Академический Документы
Профессиональный Документы
Культура Документы
ABSTRACT
GARDNER, A. S., S. STEPHENS, D. T. MARTIN, E. LAWTON, H. LEE, and D. JENKINS. Accuracy of SRM and Power Tap Power
Monitoring Systems for Bicycling. Med. Sci. Sports Exerc., Vol. 36, No. 7, pp. 12521258, 2004. Purpose: Although manufacturers
of bicycle power monitoring devices SRM and Power Tap (PT) claim accuracy to within 2.5%, there are limited scientific data available
in support. The purpose of this investigation was to assess the accuracy of SRM and PT under different conditions. Methods: First,
19 SRM were calibrated, raced for 11 months, and retested using a dynamic CALRIG (50 1000 W at 100 rpm). Second, using the same
procedure, five PT were repeat tested on alternate days. Third, the most accurate SRM and PT were tested for the influence of cadence
(60, 80, 100, 120 rpm), temperature (8 and 21C) and time (1 h at 300 W) on accuracy. Finally, the same SRM and PT were
downloaded and compared after random cadence and gear surges using the CALRIG and on a training ride. Results: The mean error
scores for SRM and PT factory calibration over a range of 50 1000 W were 2.3 4.9% and 2.5 0.5%, respectively. A second
set of trials provided stable results for 15 calibrated SRM after 11 months (0.8 1.7%), and follow-up testing of all PT units
confirmed these findings (2.7 0.1%). Accuracy for SRM and PT was not largely influenced by time and cadence; however, power
output readings were noticeably influenced by temperature (5.2% for SRM and 8.4% for PT). During field trials, SRM average and
max power were 4.8% and 7.3% lower, respectively, compared with PT. Conclusions: When operated according to manufacturers
instructions, both SRM and PT offer the coach, athlete, and sport scientist the ability to accurately monitor power output in the lab and
the field. Calibration procedures matching performance tests (duration, power, cadence, and temperature) are, however, advised as the
error associated with each unit may vary. Key Words: CALIBRATION, CYCLING, ERGOMETERS, VALIDITY
1252
METHODS
Experimental design. In the first of four experiments,
the accuracy of power data from 19 SRM and 5 PT powermeters were initially assessed using a dynamic calibration
rig (CALRIG) (Tom Stanef, SASI, Australia) (18). Powermeters were examined between the range of 50 and 1000 W.
After the initial trials, 15 of the SRM powermeters were
retested after 1 yr of use, whereas the five PT hubs were
tested again 2 d later. The second experiment used the most
accurate SRM and PT units (i.e., one of each) to examine the
influence of cadence (60, 80, 100, and 120 rpm) on accuracy
over a range of power outputs (50 1000 W). The third
experiment examined the influence of temperature and equipment warm-up on power output measurements. More specifically, power output was measured simultaneously in both
devices over one continuous hour using a CALRIG. These
trials were repeated on separate days in order to establish
repeatability. The fourth experiment examined the relationship
between power output, cadence, and speed of an SRM and PT
unit during a typical training ride in the field and a simulated
CALRIG session in the laboratory.
Equipment. The CALRIG used in this study was calibrated using four accredited masses up to 11.39 kg. For the
purposes of this study, the most popular professional version
of the SRM powermeters was used (i.e., four strain gauges).
SRM powermeters and PT were mounted to a standard
nine-speed road bicycle (Giant, Taiwan), which was attached to a stationary air-braked trainer (RX-5, Blackburn,
Sydney, Australia) (Fig. 1). The PT were all less than 1 yr
old whereas the SRM were less than 3 yr old with serial
numbers ranging between 235 and 1031.
Standard calibration procedure. After a 10-min
warm-up at a power output of 200 W, the zero offset for
the SRM was set and the PT torque was zeroed; both
procedures were carried out in accordance with the manufacturers instructions. The CALRIG load cell was calibrated with verified precision masses up to 11.39 kg and
dynamic system losses were accommodated for by repeating
unloaded zero offset procedures pre and post each trial.
