Вы находитесь на странице: 1из 9

~

..
CopyrightAmericanSocietylor TestingandMaterials(ASTM)
100 Barr Harbor Orive, West Conshohocken,
Pennsylvania 19428-2959
Unitea States ot Amerlca

Kopia wykonana przez Polski Komilet NormalizacvJny

~~r~ Designation: D 5453 - ooe1 na pOdstawielicencji udzielonejprzez ASTM An AmericanNationalStandard

Standard Test Method for


Determination ot Total Sultur in Light Hydrocarbons, Motor
Fuels and Oils by Ultraviolet Fluorescence 1
This standard is issued under the fixed designation D 5453; the number immediately following the designation indicates the year of
original adoption ar, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (E) indicates an editorial change since the last revision or reapproval.

El NOTE-Footnote l was corrected editorially in March 2001.


.
1. Scope D 4177 Practice for Automatic Sampling of Petroleum and
1.1 This test method covers the determination of total suIfur Petroleum Products3
fi D 6299 Practice for Applying Statistical Quality Assurance
in liquid hydrocarbons, boiling in the range from approxi-
maiely 25 to 400C, with viscosities bt:tween approximately Techniques to Evaluate Analytical Measurement System
Performance4
0.2 and 20 cSt (mm2/S) at room temperature.
1.2 Three separate inter-Iaboratory studies (ILS) on preci-
3. Summ ary of Test Method
sion have determined that this test method is applicable to
naphthas, distillates, oils, and motor fuels: such as gasoline, 3.1 A hydrocarbon sample is either directly injected or
oxygen enriched gasoline (M-85, RFG), diesel, biodiesel and placed in a sample boat. The sample or boat, or both, is inserted
jet fue!. SampIes containing 1.0 to 8000 mg/kg total suIfur can into a high temperature combustion tube where the suIfur is
be analyzed (Note l). oxidized to sulfur dioxide (S02) in an oxygen rich atmosphere.
Water produced during the sample combustion is removed and
NOTE I-Estimates of the pooled limit of quantification (PLOQ) for the sample combustion gases are next exposed to ultraviolet
each ofthe three precision studies were calculated. Values ranged between
(UV) light. The S02 absorbs the energy from the UV light and
less than 1.0 and less than 5.0 mg/kg (see Section 8 and 14.1).
is converted to excited suIfur dioxide (S02 *). The fluorescence
1.3 This test method is applicable for total sulfur determi- emitted from the excited S02 * as it retums to a stable state S02
nation in liquid hydrocarbons containing less than 0.35 % is detected by a photomultiplier tube and the resulting signal is
(mim) halogen(s). a measure of the sulfur contained in the sample. Warning-
lA The values stated in SI units are to be regarded as Exposure to excessive quantities of ultraviolet (UV) light is
standard.
injurious to health. The operator must avoid exposing any part
1.5 This standard does not purport to address al! oj the of their person, especially their eyes, not only to direct UV
saJety concerns, if any, associated with its use. It is the light but also to secondary or scattered radiation that is present.
~ responsibility oj the user oj this standard to establish appro-
priate saJety and health practices and determine the applica- 4. Significance and Use
..... bility oj regulatory limitations prior to use. See 3.1, 6.3, 6A, 4.1 Some process catalysts used in petroleum and chemical
8.1 and Section 7.
refining can be poisoned when trace amounts of suIfur bearing
materials are contained in the feedstocks. This test method can
2. Referenced Documents
be used to determine sulfur in process feeds sulfur in finished
2.1 ASTM Standards:
products, and can also be used for purposes of regulatory
D 1298 Test Method for Density, Relative Density (Specific contro!.
Gravity), or API Gravity of Crude Petroleum and Liquid
Petroleum Products by Hydrometer Method2 5. Apparatus 5
D 4052 Test Method for Density and Relative Density of 5.1 Furnace-An electric fumace held at a temperature
Liquids by Digital Density Meter3 (1075 ::!: 25C) sufficient to pyrolyze all of the sample and
D 4057 Practice for Manual Sampling of Petroleum and oxidize suIfur .to S02'
Petroleum Products3 5.2 Combustion Tube-A quartz combustion tube con-
structed to allow the direct injection of the sample into the
l This test method is under the jurisdiction of ASTM Committee D02 on heated oxidation zone of the fumace or constructed so that the
Petroleum Products and Lubricantsand is the dirGt responsibility of Subcommittee inlet end of the tube is large enough to accommodate a quartz
D02.030C on Electrometric Methods.
Current edition approved Jan. lO, 2000. Published March 2000. Originally
published as D 5453 - 93. Last previous edition D 5453 - 93. 4 Annuat Book or ASTM Standards, Vol 05.04.
2Annuat Book oj ASTM Standards, Vol 05.01. 5 Apparatus manufactured in several variations by Antek Instruments, Inc"
3Annuat Book oj ASTM Standards, Vol 05.02. Houston, TX has been found suitable for this purpose.

