Академический Документы
Профессиональный Документы
Культура Документы
8 Instrument Calibration
H. M. HASHEMIAN (2003)
INTRODUCTION
Span Shift
The performance of a process instrument such as a temper- Zero Shift
ature or pressure sensor normally depends on its accuracy 100
and response time. Accuracy is a qualitative term that de-
e
rv
Cu
scribes how well the instrument may measure the process
e
rv
n
tio
Cu
parameter (see Section 1.4 for an in-depth discussion), and
% Output y
ra
tic
lib
response time specifies the speed by which the instrument
ris
Ca
te
ac
al
can detect any significant change in the value of a process
ar
tu
Ch
Ac
parameter. Accuracy and response time are largely indepen-
ed
ifi
dent and are therefore identified through separate procedures.
ec
b
Sp
The accuracy of a process instrument is established m
0
through its calibration, and the response time is determined by
0 % Input x 100
exposing the instrument to a dynamic input and measuring its
Span
response time from the transient output. Response time is
discussed in more detail in Section 1.9. Calibration is done by
FIG. 1.8a
providing the instrument with a number of known and stable
Illustration of zero and span. b = zero elevation; m = gain of a
inputs to ensure that the output accurately represents the input. linear transmitter; span = the change in input (x) as it travels from
Two terms are important in instrument calibration. These 0100%; range = the change in output (y) corresponding to 0100%
terms are zero and span, as illustrated in Figure 1.8a. In this span; output @ input = 0 is called lower range value; output @
figure, the calibration of a linear transmitter is approximated input = 100% is called upper range value.
with a straight line, represented by the equation y = mx + b,
where y is the output, x is the input, m is the slope of the +Zero
line, and b is the intercept. The calibration of an instrument Shift
may change due to a change in zero, a change in span, or a Zero
change in both zero and span. A change in zero is also Shift
referred to as a bias error, DC offset, or zero shift. A zero
shift results in a change in instrument reading (either positive
Output
span shift. A span shift means an increase or a decrease in the FIG. 1.8b
slope of the instrument output line for the same input (see Illustration of a zero shift in an instrument calibration.
Figure 1.8c). Typically, calibration errors involving span shift
alone are less common than calibration errors due to both zero Calibration of Pressure Sensors
and span shifts. In Figure 1.8c, both cases are shown: span shift
without zero shift and span shift with zero shift. In pressure Calibration of pressure sensors (including both absolute and
transmitters, about 40% of the calibration changes are caused differential-pressure sensors) involves using a constant pres-
by zero shift, about 30% by span shift, and only about 20% by sure source such as a deadweight tester (see Figure 1.8d).
1
span shift alone. As for the remaining 10%, the calibration With a deadweight tester, constant pressure is produced for
changes are due to other effects, such as nonlinearity. the sensor while the sensor output is monitored and adjusted
108
2003 by Bla Liptk
1.8 Instrument Calibration 109
Span Shift adjustment is also provided (in addition to the zero and span
+ Zero Shift
potentiometers) to help achieve the best agreement between
the input pressure and the output current.
In lieu of a deadweight tester, one can also use a stable
pressure source and a precision pressure gauge as the input.
Precision pressure gauges are available in a variety of ranges
Input from a number of manufacturers (see Section 5.11). Highly
accurate digital pressure indicators can also be used for cal-
FIG. 1.8c ibration. As will be seen later, automated pressure sensor
Illustration of span shifts. calibration equipment is also available that uses digital tech-
nology to offer both accuracy and convenience.
TABLE 1.8e
FIG. 1.8d Example of an Instrument Calibration Data
Schematic of a typical deadweight tester.
Input Signal Desired Output As-Found Data As-Left Data
(% of Span) (mA) (mA) (mA)
to make the electrical output proportional to the applied pres- 0 4.00 3.93 3.99
sure. For example, a pressure sensor may be calibrated to 25 8.00 8.03 8.01
produce an output in the range of 4 to 20 mA for pressure
50 12.00 11.92 12.04
inputs covering the whole span of the sensor (0 to 100%).
