Академический Документы
Профессиональный Документы
Культура Документы
INTRODUCTION
Accurate and precise measurements of ultraviolet irradiance are required in the determination of the
radiant exposure of both total and selected narrow bands of ultraviolet radiation for the determination
of exposure levels in (1) outdoor weathering of materials, (2) indoor accelerated exposure testing of
materials using manufactured light sources, and (3) UV-A and UV-B ultraviolet radiation in terms both
of the assessment of climatic parameters and the changes that may be taking place in the solar
ultraviolet radiation reaching earth.
Although meteorological measurements usually require calibration of pyranometers and radiom-
eters oriented with axis vertical, applications associated with materials testing require an assessment
of the calibration accuracy at orientations with the axis horizontal (usually associated with testing in
indoor exposure cabinets) or with the axis at angles typically up to 45 or greater from the horizontal
(for outdoor exposure testing). These calibrations also require that deviations from the cosine law, tilt
effects, and temperature sensitivity be either known and documented for the instrument model or
determined on individual instruments.
This test method requires calibrations traceable to primary reference standards maintained by a
national metrological laboratory that has participated in intercomparisons of standards of spectral
irradiance.
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
priate safety and health practices and determine the applica- 4.1 This test method represents the preferable means for
bility of regulatory limitations prior to use. calibrating both narrow-band and broad-band ultraviolet radi-
ometers. Calibration of narrow- and broad-band ultraviolet
2. Referenced Documents
radiometers involving direct comparison to a standard source
2.1 ASTM Standards: 2 of spectral irradiance is an alternative method for calibrating
E 772 Terminology Relating to Solar Energy Conversion ultraviolet radiometers. An ASTM test method describing this
E 824 Test Method for Transfer of Calibration From Refer- procedure is under development by Subcommittee G03.09 on
ence to Field Radiometers Radiometry.
4.2 The accuracy of this calibration technique is dependent
3. Terminology
on the condition of the light source (for example, cloudy skies,
3.1 Definitions: polluted skies, aged lamps, defective luminaires, etc.), and on
3.1.1 broad-band radiometera relative term generally source alignment, source to receptor distance, and source
applied to radiometers with interference filters or cut-on/cut-off power regulation.
filter pairs having a FWHM between 20 and 70 nm and with
tolerances in center (peak) wavelength and FWHM no greater NOTE 5It is conceivable that a radiometer might be calibrated against
a light source that represents an arbitrarily chosen degree of aging for its
than 62 nm.
class in order to present to both the test and reference radiometers a
3.1.2 diode array detectora detector with from 50 to 1000 spectrum that is most typical for the type.
silicon photodiodes affixed side-by-side in a linear array and
mounted in the focal plane of the exit slit of a monochromator. 4.3 Spectroradiometric measurements performed using ei-
3.1.3 full width at half maximum (FWHM)in a bandpass ther an integrating sphere or a cosine receptor (such as a shaped
filter, FWHM is the interval between wavelengths at which TFE3, or Al2O3 diffuser plate) provide a measurement of
transmittance is 50 % of the peak, frequently referred to as hemispherical spectral irradiance in the plane of the spheres
bandwidth. entrance port. As such, the aspect relative to the reference light
3.1.4 narrow-band radiometera relative term generally source must be defined (azimuth and tilt from the horizontal for
applied to radiometers with interference filters with FWHM solar measurements, normal incidence with respect to the beam
#20 nm and with tolerances in center (peak) wavelength and component of sunlight, or normal incidence and the geometri-
FWHM no greater than6 2 nm. cal aspect with respect to an artificial light source, or array). It
3.1.5 scanning monochromatora monochromator that is important that the geometrical aspect between the plane of
uses either a single, or several interchangeable, detector(s) the spectroradiometers source optics and that of the radiom-
mounted at the exit slit, that is presented with dispersed light eter being calibrated be as nearly identical as possible.
by sweeping the spectrum across the slit to illuminate the NOTE 6When measuring the hemispherical spectral energy distribu-
detector with a succession of different very narrow wavelength tion of an array of light sources (for lamps), normal incidence is defined
light distributions. The detector may be either a photomulti- by the condition obtained when the plane of the spheres aperture is
plier tube (PMT) or silicon photodiode (visible), or a PMT or parallel to the plane of the lamp, or burner, array.
an ultraviolet-enhanced silicon photodiode (ultraviolet and 4.4 Calibration measurements performed using a spectrora-
visible), or a lead sulfide cell or other solid state detector (near diometer equipped with a pyrheliometer-comparison tube (a
infrared), etc. The dispersed spectrum is swept across the sky-occluding tube), regardless of whether affixed directly to
monochromators exit slit using a mechanical stage that rotates the monochromators entrance slit, to the end of a fibre optic
either a prism or a grating dispersive element, usually under the bundle, or to the aperture of an integrating sphere, shall not be
control of an external microprocessor or computer. performed unless the radiometer being calibrated is a true
3.1.6 spectroradiometera radiometer consisting of a pyrheliometer (that is, unless it possesses a view-limiting
monochromator with special acceptance optics mounted to the device having the approximate optical constants of the spec-
entrance aperture and a detector mounted to the exit aperture, troradiometers pyrheliometer-comparison tube).