Unless specified, all testing sessions were performed at 100
rpm and involved shifting through the full range of nine
CALIBRATION OF POWER MONITORING SYSTEMS
gears in the rear cluster using both the large and small front
chain rings. Due to technical limitations with the PT and
CALRIG download process, a second-by-second comparison between the CALRIG, PT, and SRM data could not be
made. Therefore, the protocol for recording involved a 40-s
stabilization period in each gear, followed by instantaneous
manual readings of power output from each instrument at
40, 45, 50, and 55-s periods. The four power outputs were
then averaged for further analysis. All trials were run under
standard environmental conditions (21C, 40 55% RH,
695710 mm Hg), and the same rear chain cluster (Durace,
Shimano, Japan) was used for each trial. All bicycle components were in good condition, less than 2 yr old, and the
chain was well lubricated. At least 1 h separated the trials in
order for the instrumentation to cool.
Experiment 1: accuracy and reliability of SRM
and PT data. Using the procedure outlined above, the accuracy of 19 SRM powermeters were initially evaluated
throughout a power range of 501000 W. After this, the
frequency versus torque slopes were adjusted so that the power
reading was within 2% of the CALRIG. The SRM calibration slope can be calculated via the following equation (Schoberer, U. SRM Training System Online Manual. Accessed;
November, 2003 www.srm.de/OnlineManual/index):
SRM Slope SRM Freq Hz SRM Zero offset Hz
Eleven months later, after a full racing season, the accuracy of 15 of the powermeters was again assessed. After
these trials and using the same calibration procedure, the accuracy of five PT rear-wheel hubs was assessed over a 2-d
period during repeat random trials. As PT calibration cannot be
altered by the user, no adjustment was made between trials.
Experiment 2: effects of cadence on SRM and PT
power data. Power measures from the most accurate
SRM and PT units from the above trials were compared
using the CALRIG as a reference at different cadences (60,
80, 100, and 120 rpm) over a 2-d period. The same equipment calibration, setup, and testing procedures were used as
in experiment 1.
Medicine & Science in Sports & Exercise
1253
TABLE 1. Mean percent error, SD, min, and max for SRM and PT when compared with the CALRIG at 100 rpm. Trial 1 and trial 2 for SRM were separated by 11 months (note
SRM calibration slope was adjusted following trial 1). Trial 1 and trial 2 for PT were separated by 1 d (note PT calibration cannot be adjusted). The individual devices with the
lowest errors from these trials were used in the subsequent trials.
Trial 1
SRM
PT
Trial 2
Mean
(%)
SD
(%)
Min
(%)
Max
(%)
19
5
2.3
2.5
4.9
0.5
1.0
2.0
10.4
2.9
Mean
(%)
SD
(%)
Min
(%)
Max
(%)
15
5
0.8
2.7
1.7
0.1
0.0
1.5
2.1
3.8
RESULTS
Table 1 shows mean percent error, SD, min, and max for
SRM and PT data when compared with the CALRIG at 100
rpm. During the first trial of experiment 1, the range in the
error scores for the 19 SRM powermeters was 10.4 to
1.0% compared with a range of 2.9 to 2.0% error observed for the 5 PT. Despite the range variations, the mean
percent error from all trials was similar. The best SRM
showed greater accuracy than the best PT over the range of
power outputs tested (1.0% for SRM vs 2.0% for PT). The
5 PT were, however, more consistently accurate. Furthermore, the results from the second experiment showed that
once corrected to within 0.0 2.1% accuracy, 14 of the 15
SRM powermeters retested after the 11 months period of use
stayed within that range. It was also observed that although
PT calibration cannot be altered by the user, it appears to
remain similar over a 2-d period.
FIGURE 2Modified Bland-Altman plot of percent error from the CALRIG for the most accurate SRM and PT during repeat trials at 100 rpm.
The mean percent error for SRM was 0.1 1.1 and 0.9 0.7 for T1 and T2, respectively. The mean percent error for PT was 2.1 1.3 and
1.5 0.6 for T1 and T2, respectively.
1254
http://www.acsm-msse.org
FIGURE 3Modified Bland-Altman plot of percent error from the CALRIG for the most accurate SRM and PT during four trials at different
cadences (60, 80, 100, and 120 rpm). The mean errors for the SRM and PT units were 0.1 1.0% and 1.9 0.8% at 60 rpm, 1.0 0.4%
and 2.0 0.7% at 80 rpm, 0.9 0.7% and 2.1 1.3% at 100 rpm, and 1.3 0.4% and 1.2 0.4% at 120 rpm, respectively.
