Content
1. INTRODUCTION OF RT-7600............................................................................ 4
3. FUNCTION DESCRIPTION.............................................................................. 10
2
RT-7600 Auto Hematology Analyzer service manual
3
RT-7600 Auto Hematology Analyzer service manual
1. Introduction of RT-7600
RT-7600 is composed of mechanical system, pipeline system, computer control system and software.
4
RT-7600 Auto Hematology Analyzer service manual
Sampling needle: use sampling needle to aspirate sample into counting baths
Aspiration key: in sample analysis, press this key, the instrument will aspirate sample, function the
same as key
5
RT-7600 Auto Hematology Analyzer service manual
This analyzer is using impedance principle (colorimetry for hemoglobin measurement) to categorize and
count blood cells in blood. Test items are shown in Table 1-1.
Full English name English abbreviation Unit(Default)
White blood cell count WBC 109/L
Lymphocyte count LYM# 109/L
Intermediate cell count MID# 109/L
Granulocyte Cell count GRA# 109/L
Lymphocyte percentage LYM% %
Intermediate cells percentage MID% %
Granulocyte Cells percentage GRA% %
Red blood cell count RBC 1012/L
Hemoglobin content HGB g/L
Hematocrit HCT %
Mean corpuscular volume MCV fL
Mean corpuscular hemoglobin MCH pg
Mean corpuscular hemoglobin MCHC g/L
concentration
Red cell distribution width SD RDW-SD fL
Red cell distribution width CV RDW-CV %
Platelet count PLT 109/L
Mean platelet volume MPV fL
Platelet distribution width PDW %
Plateletcrit PCT %
Plateletlarge cell ratio P-LCR %
White blood cell histogram WBC Histogram
Red blood cell histogram RBC Histogram
Platelet histogram PLT Histogram
Table 1-1
6
RT-7600 Auto Hematology Analyzer service manual
2. Installation of RT-7600
1. Unpack the instruments package and remove material for transportation. Please keep original
packing carton and packing material, in case you need to repack the instrument in the future.
2. Take out the instrument from plastic package.
3. In accordance with packing list, make sure the packing carton content includes:
z RT-7600 hematology analyzer
z Users manual
z Packing list
z Warranty card
z Power adapter
z Product COA
Note: Product packing feature is as per packing list, in case of inconformity, please
contact seller.
In order to ensure instrument work normally, please choose working place compliant with following condition to
place RT-7600 Hematology Analyzer:
z No direct sunlight;
z No massive dust or powder;
z No strong electromagnetic radiation;
z Sufficiently large flat and solid desktop.
z a.c.110V220V
z 50/60Hz
z 96VA
7
RT-7600 Auto Hematology Analyzer service manual
1 Loosen the screw which fixed the right side cover, open the cover.
2 Cut the belt which fix the sample needle carrier and installation board, take the belt out of
instrument, check whether the sample needle has been bent or the screws which fix the carrier
and track are loosen.
3 Loosen the two screws which fix the shield cover of WBC and RBC counting bath. Lift up the
cover and take it out. Check whether the WBC and RBC counting bath with bracket has been
installed correctly. Take off the adhesive tape which sealed the counting baths.
4 When the checking finished, install the shield cover and right side cover again.
2.5 Reagent
The instrument needs to use Lyse, Cleanser, diluents for measurement and maintenance. To ensure accuracy
of test result, please use reagents accompanied with instrument.
1) From reagent packing carton, take out Lyse and Cleanser respectively, open the bottle lids, and place
them in instrument reagent booth horizontally.
2) Insert plastic tube marked Lyse into the Lyse bottle and tighten the bottle cap.
3) Insert plastic tube marked Cleanser into the Cleanser bottle and tighten the bottle cap.
8
RT-7600 Auto Hematology Analyzer service manual
1) Make sure both printer and instrument have been shut down.
2) Insert one end of USB cable into printer USB interface socket.
3) Insert another end of USB cable into instrument USB interface.
4) Connect printer and AC power supply with power line accompanied to printer.
9
RT-7600 Auto Hematology Analyzer service manual
3. Function description
Analogue board is composed by two parts of front CPU board and analog signal board.
