Академический Документы
Профессиональный Документы
Культура Документы
In recent years, Glove Integrity Testing has been profoundly spores diameter), but the practical answer is much larger due
discussed and numerous articles have been written on this to technological limitations at this time. The methodologies
topic. However, many doubts still remain within the industry of testing available today are many, but few of them may be
due to the lack of norms and clear guidance available. In 2008 automated and many times they are not suitable for the routine.
Annex 1 to EU Guidelines to Good Manufacturing Practice Moreover today it is difficult to find equipment to test the
reiterated the importance of having glove integrity testing: integrity of gloves that comes with performance qualification
25. Monitoring should be carried out routinely and should (PQ) and this is mainly due to the lack of test repeatability.
include frequent leak testing of the isolator and glove/sleeve Although there are many different methods available in the
system. industry, the method used by the greatest portion of the users
Over the course of the past few years different industry to test the integrity of the gloves is the Positive Pressure Decay
specific associations have conducted statistical studies about method which follows the international standard ISO 14644-7
the frequency in which this type of testing is performed in Annex E.5.
the pharma industry and the results indicated that the test The ISO 14644 Part 7 doesnt detail specific methodologies for
is executed normally prior to each batch of a production run. the Positive Pressure Decay. In terms of acceptance criteria
However as far as performance requirements the question is many different interpretations can be reached. The box below
still open: which hole diameter should I be able to detect? shows what the ISO norm states about the test procedure and
The theoretical answer would probably be 1m (bacteria the results.
With experience, operators will be able to identify potential problems in a 10 s test period. Suspect gloves/gauntlets should be re-tested, and a longer test period
may be required to confirm results.
E.5.3.3 Results
E.5.3.3.1 Pass
If the glove/gauntlet is sound, then the reading shown on the manometer will remain static within 2 Pa to 10 Pa, subject to the small variations noted in E.5.3.2.
E.5.3.3.2 Fail
If the glove/gauntlet is damaged, then the reading shown on the manometer will fall (i.e. 500 Pa, 495 Pa, 490 Pa). This trend will be distinct and progressive. The rate
of change will be proportional to the level of damage to the integrity of the glove.
Any test showing probable damage should be repeated.
Any tests that record a distinct change in pressure should be closely investigated and the fault (e.g. incorrectly positioned cuff ring, damaged glove) either re-tested
or the suspect glove/gauntlet changed and a successful test conducted.
COMECER SPA | All rights reserved | 22/01/14
Many questions can arise from such a simple test procedure The glove materials available in the industry vary (Hypalon,
and such a wide interpretation of the results. Butadyl, Neoprene, EPDM). In particular, Hypalon is a mixture
What is the performance of this method? Which hole diameters of polymers such as plastic and rubber, with a very different
will I be able to detect? reaction to stress and elasticity. These gloves, very common in
The first step is to create a mathematical model which can aseptic applications, result in more of a challenge to integrity
describe an inflated glove with an orifice, as this will provide you testing than other types of materials. Thickness and size,
with a formula relating to the hole diameter for the expected together with the flange mounting, greatly contribute to the
pressure drop. Unfortunately the calculation is very complex determination of a customized testing procedure for each kind
due to different elastic properties of the glove material (different of glove material. Therefore an experimental approach needs to
elastomers, thickness, and sizes) so different strategies need to be strictly followed.
be followed.
Isolation Technology | White Paper
For the positive pressure decay method, performance tests (< 1000 Pa) for the hole whose performance we are trying to
may be carried out in different ways, such as calibrated hole prove and collect a statistical set of data from these tests4.
on the flange, needle inserted through the glove or pinholes on Definition of test parameters and acceptance criteria:
the glove1. Based on our experience, the calibrated hole method
From the two sets of curves the following test parameters are
is the most appropriate and precise method of the three listed
then defined:
above, so this configuration,
Inflate Pressure (between 500 and 1000Pa)
together with the Automatic
Stabilization Time
Glove Tester2, was used
Testing Time
for this study. After testing
Acceptance Criteria (P limit)
gloves of different sizes and
COMPRESSED AIR
materials currently available
The objective is to locate the value of these paremeters so that
in the market, we were
they will correspond to a significant difference between the
able to collect a significant
Pressure Drop of the Open System (Pos) and the Pressure
amount of data to support
Drop of the Closed System (Pcs), i.e. Pcs - Pos . The
the methodology described
statistical deviation of the collected data needs to be as small
below that is based on the
as possible and the accuracy of the pressure transmitters
comparison between these COMPRESSED AIR
must be considered in order to reach an effective and valuable
two systems:
acceptance criteria. Basically what we look for is a large
Closed System - brand new
difference between the slopes of the two curves based on the
glove correctly mounted on
collected data of the two systems during testing: a smaller
the flange.
difference between the two curves may result in a longer overall
Open System - brand new
test procedure time until a significant difference is observed.
glove correctly mounted
COMPRESSED AIR As such, the equipment and the test can be easily validated
with a calibrated hole device
while the specific glove and fixing mechanism (flange and
connected.
counterflange, o-ring, etc.) is characterized in terms of normal
The expected behavior of the two systems should be like the
leakage (Closed System curve). Before running the test, it is
one shown in the graph below.
important that the integrity of the machine running the test is
also verified and assured.
CLOSED SYSTEM With the method described above a variety of hole diameters
can be detected developing the correct test parameters.
However the detection limits are primarily influenced by the
precision of the adopted transmitters, statistics involved and
time constraints.
R
COMECER S.p.A. - Via Maestri del Lavoro, 90 - 48014 Castel Bolognese (RA) - Italy
Phone +39 0546656375 Fax +39 0546656353 E-mail: comecer@comecer.com http://www.comecer.com