Research article
www.scipharm.at
Open Access
Abstract
A novel stability-indicating reversed-phase Ultra-performance liquid chromatographic
(RP-UPLC) method was developed for the determination of purity of drotaverine
hydrochloride and ibuprofen in presence of their impurities and degradation
products. The method was developed using a Waters UPLC BEH C18, 1002.1mm,
1.7m column with a flow rate of 0.3 mL/min and detector wavelength at 210 nm.
The mobile phase consists of Potassium dihydrogen orthophosphate buffer and
acetonitrile. Drotaverine hydrochloride and ibuprofen were subjected to the stress
conditions of oxidative, acid, base, photolytic and thermal degradation. Degradation
products resulting from the stress studies were well resolved, thus the test method is
stability indicating. Validation of the method was carried out as per International
Conference on Harmonization
Keywords
dont touch!
Introduction
Drotaverine hydrochloride is Antispasmodic. it is Benzylisoquinoline derivative
described
chemically
as
1,2,3,4-Tetrahydro-6,7-diethoxy
-1-((3,4diethoxyphenyl)methylne)-isoquinoline hydrochloride (Figure 1E). Its empirical
formula is C24H31ClNO4 and molecular weight is 433.5 g/mol. [1-4]
Ibuprofen is a pain killer. It is chemically described as (RS)-2-(4-(2-methylpropyl)
phenyl) propanoic acid (Figure 1A). Its molecular formula is C13H18O2 and molecular
weight is 206.29 g/mol. [1-4]
A detailed literature survey revealed that there are some analytical methods reported
for the estimation of drotaverine either individually or in combination with other drugs
like HPTLC,[5] spectrophotometric[6-8] and by HPLC[813] and Ibuprofen either
individually or in combination with other drugs like HPLC[14] and by UPLC[1516].The route of synthesis of drotaverine and possible degradants resulted in four
known impurities, Impurity 1, Impurity 2, Impurity 3, and Impurity 4, which are not
reported in any of the pharmacopeia, where as ibuprofen EP reveals that Impurity A,
Impurity F, Impurity J and Impurity N are the known impurities for ibuprofen, in which
Impurity F detection reported by gas chromatography.
To date, there is no single method that has been reported for the determination of
Drotaverine and Ibuprofen combination impurities either in bulk drugs or in
pharmaceutical formulations of drotaverine along with Ibuprofen. It is necessary to
develop a stability-indicating method for Drotaverine and Ibuprofen-related impurities
in API and solid oral dosage forms which are available in combination packs.
Hence, an attempt has been made to develop an accurate, rapid, specific, and
reproducible method for the determination of Drotaverine and Ibuprofen combination
impurities (Figure 1) along with method validation as per ICH guidelines. The stability
tests were also performed on drug product as per ICH guidelines.[17-19]
There are no methods are available for simultaneous determination of drotaverine
HCl and ibuprofen in any dosage forms. However, numbers of methods are available
for quantification of related substances of plain ibuprofen and simultaneous
estimation of Ibuprofen (assay) with combination products. Two separate methods
are available for drotaverine HCl and ibuprofen. To our present knowledge, no
validated stability-indicating analytical HPLC or ultra-performance liquid
chromatography (UPLC) methods are available in literature for drotaverine
hydrochloride and ibuprofen and its impurities in oral dosage forms. Attempts were
made to develop a stability indicating UPLC method for the related substance
determination of drotaverine hydrochloride and ibuprofen in the presence of placebo.
This paper deals with the forced degradation of drotaverine hydrochloride and
ibuprofen oral solid dosage forms under stress conditions like acid hydrolysis, base
hydrolysis, water hydrolysis, oxidation and heat. This paper also deals with the
validation of the developed method for the accurate quantification of drotaverine
hydrochloride and ibuprofen impurities in oral solid dosage forms.
dont touch!
The method was developed on UPLC with low particle size (1.7 micron) and a short
100-mm column. A rapid, unique, reproducible, stability-indicating UPLC method was
developed for the quantitative determination of drotaverine hydrochloride and
ibuprofen impurities in oral solid dosage forms.
Figure 1. Chemical structure and name of Ibuprofen, Drotaverine and impurities.
Fig
Name
1A
Ibuprofen
Structure
(2RS)-2-[4-(2Methylpropyl)phenyl]propanoi
c acid.