Copyright@ASTM. 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
-

~~~ D 5453
sample boat. The combustion tube must have sidc arms for the duction from a microlitre syringe. A drive mechanism which
introduction of oxygen and canier gas. The oxidation section advances and withdraws the sample boat into and out of the
shall be large enough (see Fig. l and Fig. 2) to ensure complete fumace at a contTOlled and repeatable rate is required. See
combustion of the sample. Fig. l and Fig. 2 depict convcn- example, Fig. 4.
tional combustion tubes. Other conflgurations are acceptable if 5.8 RejTigerated CireulatormAn adjustable apparatus ca-
precision is not degraded. pable 01'delivering a coolant material at a constant temperature
5.3 Flow Control- The apparatus mus t be equipped with as low as 4C could be required when using the boat inlet
flow controllers capable of maintaining a constant supply 01' injection method (optional).
oxygen and carrier gas. 5.9 Strip Chart Recurder, (optional).
5.4 Drier Tube- The apparatus must be equipped with a 5.10 Balance, with a precision of::!:: 0.01 mg (optional).
mechanism for the removal of water vapor. The oxidation
reaction produces water vapor which must be eliminated prior 6. Reagents
to measurement by the detector. This can be accomplished with 6.1 Purity uf Reagents Reagent grade chemicals shall be
a membrane drying tube, or a permeation d/J'er, that utilizes a used in tests. Unless otherwise indicated, it is intended that all
selective capillary action for water removal. reagents shall confoll11 to the specilications of the Committee
r 5.5 UV Fluoreseenee Detector-A qualitative and quantita- on Analytical Reagents 01' the American Chemical Society,
tive detector capable 01' measuring light emitted from the where such specilications are available.6 Other grades may be
fluorescence 01'sulfur dioxide by UV light. used, provided it is flrst ascertained that the reagent is 01'
5.6 Mierolitre S)Jringe-A microlitre syringe capable 01' sufficiently high purity to perrnit its use without lessening the
accurately de]ivering 5 to 20-)lL quantities. The needle shall be accuracy 01'the determination.
50 mm (::!::5 mm) long. 6.2 fnert Gas-Argon or helium only, high purity grade
5.7 Sample Inlet System-Either 01' two types of sample (that is, chromatography or zero grade), 99.998 % minimum
inlet systems can be used. purity, moistm'e 5 ppm w/w maximum.
5.7.1 Direet 1l1;jeetion-A direct injection inlet system must 6.3 Oxygen-High purity (that is, chromatography Ol' zero
be capable 01'allowing the quantitative delivery ofthc materia I grade), 99.75 % minimum purity, moisture 5 ppm w/w maxi-
to be analyzed into an inlet carrier stream which directs the mum, dried over molecular sieves. Warning- Vigorously
sample into the oxidation zone at a controlled and repeatable accelerates combustion.
rate. A syringc drive mechanism which discharges the sample 6.4 Toluene, Xylenes, lsooctane, reagent grade (other so]-
from the microlitre syringe at a rate of approximately l )lLls is vents similar to those occurring in samples to be analyzed are
required. See example, Fig. 3. also acceptable). Conection for sulfur contribution from so]-
5.7.2 Boat Inlet System An extended combustion tube vents (solvent blank) used in standard preparation and sample
provides a seal to the inlet ofthe oxidation area and is swept by specimen dilution is required. Alternatively, use 01'a solvent
a canier gas. The system provides an area to position the
sample carrying mechanism (boat) at a retracted position
removed from the fumace. The boat drive mechanism will 6 Reagent Chemlcals. Amerlcan Chemlcal Sodety Spec!ticatlons, American
~ fully insert the boat into the hottest section ofthe fumace inlet. Chemical Socjety, Washington, DC. For suggestions on Ihe testjng of reagents not
The sample boats and combustion tube are constructed of listcd by 1he American Chcmical Socjcty, sec Analar Standards fol' Laboralory
Chemicais, BDH Ltd., Poolc, Dorsct, U.K., and thc Unlted States Pharmacopela
quartz. The combustion tube provides a cooling jacket for the and Hallonal Fonnltlm)J, U.S. Pharmacopcial Convcntion, Inc- (USPC), Rockvil1c,
area in which the retracted boat rests awaiting sample intro- NID.

4S0mm

~- ":'"'" ~
L
/ 25mm
,,""'-~
.~
~
l ~
l
90mm

77[~
~ T
I
- 12m,n
IZmm.-/ /
- 6m:n , lmm ID isee DETAILS,LOW

/ "
CAPILL~RY
riO MUST FIT 12mm S,PTUM

1- I ~/
~, \lJ ~
100mm ..j
Uf 12mm
25mm

BURNER TIP AND SEPTUM DETAJLS

FIG. 1 Direct Inject Quartz Pyrolysis Tube

2
-

~m~ D 5453

rI 59?mm

,- ~ 257mm
6mm OD x 2mr~ ID

~
12r~m

~ 2.';rnm
6rnm OD x 2rnm ID

FIG. 2 Boat Inlet


~
~.~..
5Dmrl --1

~
35mm

J
5Dmm
I

r---.
Su
AJ\t (PuR
A(h
<cif FURNACE

-INLET-CARRIER/OXYGEN MIX

DIRECT INJECT PYROTUBE

SYRINGE

PYRO OXYGEN

~ FIG. 3 Direct Inject Syringe Drive

fUANACE

COOlANT OUT
INlET .cARRIfR
ANO OXYGfN ..X

PYRO OXYGON

BOAT ORIVE

FIG. 4 Boat Inlet System

with nondetectable leve] of sulfur contamination relative to the NOTE 2-A correctiou for chemical impurity cau be required.
sulphur content in the samp]e unknown makes the blank
6.8 Quartz vViJol:
conection unnecessary. 'Warning-Flammable solvents.
6.5 Dibenzothiophene, FW] 84.26, 17.399 % (mim) S (Note 6.9 Suljilr Sto ck Solution, 1000 J.lgS/mL-Prepare a stock
2). solution by accurately weighing approximately 0.5748 g of
6.6 Butyl Suljide, FW146.29, 21.92 % (mim) S (Note 2). dibenzothiophene or 0.4652 g of butyl sulfide ar 0.4]84 g of
6.7 Thionaphthene (Eenzothiophene), FW134.20, 23.90 % thionaphthene into atared 100 mI. volumetric fiask. Di]ute to
(m/m) S (Note 2). volume with selected solvent. This stock can be further diluted

3
~~~ D 5453
to desired sulfur concentration (Notes 3-5). TABLE 2 Sulfur Standards
Curvel Curve II Curve III
NOTE3-Working standards shollld be remixed on a regular basis Sulfur, ng/L Sulfur, ng/L Sulfur, ng/L
depending upon frequency ofuse and age. Typically, stock solutions have
a usefullife of about 3 months. 0.50 5.00 100.00
1.00 25.00 500.00
NOTE4-Calibration standards can be prepared and diluted on a 2.50 50.00 1000.00
mass/mass basis when result calculations are adjusted to accommodate 5.00 100.00
them. 10.00
NCJTE5-Calibration standards fi'om commercial sources Callbe used if Injection Size Injection Size Injection Size
checked for accuracy and if precision is not degraded. 10-20 L 5-1 O L 5 L

