Forma Biological Safety Cabinets

FORMA QUALITY MEANS MAXIMUM PROTECTION


Everyday, over 500,000 laboratory workers around the world routinely
handle hazardous biological materials. The primary persons at risk of
acquiring an infection from the aerosols produced by most laboratory
procedures are the laboratory workers themselves. But the risk can
extend to students, custodial and maintenance workers who enter
laboratories, handlers of shipped materials, sanitation workers, and
all who work in or pass through buildings adjacent to the laboratories.

Biological safety cabinets are the primary containment devices used in


laboratories to prevent the escape of aerosols and to provide protection
for personnel, products, and the environment. When used with good
microbiological techniques, Class II biological safety cabinets provide
effective partial containment of airborne contaminants.

Established in 1950, Forma Scientific has grown to become one of the


worlds leading manufacturers of controlled environment laboratory All Forma Biological Safety Cabinets are ergonomically
equipment, including biological safety cabinets. Ongoing research, designed for maximum safety and comfort for lab personnel.
innovative technology, and proven performance have enabled Forma
to design and manufacture one of the most reliable lines of NSF Listed
biological safety cabinets available to you today.

CABINET CLASSIFICATIONS
Biological safety cabinet classifications are defined in NSF Standard 49. Class II refers to ventilated cabinets with open fronts that use inward
airflow for personnel protection, HEPA (High Efficiency Particulate Air) filtered downward laminar airflow for product protection, and HEPA
filtered exhausted air for environmental protection. Class II cabinets are available in the following configurations.

APPROXIMATE PERCENTAGE OF CABINET AIR MINIMUM


RECIRCULATED VENTED BACK VENTED OUT FACE VELOCITY
CLASS II WITHIN CABINET INTO ROOM TO ENVIRONMENT (FEET PER MINUTE)
Type A 70 30* 0 75 FPM
Type B2 0 0 100 100 FPM
Type A/B3 70 0 30** 100 FPM

*Cabinet can be exhausted to the environment.


**As a Type A, it can be vented back into the room.

Extensive microbiological tests are performed at our facility to ensure lab, product,
and personnel safety. Pictured above is a test procedure to study airflow in
the work area using a simulated arm with nebulizer and plates.
THEORY OF OPERATION
All biological safety cabinets contain HEPA filters for entrapment of extraneous airborne particulates or aerosols. HEPA filters are constructed
of a continuous sheet of submicron glass fibers folded back and forth over corrugated spacers that position and support the filter media.
Forma Biological Safety Cabinets have front-loading HEPA filters that are zero probed and rated 99.99% efficient for 0.3 micron particulates.
Optional ULPA (Ultra Low Penetration Air) filters, rated at 99.999% efficient for 0.12 micron particulates, are available.

CLASS II, TYPE A CABINETS


In Formas Class II, Type A cabinets, the airflow descends from the top of the cabinet
through a supply HEPA filter. As this vertical airstream approaches the work surface the
airflow splits with approximately 50% being drawn into the cabinets rear intake grille and
the rest meeting ambient room air at the front intake grille. The air is drawn into a negative
pressure plenum under the work area by the blower located in the bottom of the cabinet.
The blower pressurizes this air into a positive pressure plenum to force air to the top of the
cabinet. The pressurized air is split, with 30% being exhausted through the exhaust HEPA
filter and the remaining 70% recirculated through the supply HEPA filter. The air being
drawn into the back and front grilles, combined with the laminar HEPA filtered downflow
air, provides a sterile work area and cross contamination protection. The area where room
air and HEPA filtered air meet produces an air barrier that provides personnel and product
protection. The HEPA filtered exhaust air provides environmental protection. Class II,
Type A cabinets are suitable for work with agents assigned to biosafety levels 1, 2, or 3 in
the absence of volatile toxic chemicals and volatile radionuclides.
Class II, Type A Console
CLASS II, TYPE A/B3 CABINETS
In Formas Class II, Type A/B3 cabinets, the airflow descends from the top of the cabinet through
a supply HEPA filter. As this vertical airstream approaches the work surface the airflow splits
with approximately 50% being drawn into the cabinets rear intake grille and the rest meeting
ambient room air at the front intake grille. The air is then drawn into a plenum surrounding the
work area by the blower located in the top of the cabinet. The blower pressurizes this air into a
positive pressure plenum. The pressurized air is split, with 30% being exhausted through the
exhaust HEPA filter and the remaining 70% recirculated through the supply HEPA filter. The air
being drawn into the side, back and front grilles, combined with the laminar HEPA filtered
downflow air, provides a sterile work area and cross contamination protection. The area where
room air and HEPA filtered air meet produces an air barrier that provides personnel and product
protection. The HEPA filtered exhaust air provides environmental protection. Class II, Type B3
cabinets may be used with agents assigned to biosafety levels 1, 2, or 3 treated with minute
quantities of volatile toxic chemicals and tracer amounts of radionuclides required as an adjunct to
microbiological studies that will not interfere with the work when recirculated in the downflow
Class II, Type A/B3 Benchtop air.

CLASS II, TYPE B2 CABINETS


In Formas Class II, Type B2 biological safety cabinets, room air is pulled into the cabinet by
the blower and pushed through the supply HEPA filter, providing clean unidirectional air
onto the work surface. Approximately 50% of this work area air exits through the rear grille.
The remaining 50% meets with room air and is pulled into the front intake grille, providing a
3
protective barrier. The divided downflow and inflow air then combines and is pulled up the
rear plenum through the exhaust HEPA filter by the facility exhaust system. Class II, Type
B2 cabinets may be used with agents assigned to biosafety levels 1, 2, or 3 treated with
volatile toxic chemicals and radionuclides required as an adjunct to microbiological studies.

Ambient Air Contaminated Air


(positive pressure)

Contaminated Air HEPA Filtered Air


(negative pressure)

Class II, Type B2 Benchtop


FORMA ADVANTAGES: PROTECTION,
PERFORMANCE, AND CONVENIENCE
Forma Biological Safety Cabinets are recognized for their quality, safety, performance, and dependability. We design and manufacture our
cabinets to exceed minimum specifications defined by NSF Standard 49 and to simplify maintenance procedures. Forma Biological Safety
Cabinets incorporate many unique design features to provide maximum personnel, product, and environment protection; reliable performance;
and customer convenience.