Function description
1 The front CPU board sends the commands to each part of the instrument, and executes next
procedure after receiving the feedback signals, until the end of test. It also gives out warnings when
receiving error signal or the command has been out of time.
2 It connects with mainboard by serial port, receiving the high-level control command form the
mainboard.
3 The sub front control system is composed of C8051F020, it controls the movements of each part by
MCS 51, including X-axis motor, Z-axis motor, syringe motor, negative pressure syringe motor and 11
valves. It also performs the original position detecting of there mechanic parts.
4 It performs the acquisition and processing of different analog signals (including the pulse signal form
WBC and RBC counting baths, analog signals of from the HGB photoelectric and pressure sensor). It
transfers the pulse signals into digital signal, class them to WBC, RBC and HGB, save the signals
provisionally and transfer them to the mainboard.
5 Though the backdoor of RT-7600 controlling program in the mainboard, it adjusts the 5024 digital
Potentiometer in the analogue board, to perform the adjustment of signal gaining.
10
RT-7600 Auto Hematology Analyzer service manual
It performs part of the output voltage transfer (DC 24V input transfers into +12V and -12V for analog circuit,
DC 5V for front CPU and its circuit).
8-pin interface for mainboard and
Control signal output for valve 6-10
analogue board. Function: to
receive control command from
mainboard and transfer the
Control signal output for valve 1-5 acquired analog signals
3.2 Mainboard
EMB-3680 is a low power consumption mainboard, which is using GeodeTM LX 700/800+AMD Geode
CS5536 chip of AMD, supporting AMD GeodeTM LX 700@0.8W or GeodeTM LX 800@0.9W CPU
supporting one 200 pin DDR SO-DIMM memory (ECC not supported, 333/400MHZ DDR supported, max
1G Bytes), integrated with two RealtekRTL 8100 network card.
+5V---+24 single power supply.
Supporting VGA, LVDS, and TTL display output mode (VGA, LVDS, TTL support two individual displays,
LVDS and TTL can not be used at the same time)
11
RT-7600 Auto Hematology Analyzer service manual
About external storage, there is one Mini IDE channel providing 2 hard disk interfaces, 1 socket for
COMPACT FLASH card, and 4 USB 2.0 interfaces.
COM 1 Connect mouse
and keyboard
Standard VGA interface
LAN 1 LAN 2
COM 3
Voltage select jumper
Device protect jumper
CS5536
BIOS
COM 4
CMOS clear jumper
LVDS display LCD TFT socket LVDS display socket +5V---24Vpower interface (from left to right)
indicator interface 1+5V---24V 2/3GND 4: null
Panel interface (cover)
Figure 3-2 mainboard
Definitions of RT-interfaces
connected to analogue board
connected to printer
connected to touch panel
connected to interfaces
12
RT-7600 Auto Hematology Analyzer service manual
Backside of mainboard
CF card
DDR Memory socket
Note: to prevent from damaging the mainboard, you must not perform the cleaning of
CMOS configuration when system is working.
13
RT-7600 Auto Hematology Analyzer service manual
CF card jumperJCF
This jumper is used to set CF card and IDE2 devices (such as hard disk) as master or slave device.
JCF
OPEN Set CF card as slave device
CLOSE Set CF card as master device
If you want to set CF card as the master startup device, please set JCF jumper as CLOSE
If you want to set CF card as the slave startup device, please set JCF jumper as OPEN
Because RT-7600 use CF card as master external storage device, so the JCF jumper is set as CLOSE.
J6
VCC>12V CLOSE
VCC<12V OPEN
14
RT-7600 Auto Hematology Analyzer service manual
Mouse/Keyboard
Interface
This board takes the work of connection of different parts and providing power supply for them. It uses a
power supply chip to complete the DC/DC converse, providing +12V, two way +5V output for Mainboard,
touch panel and printer.
J8 Power output
J7 3 pin interface for mainboard
for LCD inverter
J3 3 pin interface
J10 moni board
for the printer
J6 2 pin interface for aspirate power output
button of keypad board
J2 for mainboard
15
RT-7600 Auto Hematology Analyzer service manual
This board takes the work of controlling and detecting of movement parts. It uses L297 and L298 to drive
different motors and transfer the original position signal to front CPU board.