Ibuprofen imp-A
1B
Ibuprofen imp-J
1C
1D
Chemical name
Ibuprofen
imp-N
(1Z)-1-(3,4Diethoxybenzylidene)-6,7diethoxy-1,2,3,4tetrahydroisoquinoline
Drotaerine
1E
dont touch!
Drota-IMP 1
(5,6-diethoxy-3,4dihydroisoquinolin-1-yl) (3,4diethoxyphenyl)methanone
1F
Drota-IMP 2
(3,4-dimethoxyphenyl)acetic
acid
1G
Drota-IMP 3
2-(3,4-diethoxyphenyl)-N-[2(3,4diethoxyphenyl)ethyl]acet
amide
1H
Drota-IMP 4
1I
N-ethyl-3,4-dimethoxyaniline
Experimental
Chemicals and reagents
Drotaverine hydrochloride, Ibuprofen standard and its impurities were supplied by Dr.
Reddys Laboratories (Hyderabad, India). The HPLC-grade acetonitrile analytical
grade, KH2PO4 and ortho-phosphoric acid were purchased from Merck (Darmstadt,
Germany). High-purity water was prepared by using a Millipore MilliQ Plus water
purification system.
Equipments
UPLC analysis was performed with a Waters (Milford, MA) Acquity UPLC system
equipped with a quaternary solvent manager, sample manager, column-heating
compartment and photodiode array detector. This system was controlled by Waters
Empower2 software.
dont touch!
An Acquity UPLC BEH C18 column, 1002.1mm, 1.7m (Waters) was employed for
chromatographic separation. All samples were centrifuged by a Thermo Scientific
multifuged machine. The specificity study was conducted by using heating oven
(MACK Pharmatech, Hyderabad, India) and photo stability chamber, and water baths
equipped with a Milli Volt controller (Julabo, Seelbach, Germany) were used for
hydrolysis studies.
Chromatographic conditions
The method was developed using a BEH C18, 1002.1 mm, 1.7m column as
stationary phase. The mobile phase A consists a mixture of 0.02M phosphate buffer
pH 3.0 and acetonitrile in 900:100 v/v ratio and mobile phase B consists a mixture
Buffer pH 3.0 and acetonitrile in 400:600, v/v ratio. A mixture of Milli Q water and
acetonitrile in 30:70 v/v was used for diluent to prepare solutions. The gradient
program Time (minutes)/% mobile phase B (%B) was set as 0/10, 6/50, 15/60, 18/65,
21/80, 23/100, 23.1/10 and 25/10 respectively. Before use, the mobile phase was
mixed thoroughly and degassed. The mobile phase was pumped at 0.3 mL/min. The
eluted compound drotaverine hydrochloride, ibuprofen and their impurities were
monitored at 210 nm. The column temperature was maintained at 25C. The
injection volume for samples and standards set as 1.0L.
Preparation of Standard Solution and System Suitability Solution
Stock solutions of Drotaverine 500g mL 1 and Ibuprofen 1000g mL1 was prepared
by dissolving an appropriate amount in diluent individually. Working standard solution
was prepared from the above stock solutions for related substances determination
(2g mL1 of Drotaverine and 15g mL1 of ibuprofen) in diluent.
System suitability solution was prepared by dissolving an appropriate amount of
drotaverine and ibuprofen to get concentration of 1000g mL 1 , 5000g mL1
subsequently in diluent and mix all impurities of drotaverine (2.0g mL 1) along with
ibuprofen (15g mL1) in same preparation.
Preparation of Test Solution
Twenty individual tablets (Drotaverine label claim: 80 mg per tablet, Ibuprofen 400mg
per tablet) content was weighed and the average weight of each tablet was
calculated. Tablet powder equivalent to 100 mg of the active pharmaceutical
ingredient drotaverine and 500 mg of Ibuprofen was transferred into a 100 mL
volumetric flask. To this, 70 mL of diluent was added and sonicated for 30 minutes
with intermediate shaking. The solution was then diluted to 100 mL with diluent and
centrifuged at 4000 rpm for 10 min. The supernatant was collected and used as
sample solution; this solution was filtered through a 0.2-mm Millipore PVDF filter.
Method validation
The proposed method was validated as per ICH guidelines.[33, 34]
dont touch!