EXCLUSIVE PRODUCT FEATURES


FORMA CLASS II, TYPE A CABINETS

For personnel protection, air is pulled into the 4.0" (10.2cm) wide intake grille. This air meets the uniform, downward, HEPA filtered air
and keeps unfiltered air out of the work area. Exceptional product protection is achieved because the directed airflow minimizes the
chance of cross contamination along the work surface. High velocity momentum air curtains are not required.

Forma's Vortex Channels (U.S. Patent 5,295,902) create high velocity air at the window corners to inhibit the escape of particulates. The
downward, HEPA filtered air behind the viewscreen meets the multi-directional inward flowing air. This multi-directional air at the corners
of the window opening can create turbulence, which increases the risk of particulates escaping from or entering the cabinet work area. Our
Vortex Channels, located at each end of the intake grille, provide high velocity air that sweeps the turbulent air into the return air system
before it reaches the work area. Product and personnel protection are increased.

Reliable, rocker style switches on the control panel are easy to use. Independent service organizations have indicated that customers
prefer our control panel design because it is easy to use and inexpensive to replace if service is required. Microprocessor controls used by
other manufacturers can cost hundreds of dollars to replace if a problem occurs.

The blower continues to operate when the sliding window is closed. This
feature lowers the ambient sound level and saves setup time while keeping the
work area purged and ready for immediate use. Other manufacturers blowers
automatically shut off when the window is closed.

FORMA CLASS II, TYPE A /B3 CABINETS

Unique features of Forma Type A/B3 Models include all of those listed for Type A
plus the following.

Return air slots, located just above the work surface on each side wall,
provide increased sample cross contamination protection. Forma Scientific is
the only manufacturer to offer this additional sample protection.

Dual-wall construction design allows us to surround the entire work area with
a negative pressure air plenum. This design feature provides the following
benefits: an added measure of safety to the environment; return air slots for
enhanced sample cross contamination protection; sturdy, aesthetically
appealing units; and the ability to locate all utility plumbing connections inside
the wall where they are out of the way and out of sight.
4 Console and benchtop styles
Forma Biological Safety Cabinets feature an all-metal, positive pressure plenum
(enclosed within negative pressure air) with a patented hook and loop connection
(U.S. Patent 6,036,737) to the exhaust damper. This unique design accommodates quick and easy filter changes for lower maintenance costs.
All maintenance functions, including HEPA filter replacement, can be performed from the front of the biological safety cabinet. Positive
feedback from certification companies attests to the success of Forma's innovative plenum attachment system.

Forma Type A/B3 Cabinets can handle a 100% increase in filter loading without adjustment and at least a 230% increase in filter
loading with voltage and damper adjustments, while still maintaining 90% of the total CFM.

FORMA CLASS II, TYPE B2 CABINETS

Each Forma Type B2 Biosafety Cabinet features 12.0" (30.5cm) supply air intake and exhaust ducts with adjustable, zero leak
dampers for connection to in-house systems.

Standard normally open, normally closed, and common remote alarm contacts monitor the low airflow alarm and provide an output
signal for in-house central alarm systems.
COMPLETE WORKSTATION
Forma Scientific helps you create an organized and complete workstation that is tailored to your work requirements and comfort needs. Forma
Biological Safety Cabinet design features and wide range of accessories provide maximum safety, convenience, and comfort.

Optional Exhaust
Transition with Zero
Top Accessible Linear
Leak Damper
Response Damper
(No. 191570)
Optional
Front Access
Universal Piping
HEPA Filters
(No. 191620)

Minihelic
Pressure Gauge

Externally-Mounted Two Service Valves


Fluorescent Lights and Two Outlets, one
on each side wall

Optional ADA
Control Panel
(No. 191496)

Exclusive Vortex Channels


Unique Front
Intake Grille

Optional Armrest Rounded Front


(No. 191509 or
No. 191512)
Optional Right-Hand
Optional Hydraulic Three Drawer Set
Crank-Up Stand (No. 191495)
(No. 191518 or
No. 191519)
Optional Lab Chair, meets
Class 1,000 clean air
specifications (No. 191487) 5
Optional Left-Hand
Two Drawer Set
(No. 191524)

Optional Adjustable
Footrest (No. 191127)
CLASS II, TYPE A/B3 BENCHTOP MODELS
Forma Biological Safety Cabinets bring the latest innovations to your laboratory.
For your primary containment device, choose a 4 ft. or 6 ft. Class II, Type A/B3
Benchtop Model, which is engineered and tested to exceed stringent performance
specifications. Class II, Type B3 cabinets may be used with agents assigned to
biosafety levels 1, 2, or 3 treated with minute quantities of volatile toxic chemicals
and tracer amounts of radionuclides required as an adjunct to microbiological
studies that will not interfere with the work when recirculated in the downflow air.

Interior work area and tray are constructed of heavy-gauge, type 304 stainless
steel to resist corrosion. The one-piece interior features radiused corners for
cleaning convenience. Work tray easily lifts out of the work area without tools.
Work tray supports and spill trough are welded in place for structural stability.

100V and 115V models include two GFIC protected duplex electrical outlets,
one on each side wall. 230V models have two single electrical outlets, one on
each side wall. All outlets include a splash-proof cover.

Two service valves, one on each side wall, are provided. Two plugged
penetrations, one on each side wall, are standard should additional service valves
be required. Universal piping option allows top or bottom plumbing connection.

Matched blower/motor and method of mounting, combined with heavy-duty


Model 1284 construction, provide quiet operation and low vibration.

Full front-to-back and side-to-side stainless steel filter protector/diffuser, located above the work area, creates a uniform downflow of
HEPA filtered air and provides filter protection.

Externally-mounted fluorescent lamps with a separate light switch illuminate the entire work surface for enhanced visibility. All units are
pre-wired for an optional UV light.

