Академический Документы
Профессиональный Документы
Культура Документы
Institute for Materials Research, National Bureau of Standards, Washington, D.C. 20234
(November 1, 1968)
A method is presented for measuring gaseous diffusion coefficients usin g a gas chromatographi c
technique. Diffu sion coefficients were measured for the sys tems: argon, krypton , oxy ge n , and nitroge n
d iffu sing into helium at temperatures from 77 to 400 K.
Key word s: Argon; diffusion coe fficient; gas c hromatography; helium ; krypton ; nitro ge n; oxygen.
Equation (3) defines a concentration distribution an d the small dimensionless parameter E defined by
which is instantaneously Gaussian and spreading
with tim e. In the actual system the assumption of s = Evto . (7c)
co nstant flow parameters applies only to the diffusion
When eq (6) is rewritten in terms of E and YJ, the result is ?:
column proper. However, the experimental system
was designed such that the concen tration distribution
was always instantaneously Gaussian within the dif (8)
fusion column. Equation (3) can be applied under
the se condition s if the zero of time is redefined.
Instead of taking t = 0 as the instant of sample injec Since E is small (typically, E2 < 10 - :)), eq (8) can be
tion, it must be so defined that eq (3) is consistent expanded in powers of E. Neglecting powers higher
with an a nalysis of the peaks recorded from two or than the second,
more separate detectors along the tube.
Equation (3) expresses conce ntration in terms of a
frame of reference that moves with the mean velocity
of the carrier gas stream. In order to express concentra
tion as a function of time at a fixed detector, we must This equation is quadratic in YJ. Let t:.YJ be the differ-
transform eq (3) by means of the relation
two roots. E xpan d'mg In h
ence 0 f Its ho as a power senes
.
x = veto - t), (4) in the solution of eq (8), correct through terms in
E2 , ~
E 2, can be written
where to is that time at which the centroid of the
dispersion arrives at the detector. When we make
this substitution, the concentration at a fixed detector (9)
is give n by
Taking typical values for E and t:.YJ (typically ,t:.YJ 2< 10),
(5) it can be readily seen that all but the first term on the
right side of this equation can be neglected when
accuracy to a few parts per thousand is adequate.
2.2. Peak Analysis If we let t:.t represent the width of the recorder trace
at height h, we can make use of eqs (7a) and (7b) to
Assuming linear detector res ponse , the height h of a write
recorder trace of the signal at time t is proportional to c
as expressed by eq (5). If ho is the hei ght at time to, it
follows from eq (5) that
v2 (t:.t)2
From this e quation it is seen that a straight line with to In hho = 40'2 "2 . (10)
the slope v 2 j4O' 2 should be obtained when
t ( In -
h 2
1 In-
ho+ - to)
t
If to In ~o is plotted versus (~t r, it is seen that a
straight line with the slope v 2 j4O' 2 should be obtained.
This plot should superimpose accurately on that based
is plotted versus (to - t)2. Such a plot is useful for on eq (6). It is obvious that e q (10) is easier to apply in
demonstratin g that the peak shape conforms to theory. determining a.
However, for determining th e diffusion coefficient an Since to, the time of arrival of the peak at a detector,
approximation based on eq (6) is easier to apply while appears on the left in eq (10), it is still necessary to
providing sufficient accuracy. define the zero of time. The difference of the to values
208
at two de tec tors along t he tube can be meas ured di- r2':?'277ZZZZZZZ/l
rec tly_ Their ratio can be ob tain ed by applying eq (10) GA S FLOW - r - T T " T / 7 7 7 " " " / l
l
determin e the te mperature of the carrier gas , the
type 9225 filament support (see footnote 2) to be sol- column's dime nsions bein g corrected for thermal
dered into position so that the sensing element was on expansion_ If the carrier gas was not at the tempera-
the ce ntral axis of the diffusion column. The two ture of the thermostat the flow rate was reduced until
it was.
A preliminary diffusion measurement was made for
each system to determine peak heights. The atte nua-
tion was then adjusted by means of a 10-turn, 10 ,000 n
I1...- Heli-pot so that for all the followin g meas ure me nts
-. the peak heights would be the maximum height al-
lowed by the recorde r chart pape r. Duplicate run s were
made by injectin g s uccessive amounts of sample gas
into the carrie r gas stream, th e injecti ons being at a
time inte rval s uc h th at no overlapping of peaks oc-
curred. At liquid nitroge n temperature, it was possible
for as many as four peaks to be in the diffusion column
simultaneously_ This enables four velocity measure-
s ments to be made during the course of a run. The time
2 Certai n commercial mat e rials and equipm e nt are iden tified in thi s paper in order 10
specify a dequa te ly the experimental procedure. In no c ase doe s such ide ntifi cati o," ~mpl y
FIG URE 1. Schematic diagram of flo w system for measuring diffusion reco mm e nd ation or endorsement by the National Bure au of Standards, nor does It Impl y
coefficients. that the ma terial or eq uipme nt id e ntifi ed is necessarily the best available for th e purpose.
