Foreword
Thank you for using the CIT-8920 Inductive Conductivity Transmitter
Carefully read the relevant parts of this manual before installation and operation to prevent
wrong operation, measurement error and damage of instrument. Improper installation and
unsuitable flow speed will cause big measurement error, so please read the installation passage in
detail.
This instrument is precise electrochemical measurement, and its installation and operation
should be performed by technicians with relevant professional knowledge.
Difficulties in the process of installation or using please timely contact the service
department.
Be sure to check the complete list and your received actual products, if missing or damaged
please contact the company.
Solemn commitment of the manufacturer
1. The meters quality guarantee is one year from the date of purchasing. During this period, if
the meter has quality problems, manufacturer is responsible for maintenance work for free or
changes it.
2. Manufacturer offers the maintenance service for whole life of the sold meters.
3. If the damage of the meter is caused by the following reasons, it is out of the maintenance
service:
AThe meter is burned or foundered caused by improper usage and maintenance.
BThe meter is refitted or misused without permit.
CThe meter is destroyed under the condition out of companys regulation.
DThe relevant damage caused by choosing the wrong type.
EThe cable damage and rupture caused by improper installation and usage.
F The incorrect measurement of the sensor caused by disconnecting or connecting wires
personally.
GThe inner broken wire caused by indiscreetly disassembling.
When this symbol appears in the manual, means please pay attention to something about the
safe, installation, product function and using
*Without the influence on the operation, any small change or improvement on the products by the
manufacturer will not be notified separately. Please make the object as the standard.
-2-
Content
IOVER VIEW................................................................................................................ 4
1.1 Performance and characteristics.................................................................. 5
1.2 Working Principle.......................................................................................... 6
1.3 Application......................................................................................................7
1.4 Technical parameters................................................................................... 12
IIINSTALLATION...................................................................................................... 13
2.1 Trasmitter and Sensor...................................................................................... 13
2.1.1 Integrated Installation...................................................................................13
2.1.2 Split Installation ...........................................................................................13
2.1.3 Submerged Installation................................................................................. 13
2.2 Electrical Installation....................................................................................... 14
2.3 Sensor and components.................................................................................... 14
2.3.1 Online analysis Sensor...................................................................................14
2.3.2 Sensor installation components.................................................................... 14
2.3.3 Installation Precautions.................................................................................14
2.4 Terminal block definition................................................................................. 15
2.4.1Terminal wire connection instruction....................................................... 15
2.4.2Output and device connection diagram........................................................ 15
IIIDEBUGGING..........................................................................................................16
-3-
IOver View
measurement.It is also a remote transmitter which can match with magnetic induction
electrolyte (or percent) .Widely used for continuous measurement of conductivity and
environment.
The sensor has the anti-pollution ability to oil pollution and the process coating
medium. There is no electrode plate directly touch the test solution, so there is no
polarization phenomenon. It will not be affected by the coating or the film form;
The sensor material has the advantages of anti-pollution, high temperature and
corrosion resistance;
The transmitter has a local display function, using 128x64 resolution LCD
Suitable for industrial production needs of the pipeline, sink and other forms of
installation. This sensor also has a variety of installation methods and accessories,
Can output (4 ~ 20) mA analog signal, and can choose to output in the
instrument module has self-learning function, for other unknown curve solution,
the user can calibrate the instrument through the custom function;
Materials that make up sensors and accessories have a variety of options for
different media, temperature range, strong corrosive and hygienic level options,
automatic identification;
concentration and conductivity of the relationship between the solution to achieve the
(primary coil) and a measuring coil (secondary coil). When the alternating current
flows through the primary coil, there will be alternating magnetic field in the
immersed solution, and the secondary coil will sense the alternating current generated
1.3 Applications
electroplating process;
-6-
Product testing of food, beverage, wine and dairy industries;
Consumption 2.5W
Protection
IP65
Level
Working Temperature(050)Humidity95%RHNone condensation
Environment
Storage Temperature-2060Humidity85%RHNone condensation
II. Installation
2.1 Transmitter and Sensor
-8-
Note: Before inserting the transmitter into the mounting seat, it is necessary to
immerse the front of the sensor with water. It is easy to install by means of water
lubricity, so as to avoid damage to the "O" type seal due to dry friction.
The sensor should be installed directly through the eight-core air socket with
-9-
2.1.3 Submerged Installation
The sensor and indicator can realize submerge measurement by using P16-Z
Connect the cable Conduction measurement signal can not be mixed with
at least 30 cm or shielded wire, must use high-quality power supply (after filtering
and adjustment).
Open Locked
- 10 -
1. Anticlockwise loosen PG9 waterproof lock put DC24V two core shielded
wire insert suitable length from PG9 waterproof Anticlockwise open the upper
case as the picture shows.
2. DC24V respectively 24V (A), 24V (B), internal automatic polarity recognition, EARTH end
close to the ground, to eliminate interference;
3. After the other wiring is completed, tighten the upper case clockwise, while tightening PG9
waterproof lock mother, positioning logo must be aligned.
