ARTICLE 6
2013 SECTION V
range. The indications of cracks shall be compared between blocks A and B. If the indications obtained under
the proposed conditions on block B are essentially the
same as obtained on block A during examination at
40F to 125F (5C to 52C), the proposed procedure shall
be considered qualified for use. A procedure qualified at a
temperature lower than 40F (5C) shall be qualified from
that temperature to 40F (5C).
When a liquid penetrant examination cannot be conducted within the standard temperature range of 40F
to 125F (5C to 52C), the temperature of the examination shall be qualified in accordance with this Appendix.
MATERIALS
III-640
III-641
Figure III-630
Liquid Penetrant Comparator
Scribe
line
11/2 in.
(39 mm)
III-630
11/2 in.
(39 mm)
III-610
2 in.
(50 mm)
3/ in.
8
(10 mm)
REQUIREMENTS
COMPARATOR APPLICATION
B
GENERAL NOTE: Dimensions given are for guidance only and are
not critical.
170
ARTICLE 6
171
2013 SECTION V