WireLess Health Monitoring System Using


ZigBee
T.Chandrasekhar 1, J.S.Chakravarthi 2, V.Srikanth 3

Professor , Department of Electronics and Comm. Engineering, GIET Engineering College, Rajahmundry, Andhra Pradesh, India1
Assistant Professor , Department of Electronics and Comm. Engineering, GIET Engineering College, Rajahmundry, Andhra
Pradesh, India2
U.G.Student , Department of Electronics and Comm. Engineering, GIET Engineering College, Rajahmundry, Andhra Pradesh,
India3

Abstract: Recent developments in off-the-shelf wireless embedded computing boards and the increasing need for
efficient health monitoring systems, fueled by the increasing number of patients, has prompted R&D professionals to
explore better health monitoring systems that are both mobile and cheap. This work investigates the feasibility of
using the ZigBee embedded technology in health-related monitoring applications. Selected vital signs of patients are
acquired using sensor nodes and readings are transmitted wirelessly using devices that utilize the ZigBee
communications protocols. A prototype system has been developed and tested with encouraging results.

Keywords: Zigbee, wireless health monitoring.

I. INTRODUCTION

The increasing cost of in-bed hospitalization and the recent technological advances in low-power integrated
circuit sensors coupled with the introduction of power efficient protocols such as Bluetooth and ZigBee has tempted
researchers to study the establishment and usage of wireless networks as a vehicle for transmitting patients-related
information without the need to confine them to a premise [1, 2]. Wireless-based non-confining monitoring systems
improve the quality of life for the patients while serving as a cost effective solution to the problem of health care
monitoring that is worsening with the increase in the aging population. The rapid development in the
telecommunication field and mobile technology has accelerated the introduction of telemedicine as a viable and
reliable alternative. Recent work [3, 4] includes using Bluetooth technology coupled with the GSM technology to
report signs to PDAs held by the patient or his doctor. Monitoring based on ultra wideband-based personal area
networks was reported in [5]. Sneha and others [6] presented an architectural framework for a system that utilizes
mobile techniques to wirelessly monitor the ECG of cardiac patients. The work reported in [7] discusses the
implementation issues, and describes the overall system architecture of a Bluetooth sensor network for patient
monitoring.

Compared to Bluetooth, ZigBee provides higher network flexibility and a larger number of nodes, and a
better transmission range with low power consumption. Recently, ZigBee-based wireless networks were tested in
various applications. The authors investigate the use of ZigBee and mobile phones in monitoring elderly patients
with diabetes mellitus or heart diseases. A ZigBee WiMAX nursery system for patient monitoring was proposed. An
interesting application of ZigBee-based wireless sensor nodes in the real-time monitoring of fruit logistics.
The work presented here simply attempts to probe into the applicability, usefulness, and practicality of
using wireless-ZigBee based network in monitoring the signs of patients on a hospital floor and surrounding area
without confining them to a bed. The presented solution is meant to be both simple, cost effective and hopefully
causes minimal interference with the patients mobility and comfort. A wearable sensor unit, attached to the
patients body, reads and transmits the patients fetal signs to a portable ZigBee-based receiver carried around

Copyright to IJIRSET www.ijirset.com 7021


ISSN: 2319-8753
International Journal of Innovative Research in Science,
Engineering and Technology
(An ISO 3297: 2007 Certified Organization)

Vol. 2, Issue 11, November 2013

by a nurse or doctor or to a hospital server. Added features include the storage of these readings in a central database
or access via the internet.
The rest of the paper is organized as follows: Section II describes the overall systems layout. Section III
discusses the software development, implementation and testing. The conclusion is presented in Section V.
II. SYSTEM HARDWARE DESCRIPTION

The system is designed and built using the ZigBee modules (Nodes) from Jennic Corporation [11]. Vital signs
sensors attached to the patients body are interfaced to these Nodes (hereafter P-Node). The complete P-Node is
packaged in a light form and carried by the patient. Sensed data is transmitted to a ZigBee coordinator (Z-Coor)
with a wide LCD display that is carried by the supervisor nurse on the hospital floor.A brief description of the major
components of the proposed monitoring platform is provided next:
ZigBee development board: we used the JN5139IEEE802.15.4/JenNet kit. The kit has five boards. One is used as
coordinator (Z-Coor) and the other four nodes are used as terminal nodes. Each node (P-Node) is attached to a
patient to acquire the vital signs. The node has a low cost wireless 32-bit RISC microcontroller-based board. It is
equipped with built- in resources such digital input/output ports, analog-to- digital converters, timers, RAM, EEPORM,
Serial RS232 and USB communications ports and a fully compliant 2.4GHz IEEE802.15.4 transceiver, It uses a USB
to serial bridge, as it creates a virtual COM port on the PC. The four P-Nodes and one Z-Coor allow the system to
monitor four different patients at a time which makes it distributed patients monitoring systems. Additional
four terminal units (hereafter M- Node/s) that are equipped with LCD are made available for Nurses and
Doctors to monitor the patients from within the adhoc network range (50-70 meters). These units do not have sensors.
They receive the patients data from the Z-Coor. The Z-Coor can handle up to 127 P-Nodes and M-Nodes per network.
Figure 1 shows the monitoring wireless adhoc network and Figure 2 shows a detail layout for one patient network.