Internal linear response damper is top-adjustable for easy access and eliminates the need for decontamination before adjusting.

Voltage Compensating Speed Control ensures that the motor receives constant voltage when the power line voltage fluctuates, providing
stable airflow and increasing operator, product, and laboratory safety.

Forma Type A/B3 Cabinets can handle a 100% increase in filter loading without adjustment and at least a 230% increase in filter loading
with voltage and damper adjustments, while still maintaining 90% of the total CFM.

Standard Minihelic pressure gauge monitors loading of the supply and exhaust HEPA filters. Front access panel is easily removed for
servicing the HEPA filters from the front of the cabinet.

Control panel opens within the cabinets side-to-side footprint for easy access when the unit is beside other equipment or against the wall.

Exclusive patented Forma Vortex Channels provide additional protection. These vortex channels, located on both sides of the front intake
grille, create high velocity air in the corners to inhibit the escape of particulates from the cabinet work area.
6
Audible/visual alarm alerts you when the window height exceeds the 10.0" (25.4cm) safe operating opening. Alarm by-pass switch on the
front of the control panel allows the alarm to ring back every five minutes until the window is returned to a safe level.

4.0" (10.2cm) wide front intake grille prohibits unfiltered air from entering the work area and allows a higher volume of air for maximum
personnel and product protection.

Calculated intake velocity through the 10.0" (25.4cm) work area access opening is a minimum of 105 LFPM for added safety.

Removable, cleanable paper catch has been designed into the cabinet to prevent paper and debris from being drawn into the blower.

Optional adjustable modular stands and hydraulic crank-up stands are available.

115V units are UL Listed, CSA Certified, and NSF Listed. 100V and 230V units are CSA Certified and NSF Listed.

U.S. Patents held include 5,050,943; 5,295,902; 5,380,244; and 6,036,737.




TYPE A/B3 BENCHTOP SPECIFICATIONS

MODEL 1284 1285 1288 1286 1287 1290 1291


Type 4 Ft. Benchtop 6 Ft. Benchtop
Window Sliding window of 0.25" (0.6cm) thick laminated safety glass
Work Area 49.0"W x 28.4"H x 22.7"F-B 73.0"W x 28.4"H x 22.7"F-B
(124.5cm x 72.1cm x 57.7cm) (185.4cm x 72.1cm x 57.7cm)
Exterior 54.0"W x 64.0"H x 32.1"F-B 78.0"W x 64.0"H x 32.1"F-B
Dimensions* (137.2cm x 162.6cm x 81.5cm) (198.1cm x 162.6cm x 81.5cm)
Electrical 115V, 60 Hz 230V, 50 Hz 100V, 50 Hz 115V, 60 Hz 230V, 50 Hz 100V, 50 Hz 100V, 60 Hz
14.0 FLA 11.0 FLA 14.0 FLA 15.0 FLA 8.5 FLA 13.5 FLA 16.0 FLA
Motor 3/4 HP
Breaker
Requirements 20 Amp 15 Amp 20 Amp 15 Amp 20 Amp
Exhaust 357-391 CFM** 530-580 CFM**
Requirements 8.0" (20.3cm) diameter duct connection, 0.25" w.g. pressure drop at the connection
Weight
Net: 538 lbs. (244.0 kg) 720 lbs. (326.6 kg)
Shipping: 685 lbs. 890 lbs. 892 lbs. 975 lbs.
(310.7 kg) (403.7 kg) (404.6 kg) (442.3 kg)

Continuing research and improvements may result in specification changes at any time.
*An additional 3.0" (7.6cm) must be added to the total height for the exhaust grille. Light canopy, dress panel, control panel,
and window channel can be removed for a 28.0" (71.1cm) exterior F-B dimension.
**Cubic Feet per Minute (CFM) is based on the Direct Airflow Method, under clean conditions.

FRONT VIEW

SIDE VIEW

See charts on pages 14 and 15 for details about the optional stands.
CLASS II, TYPE A CONSOLE MODELS
Safeguard yourself, your product, and your environment with Forma Class II, Type A
Console Biological Safety Cabinets. These units meet design and performance criteria
for NSF Standard 49 and are designed for use with biosafety levels 1, 2, and 3
biological or cytotoxic agents.

Low level dBA ensures comfortable operation.

Interior work area and work tray are constructed of heavy-gauge, type 304
stainless steel. Units feature unitized interior rear and side walls for durability.
Work tray lifts out of the work area without tools. Work tray supports and spill
trough are welded in place for structural stability.

100V and 115V models are equipped with two GFIC protected duplex electrical
outlets, one on each side wall. 230V models have two single electrical outlets, one
on each side wall. All outlets include a splash-proof cover.

All cabinets include two service valves, one on each side wall. Two plugged
penetrations, one on each side wall, are also available should additional service
valves be required.

Cabinets are designed with three 1.0" (2.5cm) openings for convenient plumbing
from either side or from the rear wall.

Model 1200 Maximum height of 79.1" (200.9cm) permits easy passage through doorways and
is ideal for facilities with low ceiling clearance.

Full front-to-back and side-to-side stainless steel filter protector/diffuser, located in the work area, creates a uniform downflow of HEPA
filtered air. Front access panel can be easily removed for servicing the HEPA filters from the front of the cabinet.

Exclusive patented Forma Vortex Channels provide additional product, personnel, and laboratory protection. These vortex channels,
located on both sides of the front intake grille, create high velocity air in the corners to inhibit the escape of particulates from the cabinet
work area.

Audible/visual alarm on sliding window models alerts you if the window height exceeds the 10.0" (25.4cm) safe operating opening. Alarm
by-pass switch, conveniently located on front of the light canopy, allows the alarm to ring back every five minutes until the window is
returned to a safe level.

Externally-mounted fluorescent lamps, with a separate switch, illuminate the work area.

Mounting pins for an optional IV bar are factory installed.

Unit is pre-wired for an optional UV light.

Internal linear response damper for balancing airflow is front-adjustable for easy access and eliminates the need to decontaminate the unit
before adjusting the damper.
8
Voltage Compensating Speed Control has been incorporated to ensure the supply motor receives constant voltage when power line voltage
fluctuates, providing stable airflow and increasing operator, product, and laboratory safety.