6. l O Quality Ccmtroi (QC) Sampies, preferabJy are portions


of one ar more liquid petroleum materials that are stable and
representative of the sampies of interest. These QC sampies 10.2 Flush the microlitre syringe several times with the
can be used to check the validity of the testing proces s as samplc prior to analysis. If bubbles are present in the liquid
described in Section 13. column, flush the syringe and withdraw a new sample.
10.3 A sample size recommended for the curve selected
r 7, Hazards
from Table 2 must be quantitatively measured prior to injection
7.1 High temperature is employed in this test method. Extra into the combustion tube or delivery into the sample boat for
care must be exercised when using f1ammable materials near analysis (Note 6 and Note 7). There are twa alternative
the oxidative pyrolysis furnace. teclmiques availabJe.
8. Sampling Num 6-Injection of a constant or similar sample size for allmaterials
8.1 Obtain a test unit in accordance with Practicc D 4057 or analyzed in a selected operating range promotes consistent combustion
condilions.
Practice D 4177. To preserve volatile components which are in
NOTE 7 Injcction of 10 f!L ol' the 100 ng/pL standard would cstablish
some sampies, do not uncover sampIes any longer than a calibration point cqual to 1000 ng or 1.0 ~lg.
necessary. SampIes sha11be analyzed as soon as possible after
taking from bulk supplies to prevent lass of sulfur or contami- 10.3.1 The volumetric measurement ofthe injected material
nation due to exposure or contact with sample container. can be obtained by fiJling the syringe to the selected leveL
\Varning SampIes that are colIected at temperatures below Retract the plunger so that air is aspirated and the Iower liquid
meniscus faJls on the 10 % scale mark and record the volume
room temperature can undergo expansion and rupture the
container. Far such sampIes, do not filI the container to the top; of liquid in the syringe. After injection, again retract the
leave sufficient air space above the sample to alIow room for plunger so that the Iower Iiquid meniscus t"il11son the ] O%
expansion. scale mark and record the volume of liquid in the syringe. The
8.2 If the test unit is not used immediately, then thoroughly difference between the two volume readings is the volume of
mix in its container prior to taking a test specimen. sample injected (Note 8).
Num 8~--~Anautomatic sarnpling and injection device can be used in
9. Preparatioll or Apparatus place of the described rnanual injection procedurc.
r 9.1 Assemble and leak check apparatus according to manu-
facturer's instructions. 10.3.2 FiJl the syringe as described in 10.3.1. Weigh the
device before and after injection to determine the amount of
9.2 Adjust the apparatus, depending upon the method of
sample introduction, to meet conditions described in Table l. sample injected. This procedure can provide greater accuracy
than the volume dcIivery method, provided a balance with a
9.3 Adjust the instrument sensitivity and baseline stability
precision of ::f:O.OImg is used.
and perform instrument blanking procedUl'es fo11owing manu-
10.4 Once the appropriate sample size has been measured
facturer' s guidelines.
into the microlitre syringe, promptly and quantitatively deliver
10. Calibration and Stalldardizatioll the sample into the apparatus. Again, there are two alternative
techniques available.
10.1 Based on anticipated sulfur concentration, select one of
10.4.1 For direct injection, carefu11y insert the syringe into
the suggested curves outlined in Table 2. Carefu11y prepare a
the inlet of the combustion tube and the syringe clrive. Allow
series of calibration standards accordingly. Make other volu-
metric dilutions of the stock solution to cover the various time for sample residues to be burned from the needle (Needle
Blank). Once a stable baseline has reestablished, promptly start
ranges of operation within these calibration curve guidelines.
the analysis. Remove syringe once the apparatus has retumed
The number of standards used per curve can vary, if equivalent to a stable baseline.
results are obtained.
10.4.2 For the boat inlet, quantitatively discharge the con-
TABLE 1 Typical Operating Conditions tents of the syringe into the boat containing quartz wool at a
Syringe drive (direct inject) drive rate (700-750) 1 ~L!s
sIm\' rate being careful to displace the last drop from tlle
Boat drive (boat inlet) drive rate (700-750) 140-160 mm/min syringe needle. Remove the syringe and promptly start the
Furnace temperature 1075 :': 25C analysis. The instrument baseline shaII ren13in stable untiJ the
Furnace oxygen fiowrneter setting (3.8-4.1) 450-500 mLirnin
Inlet oxygen fiowmeter setting (0.4-0.8) 10-30 mLimin
boat approaches the fumace and vaporization of the sample
tnlet carrier fiowmeter setting (3.4-3.6) 130-160 mLirnin begins. Instrument baseline is to be restabIished before tlle boat
has been completely withdrawn from tlle furnace (Note 9).

