Вы находитесь на странице: 1из 16

MANUAL

3025
FLUORESCENCE MICROSCOPE
SYSTEM

73 Mall Drive, Commack, NY 11725 631-864-1000 (P) 631-543-8900 (F)


www.accu-scope.com info@accu-scope.com
3025 FLUORESCENCE MICROSCOPE SYSTEM

CONTENTS
SAFETY NOTES .................................................................................................................. 3

CARE AND MAINTENANCE ................................................................................................ 3

INTRODUCTION .................................................................................................................. 4

UNPACKING AND COMPONENTS ..................................................................................... 4

EPI FLUORESCENCE ATTACHMENT DIAGRAM .............................................................. 4

EPI FLUORESCENCE ATTACHMENT ASSEMBLY ............................................................ 5

DETAILED ASSEMBLY PROCEDURE ................................................................................ 6

ADJUSMENT & OPERATION


COMPONENTS DIAGRAM................................................................................................. 8
ADJUSTING THE ROTATION TENSION
OF THE FLUORESCENCE MIRROR BLOCK (FILTER BLOCK) ..................................... 9
PREPARATION ............................................................................................................... 10
CENTERING THE FIELD IRIS DIAPHRAGM ..................................................................... 11
ADJUSTING THE FIELD IRIS DIAPHRAGM ...................................................................... 11
CENTERING THE APERTURE IRIS DIAPHRAGM ............................................................. 12
ADJUSTING THE APERTURE IRIS DIAPHRAGM ............................................................. 12
CENTERING THE MERCURY BURNER ........................................................................... 13
CENTERING THE MIRROR REFLECTED IMAGE ............................................................. 14
MOUNTING ND FILTER................................................................................................... 14

TROUBLESHOOTING........................................................................................................ 15

MAINTENANCE.................................................................................................................. 16

SERVICE ............................................................................................................................ 16

WARRANTY ....................................................................................................................... 16

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 2


3025 FLUORESCENCE MICROSCOPE SYSTEM

SAFETY NOTES
1. Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from
dropping and being damaged.
2. Do not discard the molded Styrofoam container; the container should be retained should the
microscope ever require reshipment.
3. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.
Ensure the microscope is located on a smooth, level and firm surface.
4. If any specimen solutions or other liquids splash onto the stage, objective or any other component,
disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be
damaged.
5. CAUTION: the lamp, lamp housing and adjacent parts will become very hot. Do not touch these parts
until they have completely cooled. Never attempt to handle a hot halogen bulb.
6. All electrical connectors (power cord) should be inserted into an electrical surge suppressor to
prevent damage due to voltage fluctuations.
7. For safety when replacing the halogen lamp or fuse, be sure the main switch is off (O), remove the
power cord, and replace the halogen bulb after the bulb and the lamp house has completely cooled.
8. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The
use of a different input voltage other than indicated will cause severe damage to the microscope.

CARE AND MAINTENANCE


1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces
should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap
solution. Do not use organic solvents for cleansing.
3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from
an air bulb. If dirt remains on the optical surface, use a soft cloth or cotton swab dampened with a
lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a
circular motion. A small amount of absorbent cotton wound on the end of a tapered stick such as
cotton swabs or Q-tips, makes a useful tool for cleaning recessed optical surfaces. Avoid using an
excessive amount of solvents as this may cause problems with optical coatings or cemented optics
or the flowing solvent may pick up grease making cleaning more difficult. Oil immersion objectives
should be cleaned immediately after use by removing the oil with lens tissue or a clean, soft cloth.

4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not
in use.
5. ACCU-SCOPE microscopes are precision instruments which require periodic preventative
maintenance to maintain proper performance and to compensate for normal wear. An annual
schedule of preventative maintenance by qualified personnel is highly recommended. Your
authorized ACCU-SCOPE distributor can arrange for this service.

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 3


3025 FLUORESCENCE MICROSCOPE SYSTEM

INTRODUCTION
Congratulations on the purchase of your new ACCU-SCOPE fluorescent microscope. ACCU-SCOPE
microscopes are engineered and manufactured to the highest quality standards. Your microscope will
last a lifetime if used and maintained properly. ACCU-SCOPE microscopes are carefully assembled,
inspected and tested by our staff of trained technicians in our New York facility. Careful quality control
procedures ensure each microscope is of the highest quality prior to shipment.

