Академический Документы
Профессиональный Документы
Культура Документы
Object
Qualitative use. With the use of standards and knowledge of sample source,
analyte identification can be made by retention times. With a scanning UV,
electrochemical, or mass spectroscopic detector analytes can be identified by
spectral interpretation.
Sensitivity. Major and trace species can be determined. Sensitivity varies with
type of detector used and analyte assayed. Detection limits range from micro- to
subnano- grams.
Introduction
Capsaicinoids are the group of compounds that give chili peppers their
characteristic, pungent taste. More novel uses of capsaicinoids include herpes
treatment, tear gas (pepper spray), squirrel-resistant bird seed (birds do not have
capsaicin receptors, but squirrels do), and as over-the-counter arthritis creams.
The compounds responsible for the pungent taste of the red pepper, Capsicum
annuum, are listed below. Capsaicin is N-[(4-hydroxy-3-methoxyphenyl)methyl]-
8-methyl-6-nonenamide.
2 HPLC
Columns. Both HPLC and GC columns are similar, although HPLC columns are
packed more densely and are difficult to manufacture whereas packed GC
columns are easily homemade. Both have stationary liquid phases coated or
bonded onto a solid support that should be immiscible with the mobile phase.
HPLC columns are not usually in thermostated ovens for temperature
programming. Temperature programming in HPLC is not common because of
the difficulty of heating or cooling liquids rapidly plus the fact that liquid-liquid
distribution coefficients are not highly temperature dependent. Columns might
be thermostated isothermally however to reduce any chance temperature effects
on the separation of interest. The lack of temperature control to affect
separations is more than adequately compensated by the range of distribution
coefficients available using a variety of stationary phases. In this experiment,
you will use a C18 reversed-phase column. The stationary phase is
octadecylsilane (C18SiH2) bonded to a 5 m spherical silica gel support. The C18
column is a non-polar column, and, in a polar mobile phase, polar analytes will
tend to elute faster than non-polar substances. The descriptor of reversed
phase originates from the fact that most columns are made with polar stationary
phases which means polar materials elute after the non-polar. Here in HPLC as
in GC, the basis the separation is based on the magnitude of the distribution
coefficient describing the partitioning of the analyte between the mobile and
stationary phase.
Detectors. Once the column has provided a separation, the analyte still must be
detected. The three types of detectors most often used are the index of refraction
(RI) detector, the UV/VIS detector, and the fluorescence detector. The first is
non-specific in that the method is based on the observation that the refractive
Chemistry 112, Spring 2014
4 HPLC
index of the mobile phase is changed when an analyte is dissolved in it. The RI
detector is as general as the thermal conductivity detector is in GC. A UV/VIS
detector can be a simple single-beam, single wavelength photometer or a
sophisticated spectrophotometer. In fact, practically any other instrument can be
connected in-line with the HPLC and used as a detector. A difficulty is, however,
that the mobile phase usually has to be removed. Modern, hyphenated (tandem)
techniques include LC-MS and LC-NMR.
Microprocessor control. Modern HPLC instruments control the injector and the
solvent pump with a computer. Software is used to program the computer. If the
mobile phase has a constant composition, it is called isocratic or of constant
composition. Most HPLC instruments allow gradients in the mobile phase to be
generated as functions of time so that the elution can be made with any solvent
combination desired (at least up to mixtures of three solvents). The HP 1090 can
program changes, that is, gradients, in the composition of the mobile phase
during elution. Gradient elution offers to HPLC analyses the advantages that
temperature programming offers to GC analyses.
Methods
Data analysis. Here the parameters to control data processing are set in the
subsections: Signal Details; Integration Events; Peak Identification; Peak
Quantification; Calibration; and Report.
Run time check list. The check list provides which and in which order the
instructions of the method will execute. The check list can: acquire, store and
process data for a report; execute only a portion of the method; acquire data
without analysis; reanalyze existing data; invoke user written macros for data
analysis.
The condensed instructions given below will take you through the software
procedures necessary to start collecting data. The procedures necessary to
construct calibration tables and curves or determine peak purities are outlined
below also but details will found in the printed How to and Help files you will
find in the binder by the instrument.
Apparatus/chemicals
The HP 1090 will be found on a cart in Keck 2334. Glassware for solution
making will be available by the instrument. Syringe filters for use in filling the
sampling vials will be available. Sampling vials are heavy walled glass
containers about 2 dram in size and capable of being capped with a metal/rubber
serum cap. Vials and the special capping tool will be located with the other
glassware.
