SECTION 13
BASIC FAULT FINDING
13-1
GFFTM
13-2
Contact MOLLEX.cdr
13-3
FAR 12 2300
P m ax=16bar
STAU BLI
P r e s s u r e R e g u la t o r @ 3 .5 0 b a r.c d r
GFFTM
13-4
13.7. Summary
Do not leave in place the manual injector tubing at all time (ball valve open).
It is possible to verify the hard disk system free space on the Settings screen,
press on More Info.
To prevent windows integration troubles due to chromato retention time instability,
it is recommended to work with the Autotracking function permanently activated.
When you want to print a report, go to playback screen and dont do this during
drilling operation. In fact, the print takes a very long time and the gas
chromatography is interrupted during the print. Any data files are saved during the
print.
To prevent any software screen freezing in Real Time screen, Playback screen or
Chromatographic screen: do your choice line by line and wait a few seconds
between each operations. A message appears at the left bottom of screen
"Drawing, please wait".
Total Gas Information screen zero offset adjustment: click on Update when the
GFF is completely clean and purged. Do not update zero when you are in
presence of gas.
If you detect an unidentified peak, it is possible that it is a pollution problem in the
gas circuit or in the air circuit of compressor. So, purge and decontaminate the
GFF (Oil, spray, contact cleaner...)
WARNING: The pump is not stopped by exiting the GFF application.
WARNING: Never exit GFF application when FID Cell coils are activated (Front
panel FID PCB, red led on) because the FID coils will stay permanently supplied
and will be burned and damaged.
If you only switch off the GFF while the GFF Computer is still running, you will
have red dotted lines on the screen. When you restart the GFF just after, you will
notice some disjointedness (like the cycle timer which stays permanently in red ).
This is completely normal. In fact you have to wait a complete cycle before having
a complete communication reset. Note that the Peltier cell can be blocked. To
restart cooling, you must return to the Settings Screen and re-enter the % power.
Caution : When you have troubles with the GFF, send an e-mail to Geoservices Hotline
explaining the trouble and if possible print a report by pressing this icon
are on Manual Control Screen (this icon
). It is
when you