SD BIOLINE
Dengue Duo Dengue IgG/IgM
Disposable capillary pipette (10l)
Dengue NS1 Ag Ab Combo
1 2 3 4
Check Point !
Capillary Pipette
YYYY/MM/DD
Check the expiry date. Put on the gloves. Open the test pouch and write the patients name on the Prepare the serum, plasma or whole blood samples.
If expiry date has passed, use another kit. test device. Too little Too much
Assay diluent
A. Dengue NS1 Ag (100l)
5 6
3x
Hold the assay diluent vertically.
Overview of the Device
Dengue NS1 Ag Dengue IgG/IgM
Use the disposable dropper to collect sample. Add 3 drops of sample into the sample well. DENGUE
W/B
C
M
B. Dengue IgG/IgM Whole Blood (10l) G Sample
Well
5 6 7 S
Assay
Fill Line 4x 90
Wait 15~20 minutes.
Dont read test results after 20 minutes ;
Sample
Well
Diluent
Well
8
Caution:
Use the disposable capillary pipette to collect sample. Add sample into the square sample well(s). Add 4 drops of assay diluent to round well. Each test can be used ONLY ONE TIME.
Using a disposable capillary pipette, draw sample to fill line. Do not reuse the test device.