Standard Minihelic pressure gauge monitors the loading of the supply and exhaust HEPA filters.

Removable, cleanable paper catch has been designed into the cabinet to prevent paper and debris from being drawn into the blower.

Optional exhaust transition allows the cabinet to be exhausted to the outside. A facility exhaust blower must be used.

115V units are UL Listed, CSA Certified, and NSF Listed. 100V and 230V units are CSA Certified and NSF Listed.

U.S. Patents held include 5,295,902 and 5,380,244.





TYPE A CONSOLE SPECIFICATIONS

MODEL 1200 1201 1206 1220 1214 1205 1210 1230


Type 4 Ft. Console 6 Ft. Console
Window 1200/1201/1220/1214/1205/1230: Sliding window of 0.25" (0.6cm) thick laminated safety glass
1206/1210: Hinged window of 0.25" (0.6) thick laminated safety glass
Work Area 46.5"W x 25.3"H x 25.0"F-B 70.5"W x 25.3"H x 25.0"F-B
(118.1cm x 64.3cm x 63.5cm) (179.1cm x 64.3cm x 63.5cm)
Exterior 51.5"W x 79.1"H x 34.5"F-B 75.5"W x 79.1"H x 34.5"F-B
Dimensions* (130.8cm x 200.9cm x 87.6cm) (191.8cm x 200.9cm x 87.6cm)
Electrical 115V, 60 Hz 230V, 50 Hz 115V, 60 Hz 100V, 50 Hz 115V, 60 Hz 230V, 50 Hz 115V, 60 Hz 100V, 50 Hz
15.0 FLA 10.0 FLA 15.0 FLA 14.0 FLA 16.0 FLA 11.5 FLA 16.0 FLA 15.0 FLA
Motor 3/4 HP
Breaker
Requirements 20 Amp 15 Amp 20 Amp 15 Amp 20 Amp
Exhaust 242-274 CFM** 391-440 CFM**
Requirements 8.0" (20.3cm) diameter duct connection, 0.25" w.g. pressure drop at the connection
Weight
Net: 604 lbs. (274.0 kg) 770 lbs. (349.3 kg)
Shipping: 730 lbs. (331.1 kg) 945 lbs. (428.6 kg)

Continuing research and improvements may result in specification changes at any time.
*Height reflects cabinet with feet in minimum position. An additional 3.0" (7.6cm) must be added to the total height for the
exhaust grille. Deduct 3.5" (8.9cm) from the F-B dimension for the light canopy.
**Cubic Feet per Minute (CFM) is based on the Direct Airflow Method, under clean conditions.

FRONT VIEW

SIDE VIEW

9
CLASS II, TYPE A SPACE SAVING BENCHTOP MODEL
The 2 ft. Forma Class II, Type A Benchtop Model is designed for use with biosafety levels 1, 2,
and 3 biological or cytotoxic agents. The unit provides ultra-clean air conditions at the work
surface. The Model 1118 is the unit of choice in chemotherapy and biological applications
when space is limited.

Interior work area and tray are constructed of heavy-gauge, type 304 stainless steel for
resistance to corrosion. Unit features unitized interior side and rear wall construction. Work
tray easily lifts out of the work area without tools. Work tray supports and spill trough are
welded in place for structural stability.

Work area includes a built-in drain valve, one service valve, and one plugged penetration.

Electrical outlet is located on the front control panel. Unit incorporates two separate circuit
breakers and two line cords, one for the blower/motor and interior lights and one for the
electrical outlet.

Motor size and method of mounting, combined with heavy-duty construction, provide
exceptionally quiet operation and low vibration.

Full front-to-back and side-to-side stainless steel filter protector/diffuser, located in the work
area, creates a uniform downflow of HEPA filtered air and provides filter protection.

All contaminated plenums are surrounded by negative pressure for maximum laboratory,
product, and personnel safety.

Externally-mounted fluorescent lamps with a separate light switch illuminate the entire work
surface.
Model 1118
Internal damper for balancing the units airflow is adjustable from inside the work area.

Voltage Compensating Speed Control ensures that the motor receives constant voltage when the
power line voltage fluctuates, providing stable airflow and increasing operator, product, and
laboratory safety.

Standard Minihelic pressure gauge monitors loading of the supply and exhaust HEPA filters.

Unit has a recessed, hinged, front window of 0.25" (0.6cm) thick laminated safety glass with a
10.0" (25.4cm) operational access opening.

Pneumatic cylinder on the window provides smooth and effortless opening and closing.
Cylinder allows window to remain open at a minimum of 45 degrees above the horizontal plane
without assistance and ensures that it is completely closed, eliminating the need for latches.

Front access panel is easily removed for servicing HEPA filters from the front of the cabinet.
10 Class II, Type A Benchtop
Small cabinet footprint provides 2.9 sq. ft. (0.3 sq. m) of work area while requiring only
5.4 sq. ft. (0.5 sq. m) of valuable laboratory space.

Optional heavy-duty, tubular steel, modular stands in either 30.0" or 36.0" (76.2cm or Ambient Air
91.4cm) heights convert the benchtop model to a freestanding unit. Stands include
1.5" (3.8cm) leveling feet. Contaminated Air
(negative pressure)
Optional exhaust transition allows the cabinet to be exhausted to the outside. A
Contaminated Air
facility exhaust blower must be used. (positive pressure)

Model 1118 is CSA Certified for protection against electrical shock and fire hazards. HEPA Filtered Air

U.S. Patent held is 5,380,244.