J11
J12
J1
J17
Warning: Be careful when you are connecting the lines, if there is any wrong connection,
it may cause damage to the sampling needle and counting baths, the
instrument can not work normally.
16
RT-7600 Auto Hematology Analyzer service manual
There are two status indicator and two buttons on this board. The indicators indicate the status of
instrument and the two buttons are paper feeding and aspiration button.
Front side Back side
Indicator
Aspiration button
17
RT-7600 Auto Hematology Analyzer service manual
Pressure
Cleaning chamber
Valve
The pipeline system of RT-7600 is composed of 4 three-way valves, 7 two-way valves, WBC counting
bath, RBC counting bath, two buffering isolation chambers, sampling needle, cleaning chamber, reagent
dispensing pump, pressure syringe, three reagent (LYSE, DILUENT, CLEANSER), waste bottle,
temperature sensor, devices to Eliminate Static Electricity and so on.
18
RT-7600 Auto Hematology Analyzer service manual
As shown in figure 4-1, there are 11 valves in RT-7600, there functions are as following:
Valve 1: Controls the dispensing of LYSE.
Valve 2: Release pressure of the pressure syringe, aspirate air.
Valve 3: Add CLEANSER to WBC counting bath.
Valve 4: Drain waste from the waste chamber of pressure syringe to waste bottle.
Valve 5: Works when WBC/RBC counting bath is counting, it controls the negative pressure.
Valve 6: Add DILUENT into WBC counting bath.
Valve 7: Drain the waste from the cleaning chamber when washing the sampling needle.
Valve 8: Add DILUENT into sampling needle.
Valve 9: Controls the dispensing of DILUENT.
Valve 10: Drain WBC counting bath.
Valve 11: Drain RBC counting bath.
Please refer to the bellowing sketch map which shows the working principle of solenoid valve. At left side
shows the status when it is not electrified, at right side shows the status when it is electrified, the direction
of magnetic force is right.
19
RT-7600 Auto Hematology Analyzer service manual
2-way valve
This symbol indicates that it is a 2-way solenoid valve which is normally closed. It is closed when not
electrified at normal conditions.
The above shows the figure of a 2-way solenoid valve in the piping diagram of RT-7600. There are two
parts of this figure. The right side indicates the status when the valve is not electrified, you can see that it
is closed. The left side indicates the status when the valve is electrified and open; the arrowhead indicates
the direction of liquid flowing.
3-way valve
3
2 1
This symbol indicates that it is a 3-way solenoid valve. When it is not electrified, common end 3 and 1 is
connected, the direction of liquid flowing is from end 1 to 3. Common end 3 and 2 is not connected.
The above shows the figure of a 3-way solenoid valve in the pipeline diagram of RT-7600. There are two
parts of this figure. The right side indicates the direction of the liquid flowing and the common ends it
passes when the valve is not electrified. The left side indicates the direction of the liquid flowing and the
common ends it passes when the valve is electrified.
At 3-way solenoid valve, we usually use COM as symbol of the common ends, abbreviated as C. NC
indicates that this end is closed at normal conditions, NO indicates that it is open at normal conditions.
When the valve is not electrified, NO connected with end C and NC not connected.
20
RT-7600 Auto Hematology Analyzer service manual
Introduction
1. The cycle has been marked with different color.
2. Without special explanation, other 2-way valves which are not related to the cycle are considered as close,
the triple valves are considered as 1-3 connected.
3. At left side from top to bottom is valve 1-5; at right side from top to bottom is valve 6-11.
21
RT-7600 Auto Hematology Analyzer service manual
Cleaning chamber
End 2
`
Figure 4-3 Diluent cycle
1 Aspirate liquidValve 9 is not electrified (1, 3 connected), the piston of reagent dispensing pump is pulled
down and aspirate the Diluent.
2 Rinse the surface of sampling needle:
A. Create Diluent priming cycle
The Diluent from reagent dispensing pump passes through 3-2 end of valve 9, 3-1 end of valve 8, 3-1 end
of valve 6, reaches at end 2 of the cleaning chamber.
B. Create negative pressure aspiration cycle.
After valve 7 is open, the piston of pressure syringe is pulled up to create negative pressure; the Diluent is
aspirated from end 1 of the cleaning chamber.