UNPACKING AND COMPONENTS


Your microscope arrived packed in a molded Styrofoam container. Do not discard the container: the
Styrofoam container should be retained for reshipment of your microscope if needed. Avoid placing the
microscope in dusty surroundings or in high temperature or humid areas as mold and mildew will form.
Carefully remove the microscope from the Styrofoam container by its arm and base and place the
microscope on a flat, vibration-free surface.

EPI FLUORESCENCE ATTACHMENT DIAGRAM

Main body of the EPI Fluorescence Attachment

Power Supply Unit

Power Cord (ONLY use the power cord provided)

A Mercury Lamp

Fuses (8A)

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 4


3025 FLUORESCENCE MICROSCOPE SYSTEM

EPI FLUORESCENCE ATTACHMENT ASSEMBLY


The diagram below shows how to assemble the various modules. The numbers indicate the order of
assembly.

When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching
any parts or touching glass surfaces.

Loosen the set screw and remove the trinocular viewing head from the body of the microscope.

Insert the EPI Fluorescence attachment into the microscope and tighten the set screw until it is firmly
in place.

Insert the trinocular viewing head into the EPI Fluorescence attachment as shown and tighten the set
screw until it is firmly in place.

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 5


3025 FLUORESCENCE MICROSCOPE SYSTEM

ASSEMBLY (continued)
Mounting the Mercury Burner
(Fig.1 & Fig. 2)

Loosen the lamp housing (burner socket)


clamping screw , and remove the housing
cover.Fig.1

Remove the foam backstop and insert lamp


by using a soft tissue into the (positive) + pole
(the wide head) of the mercury burner to the
lower terminal first and then the (negative)
pole (the thin head) to the upper terminal, then
Fig. 1 tighten the two socket clamping screws .

Slide the lamp housing cover back on, and


tighten the housing clamping screw .

Be sure to ONLY use a 100 watt mercury


lamp.
Be sure to mount the positive + pole (the
wide head) before the negative pole, or
damage to the lamp may occur.
Never use excessive force when mounting
the mercury lamp.

Avoid touching or soiling the mercury lamp. If
this happens, gently wipe it with lens tissue or
a soft, clean cloth.
Fig. 2 To prevent hazard, always turn the main
switch on the power supply unit to O (OFF)
position, unplug the power cord plug from the
main outlet, and wait for a minimum of 10
minutes before replacing the lamp.

Mounting Filter Blocks (Fig.3 and Fig.4)


Unscrew the Allen screws with the
included Allen wrench and take out the filter
block turret .

Allen
screw

Fig. 3

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 6


3025 FLUORESCENCE MICROSCOPE SYSTEM

Invert the filter block turret , (several model


blocks can be found.) Loosen the Allen screw
to take one of the blocks out.
Mount a filter cube (Fig. 3) into the top of the
model block and tighten the screw . Beside the
screw, you can see a number on the turret
indicating G-excitation. To help identify mirror
blocks, place a label below the corresponding
number on the front side of the mirror turret.
Repeat procedure to mount other blocks.

Fig. 4 Push the filter block turret back into the rail slot
and tighten the Allen screw.

Mounting Protection Barrier (Fig.5)


Install the protection barrier on the attachment by
tightening the screws.

Assembly of the Fluorescent Attachment


(Fig.6)
Fig. 5 Mount the lamp housing into the other end of
the attachment and secure it with two screws .

Cable & Cord Connections


(Fig. 7 & Fig. 8)
Make sure that the main switch of the power
supply is set to O (OFF) position before
connecting cables.
Plug the connector from the lamp socket
securely into the connector on the power supply
Fig. 6 unit.
Plug the power cord connector into the power
supply unit.
Verify that the voltage and the frequency of the
AC main outlet match the setting of the voltage
switch and the frequency switch on the rear of the
power supply unit as improper setting may result in
poor lamp performance, or in rare instances cause
the lamp to explode.
Only use the power cord provided by the
manufacturer. Should you lose or damage the
power cord, contact the manufacturer or your
Fig. 7 distributor.

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 7


3025 FLUORESCENCE MICROSCOPE SYSTEM

Fuse Replacement (Fig.7 & Fig. 8)


Set the main switch to O (OFF) position and
unplug the power cord before replacing fuses.