4
~~~ D 5453
Once the boat has reached its fully retracted position, allow at using one of the procedures described in 10.2-10.4.
least l min for cooling before the next sample injection (Note 11.3 Inspect the combustion tube and other ilow path
9). components to verify complete oxidation of the test specimen.
NOTE 9-S1owing boat speed or briefly pausing the boat in the furnace 11.3.1 Direet lnjeet Systems Reduce the sample size Ol'the
can be necessary to ensure complete sample combustion. Direct injection rate of injection, Ol' both, of the specimen into the fumace if
can ease sample handling and improve sample combustion characteristics coke or sooting is observed.
for materiaIs containing very volatile sulfur compounds. 11.3.2 Boat Inlet Systems Increase the residence time for
10.4.3 The level of boat cooling required and the onset of the boat in the fumace if coke or soot is observed on the boat.
sulfllr detection tollowing sample injection are directly related Decrease the boat drive introduetion rate Ol'specimen sampIc
to the volatility of the materials analyzed. For volatile materi- size, Ol'both, if eoke or soot is observed on the exit end of the
combustion tube.
aIs, effective cooling of the sample boat prior to sample
injection is essentiaL The use of a refrigerated circulator to 11.3.3 Cleaning and Reealibration Clean any coked or
minimize the vaporization of the sample until the boat begins sooted parts per manufacturer's instructions. After any cIean-
approaching the furnace Ol'an increased time tlx boat cooling ing Ol' adjustment, assemble and leak che ck the apparatus.
I' can be required. Repeat instrument calibration prior to reanalysis of the test
10.5 Calibrate the instrument using one ofthe folIowing two specimen.
techniques. 11.4 To obtain one result, measure each test speeimen
10.5. l Perform measurements for the calibration standards solution three times and calculate the average detector re-
and blank using one of the procedures described in 10.2- l 0.4. sponses.
Measure the calibration standards and blank three times. 11.5 Density values needed for calculations are to be
Subtract the average response ofthe blank injections from each measured using Test Methods D 1298, D 4052 Ol'equivalent, at
caIibration standard response. Then determine the average the temperature at which the sample was tested.
integrated response of each concentration (see 6.4). Construct
12. CalcuIation
a curve plotting of thc average integrated detector response
(v-axis) versus micrograms suIfur injected (x-axis) (Note 10). 12.l For analyzers caIibrated using a standard curve, ealcu-
This curve shall be linear and system performance must be late thesulftlr content of the test specimen in parts per miJJion
checked each day of use. See Section 13. (ppm) as follows:
NOTE l OOthcr calibration curve techniqucs Call be uscd when accu- (I - Y)
( l)
racy and precision are not degraded. Sulfur.ppm (Ilg/g)= S x M x Kg
10.5.2 If the apparatus features self calibration routine, Ol',
measure the calibration standards and blank three times using
one of the procedures desclibed in 10.2-10.4. If blank con-ec- Suifur, ppm (Ilg/g)= (J - y (1(00) (2)
tion is required and is not an available instrument option (see
where:
6.4 or 10.5.1), calibrate the analyzer in accordance with
D = density of test specimen solution, g/mL,
/'"' manufacturer's instructions to yield results expressed as nano-
I = average of integrated detector response tor test
grams of sulfur (Note 10). This curve shall be linear and system
performance must be checked with each day of use (see specimen solution, counts,
Section 13). Kg = gravimetric dilution factor, mass of test specimen!
10.6 1f analyzer calibration is pertom1ed using a different mass of test specimen and solvent, g/g,
Kv - volumetric dilution factor, mass of test specimen!
caIibration curve than Iisted in Table 2, select an injection size
based on the curve closest in concentration to the measured volume of test specimen and solvent, g/mL,
,'1;1 = mass of test specimen solution injected, either
solution(s). Construct the caIibration curve to yield values that
measured directly or calculated from measured
can be used to report sulful' content on a mass/mass basis.
volume injected and density, V X D, g,
11. Procedure s = slope of standard curve, counts//lg S,