FIGURE 4 Modified Bland-Altman plot of percent error from the CALRIG for the most accurate SRM and PT units during four 1-h and one
30-min steady-state trials (300 W). Repeated measurements were taken at 1, 3, 5, 7, 10, and then every 5 min until the 1-h mark after 6 h at standard
Lab conditions (21C, 40 55% RH) or Cool conditions (6C, 60% RH). One trial was also completed after the standard warm-up (10 min at
200 W). Note that for the 1-h trials, the zero offset of both the PT and SRM was set prior without warm-up; only the CALRIG was prewarmed for
these trials.
CALIBRATION OF POWER MONITORING SYSTEMS
1255
FIGURE 5Direct comparison between power (W) and cadence (rpm) for SRM and PT during dynamic CALRIG trials.
was similar, with 27.6 3.5 kmh1 for SRM and 27.7
4.1 kmh1 for PT (Fig. 6c).
DISCUSSION
The primary purpose of this study was to establish
whether the SRM powermeter and PT hub accurately measure power output during simulated high-intensity cycling
under a variety of conditions. The present data show that
using average data, both SRM and PT are usually within
their manufacturers specifications. However, when the data
are expressed using modified Bland-Altman plots, there is a
considerable scatter around the mean percent error. Acknowledging the range of scatter has important implications
when trying to interpret laboratory intertrial reliability or
technical error data using this type of power monitoring
system. The present data also show that during trial 1 (with
FIGURE 6 Direct comparison of power (W), cadence (rpm), and speed (kmh1) for SRM and PT during a training ride.
1256
http://www.acsm-msse.org
1257
that the exported time base (1.26 s) for the PT hub is much less
flexible than the SRM.
In conclusion, both SRM and PT are valuable instruments
for the monitoring of power output during cycling. However, researchers and scientists should be aware of the
limitations in both models of SRM and PT tested here when
trying to detect performance changes of less than 2%, common among elite athletes (7,8). Researchers should also note
that setting the zero offset does not substitute for a standardized calibration. In summary, accuracy may be different
between and within SRM and PT units and both units are
affected by temperature. It is possible that with continual
refinements by the manufacturers, improvements will, and
may have already been made to both devices.
The authors would like to extend gratitude to Prof. Allan Hahn and
the staff of the Physiology Department at the Australian Institute of
Sport for providing facilities and support in order to make this
project possible. Thanks must also go to Dr. Inigo Mujika for his
advice and critique of the manuscript.
REFERENCES
1. ALTMAN, D. G., and J. M. BLAND. Measurement in medicine: the
analysis of method comparison studies. Stat. 32:307317, 1983.
2. BALMER, J., R. C. DAVISON, and S. R. BIRD. Reliability of an
air-braked ergometer to record peak power during a maximal
cycling test. Med. Sci. Sports Exerc. 32:1790 1793, 2000.
3. BALMER, J., R. C. DAVISON, D. A. COLEMAN, and S. R. BIRD. The
validity of power output recorded during exercise performance tests
using a Kingcycle air-braked cycle ergometer when compared with
an SRM powermeter. Int J. Sports Med. 21:195199, 2000.
4. BASSETT, D. R., JR., C. R. KYLE, L. PASSFIELD, J. P. BROKER, and
E. R. BURKE. Comparing cycling world hour records, 19671996:
modeling with empirical data. Med. Sci. Sports Exerc. 31:1665
1676, 1999.
5. BROKER, J. P., C. R. KYLE, and E. R. BURKE. Racing cyclist power
requirements in the 4000-m individual and team pursuits. Med.
Sci. Sports Exerc. 31:16771685, 1999.
6. GOLICH, D., and J. BROKER. SRM bicycle instrumentation and the
power output of elite male cyclists during the 1994 Tour Dupont.
Performance Conditioning for Cycling 2:2, 1996.
7. HOPKINS, W., H. J. A. HAWLEY, and L. M. BURKE. Design and
analysis of research on sport performance enhancement. Med. Sci.
Sports Exerc. 31:472 485, 1999.
8. HOPKINS, W. G., E. J. SCHABORT, and J. A. HAWLEY. Reliability of
power in physical performance tests. Sports Med. 31:211234, 2001.
9. JEUKENDRUP, A., and A. VAN DIEMEN. Heart-rate monitoring during
training and competition in cyclists. J. Sports Sci. 16:S91S99, 1998.
1258
http://www.acsm-msse.org