
9.
10.
11.
7
10
11
3
6
2
1
1.
2.
3.
4.
5.
Fresh gas in
6.
7.
8.
9.
8
9
2
1
43
USER NOTES
44
USER NOTES
45