75 16.00 16.09 15.96
For most pressure sensors, with no pressure applied, the
transmitter output is adjusted to produce a 4-mA signal. Next, 100 20.00 20.12 19.97
calibrated by systematically applying a series of input signals sensor may be calibrated using input signals in this sequence:
and making zero and span adjustments as necessary to bring 0, 25, 50, 75, 100, 75, 50, 25, and 0% of span.
the sensor within acceptance limits or criteria. The input/out- Normally, a manufacturers specification of an instru-
put data after this calibration is referred to as the as-left data, ments accuracy is arrived at considering hysteresis, linearity,
and the difference between the as-found and as-left data is repeatability, and other factors that can affect the instruments
often termed the calibration drift or calibration shift. input/output relationship (see Section 1.3 for the definitions
The acceptance criterion for calibration is normally of these terms).
established by the user, based on the accuracy requirements
for the instrument and/or on the manufacturers specification. Calibration Traceability
Typically, manufacturers state the inaccuracy (the error or
To be valid and acceptable, all calibrations should be traceable
uncertainty) of an instrument in terms of a percentage of span
to a national standard or to a known physical phenomenon.
throughout the range of the instrument. For example, pressure
In the United States, equipment calibrations are expected to
sensor errors can be as low as about 0.05% of span in high-
be traceable to the National Institute of Standards and Tech-
precision sensors or as high as about 1% of span in sensors
nology (NIST), in Gaithersburg, MD. Each country has its
for general applications.
own standard calibration laboratory for a variety of parameters
The high-precision performance can only be obtained if
such as pressure, temperature, voltage, resistance, weight, time,
the sensor is properly calibrated, because overall accuracy
and so on. The national standards laboratories are normally
depends as much on the quality of the calibration process as
charged with calibration of primary standards. A primary stan-
it does on the ability of the sensor to provide a particular
dard is an instrument that is calibrated at the national standard
accuracy. Reliable calibration requires the availability of
laboratory like the NIST and used to calibrate other equipment.
trained personnel, written procedures, and accurate calibra- For example, the primary standard for calibration of resistance
tion equipment. Obviously, the accuracy of the calibration temperature detectors (RTDs) is a precision RTD that is
equipment must be much better than the equipment being referred to as a standard platinum resistance thermometer or
calibrated, and the calibration personnel must understand the SPRT. Each RTD calibration laboratory has one or more SPRTs
calibration process and exercise care in performing the cali- that are sent to NIST periodically to be calibrated.
bration task, reading the calibration data, and documenting To be calibrated by a national standard laboratory such
the results. If an instrument is not properly calibrated, it will as NIST, a primary standard must be in good working con-
not produce accurate and reliable measurements, even if the dition and meet certain requirements. Otherwise, NIST may
quality of the instrument itself is high. not calibrate it. National standards laboratories are charged
with helping to maintain primary standards and therefore do
Hysteresis not calibrate common-purpose equipment, nor do they repair
equipment that is not in good working order.
Hysteresis is a phenomenon that causes an instruments output To protect the primary standard equipment, a secondary
to vary, depending on the direction of the applied input signals, standard (also called transfer standard) may be used. In this
i.e., whether it is increasing or decreasing (see Figures 1.3b case, the primary standard is used to calibrate the secondary
and 1.8f). To account for hysteresis, instruments are some- standard, which is then used for calibration of other instru-
times calibrated using both increasing and decreasing input ments. The advantage of this approach is that it preserves the
signals, and the results are averaged. For example, a pressure integrity as well as the accuracy of the primary standard by
minimizing its handling and use. The disadvantage is that,
every time an instrument is calibrated, the accuracy that can
be claimed is below the accuracy of the standard that is used
in its calibration. Therefore, in going from a primary standard
to a secondary standard, a loss of accuracy results.
A common rule is that an instrument shall be calibrated
using a standard that is at least four times more accurate than
the instrument being calibrated. This requirement is some-
Output
RTD SPRT drift randomly within a certain band. Typically, the band is
the largest at the beginning of the life cycle of the instrument,
then it narrows and remains stable for a long period before
it widens again as the instrument ages.
Assessment of Accuracy
For fieldbus devices, ranging is called scale to distinguish it that this is ranging and not calibration. Calibration is when
from calibration. Ranging is accomplished by selecting input at zero differential pressure the transmitter reading is found
values at which the outputs are 4 mA and 20 mA, while to be 1 inch of water and it is corrected to read 0 inches of
calibration is performed to correct the sensor reading when water.
it is inaccurate.
For example, if a differential-pressure sensor has a lower
range limit (LRL) of 200 inches of water and an upper References
range limit (URL) of +200 inches of water, and if one wants
to use it to measure a differential pressure in the range of 0 1. EPRI Topical Report, On-Line Monitoring of Instrument Channel Per-
to 100 inches of water, then both the LRV and the URV formance, TR-104965-R1 NRC SER, Electric Power Research Insti-
settings must be changed. This adjustment would range tute, Final Report, Palo Alto, CA, September 2000.
2. Hashemian, H. M. et al., Aging of Nuclear Plant Resistance Temper-
(increase) the lower range value (LRV) to equal 0 and lower
ature Detectors, U.S. Nuclear Regulatory Commission, Report Num-
the upper range value (URV) to be 100 inches of water. Note ber NUREG/CR-5560, Washington, DC, June, 1990.