Specificity
Specificity is the ability of the method to measure the analyte response in the
presence of its potential degradants. Stress studies were performed for tablets to
provide an indication of the stability-indicating property and specificity of the
proposed method. Intentional degradation was attempted to determine the stress
conditions of heat (60C), photolytic sunlight for approximately 1.2 million lux hours
and UV light, both at shorter and longer wavelengths for approximately 200 Wh/m 3,
acid (2N HCl), base (2N NaOH) and oxidation (3% H2O2) to evaluate the ability of
the proposed method to separate drotaverine and ibuprofen from their degradation
products. For the acid, base, water, hydrolysis and oxidation, the study period was
3hrs in water bath for thermal degradation it was stressed to 8 hr. Humidity exposed
to 90%RH for 7 days. Peak purity test was carried out for the drotaverine and
ibuprofen peaks by using photodiode array (PDA) detector in stress samples.
METHOD VALIDATION PARAMETERS
Precision
The precision of the related substances method was checked by injecting six
individual preparations of drotaverine hydrochloride (1000g/mL), ibuprofen
(5000g/mL) and its impurities with 0.20% of imp-1, imp-2, imp-3 and imp-4 for
drotaverine, imp-A, imp-J and imp-N for ibuprofen with respect to analyte
concentration. The relative standard deviation (RSD) of the area for each impurity
was calculated.
Intermediate precision of the method was also evaluated using different analysts and
different instruments in the same laboratory
Limit of Detection (LOD) and Quantification (LOQ)
Limits of detection and quantification the limit of detection (LOD) and limit of
quantification (LOQ) for impurities and analyte (with respect to unknown impurities)
were determined at signal-to-noise ratios of 3:1 and 10:1 respectively, by injecting a
series of dilute solutions with known concentrations. A precision study was also
carried out at the LOQ level by injecting six individual preparations of ibuprofen and
its known impurities (imp-1, imp-2, imp-3 and imp-4) and drotaverine hydrochloride
and its known impurities (imp-A, imp-J and imp-N) and calculating the RSD of the
area.
Linearity
Linearity test solutions for the related substance method were prepared by diluting
stock solutions (described previously) to the required concentrations. The solutions
were prepared at five concentration levels from LOQ to 250% of the specification
level 0.2% of the respective analyte concentrations of drotaverine hydrochloride
(1000g/mL) and ibuprofen (5000g/mL). Correlation coefficient, value for the slope,
Y-intercept and percent bias of the calibration curve was calculated.
dont touch!
Accuracy
The accuracy study of all impurities was carried out in triplicate at LOQ, 50%, 100%,
150% and 250% of the target concentration level: 0.2% of respective analyte
concentrations of drotaverine hydrochloride (1000g/mL) and ibuprofen
(5000g/mL). The percentages of recoveries for impurities were calculated.
Robustness
To determine the robustness of the developed method, experimental conditions were
deliberately altered and the resolutions between all peaks were recorded. The flow
rate of the mobile phase was 0.30 mL/min. To study the effect of flow rate on the
resolution, flow was changed by 0.03 units from 0.27 to 0.33 mL/min. The effect of
the column temperature on resolution was studied at 20 and 30C instead of 25C.
The effect of the percent organic strength on resolution was studied by varying
acetonitrile by +10% with constant ratio of buffer. While other mobile phase
components was held constant as stated previously.
dont touch!
The stationary phases like BEH C8, HSS T3 and Phenyl columns were tried but
ibuprofen, drotaverine and its impurities were not well separated from main peaks
and symmetry is not adequate as shown in Fig.2A. On the other hand the peak
shapes for all the two components are good and all the impurities peaks are well
separated from each other and their degradents. So that the column with BEH C18,
1002.1mm, 1.7m stationary phase is selected.
Selection of mobile phase buffer:
The buffers like Phosphate, Acetate buffers With Buffer with pH 4.5, pH 7.5 and the
perchlorate buffer with pH 2.5 and 3.2 were tried as mobile phase buffer. But with pH
4.5, pH 7.5 phosphate and per chlorate buffer with pH 2.5 the results were not
enough good in terms of peak separation and peak shapes. On the other hand the
potassium di hydrogen phosphate buffer with pH 3.0, the results were found
satisfactory so that the mobile phase with pH 3.0 phosphate buffer is finalized.
Optimization of gradient program:
Before going to gradient program, an isocratic program with organic modifier
methanol and acetonitrile were tried but all the two components are eluted at the
same retention time with both methanol and acetonitrile. Hence gradient method was
used to separate all components from its known impurities. As the column
temperature was maintained at 30C, acetonitrile is preferred as an organic modifier
rather than methanol. The different gradient programs tried are shown in the Fig. 2B.