Sensor Array: four sensors are interfaced with each P-Node. An LM35 temperature sensor is used to measure the
surrounding area temperature. A thermocouple k-type sensor was used to record the patients body temperature.
Thermocouples generate very low-voltage (mV), an AD625 programmable gain instrumentation amplifier from
Analog Devices is used to enhance the signal quality. A heartbeat sensor from sunroom technologies was selected to
measure the beat rate. This sensor is designed to produce a digital output equivalent to the heart beat when a finger is
placed inside it. This sensor has an LED that flashes in unison with each heart beat. The digital output can be connected
to a microcontroller directly to measure the Beats Per Minute (BPM) rate. It works using the principle of light
modulation by blood flow through the finger at each pulse.
For measuring the patients activity and guarding against the possibility of falling down, an ADXL213
Accelerometer Sensor is used. Its low cost, 1.2g Dual and measures both static (Gravity) and dynamic
acceleration (Vibration). The accelerometer is used in order to determine whether the patient is stable and is in good
position (standing or sitting) or has fallen down (sudden vertical change of the position). This sensor provides a
digital output and can be connected directly to the microcontroller, without the need for a conditioning circuit. The
Duty cycles of the digital signals are proportional to acceleration and the typical noise floor is 160 g/ Hz.The
complete hardware unit (sensors integrated with the ZigBee module and a battery pack) was implemented and housed
inside an enclosure with dimensions (19 x 9 x 5 cm) as shown in Figure 3.
The XBee Gateway: shown in Figure 2 is used to provide gateway functionality between the ZigBee network and the
Ethernet. This gateway device collect data from the coordinator packetize it and via the TCP/IP layer, data is sent and
stored in the main server where a database is used to keep records of the patients history.

Copyright to IJIRSET www.ijirset.com 7022


ISSN: 2319-8753
International Journal of Innovative Research in Science,
Engineering and Technology
(An ISO 3297: 2007 Certified Organization)

Vol. 2, Issue 11, November 2013

Figure 1. Monitoring Network

III.SYSTEM SOFTWARE DEVELOPMENT AND TESTING

The P-Nodes (units carried by the patients) together with M-Nodes (units carried by the nurse) and Z-Coor
work in a star topology formation. Using this topology, every device in the network can communicate only with
Z-Coor. All units are programmed using C. Appropriate code was developed to carry out various functions such
as configuring and enabling the A/D converters, reading of data via specific ports or channels, programming
the timers, calculating the acceleration, acquiring sensed data and formatting of payloads for transmission, etc. Each
P-Node carried by a patient has a unique ID (Node ID). A frame that includes this nodeID, Room Temperature, Heart
Rate, Body Temperature, and computed Acceleration is created and sent to the Z- Coor. Table 1 shows the frame
format and content. The OK indicator corresponding to the accelerometer field indicates that the patients
orientation is normal.

On the receiving side, the M-Node unit (carried by the Nurse or Doctor) extracts the readings from the
received frame and displays them on a wide LCD. Figure 3(b) shows an actual screen shot for the M- Node
during the testing. Based on these readings, the nurse can assess the condition of a given patient and decide if a
rectifying action needs to be taken or not.
Also, the system can send an SMS message to the mobile of the on call physician. A screen shot from the
physicians mobile is shown in Figure 4.
On the server side, PHP and MySql were used. PHP code is developed to implement the web pages. A
nurse or physician can register as many patients as possible and save this in a database. Moreover, the nurse or
physician can monitor patients using the PHP web pages. The PHP pages will retrieve data from the database and post
them on the web page.

Copyright to IJIRSET www.ijirset.com 7023


ISSN: 2319-8753
International Journal of Innovative Research in Science,
Engineering and Technology
(An ISO 3297: 2007 Certified Organization)

Vol. 2, Issue 11, November 2013

TABLE I. P-Node Frame

Figure 2. System Hardware Architecture

Figure 3. (a) P-Node unit and (b) M-Node screen.