usually provided with electronic or computer encoding of 4.5 Spectroradiometric measurements performed using
wavelength and radiometric intensity. The monochromator of source optics other than the integrating sphere or the stan-
dard pyrheliometer comparison tube, shall be agreed upon in
advance between all involved parties.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3
Standards volume information, refer to the standards Document Summary page on Tetrafluoroethylene such as a special grade of Teflont or an equivalent
the ASTM website. material, has been found suitable.
son by measurement of a reference lamp source that is either 5.3 Instrument Mounts:
managed, sponsored, or sanctioned by a national metrological 5.3.1 Equatorial MountAn altazimuthal or equatorial,
laboratory, or another appropriate body. follow-the sun mount that is equipped with a platform on
5.1.2 If a linear diode-array spectroradiometer is used as the which the spectroradiometer is mounted is required for all
reference, it shall possess focusing optics internal to the hemispherical normal-incident and direct (beam) calibrations
monochromator and a linear diode array detector with a measurements.
sufficient number of diodes that, together, result in a resolving 5.3.2 Tilt TableA stable, adjustable tilt table having tilt
power of 1 nm or less. The monochromators dispersive and azimuth adjustments is required for global solar radiation
element shall be a holographic grating, and the spectroradiom- measurements (for example, at horizontal orientation) and
eters acceptance optics shall consist of either an integrating hemispherical measurements at specified azimuthal and tilt
sphere with appropriately sized and oriented light entrance positions.
port, or a shaped translucent diffuser plate (such as a TFE3 or NOTE 9An altazimuthal mount so equipped also may be used as the
Al2O3 wafer) whose deviation from true cosine response is tilt table.
small and known. A further requirement is that the stray light
rejection be determined for any diode-array spectroradiometers 5.3.3 Optical PlatformA stable, platform equipped with
used to perform this test method and that it be 105 or greater in height adjustment is required for use in measuring the calibrat-
the spectral region for which calibration is required. ing radiometers against light sources such as arrays, solar
5.1.2.1 A diode-array spectroradiometer shall not be used as simulators, special lamps, and burners, etc.
the reference instrument below 300-nm wavelength. Further, NOTE 10When using a fiber-optic/integrating sphere source configu-
when used in the wavelength region between 300 and 320-nm ration to calibrate radiometers, for example, against Xenon-arc lamps,
wavelength, evidence shall be presented with the calibration carbon arcs, and other burners employed in indoor exposure cabinets,
reports, or certificates, showing that the stray light has been special fixtures may be required to rigidly mount and present the source
optics to the source of irradiance. For UV-A and UV-B calibrations, the
eliminated by a combination of internal baffeling, merging of fiber-optic bundle must be constructed of quartz fibers.
two determinations in which the wavelength region below
6. Procedure
6.1 Calibrate the spectroradiometer in accordance with the
4
The Spectrodiometric Measurement of Light Sources, Publication No. 63, The manufacturers instructions and CIE Publication 634 unless the
International Commission on Illumination (CIE). spectroradiometers calibration is known to be stable within 30
than 30 of the sun during any one measurement sequence. where: Es (j) is the integrated irradiance for the measurement
6.2.5 Determine the spectral irradiance distribution of the series j defined by the test radiometer being calibrated, l1 and
sun in conformance with the procedures specified in CIE l2 are the wavelength limits of integration defined above, and
Publication 63.4 Perform not less than five spectral irradiance Es,l(ij) is the spectral irradiance readings i in the wavelength
measurements separated by at least 30 min. Ensure that at least interval dl.
one measurement is taken at, or not greater than 30 min from NOTE 12The wavelength bands to which a radiometer is calibrated
solar noon. may be slightly larger, or slightly smaller than the advertised band-pass
6.2.6 Record the instantaneous voltage signals of the radi- for the radiometer. The essential requirement is that the out-of-band
ometer, or radiometers, being calibrated at a frequency not less spectrum of the reference light source, and, hence, the field source, must
than every minute during the time period subtended by the not represent a significantly greater irradiance than the average in-band
irradiance, and the out-of-band irradiance must not exhibit poorer tempo-
spectral irradiance measurements.
ral stability than the average in-band irradiance.
6.3 Calibration Against Manufactured Light Sources:
6.3.1 Mount the radiometer to be calibrated in the geometri- 6.4.2 For each value of integrated spectral irradiance Es (j),
cal configuration and aspect that will be employed in its compute the average voltage Vr (j) measured with the test
end-use application. Ensure that the receptor aperture (for radiometer in the interval j corresponding to the time interval
example, entrance port) of the spectroradiometers sphere is at of the reference spectral measurement:
exactly the same distance from the light source (see section n
5.2.2) as the radiometer being calibrated, and ensure that the ( Vr~ij!
i51
Vr ~ j ! 5 (2)
reference spectroradiometer and radiometer being calibrated n~j!
--````,``,,,`,````,,`,`,,,`,``,-`-`,,`,,`,`,,`---
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).
--````,``,,,`,````,,`,`,,,`,``,-`-`,,`,,`,`,,`---