V = volume of test specimen solution injected, either
11.1 Obtain a test specimen using the procedure described
measured directly Ol' calculated from measured
in Section 8. The sulfur concentration in the test specimen must
mass injected and density, Al/D, /lL, and
be less than the concentration of the highest standard and y = y-intercept of standard curve, eounts,
greater than the concentration of the lowest standard used in lOOO - factor to convert /lL to mL.
the calibration. If required, a dilution can be perfomled on 12.2 For analyzers eaIibrated using self calibration routine
either a weight Ol' volume basis.
with blank correction, calculate the sulfur in the test specimen
11.1.1 Gravimetrie Dilution (mass/mass) Record the mass in parts per milIion (ppm) as tolIows:
of the test specimen and the total mass of the test specimen and
solvenL. G x 1000
(3)
11.1.2 Volumetrie Dilution (mass/volume) Record the Sulfur.ppm (Ilg!g)= M x Kg
mass of the test specimen and the total volume of the test or,
specimen and solvenL. G x 1000
11.2 Measure the response for the test specimen solution SulflIr, ppm (Ilg/g) = - (4)

5
~~~ D 5453
where: under constant operating conditions on identical test materia l
D density of test specimen solution, mg/IlL (neat would, in the long run, in the nOffi1aland correct operation of
injection), or concentration of solution, mg/~lL the test method, exceed the following values in only 1 case in
(volumetric dilute injection), 20, where x = the average of the two test results.
Kg - gravimetric dilution factor, mass of test specimen/ Less than 400 mg/kg
mass of test specimen and solvent, g/g, r 0.1788 X(075)
lvI (5)
= mass of test specimen solution injected, either
measured directly or calculated from measured Greater than 400 mg/kg
volume injected and density, V X D, mg, r = 0.02902X (6)
v - volume of test specimen solution injected, either
14.1.2 Reproducibilzty-.The difference between two single
measUl'ed directly or caIculated from measured
and independent resuIts obtained by different operators work-
mass injected and density, MlD, ~lL,
ing in different laboratories on identical test material would, in
G - sulfur found in test specimen, Ilg, and
1000 = factor to convert ~lg/mg to Ilg/g. the long run, in the nonnal and correct operation of the test
method, exceed the following values in only l case in 20,
r 13. Quality Control where x = the average of the two test results.
13.1 Confinn the perfonnance of the instrument or the test Less than 400 mg/kg
procedure by analyzing a quality control (QC) sample (6.10) R =, 0.5797 X(O.7SJ (7)
after each calibration and at least each day of use thereafter Greater than 400 mg/kg
(see 10.5).
R ,= 0.1267 X (8)
13.1.1 When QC/Quality Assurance (QA) protocols are
already established in the testing facility, these can be used 14.2 Bias.--The bias of this test method was determined in
when they confinn the reliability of the test result. a 1992 research report,7 by analysis of standard reference
13.1.2 When there is no QC/QA protocol established in the materials (SRMs) containing known levels of sulfur in hydro-
testing facility, Appendix Xl can be used as the QC/QA carbon. This report indicated that the results obtained on the
system. SRMs were witI1in the repeatability of the test method.
14.3 Examples ofthe above precision estimates for sampies
14. Precision
containing less than 400 mg/kg are shown in Table 3.
14.1 The test method was examined in three separate
inter-1aboratory studies.7 The precision of the test method, as 15. Keywords
obtained by statistical analysis of test resuIts, is as follows 15.1 analysis; diesel; ftuorescence; gasoline; jet fue!; kero-
(Note II). sene; M-SS; RPG; sulfur; uItraviolet
NOTE11 Volatile matcrials can causc a dctcrioration in prccision
whcn not handlcd with care (see Section 8 and 10.4). TABLE 3 Repeatability (tj and Reproducibility (R)
Concentration
14.1.1 Repeatability The difference between two test re- R
~ (mg/kg S)
suIts obtained by the same operator with the same apparatus 1 0.2 0.6
5 0.6 1.9
10 1.0 3.3
7 Supp0I1ing data available trom ASTM Headqum1ers. Request RR:D02-1307 50 3.4 10.9
(1992) original with multiple matrices; RR:D02-1456 (1999) lJVF/X-ray equiva- 100 5.7 18.3
lence study; RR:D02-1465 (1997) gasoline and RFG only; RR:D02-1475 (1998) 400 16.0 51.9
low level gasoline, diesel, and biodiesel.