Finalization of Chromatographic conditions
By considering all the experiments the chromatographic conditions are finalized as
BEH C18, 1002.1mm, 1.7m column with mobile phase containing a gradient
mixture of solvents A and B. Phosphate buffer (0.02M) with pH 3.0. Buffer and
acetonitrile in 900:100, v/v ratios were used as mobile phase A, Buffer and
acetonitrile in 400:600, v/v ratios were used as mobile phase B. The gradient
program time (minutes)/% mobile phase B (%B) was set as 0/10, 6/50, 15/60, 18/65,
21/80, 23/100, 23.1/10 and 25/10 respectively .The mobile phase was pumped at 0.3
mL/min. The eluted compound Drotaverine hydrochloride, Ibuprofen and their both
the impurities were monitored at 210 nm. The column temperature was maintained at
25C. The injection volume for samples and standards was 1.0L.
The chromatograph with finalized chromatographic condition was shown in the Fig.
2C.
Figure 2. Method development chromatograms. (A) Selectivity differences of
Ibuprofen, Drotaverine and its impurities by using different stationary phases. (B)
Retention of Ibuprofen, Drotaverine and its impurities with gradient elusion. (C)
Blend chromatogram of Ibuprofen, Drotaverine and its related impurities in final
chromatographic conditions.
A.
dont touch!
B.
0.050
0.040
0.030
AU
0.020
0.010
0.000
-0.010
-0.020
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.00
Minutes
11.00
12.00
13.00
14.00
15.00
16.00
17.00
18.00
19.00
C.
IBUPROFIN- 17.933
IBU-IMP-A- 17.387
DROTA-IMP-4- 14.144
IBU-IMP-N- 10.379
DROTA-IMP-3- 11.448
0.02
DROTA-IMP-2- 6.201
0.04
DROTA-IMP-1- 2.416
AU
0.06
IBU-IMP-J - 8.593
0.08
DROTAVERINE- 7.742
0.10
0.00
-0.02
0.00
2.00
4.00
6.00
8.00
10.00
12.00
Minutes
14.00
16.00
18.00
20.00
22.00
24.00
20.00
10
dont touch!
RT
Drotaverine 7.742
Imp-1
2.416 0.31
Imp-2
6.201
0.8
Imp-3
11.448 1.48
Imp-4
14.144 1.83
1.1
1.02
1.05
1.14
1.11
1.01
0.99
0.99
1.01
1.01
B) Ibuprofen
Compound
Name
RT
RRT
Ibuprofen
Imp-A
Imp-J
Imp-N
17.933
17.387
8.593
10.379
0.97
0.48
0.58
Tailing Similarity
Factor
Factor
1.2
1.14
1.01
1.11
0.99
0.99
1.01
0.99
Specificity
All forced degradation samples were analyzed at an initial concentration of
drotaverine hydrochloride and ibuprofen with previously described UPLC conditions
using PDA detector to ensure the homogeneity, purity of the drotaverine and
ibuprofen peaks. Significant degradation of drotaverine and ibuprofen was observed
in heat (60C for 3hr), photolytic UV light (200 Wh/m 3), sun light (1.2 million lux
hours), oxidation (3.0% H2O2 at 24Hr RT), acid (2N HCl at 60C for 3Hrs) base (2N
NaOH at 60C for 3Hr) and humidity exposed to 90%RH for 7 days. The percent
degradation values are presented in Tables 2, Figures 3 A to F.
dont touch!
18.723
17.200
10.815
9.967
10.204
8.593
6.877
4.698
5.110
5.558
5.850
6.196
0.50
13.404
1.00
14.108
AU
1.50
11.529
DROTAVERINE - 7.672
2.00
21.066
21.516
IBUPROFINE - 17.761
2.50
0.00
0.00
2.00
4.00
6.00
8.00
10.00
12.00
Minutes
14.00
16.00
18.00
20.00
22.00
24.00
B.
2.40
IBUPROFINE - 17.837
2.20
2.00
1.60
AU
1.40
1.20
1.00
0.80
11.531
DROTAVERINE - 7.836
1.80
22.124
21.087
21.514
19.685
20.411
18.690
17.196
15.925
13.380
11.166
8.578
9.019
5.109
5.542
0.20
2.330
0.40
6.869
7.162
7.540
0.60
0.00
0.00
2.00
4.00
2.00
4.00
6.00
8.00
10.00
12.00
Minutes
14.00
16.00
18.00
20.00
22.00
24.00
20.00
22.00
24.00
C.