TYPE A BENCHTOP SPECIFICATIONS

MODEL 1118

Type 2 Ft. Benchtop


Window Hinged window of 0.25" (0.6cm) thick laminated safety glass
Work Area 22.5"W x 26.5"H x 19.5"F-B (57.2cm x 67.3cm x 49.5cm)
Exterior Dimensions* 28.5"W x 58.5"H x 27.5"F-B (72.4cm x 148.6cm x 69.9cm)
Electrical 115V, 60 Hz, 6.7 FLA
Motor 1/3 HP
Breaker Requirements 15 Amp
Exhaust Requirements 136-151 CFM**
8.0" (20.3cm) diameter duct connection, 0.25" w.g. pressure drop at the connection
Weight
Net: 302 lbs. (137.0 kg)
Shipping: 368 lbs. (166.9 kg)

Continuing research and improvements may result in specification changes at any time.
*An additional 3.5" (8.9cm) must be added to the total height for the exhaust grille.
Deduct 2.0" (5.1cm) from the F-B dimension for the light canopy.
**Cubic Feet per Minute (CFM) is based on the Direct Airflow Method, under clean conditions.
220V, 50 Hz unit is available upon request (not CSA Certified).

FRONT VIEW

SIDE VIEW

11

See chart on page 14 for details about the optional stand.


CLASS II, TYPE B2 BENCHTOP MODELS
Forma Class II, Type B2 Benchtop Models provide maximum personnel and
product protection by maintaining ultra-clean laminar airflow and total exhaust of
the entire work zone air. Total exhaust makes units ideal for work that generates
non-explosive chemical vapors and gases. Type B2 units may be used for work
with biosafety levels 1, 2, and 3 biological agents, carcinogens, anti-neoplastic
drugs, and low level radioisotopes.

Interior work area and tray are constructed of heavy-gauge, type 304 stainless
steel to resist corrosion. Work tray, which provides a full 19.0" (48.3cm) front-
to-back work surface, easily lifts out of the work area without tools. Tray
seams, supports, and spill trough are welded in place for structural stability.

Interior side and rear walls feature unitized construction and metal plenums.
All contaminated plenums are under negative pressure to prevent leakage into
the environment.

Each cabinet is equipped with a built-in drain valve. Two service valves and
two plugged penetrations are provided, one on each side wall. Service valves
can be plumbed from the side or rear of the cabinet.

Two 12 Amp duplex electrical outlets with splash-proof covers are located
inside the work area. One outlet is on each side wall.

Unit incorporates two separate circuit breakers and two line cords, one for the
blower/motor and interior lights and one for the electrical outlets.
Model 1148
115V outlet is located on the exterior back wall to allow an interlock between
the cabinet and the in-house exhaust blower system control relay.

Motor size and method of mounting, combined with heavy-duty construction, provide exceptionally quiet operation and low vibration.

Two HEPA filters, one supply and one exhaust, are accessible from the front of the unit.

Full front-to-back and side-to-side stainless steel filter protector/diffuser, located in the work area, creates a uniform downflow of HEPA
filtered air and filter protection. This uniform downflow air provides turbulent-free product protection across the work area.

Unique design of recessed front grille provides additional protection to laboratory personnel.

All maintenance/service procedures can be performed conveniently from the front of the unit.

Adjustable, zero leak damper is provided in the 12.0" (30.5cm) supply air intake and the exhaust duct. Pressure differential switch
continuously monitors the exhaust pressure. In the event of inadequate exhaust, the pressure switch interrupts power to the supply
blower/motor to avoid blowing contaminated air onto you or into the lab.

Audible/visual alarms alert you to low exhaust flow conditions. When exhaust pressure returns to normal, the entire circuit automatically
12 resets. Normally open, normally closed, and common (NO, NC, COM) contacts are provided for remote monitoring.

Work area is easily accessible through a 19.0" (48.3cm) opening provided by the vertical sliding window. Audible alarm warns you if the
window is raised above the recommended operating height of 8.0" (20.3cm).

Voltage Compensating Speed Control ensures that the supply motor receives constant voltage when the power line voltage fluctuates,
providing stable airflow and increasing operator, product, and laboratory safety.

Standard Minihelic pressure gauge monitors loading of the supply filter.

Optional adjustable modular stand is available to convert the benchtop cabinet into a freestanding unit.

All 100V and 115V units are CSA Certified and NSF Listed.

U.S. Patent held is 5,380,244.



TYPE B2 BENCHTOP SPECIFICATIONS

MODEL 1148 1150 1168 1170


Type 4 Ft. Benchtop 6 Ft. Benchtop
Window Sliding window of 0.25" (0.6cm) thick laminated safety glass
Work Area 46.5"W x 25.3"H x 25.0"F-B 70.5"W x 25.3"H x 25.0"F-B
(118.1cm x 64.3cm x 63.5cm) (179.1cm x 64.3cm x 63.5cm)
Exterior Dimensions* 50.5"W x 69.8"H x 35.6"F-B 74.5"W x 69.8"H x 35.6"F-B
(128.3cm x 177.3cm x 90.4cm) (189.2cm x 177.3cm x 90.4cm)
Electrical 115V, 60 Hz, 6.0 FLA 100V, 50 Hz, 6.5 FLA 115V, 60 Hz, 10.0 FLA 100V, 50 Hz, 10.0 FLA
Motor 1/3 HP 3/4 HP 1/2 HP 3/4 HP
Breaker Requirements 15 Amp
Exhaust Requirements** 712-821 CFM 1,105-1,271 CFM
12.0" (30.5cm) diameter duct connection, 1.5 to 2.0" w.g. pressure drop at the connection
Weight
Net: 604 lbs. (274.0 kg) 904 lbs. (410.1 kg)
Shipping: 750 lbs. (340.2 kg) 975 lbs. (442.3 kg) 1,100 lbs. (499.0 kg) 1,175 lbs. (534.0 kg)

Continuing research and improvements may result in specification changes at any time.
*Deduct 3.5" (8.9cm) from the F-B dimension for the light canopy.
**Units require large volumes of supply and exhaust air, critical to the proper operation of the Class II, Type B2 Biological
Safety Cabinets. It is your responsibility to supply the correct volumes of make-up air.
220V, 50 Hz units are available upon request (not CSA Certified or NSF Listed).

SIDE VIEW

FRONT VIEW

13

See chart on page 14 for details about the optional stand.


ACCESSORIES FOR FORMA CLASS II
BIOSAFETY CABINETS
Forma Scientific provides a complete line of biological safety cabinet accessories. When combined with your biosafety cabinet, these
accessories create a complete workstation that provides comfort and convenience as well as functionality.