APPENDIXES

(Nonmandatory lnformation)

Xl. QUALlTY CONTROL

Xl.I Confirm the performance of the instrument or the test controllimits of the QC sample (see Test Method D 6299 and
procedure by analyzing a quality control (QC) sample. MNL 7).8

Xl.2 Prior to monitoring the measurement process, the user XI.3 Record the QC resuIts and analyze by control charts
or other statistically equivalent techniques to ascertain the
of the test method needs to determine the average value and
statistical control status of the total testing process (see Test

, ASTM MN!' 7, Manual on Presentation ofData C,mlrol Chart Analysis, 6th ed.,
available from ASTM Headquarters.

6
~t D 5453

Method D 6299 and MNL 7). Any out-of-control data should when it is demonstrated that the test ing is lInder statistical
trigger investigation for root cause(s). The results of this control, the QC testing frequency may be reduced. The QC
investigation may, but not necessarily. result in instrument sample precision should be checked against the ASTM test
re-calibration. method precision to ensure data quality.

X1.4 In the absence of explicit requirements given in the Xl.5 It is recommended that, if possible, the type of QC
test method, the frequency of QC testing is dependent on the sample that is reglllarly tested be representative of the material
criticality of the quality be ing measured, the demonstrated routinely analyzed. An ample supply of QC sample material
stability of the testing process, and customer requirements. should be available for the intended period ofuse, and must be
Generally, a QC sample is analyzed each testing day with homogeneous and stable under the anticipated storage condi-
routine sampies. The QC trequency should be increased if a tions. See Test Method D 6299 and MNL 7 for further guidance
large number of sampIes are routinely analyzed. However, on QC and control charting techniques.

X2. IMPORTANT FACTORS IN DTRECT IN.JECTTON ANALYSIS OF HYDROCARBONS USING TEST METTTOD D-5453
f'. (SULFUR)

X2.l FurlUlce Temperature. A temperature of 1075 ::i:: X2.7 Flow Path, Leak Check, and Back Pressure The
25C is required for sulfur. The llse of qualiz chips in the sample f10w path must be leak free when pressure tested in
combustion zone of the pyrotube is required. accordance with the manufacturers recommended procedure
(2-3 psi). Flow path back pressure during nonnal operation can
X2.2 Heedle Tip Position during Injection- The needle tip
range from 0.75 to 2.00 psi.
should be presented fully into the hottest part of the inlet area
of the furnace. Assembly of apparatus to manufacturer's X2.8 Gas Flow Settings-Gas supplies to various points in
specification and full insertion of the needle will ensure this. the sample path must be consistently controlled to allow for
X2.3 Injection PeaklHeedle Blank-Avoid integration of smooth, complete combustion of the sample. See Table X2.1.
any baseline upset caused by the needle penetration of the
septum. After the sample specimen has been measured into the X2.9 lvlembrane Dryer Purge Water produced during the
syringe, retract the plunger to form an air gap up to approxi- combustion of the sample is removed by the membrane dryer.
mately the 10 % scale mark of the syringe barrel. Insert the This water must then be purged from the membrane dryer. For
syringe needle into the injection inlet and allow the needle/ an apparatus that utilizes a desiccant scrubber (flow recycle) to
septum blank to dissipate. Reset the instrument baseline or provide the membrane dryer purge gas, replace the drying
enable integration, if required, prior to the injection of the agent when color change (blue to pink) indicates. When an
syringe contents. auxiliary gas flow is used, set membrane dryer purge flow at
200 to 250 mL/min.
X2.4 Residence Time 0/ Heedle in Furnace-Residence
~ time of the needle in the fumace m,ust be consistent foJlowing X2.10 Sample Homogeneity/Calibration Response-Prior
the injection of the sample. For direct injections it is recom- to analysis, mix sampies and calibration materials well. Mini-
mended that the needle remain in the furnace until the mum detector response; (Model 7000) should be no less than
instnunent retums to baseline and the analysis of the injected 2000 to 3000 counts, (Model 9000) should be no less than 200
material is complete. to 300 counts or three times baseline noise, for the lowest point
on the calibration curve. The highest point on the curve is
X2.5 Injection Size As a general rule, larger sample sizes
below the saturation point of the detector; use a maximum
are required for measurement of lower levels of sulfur. While
response of 350000 to 450000 counts (Model 7000) as a
detennining the best sample size, frequently check for evi-
guideline. The Model 9000 should not have flat-top peaks.
dence ofincomplete combustion (sooting) that may be present Adjust Gain Factor, PMT voltage or sample size, or both,
in the sample path. Control sooting by slowing the injection
accordingly.
rate of the sample from the syringe, or increasing the pyro-
oxygen or inlet oxygen supply, or a combinatioll thereof. X2.ll Baseline Stability-Prior to analysis, especiallywhen
Suggested sample sizes are as follows: analyzing low levels, be celiain that the detector baselines are
Traceto 5 mg/kg 10to 20 ~L stable and noise free. For a given gai n factor, photomultiplier
5 ppmto 100mg/kg 5 to 10~L
100mg/kgto % 5 ~L
X2.6 Injection Rate and Frequency Discharge contents of
TABLEX2.1 Gas Flow Settings-Direct Injection Analysis