2.00
0.00
0.00
6.00
8.00
10.00
12.00
Minutes
14.00
17.280
13.503
11.529
10.933
6.888
5.582
5.878
6.217
4.734
0.50
8.644
1.00
DROTAVERINE - 7.761
AU
1.50
18.878
IBUPROFINE - 17.990
2.50
16.00
18.00
AU
0.00
F.
2.40
2.20
2.00
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
0.00
2.00
4.00
2.00
4.00
6.00
8.00
8.00
10.00
10.00
12.00
Minutes
12.00
Minutes
14.00
14.00
16.00
18.00
18.00
20.00
20.00
22.00
22.00
22.951
22.123
20.00
21.064
21.533
18.00
20.428
16.00
19.744
IBUPROFINE - 17.863
18.673
17.191
IBUPROFINE - 17.795
1.50
18.777
17.232
14.00
15.937
14.955
11.602
10.762
9.979
8.562
8.801
AU
2.00
18.831
IBUPROFINE - 17.960
6.00
12.00
Minutes
17.282
0.50
13.129
13.431
1.50
10.00
13.481
2.00
11.476
8.00
11.548
6.00
DROTAVERINE - 7.747
0.00
8.633
2.50
DROTAVERINE - 7.784
5.102
5.547
5.846
6.182
6.443
6.864
0.50
8.364
8.605
9.033
9.335
9.807
9.975
4.00
6.886
7.170
7.527
2.00
5.889
6.206
2.447
1.00
DROTAVERINE - 7.769
5.580
5.878
6.215
6.460
6.892
0.00
4.735
5.115
2.334
AU
12
dont touch!
D.
2.50
22.00
24.00
E.
1.00
0.00
24.00
24.00
dont touch!
1.1
3.4
1.7
0.267
0.335
0.258
1.957
1.789
1.57
99.8
99.1
100.2
2.1
2.5
1.4
0.524
0.325
0.245
1.524
2.524
1.85
97.2
96.5
98.8
0.9
0.4
0.424
0.425
1.794
2.241
98.9
99.3
1.2
0.8
0.534
0.248
1.245
1.662
98.3
98.7
0.6
0.344
1.883
99.2
0.4
0.254
1.852
99.5
Precision
The RSD for the area of drotaverine imp-1, imp-2, imp-3 and imp-4 and ibuprofen
imp-A, imp-J and imp-N in related substances are presented in Table 3. The method
precision and intermediate precision study found to be less than 2% shown in the
Table 4 (which should be less than 10.0%), conforming good precision of the
method.
Table.3 Precision of the proposed HPLC method
Prepar
ation
Prep-1
Impu Impurit
rity 1 y 2
0.201
0.198
Prep-2
0.204
0.196
Prep-3
0.207
0.194
Prep-4
0.201
0.197
Prep-5
0.202
0.192
Prep-6
0.201
0.196
Avg
0.203
0.196
Stdev
0.002
0.002
% RSD 1.2
0.2
Impur Ibupr
ity N
ofen
0.202 0.198
0.200
0.205
0.204
0.204
0.200
0.205
0.202
0.208
0.200
0.204
0.201
0.204
0.002
0.003
0.8
0.3
14
dont touch!
Impu
rity 1
0.025
0.026
0.025
0.024
0.027
0.026
0.025
4.1
Impu
rity 2
0.023
0.022
0.025
0.024
0.023
0.022
0.023
5.3
Impu
rity 3
0.023
0.025
0.024
0.025
0.024
0.022
0.024
4.9
Impu
rity J
0.016
0.017
0.015
0.015
0.016
0.016
0.015
3.9
Impu
rity N
0.019
0.019
0.021
0.02
0.021
0.02
0.02
4.1
Ibupr
ofen
0.013
0.014
0.013
0.015
0.013
0.014
0.015
4.9
Linearity
The result shows that an excellent correlation existed between the peak area and
concentration of the analyte. A linear calibration plot for the related substance
method was obtained over the calibration ranges tested, i.e., LOQ to 250% for
drotaverine and its impurities (imp-1, imp-2, imp-3 and imp-4) and ibuprofen and its
impurities (imp-A, imp-J and imp-N). The correlation coefficient obtained was greater
than 0.997 (Table 6). These results show that an excellent correlation existed
between the peak area and the concentration. The percent bias was also calculated
for all related compounds and main analytes and found to be less than 2.5% (Table
6).