STANDS FOR FORMA BENCHTOP MODELS


Customer installed stands for benchtop models allow you to convert a benchtop model to a freestanding unit. Three types of Forma stands
available for our benchtop models include the Fixed Modular Stands for the Type A Benchtop, Adjustable Modular Stands for all Type A/B3
and B2 Models, and Hydraulic Crank-up Stands for Type A/B3 Models.

FIXED STAND FOR THE TYPE A BENCHTOP MODEL

NUMBER DESCRIPTION
191386 30.0" (76.2cm) High Modular Stand
191387 36.0" (91.4cm) High Modular Stand

MINIMUM HT. (LEVELER IN) MAXIMUM HT. (LEVELER OUT)


HEIGHT NO. 191386 NO. 191387 NO. 191386 NO. 191387
Stand Height 25.5" (64.8cm) 31.5" (80.0cm) 26.8" (68.1cm) 32.8" (83.3cm)
Work Surface Height 28.3" (71.9cm) 34.3" (87.1cm) 29.6" (75.2cm) 35.6" (90.4cm)
Total Height of Cabinet with Stand 84.0" (213.4cm) 90.0" (228.6cm) 85.3" (216.7cm) 91.3" (231.9cm)

ADJUSTABLE MODULAR STAND FOR TYPE A/B3 AND B2 BENCHTOP MODELS


Adjustable modular stands provide a work surface height from 26.5" to 35.8" (67.3cm to 90.9cm) heights in 1.0" increments. These stands are
designed in accordance with American Disabilities Act (ADA) guidelines.

14

NUMBER DESCRIPTION
191550 Adjustable Modular Stand for 4 Ft. Type A/B3 Benchtop Models
191551 Adjustable Modular Stand for 6 Ft. Type A/B3 Benchtop Models
191580 Adjustable Modular Stand for 4 Ft. Type B2 Benchtop Models
191581 Adjustable Modular Stand for 6 Ft. Type B2 Benchtop Models

HEIGHT MINIMUM HT. (LEVELER IN) MAXIMUM HT. (LEVELER OUT)


Stand Height 23.4" (59.4cm) 32.7" (83.1cm)
Work Surface Height 26.5" (67.3cm) 35.8" (90.9cm)
Total Height of Cabinet with Stand
Type A/B3:* 90.4" (229.6cm) 99.7" (253.2cm)
Type B2:** 93.2" (236.7cm) 102.5" (260.4cm)
*Recommended minimum clearance is 5.0" (12.7cm) additional above the unit for certification.
**Includes the height of the exhaust duct.
HYDRAULIC CRANK-UP STAND FOR TYPE A/B3 BENCHTOP MODELS
With a vertical travel range of 7.0" (17.8cm), the Forma Hydraulic Crank-up Stand allows you to define your own level of comfort and
convenience by adjusting the work surface height anywhere between 27.6" and 35.8" (70.1cm and 90.9cm). The stand includes 1.3" (3.3cm)
adjustable levelers. The actuator handle folds conveniently out of the way.

NUMBER DESCRIPTION
191518 Hydraulic Crank-up Stand for 4 Ft. Type A/B3 Benchtop Models
191519 Hydraulic Crank-up Stand for 6 Ft. Type A/B3 Benchtop Models

HEIGHT MINIMUM HT. (LEVELER IN) MAXIMUM HT. (LEVELER OUT)


Stand Height 24.6" (62.5cm) 32.8" (83.3cm)
Work Surface Height 27.6" (70.1cm) 35.8" (90.9cm)
Total Height of Cabinet with Stand 91.6" (232.7cm) 99.8" (253.5cm)
*Recommended minimum clearance is 5.0" (12.7cm) additional above the unit for certification.

DRAWER ASSEMBLIES
Customer installed drawer assemblies provide convenient storage for pipettes, gloves, safety glasses, and other items used in the laboratory. Two,
three, and four position drawer assemblies mount directly on our 4 ft. and 6 ft. hydraulic crank-up and adjustable modular stands or can be used
as an island storage unit when mounted on the optional roller base. The drawer dimensions are 6.8"W x 3.3"H x 20.1"F-B (17.3cm x 8.4cm x
51.1cm).

15

NUMBER DESCRIPTION
191494 Left Side, Four Drawer Assembly for Adjustable Modular Stand Only
Fits Type A/B3 Benchtop and Type B2 Benchtop Models
Dimensions: 9.0"W x 18.0"H x 21.5"F-B (22.9cm x 45.7cm x 54.6cm)
191495 Right Side, Three Drawer Assembly for Hydraulic Crank-up or Adjustable Modular Stand
Fits Type A/B3 Benchtop and Type B2 Benchtop Models
Dimensions: 9.0"W x 18.0"H x 21.5"F-B (22.9cm x 45.7cm x 54.6cm)
191524 Left Side, Two Drawer Assembly for Hydraulic Crank-up Stand Only
Fits Type A/B3 Benchtop Models
Dimensions: 9.0"W x 10.1"H x 21.5"F-B (22.9cm x 25.7cm x 54.6cm)
191591 Roller Base for Drawer Assembly
Adds only 3.0" (7.6cm) to the drawer assembly height
Dimensions: 9.4"W x 3.6"H x 21.9"F-B (23.9cm x 9.1cm x 55.6cm)
ULTRAVIOLET (UV) LIGHTS
Ultraviolet lights are available for all Forma Biological Safety Cabinets.

NUMBER DESCRIPTION
191069 15 Watt Portable UV Light for the Type A Benchtop Model
191070 30 Watt Portable UV Light for 4 and 6 Ft. Cabinets
191418 15 Watt UV Light for the Type A Benchtop Model, Customer Installed
191419 30 Watt UV Light for 4 and 6 Ft. Cabinets, Customer Installed

EXHAUST TRANSITIONS
Customer installed exhaust transitions with an 8.0" (20.3cm) diameter duct and zero leak damper allow you to vent your cabinet to the outside
environment.