The transmitter controller and the inductive sensor are the basic ordering unit,
and the sensor mounting kit is the optional unit. Select the sensor and installation kit
for your sensor according to the measurement object and the scene. Please install
the service before ordering. Please design separately according to your needs.
- 11 -
Note: Split sensors are only available for tubing mounting components, and submerged sensors
are available for wall-mounted components, tank-type mounting assemblies and eco-mount
mounting components.
- 12 -
Inductive Conductivity Sensor(Submerged) Inductive Conductivity Sensor(Split)
- 13 -
3. Ensure the consistency between sensor deflector hole direction and water
direction in the pipeline during pipe installation, and the gap between the sensor and
4. Select flange protective tube while installing pot kettle with stirrer device, avoid
- 14 -
2.4.1 Terminal wire connection instruction
485A/485B/485
RS485 communication interface
G
- 15 -
2.4.2 Output and device connection diagram
. Debugging
Key operation
- 16 -
Key Name Function
Select the bits cyclically under parameter
setting;
Bit key
Check measured parameters;
Return to previous menu or main interface.
adjust value of the selected bit under parameter
Add key setting;
Select up or down under parameter setting.
1. Enter login interface;
Enter
2. Confirm and save parameters and enter the
key
next menu.
Instrument has undergone parameter setting before delivery. In case test conditions
change (such as, replacement of sensor, change of alarm setting, etc.), check data
In the main interface, press to enter output interface; specific meaning and
Output configuration
Relay configuration
Check current configuration and control value of relay in this menu according to the
unit of ON and OFF: the unit of conductivity means that current configuration of
the relay is the conductivity; the unit of temperature means that current
Check corresponding value of 4mA and 20mA of transmission output in this menu
- 17 -
Note: under view status, press or 5s in any interface and there will be
no action, the instrument returns to the main interface automatically
Measurement pattern
Setting pattern
Measured parameters
Control parameters
Output setting
Backlight setting
Password setting
Calibration
Language selection
Manufacturer information
Engineering user
Ordinary user
Note: set and check any parameter and calibrate the instrument under
password is 1999); only instrument parameters can be set and ordinary user
- 18 -
password can be set under ordinary user right. Ordinary user default password is
0000.
Unit of main measurement and temperature can be set in this menu. After unit
main interface and unit of ON/OFF values of control OC door will be correlated
Temperature unit
Conductivit
Measure
ment unit
Temperature unit can
and convert the measured value of conductivity into the measured value of
measured. First, prepare the solution with known concentration; then measure the
conductivity value of solution using the instrument; finally input values of solution
according to hints on the interface. The instrument will prepare the concentration
curve according to input data. After inputting and saving data and correctly set the
warming coefficient of the medium solution, the instrument can measure the
concentration.
Note: user-defined concentration must contain 2-8 data points (the greater the
number of data points is, the higher the accuracy is). Each point must have a
conductivity value of each continuous data point must be monotone increasing, and
(that is, concentration values in the data sheet must be monotone increasing or
- 20 -
3.2.2 Control parameters
Control parameters
Conductivity:Normally open
Normally close
Pulse
Normally open
Minimum pulse
Times/min
Corresponding value
Maximum pulse
Corresponding value
and pulse.
of alarm;
case of alarm;
- 21 -
Pulse: open and close according to set frequency and correspondence;
Conductivity Conductivity
Output setting
mA output: 4mA
20mA
Local address
Set Baud rate and local address used in 485 communication in this menu
- 22 -
Baud rate: 9600, 4800, 2400, 1200
Note: select mA output: I+/485B and I-/485B Ports are used for mA output
Select 485output: I+/485B and I-/485B Ports are used for 485 communication
Delayed close
Automatic close of screen backlight can be set in this menu when the
Password setting:
Ordinary user
Engineering user
- 23 -
Press and in this menu to set ordinary user password, then
3.2.6 Calibration
CalibrationZero calibration.
Keep sensor dry and suspend it, then press and the instrument starts
conductivity.
Calibrate the instrument using standard liquid in this menu. Prepare the
- 24 -
3.2.7 Language setting
Language selection :
Chinese
English
. Failure predication
Phenomen
Possible Factor Elimination Method
on
DC24V is not
transportation and
- 25 -
storage result in
damage
Constant
setting error
Bubble on
Measuring
through-hole is
blocked by sludge
of mA sending
. Product completeness
Transducer controller 1 set
. Ordering instruction
Reserve cable length of split type sensor during order placement; in case of no
For the best material quality, indicate measuring medium or application field
- 27 -
Appendix: Sensor application guide
Main Application
Application Case
Material Industry
PP scrubbing solution
concentration
Steam
Boiler blow-off, fuel gas scrubber
production
- 28 -
idene Chemistry and
Glycerol, humidity detection, scrubber,
Fluoride) petroleum refining
sewage
industry
Metal extracting
Metal recovery, copper flotation
and mining industry
Chemistry and
Petroleum drilling fluid, leakage detection,
PEEK petroleum refining
spill detection
(Polyether industry
treatment detection
- 29 -
Executive standard: Q /HKY01
Website: http://www.createc.cn
Email: webmaster@createc.cn
Address: 1F, Building 2, Jinshui Industrial Park, No. 368, Xinshi North
service)
- 30 -