the syringe into the fllmace at a slow rate, approximately one TypicalGasFlows FlowmeterBali MFC
ftL/s (Model 735 Sample Drive rate of 700 to 750). Frequency Inlet carrier flowmeter settingsA 3.4-3.6 140-160 mUmin
Inlet oxygen flowmeter setting 0.4-0.6 10-20 mUmin
of injection can vary depending upon sample and syringe F urnace oxygen flowmeter setting 3.8-4.1 450-500 mUmin
handling techniques, rate of injection and needle in furnace Ozone generator flowmeter settingB 1.5-1.7 35-45 mUmin
residence time. Typical injection frequency allows at least 3.5 AHelium or argon may be used as a carrier gas.
min between injections. BFlow to ozone generator (optional).

7
~m~ D 5453
tube voltage may be adjusted to ensure maximum sensitivity impurity of source materia!. U se calibration curves that bracket
while maintaining a stable, noise-free baseline. Model 9000 the expected levels in the sample unknown. Do not force the
users can utilize the baseline evaluate and peak threshold calibration curve through the 0,0 axis, unnecessarily. Construct
functions to reduce baseline noise. standard coneentrations that will yield a calibration curve that
X2.l2 Calibration Materials/Standard Curve is linear and that does not exceed the dynamie range of the
Construction ..Prepare calibration standard s with solvent ma- detector [use a couelation coefficient of .999 and l to 2 orders
terials that have minimum or no sulful' contamination relative of magnitude (example: 5 to 100 mg/kg) as a guideline). The
to the concentration anticipated in the samp]e unknown. curve should yield an estimated value that can be used to
Couect for sultur contribution trom solvent materia]s and calculate content in the sample on a mass/mass basis.

X3. IMPORTANT FACTORS IN BOAT-INLET ANALYSIS OF HYDROCARBONS USING TEST METHOD D-5453 (SUI_FUR)

X3.1 Furnace Temperature A temperatlll'e of 1075 :t:: X3.7 Sample Flow Path: Leak Check and Back Pressure
0. 25C is required for sulfur. The use of quartz chips in the The sampIc flow path must be leak free when pressure tested in
combustion zone of the pyrotube is required. accordance with the manufacturer' s recommended proeedure
X3.2 Boat Path- The boat shou]d be presented fully into (2 to 3 psi). Flow path back pressure during nonnal operation
the inlet area of the furnace. Assembly of the apparatus to the can range from 0.75 to 2.00 PSI, for non-atmospheric-vent
manufacturer's specification ensures this. systems.