NUMBER DESCRIPTION
191217 Exhaust Transition Thimble without Damper for the Type A Benchtop Model
191478 Airtight or Thimble Exhaust Transition with Zero Leak Damper for Type A Console Models
191570 Airtight or Thimble Exhaust Transition with Zero Leak Damper for Type A/B3 Benchtop Models

LOW AIRFLOW ALARMS


Customer installed low airflow alarms monitor the airflow in the biosafety cabinets. Alarms feature standard NO, NC, and COM dry contacts,
rated at 115V nominal.

16

NUMBER DESCRIPTION
191168 120V, 50/60 Hz Low Airflow Alarm
191380 230V, 50/60 Hz Low Airflow Alarm
IV HANGER AND RODS
Heavy-duty, stainless steel IV rods are pre-assembled for easy installation without the use of tools.

NUMBER DESCRIPTION
191190 IV Bag Hanger
191571 4 Ft. IV Rod with Four Hangers
191572 6 Ft. IV Rod with Four Hangers
191576 2 Ft. IV Rod with Four Hangers

ADDITIONAL ACCESSORIES
Additional accessories also complement Forma Biological Safety Cabinets to create complete workstations that meet your daily work performance
requirements as well as your comfort and convenience needs.

NUMBER DESCRIPTION
191127 Adjustable Footrest for All Models
191416 Locking Caster Set for Type A Console Models (adds 3.3" (8.4cm) to the maximum height of the cabinet), Customer Installed
191509 Front Cabinet Armrest for 4 Ft. Type A/B3 Benchtop Models, Customer Installed
191512 Front Cabinet Armrest for 6 Ft. Type A/B3 Benchtop Models, Customer Installed
191486 Lab Chair with Arms for All Models (meets Class 1,000 clean air specifications)
191487 Lab Chair without Arms for All Models (meets Class 1,000 clean air specifications)
191275 Service Valve Kit for All Models, Side Mounted, Customer Installed
191620 Universal Piping Option for Type A/B3 Benchtop Models, Factory Installed
191597 Service Valve Kit for Type A/B3 Benchtop Models (used with No. 191620), Customer Installed
800045 Rear Piping Kit for Type A Console and Type B2 Benchtop Models (includes street ell and spacer), Customer Installed
191496 ADA Control Panel for Type A/B3 Benchtop Models, Factory Installed
188009 Waste Bag Holder Assembly for All Models
921025 Clean Wastebasket Poly Bags, 100 Bags, 6.0" x 6.0" x 12.0" (15.2cm x 15.2cm x 30.5cm)
921036 Absorbent Mats, 400 Disposable 18.0" x 20.0" (45.7cm x 50.8cm) Mats
191056 Ground Fault Interrupt (GFI) Duplex Outlet for Type B2 Benchtop Models, Factory Installed
191140 Extra Thru-Wall Port for Type A Benchtop and Console Models and Type B2 Benchtop Models (specify left or right side
wall mount), Factory Installed
760190 Supply and Exhaust ULPA Filters for 4 Ft. Type A Consoles, Factory Installed
760191 Supply and Exhaust ULPA Filters for 6 Ft. Type A Consoles, Factory Installed
760192 Supply and Exhaust ULPA Filters for 4 Ft. Type A/B3 Benchtop Models, Factory Installed
760193 Supply and Exhaust ULPA Filters for 6 Ft. Type A/B3 Benchtop Models, Factory Installed
760197 Supply and Exhaust ULPA Filters for 4 Ft. Type B2 Benchtop Models, Factory Installed
6001284 Installation and Operational Qualifications (IQ/OQ) Protocol for Biological Safety Cabinet Models 1148, 1168, 1200,
1214, 1284, and 1286

17

Waste bag holder assembly, absorbent mat, and armrest Adjustable footrest and lab chair without arms
RECOMMENDED BIOSAFETY LEVELS
Table 1. Summary of recommended biosafety levels for infectious agents1
BIOSAFETY SAFETY EQUIPMENT FACILITIES
LEVEL AGENTS PRACTICES (PRIMARY BARRIERS) (SECONDARY BARRIERS)

1 Not known to consistently cause Standard microbiological practices None required Open benchtop sink required
disease in healthy adults

2 Associated with human disease. BSL-1 practice plus: Primary barriers = Class I or II BSCs or BSL-1 plus:
Hazard: percutaneous injury, Limited access other physical containment devices used Autoclave available
ingestion, mucous membrane Biohazard warning signs for all manipulations of agents that cause
exposure Sharps precautions splashes or aerosols of infectious materials;
Biosafety manual defining any PPEs: laboratory coats; gloves; face
needed waste decontamination protection as needed
or medical surveillance policies

3 Indigenous or exotic agents with BSL-2 practice plus: Primary barriers = Class I or II BSCs or BSL-2 plus:
potential for aerosol transmission; Controlled access other physical containment devices used Physical separation from access
disease may have serious or Decontamination of all waste for all manipulations of agents; corridors
lethal consequences Decontamination of lab clothing PPEs: protective lab clothing; gloves; Self-closing, double door access
before laundering respiratory protection as needed Exhausted air not recirculated
Baseline serum Negative airflow into laboratory

4 Dangerous/exotic agents that BSL-3 practice plus: Primary barriers = All procedures conducted BSL-3 plus:
pose high risk of life threatening Clothing change before entering in Class III BSCs or Class I or II BSCs in Separate building or isolated zone
disease, aerosol-transmitted lab Shower on exit combination with full-body, air-supplied, Dedicated supply and exhaust, vacuum,
infections, or related agents with All material decontaminated positive pressure personnel suit and decontamination systems
unknown risk of transmission on exit from facility Other requirements outlined in the text

Table 2. Summary of recommended biosafety levels for activities in which experimentally or naturally infected vertebrate
animals are used1
BIOSAFETY SAFETY EQUIPMENT FACILITIES
LEVEL AGENTS PRACTICES (PRIMARY BARRIERS) (SECONDARY BARRIERS)

1 Not known to consistently cause Standard animal care and manage- As required for normal care of each species Standard animal facility
disease in healthy adults ment practices, including appropriate Non-recirculation of exhaust air
medical surveillance programs Directional airflow recommended
Handwashing sink recommended