Copyright to IJIRSET www.ijirset.com 7024


ISSN: 2319-8753
International Journal of Innovative Research in Science,
Engineering and Technology
(An ISO 3297: 2007 Certified Organization)

Vol. 2, Issue 11, November 2013

Figure 4. Screen shot from the Physicians mobile phone.

MySQL database is the world's most popular open source database. MySql has many advantages that
make it popular and desired, e.g. fast performance, high reliability, ease of use, and cost savings. A Database
called hospitalgroup was created with simply two tables. The first table is called patientinfo and stores data
such as patients ID, name, room number and threshold values for temperature and heart rate. These values are
determined by the patients physician and if the patients readings exceed these values the system will
automatically send an alarm SMS using the GSM network to the doctor. The other table is called records and
stores the patients records or history of readings of the various signs. An Apache webserver was used in the
experimental set up. For SMS notifications, a GSM modem was used, Eclipse Java was used to program this
modem. Figure 4 depicts a sample display of a patient with ID number 16 showing his temperature and heart rate,
indicating that these two sensed values have exceeded the pr-specified stored thresholds for this patient and that
therefore an SMS alert message is sent to the physician. Figure 5 is flowchart for the data flow analog with the
experimental setup displayed in Figure 6.

Figure 5. Data Flow Flowchart

Figure 6. System Experimental Setup.

Copyright to IJIRSET www.ijirset.com 7025


ISSN: 2319-8753
International Journal of Innovative Research in Science,
Engineering and Technology
(An ISO 3297: 2007 Certified Organization)

Vol. 2, Issue 11, November 2013

IV.FUTURE WORK

The proposed prototype was tested in a laboratory setting by monitoring the heart beat, temperature and
acceleration of few subjects. The readings obtained using the ZigBee wireless network were compared against reading
collected using regular static devices such as a thermostat placed under the patients tongue, or by using a stethoscope
to measure the beat for calibration purposes. At this stage of the project implementation, the differences are tolerable,
further fine tuning is needed in the heart beat measurement. The tested candidates felt that the enclosed device was light
and did not cause them any discomfort. A major limitation was the ZigBee transmission/ reception range. The
SMS alerting mechanism using the GSM network worked reliably.

V. CONCLUSION
The paper discusses the implementation of a monitoring system targeting patients on a hospital floor. The system
consists of end units carried by patients that collect sensed data (health sign readings) and transmit it to a coordinator
unit at the hands of the floor nurse. This is implemented using a ZigBee-based network. Such wireless environments
provide patients with the ability to move around and feel confident when in the mood to stride for a walk or a
stretch within a pre-defined vicinity. They can also contribute to cost reduction efforts in health care.

REFERENCES

[1] A. Milenkovic, C. Otto, and E. Jovanov, Wireless Sensor Networks for Personal Health Monitoring: Issues and an Implementation, in
Computer Communications, vol. 29 (13-14), August 2006.

[2] Y. Shieh, Y. Tsai, A. Anavim, M. Shieh, and M. Lin, Mobile Healthcare: Opportunities and Challenges, in International Journal of
Electronic Healthcare, 4(2), 208-219, 2008.

[3] F. Tay, D. Guo, L. Xu, M. Nyan, and K. Yap, MEMS Wear- biomonitoring System for Remote Vital Signs Monitoring, in Journal of the
Franklin Institute, 346(6), 531-542, August2009.

[4] A. Sagahyroon, H. Raddy, A. Ghazy, and U. Suleman, Design and Implementation of a Healthcare Monitoring System, in International.
Journal of Electronic Healthcare, 5(1), 68-86,2009.

[5] K. Takizawa, Huan-Bang, L. Kiyoshi, H. Kohno, Wireless Vital Sign Monitoring using Ultra Wideband-Based Personal Area Networks,
in Proc. of the International Conference of the IEEE Engineering Medicine in Biology Society, 1798-1801, August 2007.

[6] S. Sneha and U. Varshney, A Wireless ECG Monitoring System for Pervasive Healthcare, in International Journal of Electronic
Healthcare, 3(1), 32-50, 2007.

[7] Y. Zhang and H. Xiao, Bluetooth-Based Sensor Network for Remotely Monitoring the Physiological Signals of Patient, in IEEE Trans. on
Information Technology in Biomedicine, 13(6),1040-1048, November 2009.

Copyright to IJIRSET www.ijirset.com 7026