209
between peaks was c hecked with a stop watch. The There are no indications of skewing or tailing of either
heights and widths of the peaks were measured to a peak.
precision of one part in two hundred fifty with an Values for the diffusion coefficients at different
engineer's scale. temperatures are given in table 1. The pressure of the
The quantity 0: 2 was calculated from the expression carrier gas in the diffusion column varied from 1.1 to
1.6 atm for these determinations. These values have
been corrected to standard atmospheric pressure.
(13) Each diffusion coefficient was obtained as an average
of eight separate determinations. The values of 0.732
where 52 and 51 are the slopes of the In Ho/h-(/lt)2 and 0.688 cm2/s for the diffusion of argon in helium and
plots and (t02 - tOI) is the time required for the peak nitrogen in helium at 25C res pectively compare
favora bly with values of 0.725 and 0.688 cm2 /s obtained
maximum to travel the distance between de tectors.
by Walker and Westenberg [1] for the same gas pairs. )
The quantity 0:2 was also calculated from
TABLE 1. Effect of Temperature on the Diffusion Coefficient
(14)
Gas pair Temp. K D(cm'!s )
where A2 and A I are the areas under the second and HeA 77.2 0.0710 0.0009'
first peak respectively and v the velocit y of th e carrier He A 298 .7335 .004
HeA 334 .8890 .006
gas in the diffusion column. Equation (14) which follows HeA 357 .991 7 .006
from eq (11), gave the same values for 0: 2 as eq (13). He-Nt 77.2 .0725 .001
He-Nt 296 .678 + .0005
He- Nt 321 .817 1 .005
He-Nt 348 .925 1 .007
HeN, 370 1.030 .006
3. Results and Discussion He-Kr
HeK r
298
322
.6491
.7372
.004
.004
He- Kr 341 .813 .005
HeKr 366 .904 .006
In fi gure 3 are shown plots of eqs (6) and (10) for one He-Oz 298 .7361 .005
HeO, 320 .8472 .006
experiment on the diffusion of nitrogen in helium at HeO, 3605 1.041 .008
30 psia and 25. The filled squares represent values
calcu lated from eq (10) for the first peak with tOI equal *90% confidence limit s fnr the mean.
to 23.59 min and the open squares are values obtained
from the second peak with t02 equal to 48.81 min. The A least square calculation for the expression
open circles represent values calculated from eq (6)
from the leadin g half of the first peak while the filled Io g
298
Iog--=m
D298
-
circles are those of the leading half of the second peak. DT T
The halffilled circles righ t represe nt values calcu
lated from eq (6) from the trailing half of the first peak wh ere D298 is the diffusion coefficient at 298 K, DT is
while the half-filled circles left are the corresponding the diffusion coefficient at temperature , T, and m is a
values for the second peak. All experim ental points constant. The following values for m were obtained
fit the theore tical lin e within a precision of 0.5 percent. with indicated precision being for a 90 percent confi-
dence limit: He-N 2 , 1.69 0.02 ; He-02' 1.67 0.06;
( ~t)2.
"2 (minutes) -2 He-Kr, 1.71O.06 and H e-Ar , 1.720.006. These
50 100 150 200 values may be compared with values obtained by
Seager, Geerston, and Giddings [5]: He-N 2, 1. 75 0.12;
150 150
He-0 2, 1.700.08 and He-Ar, 1.730.07. The two sets
agree within the given experimental errors. In figure
4 is shown the log diffusion coefficient-log temperature
plot for argon diffusing into helium.
There are several advantages to this method.
100 1
E
(a) The use of two detectors instead of one eliminates
the n eed for a zero time correction , since it is not
J
_IN
+ necessary to know the shape of the peak at the point
~I.c of sample injection. The only requirements are that
c
'? 50 50 the conce ntration profile b e Gaussian when the peak
reaches the first detector, that the thermal conductivity
detectors do not distrub the concentration profile,
and that the carrier gas be at thermostat temperature
since the sample injection value is at room temperature.
50 100 150 200
(b) Diffusion coefficients over a wide range of tem-
00 - t)2 (minutesf 2
perature may be measured since the apparatus may be
FIGURE 3. Plot of eqs (6) and (10) for the diffusion of nitrogen in used in a liquid nitrogen bath or in an oven without
helium at 30 psia and 25C. any c hange in design.
210
term in the diffusion equation it is important to be
1.0
able to measure this quantity accurately. This can be
done by measuring the time required for the peak
.8 maxima to travel the measured distance between the
two de tectors. By inj ectin g more than one peak into
I .6 the diffusion column it is possible to determine whe-
:, ther the velocity of the carrier gas has changed during
a measureme nt.
.4 (d) Deviations of experiment from the assumed
model may be determined. Equation (6) provides a
means for analyzing the peak for any disturbance due
to gas absorption on the walls of the diffusing column
or for any disturbance to the concentration profile
~
.2 due to the gas injection and detec tion sys te ms .
Ne It is conluded that thi s is a simple, rapid, widely
0 applicable me thod for th e determination of gaseous
diffu sion coe fficients .
.10
4. References
(> .08
211