2 Associated with human disease. ABSL-1 practices plus: ABSL-1 equipment plus ABSL-1 facility plus:
Hazard: percutaneous exposure, Limited access primary barriers: containment equipment Autoclave available
ingestion, mucous membrane Biohazard warning signs appropriate for animal species; Handwashing sink available in the
exposure Sharps precautions PPEs: laboratory coats, gloves, face and animal room
Biosafety manual respiratory protection as needed Mechanical cage washer used
Decontamination of all infectious
wastes and of animal cages prior
to washing

3 Indigenous or exotic agents with ABSL-2 practices plus: ABSL-2 equipment plus: ABSL-2 facility plus:
potential for aerosol transmission; Controlled access Containment equipment for housing Physical separation from access
disease may have serious health Decontamination of clothing animals and cage dumping activities corridors
effects before laundering Class I or II BSCs available for Self-closing, double door access
Cages decontaminated before manipulative procedures (inoculation, Sealed penetrations
bedding removed necropsy) that may create infectious Sealed windows
18 Disinfectant foot bath as needed aerosols; Autoclave available in facility
PPEs: appropriate respiratory protection

4 Dangerous/exotic agents that ABSL-3 practices plus: ABSL-3 equipment plus: ABSL-3 facility plus:
pose high risk of life threatening Entrance through change room Maximum containment equipment Separate building or isolated zone
disease, aerosol transmission, or where personal clothing is (i.e., Class III BSC or partial containment Dedicated supply and exhaust,
related agents with unknown risk removedand laboratory clothing equipment in combination with full body, vacuum, and decontamination systems
of transmission is put on; shower on exiting; air-supplied positive pressure personnel Other requirements outlined in
All wastes are decontaminated suit) used for all procedures and activities the text
before removal from the facility

1. CDC NIH, 4th Edition, Biosafety in Microbiological and Biomedical Laboratories, U.S. Government Printing Office, Washington, D.C., 1999, pp. 52, 75.
LEADER IN QUALITY AND SAFETY
Since 1950, the name Forma has represented high quality, reliable laboratory equipment for a wide variety of applications. As the leader in
quality, Forma Biological Safety Cabinets are manufactured to conform with the intense guidelines of third party testing facilities for your
personal safety and product protection. Formas commitment to you is reinforced by our adherence to world quality standards.

UL LISTING, CSA CERTIFICATION, AND NSF LISTING


UL Listing (Standard UL 1262), CSA Certification (Standard C22.2 No. 151-M1986), and NSF Listing (Standard NSF 49) attest to the
electrical, mechanical, and biological safety of our biological safety cabinets. Electrical requirements are in compliance with Underwriters
Laboratories and/or CSA International. NSF International Listing attests to the biological safety of the cabinets. Stringent testing is
performed on Forma Biological Safety Cabinets before they carry the UL, CSA, and NSF markings. Copies of UL, CSA, and NSF
Listing/Certification are available from Forma Scientific for compliance verification.

FACTORY PRE-TEST
Before shipment, each Forma Biological Safety Cabinet undergoes a number of tests to ensure that the unit operates as designed and meets or
exceeds required specifications. These tests include visual inspection, electrical safety, unit performance, HEPA filter integrity, cabinet
integrity, and airflow specifications. For additional customer comfort and safety, cabinets are also tested to NSF Standard 49 for sound, light,
and vibration. A copy of the test report is supplied with each unit.

ISO 9001
As a global leader in innovation, Forma Scientific continues its pursuit of excellence by maintaining ISO 9001 registration. This advanced
quality system reinforces Formas commitment to provide the highest quality products available. The ISO 9000 series is a set of international
requirements for establishing and maintaining quality management systems. It is based on the widely recognized quality assurance standards
of the British Standards Institute (BSi). Continued compliance is assured by ongoing internal audits and periodic audits by the assessing
agency. Forma Scientific is proud of the fact that we are ISO 9001 registered.

3 YEAR WARRANTY
All Forma Biological Safety Cabinets feature a three year warranty on parts and labor, enabling us to offer the most comprehensive warranty
of any manufacturer in the industry.

CERTIFIERS AND SERVICE


To ensure maximum personnel safety, Forma recommends the inspection and certification of all biological safety cabinets by a qualified
certifier. Forma recommends that a cabinet should be certified under the following conditions: new installations, annually, if the cabinet has
been moved, or after service work has been performed inside the cabinet. A partial list of qualified certifiers is provided in the operation
manual and on our homepage.

To save you both time and expense, our toll-free Technical Service Department (1-888-213-1790) provides access to qualified telephone
technicians who can help you troubleshoot and make adjustments to your Forma equipment. In an ongoing effort to better serve you, our
Technical Service Department and Sales Support Department (1-800-848-3080) are available from 8:00 a.m. to 7:00 p.m. (EST).
19
For added coverage, we offer an Extended Warranty to repair or replace component parts proven to be non-conforming once the original
warranty period has ended. For additional convenience, we also provide Preventive Maintenance Agreements. Contact our Technical
Service Department for details.

OTHER FORMA LAMINAR AIRFLOW PRODUCTS


In addition to our line of Forma Biological Safety Cabinets, we offer a complete line of Laminar Airflow Equipment, including Benchtop and
Console Workstations, Small Animal Isolators, and Custom Enclosures.
Forma offers the following products:
Cell Culture Incubators Orbital Shakers

In Vitro Fertilization (IVF) Incubators Monitor/Alarm System

Environmental Chambers Programmable Centrifuges

Ultra-Low Temperature Freezers Lab Washers and Dryer

Laminar Airflow Equipment Lab Refrigerators

Biological Safety Cabinets Lab Freezers

Cryopreservation Equipment Chromatography Refrigerators

Blood Bank Equipment Custom-Designed Equipment

P.O. Box 649, Marietta, OH 45750


1-740-373-4763 Fax: 1-740-373-6770
USA and Canada 1-800-848-3080
International Distributors Worldwide
An ISO 9001 Company
http://www.forma.com

A Thermo Electron Company


Rev. 0
7/00