Table.6 Linearity of the proposed HPLC method
Substance
Linearity
range(g
mL1)
Correlation
coefficient (R2)
Y-intercept
Bias in %
Slope
Impurity 1
Impurity 2
Impurity 3
Impurity 4
Drotaverine
Impurity A
Impurity J
0.02515.033
0.02315.995
0.02305.089
0.03345.011
0.03245.050
0.0251-25.233
0.03125.195
0.9998
0.9996
0.9987
0.9992
0.9984
0.9993
0.9997
3
1.7
1.8
1.8
1.5
1.1
2.3
39419.7
31872.2
41923.2
49327.6
37977.5
37223.9
21139.7
dont touch!
Impurity N
0.0210-25.189
0.9995
1.6
26678.3
Ibuprofen
0.154-24.9980
0.9994
1.3
30278.9
Accuracy
The percentage recovery of drotaverine hydrochloride and ibuprofen impurities
varied from 85 to 115% and RSD of three samples at each level was found to be less
than 15 % at 50%, 100%, 150% and 250% levels of the target 0.2% level of the
target concentrations, respectively. An LC chromatogram of the spiked sample at the
0.2% level of all seven impurities and two main peaks in the oral solid dosage
sample solution is shown in Figure 2. Percent recovery values for impurities are
presented in Table 7(Percent recovery should be between 90 and 110%).
Table. 7 Accuracy of the Proposed HPLC Method
Spike Impur Impur
Level
ity 1
ity 2
0.10% 101.6 102.4
0.4
0.7
0.20% 100.7 103.4
0.5
1.2
0.50% 102.6 105.5
1.0
0.3
0.75% 100.9 102.6
0.9
1.1
1.00% 100.7 103.2
0.3
1.9
Impu
rity 3
101.2
1.1
102.6
0.2
103.1
0.4
98.6
0.3
103.0
0.5
Impuri
ty 4
100.9
1.0
98.9
1.2
97.2
0.8
96.7
0.4
98.9
0.9
Drotave
rine
98.5
1.4
100.5
0.9
100.1
0.5
102.5
0.5
99.2
0.8
Impurit
yA
98.5
0.4
105.5
0.10
100.2
0.6
99.5
0.6
99.2
0.9
Robustness
In all of the deliberate varied chromatographic conditions (flow rate, column
temperature and composition of organic solvent), the resolution between critical pairs
was greater than 2.0, illustrating the robustness of the method.
Conclusions
The gradient UPLC method was developed for drotaverine hydrochloride and
ibuprofen related substances in oral solid pharmaceutical dosage forms is found to
be precise, accurate, linear, robust, rugged and specific. Satisfactory results were
obtained from validation of the method. Hence, the method is stability-indicating and
can be used for routine analysis of production samples and to check the stability
samples of Oral solid dosage form.
16
dont touch!
Acknowledgments
The authors wish to thank the management of Dr. Reddys group for supporting this
work.
Authors Statement
Competing Interests
The authors declare no conflict of interest.
REFERENCES
[1]
[2]
Oneil MJ, Smith A, Heckelman PE. The Merck Index (Thirteenth Edition).
2001; 349
[3]
[4]
[5]
[6]
[7]
[8]
dont touch!
Syal PK, Sahoo M, Ingale KD, Ingale SS, Choudhari VP, Kuchekar BS.
Development and validation of a RP-HPLC-PDA method for simultaneous
estimation of Drotaverine and Etoricoxib in tablet and its application for
dissolution studies.
Der Pharma Chemica.2010; 2:93102.
[10]
[11]
Patel JD, Patel BA, Vaghela VM, Patel LJ, Patel PU.
RP-HPLC method for the simultaneous determination of mefenamic acid and
drotaverine hydrochloride in tablet dosage form.
Int J Pharm Res.2009; 1:3539.
[12]
[13]
[14]
[15]
[16]
18
dont touch!
[17]
[18]
ICH Q1A (R2) (2003). Stability Testing of New Drug Substances and
Products.
[19]
ICH Q1B (1996) Photo Stability Testing of New Drug Substances and
Products.