4. What is the brightness gain for a 17- cm image intensifier tube with a
flux gain of 120 and a 2.5 cm output phosphor
( d input )2
Minification gain=
(d output )2
2
(17)
Minification gain=
(2.5)2
289
Minification gain=
6.25
Minification gain=46.24
Brightness gain=5,549
( d input )2
Minification gain=
(d output )2
2
(23)
Minification gain=
(10)2
529
Minification gain=
100
Minification gain=5.29
6. What is the change is dose (how much higher) when switching from a
23cm FOV to a 17cm FOV?
Use same principal as in question 5
( d input )2
Minification gain=
(d output )2
2
(23)
Minification gain=
(17)2
529
Minification gain=
289
Minification gain=1.83
15 cm
X 0.5 mGy t=New ESD
25 cm
[CLASSROOM} FLUOROSCOPY SCENARIOS AND EQUATIONS
( d input )2
Minification gain=
(d output )2
2
(23)
Minification gain=
(2.5)2
529
Minification gain=
6.25
Minification gain=84.64
Brightness gain=84.64 X 75
Brightness gain=6,348
9. Create a scenario using one of the equations above with the specific
factors required. Be sure to show your work and include the answer.
(18)2
Minification gain=
(1 .5)2
324
Minification gain=
2.25
[CLASSROOM} FLUOROSCOPY SCENARIOS AND EQUATIONS
Minification gain=144
Brightness gain=144 X 13 0
Brightness gain=18,720