Piston of vacuum is pulled up to Val 9 is open, Com 3 and 2 is Reagent dispensing pump
create negative pressure connected works and dispenses Diluent
Valve 7 is open
Create Diluent priming circuit
Create negative pressure aspiration circuit
22
RT-7600 Auto Hematology Analyzer service manual
Valve 8 and 9 is electrified, the Diluent flows though the sampling needle to rinse its inside surface.
Part of the Diluent flows into the counting chamber with the samples at inside surface of the sampling needle.
Flow chart
Vacuum create
negative pressure Piston of reagent dispensing Valve 9 open, end 3-2 connected Valve 6 open, end 3-2 connected
pump pushes up completely
Valve 5 is open
Create the circuit from the reagent dispensing pump to back
chamber of counting chamber
23
RT-7600 Auto Hematology Analyzer service manual
24
RT-7600 Auto Hematology Analyzer service manual
There are two main steps of Lyse cycle, step 1 is to aspirate Lyse and step 2 is to add Lyse to WBC counting
bath.
Aspirate Lyse: Valve 1 is close (The reagent dispensing pump is connected with Lyse bottle), the piston of
reagent dispensing pump is pulled down and aspirate Lyse form the bottle.
Add Lyse to WBC counting bath: Valve 1 is open (The reagent dispensing pump is connected with WBC
counting bath), the piston of reagent dispensing pump is pushed up and add Lyse into WBC counting bath.
25
RT-7600 Auto Hematology Analyzer service manual
Time of creating counting circuit: After dilution of RBC and WBC, Lyse has been added and air bubble mixing
has been finished.
26
RT-7600 Auto Hematology Analyzer service manual
The pressure syringe aspirates the waste liquid into the pump and drains it into the waste bottle by switch
between positive and negative pressure.
The pressure pump release pressure Valve 10 and 11 is closed and valve 4 is open
27
RT-7600 Auto Hematology Analyzer service manual
28
RT-7600 Auto Hematology Analyzer service manual
RT-7600 is an instrument with high precision; there are no parts which can be repaired by users. When
there are problems which can not be solved by users, you shall ask professional service engineer for
repair.
After several times of back flush and soaking with concentrated Cleanser, if you still can not eliminate the
problem of aperture block, you shall take out the aperture following the below steps strictly. You can clean
the aperture or replace a new one.
1) At first, remove the two screws which fix the side covers, also remove the screw at backside,
and remove the side covers. Loosen the two screws which fix the shielding cover of counting
bath, lift the shielding cover and remove it, you can see WBC/RBC counting bath assembly,
as shown in figure 5-1:
29
RT-7600 Auto Hematology Analyzer service manual
2) First, take off the signal cable of counting bath from PCB, pull out the counting bath slowly
from the bracket. Take the counting bath out of the instrument and disassembly it as
following figure:
When you want to replace LED and photocell, just take out the fixing block with LED or photocell from the
counting bath, replace a new one, then install it again.
Warning: It is not allowed to take the aperture with forceps. You shall use
your forefinger, dip it into clean water and take out, then use the
forefinger to take the aperture and install it into the bath.
30
RT-7600 Auto Hematology Analyzer service manual
If there is air leakage in the pressure syringe (can be detected by pressure detector), you shall replace the
airproof washer (O-type) of the pressure syringe. Please follow bellow procedures to make the
replacement.
1) Pull the piston of the pressure syringe to extreme position, then loosen the three screws which fix the
syringe, disassemble the body of the pressure syringe from the inside board, as figure 5-3:
2) Loosen the 4 screws on the installation board of pressure syringe, take out the broken airproof
washer (O type), install a one to the original position and fix it (smear the inside and outside of the
washer with airproof lubricating grease), as shown in figure 5-4.
3) Install the pressure syringe.
31
RT-7600 Auto Hematology Analyzer service manual
The washers (O type) for reagent dispensing syringe need to be replaced regularly. Please following the
below procedures to finish the replacement:
1 Loosen the 4 screws which fix the reagent dispensing syringe assembly, take off the syringe from the
inside board, as shown in figure 5-5:
32
RT-7600 Auto Hematology Analyzer service manual
Figure 5-6
33
RT-7600 Auto Hematology Analyzer service manual
3 Take off the broken washer and replace new washer for each piston, smear the inside and outside of
the washers with airproof lubricating grease and fix the syringe again, As show in figure 5-7:
Figure 5-7
5.3.3 Replace the airproof washer for sampling needle cleaning chamber
If the airproof of sampling needle is not good, you should replace the airproof washer (O type) at the
sampling needle cleaning chamber, please follow the below procedures to finish the replacement, as
shown in figure 5-8:
1 Loosen the 2 screws which fix the sampling needle, then loosen the 2 screws which fix the cleaning
chamber assembly.