Experiment
Column choice (actually you will not have choice here but, in practice,
yes.)
Column temperature
Flow rate
Injection amount
Detection wavelength: fixed or variable with analyte peak
Solvent composition: isocratic or gradient elution
Maximum and minimum pressure
Stop and post-analysis times
Locate a printed copy of the method analges.M appended to this experiment and
discuss it with your instructor.
Chemistry 112, Spring 2014
6 HPLC
Please remember to pass each solution through a 0.2 micron syringe filter
when filling the sample vial with that solution.
All methods and data files should be stored in the c112 directory.
Wear latex gloves for handling all samples! Add 2-5 g of accurately-
weighed hot sauce, 2 g of arthritis cream, or 1 g of dehydrated, crushed pepper to
exactly 10 mL of methanol and mix thoroughly for two min. Allow the sample to
settle, then remove exactly 1mL of the methanol and add to approximately 9 mL of
water.
Aqueous solutions may be poured down the sink with water rinse.
Organic solutions should be disposed of in the appropriate waste container.
Glassware can be washed with soap and water and air dried. Capped vials
should be uncapped, solutions disposed of, and empty vials discarded in a waste
glass bin. Used metal-rubber septum caps can go in normal trash cans. SPE
cartridges should be returned to the counter.
Data treatment/calculations
References
Skoog, D.A.; Holler, F.J.; Nieman, T.A. Principles of Instrumental Analysis, 5th Ed.;
Harcourt Brace & Company: Philadelphia, 1998; Chapter 28.
Batchelor, James D.; Jones, Bradley T. Determination of the Scoville Heat Value
for Hot Sauces and Chilies: An HPLC Experiment. J. Chem. Educ., 2000, 77,
266-267.
Huang, Jiping; Mabury, Scott A.; Sagebiel, John C. Hot Chili Peppers: Extraction,
Cleanup, and Measurement. J. Chem. Educ., 2000, 77, 1630-1631.
Addenda
The next several pages are either brief instructions for using methods or copies of
various outputs from the Chemstation software. In order of appearance:
Note: These instructions assume that you are running a preprogrammed METHOD from
disk.
Startup
Turn on the helium cylinder to 50 psi. Turn on the nitrogen cylinder to 65 psi. Open
the valve on the back-left of the HP-1090 to purge the solutions with He; after
purging for 5 min close the valve on the back of the HP-1090.
Turn on the main power to the HP-1090. The switch is underneath the helium purge
valve on the back left of the instrument. Turn on the front panel power by pressing
the ON button located at the bottom right front of the instrument.
Turn on the HP-5 laser printer. The switch in on the lower right front. Turn on the HP-
Vectra computer. The switch is on the center front. Turn on the monitor. The
switch is on the lower center front.
After Windows 95 starts, double-click on the HP-1090 HPLC icon to launch the
acquisition and analysis software.
- To run another single sample, return to the METHOD & RUN CONTROL menu
under VIEW, and click on START.
Shut-down
From the INSTRUMENT menu, select SYSTEM OFF. From the Windows 95 START
button, select SHUT DOWN, then SHUT DOWN THE COMPTUER. Turn off the
main power on the back pannel of the HPLC. Turn off the gas cylinders.
Method Analges.M
Below is a partial printout of analges.M, a method for the separation of analgesics in water/methanol
mixtures on a 100 x 2.1 mm Hypersil ODS 5 microM column with single wavelength monitoring at 225 nm;
spectra acquired from 200-400 nm at 2 nm resolution. Method first written by H. Van Ryswyk, 31 Dec 96,
modified 31 Mar 200 by G.R. Van Hecke
LC 1090
Pump (DR5):
Stop time 4.00 min
Post time 3.00 min
Flow 0.500 mL/min
Store Flow time Yes
Min. Pressure 0 bar
Max. Pressure 350 bar
Store Pressure Curve Yes
Oven Temperature Off
Store Temperature Curve Yes
Solvent A 78.0% (bottle A) water
Store Solvent A Curve Yes
Solvent B 22.0% (bottle B) methanol
Store Solvent B Curve Yes
Solvent C Off (bottle C) 20/80 water/methanol
Store Solvent C Curve Yes
Injector:
Injection Volume 4.0 L
Draw Speed 83.8 L/min
Contacts:
Contact 1 Off
Contact 2 Off
Contact 3 Off
Contact 4 Off
Time Table: empty