Using a flat-head screwdriver to loosen and


remove each of the fuse holders .

Replace both fuses with new ones.


Fig. 8
Always use the designated fuse (8A). And
make sure the voltage of the fuse matches
the voltage of the AC main outlet.

ADJUSTMENT & OPERATION


Components Diagram

Fig. 9

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 8


3025 FLUORESCENCE MICROSCOPE SYSTEM

Fig. 10

Adjusting the Rotation Tension of the


Fluorescent Mirror Block (Filter Block)
There are 6 fluorescent mirror blocks (filter
blocks) mounted in the filter block turret.
(Fig.11).

A mirror block includes a dichrohic mirror, a


barrier filter, and on excitation filter. Do not take
apart the filter block.

A wide variety of filter sets are available. Please


consult with your ACCU-SCOPE distributor to
select the appropriate set for your application.
Fig. 11

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 9


3025 FLUORESCENCE MICROSCOPE SYSTEM

OPERATION
Preparation
Verify that the voltage and the frequency of the AC The system employs high-intensity excitation light
main outlet match the setting of the voltage switch to enable bright observation of dark fluorescence
and the frequency switch on the rear of the power specimens. As a result, if high-power objectives
supply unit. are used frequently, color fading of the specimen
occurs early, degrading the view (contrast) of
Make sure the cord is connected firmly. fluorescence images. So it is effective to use the
When transmitted light observation is required, pull shutter frequently to avoid illuminating the
out the filter system and rotate the blank opening specimen for a longer period than required.
into the light path. A ND (neutral density) filter and small aperture
Adjust the field diaphragm to match the field edge. diaphragm can help weaken the intensity of the
If it is not centered, use the Allen wrench to adjust excitation light. Also, it is useful to use the light
the screw. shutter to reduce the specimen color fading.
Color fading of the specimen can also be delayed
When it is required to interrupt observation for a using commercially available color fading
short period, use the shutter mechanism on the preventing agent (DABCO, etc). The use of color
filter wheel to block the light accessorial excitation fading preventing agent is recommended when
filter part. (Repeated on-off of the mercury lamp you perform high-magnification observation
will shorten its service life considerably). frequently.
Precautions on the specimen color fading: NOTE: color fading preventing agent cannot be
used with certain specimens

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 10


3025 FLUORESCENCE MICROSCOPE SYSTEM

Centering the Field Iris Diaphragm (Fig.1)


Switch the light shutter to closed position.

Turn filter block turret to engage the B-excitation


mirror in the light path.

Switch the light shutter to O open position.


Rotate the 10 objective in the light path, and
place the specimen on the stage and bring into
approximate focus.
Fig. 1
Pull the field iris diaphragm lever out until
the diaphragm comes into the smallest state.

Use the Allen wrench to adjust the two field


iris diaphragm centering screws alternately
to move the image of the diaphragm to the
center. (Fig. 2 shows the adjustment of
diaphragm)

Move the field diaphragm lever to open the


diaphragm. As this makes slight deviation
noticeable, adjust the centering precisely.

Enlarge the diaphragm until it just


circumscribes the field of view.

Adjusting the Field Iris Diaphragm


(Fig. 2)
The field diaphragm adjusts the diameter of the
illuminating beam to obtain good image contrast.

Keeping the field diaphragm stopped down to


the smallest required area for each observation
makes it possible to prevent color fading of
areas outside the observation target region.

According to the objective in use, adjust the


Fig. 2 diaphragm image using the field diaphragm
lever so that the field of view is circumscribed by
the field diaphragm to exclude stray light.

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 11


3025 FLUORESCENCE MICROSCOPE SYSTEM

Centering the Aperture Iris Diaphragm


(Fig. 3)
Switch the light shutter to close position to
turn off the light path.
Rotate the filter block turret to engage the G-
excitation mirror block or another into the light
path.
Fig. 3 Switch the light shutter toO open position
to open the light path.
Rotate the 10 objective in the light path, and
place the centering plate (a white plate with a
cross) on the stage and bring into approximate
focus.
Move the cross of the centering plate to the
center of the field of view.

Pull out the aperture diaphragm lever to


adjust the aperture iris diaphragm to the
smallest diameter.