34
RT-7600 Auto Hematology Analyzer service manual
2 Take off the sampling needle and cleaning chamber assembly from the carrier, then take out the
cleaning chamber cover and sampling needle from the cleaning chamber assembly, you can take off
the airproof washer and replace a new one. Please remember to smear the inside and outside of the
washers with airproof lubricating grease and fix the cleaning chamber assembly again, as shown in
figure 5-9:
WarningYou must smear the inside and outside of the washers with airproof lubricating
grease when you replace a new washer.
The procedures of replacing sampling needle are similar to the procedures of replacing airproof washer
for sampling needle. First loosen the 4 screws as shown in figure 5-8, take out the sampling and cleaning
chamber assembly, take off the PVC tubes connected to the sampling needle. At last take out the sampling
needle from the cleaning chamber, take out the cleaning chamber cover and airproof washer from the
sampling needle, replace a new needle and fix the cleaning chamber again.
35
RT-7600 Auto Hematology Analyzer service manual
6. Maintenance of RT-7600
RT-7600 is a precise auto hematology analyzer. To ensure its good working status, it is very important to
perform the daily maintenance. The procedure is easy but it must be done according to requirements.
Warning: do not use any solvent, fat, corrosive substance to clean instrument.
The WBC/RBC counting bath must be kept clean to ensure the accuracy and stability of instrument.
1. Clean the surface of counting bath
If the shield cover of counting bath is dirty, you must take off the cover and clean it with clean wet cloth,
this is to prevent unexpected objects entering the counting bath and block the aperture.
2. Clean the inside of counting bath
1) During daily maintenance, you must rinse the pipeline before shut down the instrument. If the
instrument is not shut down normally (sudden electricity loss during operation), you shall perform
the rinse procedure when you turn on the instrument again. This is to avoid blockage in aperture.
2) If the counting bath is dirty, the aperture is blocked. After several operation of removing blockage,
it is still not working, please disassembly the aperture and clean it under the instructions in
chapter 5.2.
If there is residue at the surface of sampling needle can not be cleaned automatically, you can wipe it by
soft cloth dipped with absolute alcohol.
WarningYou must drain the counting bath if the instrument will not be used for more
than one week.
36
RT-7600 Auto Hematology Analyzer service manual
AEvery day
a) Before shut down the instrument, you should use a soft cloth or clean paper dipped with a little
concentrated cleanser, to clean the sampling needle. You should from top to bottom, clean the 2cm area
which is above the needlepoint. After cleaning, get into the menu of Maintenance; perform the
b) If there is liquid in the plate which is at the bottom inside the instrument, please wipe it with clean cloth.
c) If you find the test results abnormal, please perform blank test and check if the bank values are high. If
d) If the shield cover of counting bath is dirty, please take it off and wipe with wet clean cloth. This is to
avoid unexpected objects like dust entering the counting bath and block the aperture.
BEvery week
a) Every week you should perform the operation of Concentrated cleanser soaking at least twice. Add
2ml concentrated cleanser into the counting bath. You can also increase times of soaking according to
b) If the instrument will not be used for a week, you should perform the operation of Drain chamber and
CTwo weeks
If the instrument will not be used for more than two weeks, you should perform the operation of Stop use in
maintenance menu to clean and drain counting bath and pipeline.
37
RT-7600 Auto Hematology Analyzer service manual
Trouble Solutions
38
RT-7600 Auto Hematology Analyzer service manual
NoteIf you meet problem that can not be solved, or some errors continue, please contact
the distributor.
NoteThis service manual may contain declaration and information of copyright. They
must be complied. Although we have paid much attention, some errors in
interpretation or printing can not be avoided.
This manual can be modified or updated without notification, Rayto reserve the
rights to improve or modify the instrument or program described in this manual
without special notification.