X3.3 Boat Entry Rate and Residence Tzme ()f Sample in X3.8 Gas Flow Settings ... Gas supplies to various points in
Furnace-lnsel1 the boatinto the fumace using a drive rate of the sampIc path must be consistently eontrolled to allow for
140 to 160 mm/min (Model 735 setting of 700-750). Addi- smooth, complete combustion of the sampIc.
tional slowing of boat speed Ol'a brief pause of the boat in the X3.9Membrane Dryer Purge-Water produced during the
fumace may be nec essary to ensure complete samp]e combus- combustion of the sampIc is removed by the membrane dryer.
tion. The boat should emerge from the furnace soon after This water must then be purged. For an apparatus that utilizes
detection is comp]ete. Boat in fumace residence times can vary a desiccant scrubber (flow recycJe) to provide the membrane
depending on sample volatility and levels of element mea- dryer purge gas, replace the drying agent when color change
sureLLTypical boat in fitmace residence times range between (blue to pink) indicates. When an auxiliary gas flow is used, set
15 to 60 s.
membrane wyer purge flow at 200 to 250 mLlmin.
X3.4 lnjection Size-As a general rule larger samp]e sizes
X3.l0 Sample Homogeneity/Calibration Response Prior
may be required for measurement of lower concentration
to analysis, mix sampIes and calibration materials welI. Mini-
levels. While determining the best sampIc size, frequently mum detector response; (Model 7000) should be no less than
cheek for evidence of incomplete combustion (sooting) that 2000 to 3000 counts, (Model 9000) should be no less than 200
~. may be present in the sample path. Control sooting by slowing to 300 counts or three times baseline noise, for the lowest point
boat speed into the fumace, increasing the length of time the on the calibration curve. The highest point on the eurve is
boat is in the fumace Ol'increasing the pyro-oxygen supply, Ol' below the saturation point of the (Model 9000) detector; use a
both. Suggested sampIc sizes are as follows:
maximum response of 350 000 to 450 000 eounts (Model
Trace to 5 mg/kg 10 to 20 ~L
5 ppm to 100 mg/kg 5 to 10 ~L
7000) as a guideline. Adjust Gain Factor, PMT Voltage and/ol'
100 mg/kg to % 5 ~L sampIc size accordingly.
X3.5 lnjection Rate and Frequencyu Discharge contents of X3.ll Boat Blank/Baseline Stabilit;y-Prior to analysis, es-
the syringe into the boat at a slow rate (approximately l ~tLls) pecially when analyzing low ]evels, advance the empty boat
being eareful to discharge the last drop. Use of quartz woal in into furnace to ensure that no contamination is present in the
the sampic boat to aid quantitative delivery is advised. Fre- boat ar on the inside areas of the pyrotube near the injection
quency of injection can vary depending upon boat speed, level area. Reat empty boat in the furnaee to ensure that boat is
of sulfur being determincd, fumaee residence time, and cooling dean, then rapidly move boat out to injection area.
capaeity of the boat loading area. Typical injection frequency
NOTl" X3.l-..Ifthe hot boa t being retumed to tbe injection area eauses
allows at least 2.5 min between injections. baseline upset, repeat the boat in and out cycle, until no sui fur is
X3.6 Boat Temperature at Time of Sample lntroduction
SampIc volatility must be addressed; ensure boat temperature TABLE X3.1 Gas Flow Settings-Boat Inlet Analysis

has retumed to ambient or sub-ambient temperatures prior to --.-... Typical Gas Flows Flowmeter Bali MFC
introduetion of sampIc into boa1. Let boat rest at least 60 s in Inlet carrier ftowmeter settingsA 3.4-3.6 130-160 mUmin
Inlet oxygen fiowmeter setting 0.4-0.6 10-20 mUmin
coolant jacket or cooling area between injections. Some sulfur 3.8-4.1 450-500 mUmin
Furnace oxygen fiowmeter setting
may be measured as the sampIc evaporates when the boat Ozone generator fiowmeter settingB 1.5-1.7 35-45 mUmin
approaches the fumace. Sub-ambient temperature can reduce AHelium or argon may be used as a carrier gas.
this evaporation. BFlow to ozone generator (optional).

8
~~J~ D 5453
measured. For a given gain factor, photomultipJier tube voitage, can be impurity of sulfur source material. Use calibration curves that
adjusted to ensure l11axil11ul11sensitivity while l11aintaining a stable, bracket the expected levels in the sample unknown. Do not
noise-free baseline. Model 9000 users can utilize the baseline evaluate and
force the calibration curvc through thc 0,0 axis, unnccessarily.
peak threshold functions to reduce baseline noise
Construct standard concentrations that will yield a caIibration
X3.l2 Calibration Materials/Standard Curve curvc that is lincar and that does not cxceed the dynamic rangc
Construction Prepare calibration standards with solvent ma- of the detector (use a conelation coefficient of .999 and l to 2
terials that have minimum or no sulfiIr contamination relative orders ofmagnitude (example: 5 to 100 mg/kg) as a guideline).
to the concentration anticipated in the sample unknown. The curve should yield an estimated value that can be used to
Correct for sulfur contribution from solvent materials and calculate content in the sample on a mass/mass basis.

SUMMARY G.F CHAN GES

Sub-Committee D02.03 has identified the location of selected changes to this standard since thc last issue
~ (D 5453 93) that may impact the use of this standard.

(a) Updated and cxpanded applicable materials described in (c) Inscrtcd quality contral statements in Sections 6 and 12.
Section ]. (d) Updated precision statements to inc1ude two recent inter-
(h) Modified referenced documcnts in Section 2. laboratory studies, Section 14.

The American Society for Testing and Materiais takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are express/y advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entire/y their own responsibility.

This standard is subject to revision at any time by the responsib/e technica/ committee and must be reviewed every five years and
if not revised. either reapproved or withdrawn. Your comments are invited either for revision ofthis standard or for additiona/ standards
and shou/d be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsib/e
technical committee, which you mayattend. If you feel that your comments have not received a fair hearing you shou/d make your
views known to the ASTM Committee on Standards, at the address shown be/ow.

This standard is copynghted by ASTM. 100 Barr Harbor Dnve, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Individua/ reprints (sing/e or multip/e copies) of this standard may be obtained by contacting ASTM at the above address or at
610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website (www.astm.org).

Вам также может понравиться