Pull out the field iris diaphragm lever to adjust


the field iris diaphragm to the smallest diameter.
The image of aperture iris diaphragm can be
found on the centering plate.
Adjust the aperture iris diaphragm centering
screws with attached wrench to superpose
the image of aperture iris diaphragm on the
cross of centering plate.

Adjusting the Aperture Iris Diaphragm (Fig. 3)


The aperture iris diaphragm adjusts image resolution and contrast.

For fluorescence observation, push in the aperture iris diaphragm lever .

Both ND (neutral density) filter and small aperture diaphragm can help weaken the intensity of the
excitation light to delay color fading of the specimen

According to the objective in use, adjust the diaphragm image using the field iris diaphragm lever so
that the field of view is circumscribed by the field iris diaphragm to exclude stray light.

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 12


3025 FLUORESCENCE MICROSCOPE SYSTEM

Centering the Mercury Lamp (Fig. 4-6)


NOTE: Before proceeding to center the lamp, wait
for the arc image to stabilize to protect against
glare during arc image centering, it should be
viewed across the excitation light protective shield.

Switch the light shutter to close position to


shut off the light path.

Rotate the filter block turret to engage the green or


blue excitation filter block into the light path. If U/V
Fig. 4 excitation filter block is used, be sure to use the
protective shield.

Revolve the nosepiece to engage 10 objective


into the light path. Place the centering plate on
stage, through transmission observation; adjust the
stage until the cross is in the centre of the field of
view.

Remove the objective from the revolving


nosepiece position and engage this position in the
light path.

Pull out the field iris diaphragm lever to close


the iris diaphragm and push in the aperture iris
diaphragm lever to open the iris diaphragm to
the large limit.

Switch the light shutter to O open position to


open the light path.

Turn the collector adjusting knob to project the


arc image on the centering plate and sharpen it.
(A)

Turn the lamp adjusting knob to move the arc


image and the mirror reflected arc image in the
symmetrical position. (B)

Adjust the mirror focusing knob (Fig.6) to


sharpen the mirror reflected arc image. (C)

Turn the burner adjusting knob to overlap the


arc image with the mirror reflected arc image.(D)

Fig. 5 Turn the collector adjusting knob to make the


field of view as bright as regular as possible..

Maintain this condition until the next time the lamp


is replaced.

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 13


3025 FLUORESCENCE MICROSCOPE SYSTEM

Centering the Mirror Reflected Image


(Fig. 6)
The mirror reflected image has been centered at
the factory. Do not adjust the knob -- only if
necessary. Only when the burner has been
centered precisely, can the knob be adjusted.

NOTE: Once the knob is adjusted, the reflected


mirror cannot be reconverted to the factory
setting.


Knob Control (Fig. 6
Fig. 6 The middle knob is the mirror reflected image
focusing knob which can focus the reflected
image.

The knobs on both sides can adjust the


up/down or left/right position of the mirror
reflected image.

Mounting the ND (Neutral Density) Filter


(Fig. 7)
The ND filter can reduce the excitation light
intensity to delay color fading of the specimen.
Use the ND filter it if does not hinder
observation.

There are two kinds of ND filters available: ND6


and ND25 for position and respectively
Fig. 7 (Fig. 7). To prevent the ND filter from being
damaged, insert the filter with the indication
surface facing the observation side.

When the filter is inserted, you will hear two


clicks. The filter is in the light path on the second
click.

CAUTION
When the mercury lamp is lit for a long
period while an ND filter is inserted, the filter
and its metallic frame may become very hot.
Wait until the ND filter cools down before
replacing it.

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 14


3025 FLUORESCENCE MICROSCOPE SYSTEM

TROUBLESHOOTING
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a
problem occurs, please review the following list and take remedial action as needed. If you cannot solve the
problem after checking the entire list, please contact your local dealer for assistance.

Optics
PROBLEM CAUSE SOLUTION
Switch the light shutter to O open
The light shutter closes the light path
position
The ND filter is engaged in the light path. Pull out ND filter to open the position
Although the mercury lamp
illumination is on, the field of The fluorescence mirror block is
Engage it properly
view is invisible or dark. improperly engaged in the light path
Open the aperture iris diaphragm fully;
The aperture iris diaphragm and field iris
adjust the field iris diaphragm to
diaphragm are not open enough
circumscribes the field of view
The objective or filter is dirty Clean them thoroughly

Visibility is poor. Image is The aperture iris diaphragm and field iris
Open these iris diaphragms fully
not sharp. Contrast is poor. diaphragm are adjusted improperly
The fluorescence mirror block is not
Use proper mirror block
proper for the specimen
The objective is improperly engaged in Make sure the nosepiece clicks properly
the light path into place
The fluorescence mirror block is
Engage it properly in the light path
improperly engaged in the light path
The edge of the field of view
is obscured or not evenly The field of view doesnt open fully Open it fully
illuminated
ND filter is stopped in halfway in the light Pull in the filter slider until it clicks into
path place
The mercury lamp is not centered. Center it
The collector focus position is not correct Adjust it to an optimum position
Shadow exists in the field of
The burner or collector is dusty or stained Clean them thoroughly
view

Electrical System
The main switch cannot The power cord is connected improperly Connect it properly
supply power to the system A fuse is blown Replace the fuses
The lamp housing connecting cord is
Connect it properly to the connectors
connected improperly
The main switch can be set The mercury lamp is not mounted Attach a mercury lamp
to ON but the burner doesnt
ignite Set the main switch of the power
The auto ignition system is supply unit to OFF then on again.
malfunctioning (Repeated ON-OFF is possible in this
case)
The phenomenon is observed in a short
The mercury lamp flickers or Wait for 10 minutes or more after ignition
period after ignition
the brightness is low
The lamp life has expired Replace the mercury lamp

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 15


3025 FLUORESCENCE MICROSCOPE SYSTEM

MAINTENANCE
Please remember to never leave the microscope with any of the objectives or eyepieces removed and
always protect the microscope with the dust cover when not in use.

SERVICE
ACCU-SCOPE microscopes are precision instruments which require periodic servicing to keep them
performing properly and to compensate for normal wear. A regular schedule of preventative maintenance
by qualified personnel is highly recommended. Your authorized ACCU-SCOPE distributor can arrange
for this service. Should unexpected problems be experienced with your instrument, proceed as follows:

1. Contact the ACCU-SCOPE distributor from whom you purchased the microscope. Some
problems can be resolved simply over the telephone.

2. If it is determined that the microscope should be returned to your ACCU-SCOPE distributor or to


ACCU-SCOPE for warranty repair, pack the instrument in its original Styrofoam shipping carton. If you
no longer have this carton, pack the microscope in a crush-resistant carton with a minimum of three
inches of a shock absorbing material surrounding it to prevent in-transit damage. The microscope should
be wrapped in a plastic bag to prevent Styrofoam dust from damaging the microscope. Always ship the
microscope in an upright position; NEVER SHIP A MICROSCOPE ON ITS SIDE. The microscope or
component should be shipped prepaid and insured.

LIMITED MICROSCOPE WARRANTY

This microscope is warranted to be free from defects in material and workmanship for a period of five years from
the date of invoice to the original (end user) purchaser. The mercury power supply is warranted for a period of one
year from the date of invoice to the original (end user) purchaser. This warranty does not cover damage caused in-
transit, misuse, neglect, abuse or damage resulting from improper servicing or modification by other then ACCU-
SCOPE approved service personnel. This warranty does not cover any routine maintenance work or any other
work, which is reasonably expected to be performed by the purchaser. Normal wear is excluded from this warranty.
No responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as
humidity, dust, corrosive chemicals, deposition of oil or other foreign matter, spillage or other conditions beyond the
control of ACCU-SCOPE INC. This warranty expressly excludes any liability by ACCU-SCOPE INC. for
consequential loss or damage on any grounds, such as (but not limited to) the non-availability to the End User of
the product(s) under warranty or the need to repair work processes. Should any defect in material, workmanship or
electronic component occur under this warranty contact your ACCU-SCOPE distributor or ACCU-SCOPE at (631)
864-1000. This warranty is limited to the continental United States of America. All items returned for warranty
repair must be sent freight prepaid and insured to ACCU-SCOPE INC., 73 Mall Drive, Commack, NY 11725
USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of
America, for all foreign warranty repairs return freight charges are the responsibility of the individual/company who
returned the merchandise for repair.

ACCU-SCOPE is a registered trademark of ACCU-SCOPE INC., Commack, NY 11725

ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 631-864-1000 www.accu-scope.com 16

Вам также может понравиться