AT O M I C A B S O R P T I O N

P R O D U C T
MHS-15 Mercury/Hydride System

N O T E
Minimum investment and Simple operation Key Benefits
low operating costs
The MHS-15 system includes a Minimum investment
The MHS-15 Mercury/Hydride reaction assembly and a quartz- Low operating costs
System is a manual accessory for cell assembly. The analyzer is free-
high-sensitivity determination Simple operation
standing and is placed adjacent
of mercury and hydride-forming to the AA spectrometer sample Works with all PerkinElmer
elements, such as As, Se, Sb, Te, atomic absorption spectrometers
compartment. It includes a reaction
Bi and Sn, by flame atomic absorp- flask, a reservoir for the reducing
tion (AA) spectroscopy. Mercury/ agent and all pneumatic compo-
hydride atomic absorption is the nents for carrier transportation
preferred technique for trace deter- of the metallic vapors to the
mination of mercury and the metal- quartz cell.
lic hydride-forming elements,
because it offers the best possible The cell mounts on a standard
detection limits, down to the ng 10-cm PerkinElmer® burner head,
range, while minimizing capital permitting the cell to be heated in
investment and operating costs. the spectrometer flame. The quartz
The technique is simple, easy-to- cell is positioned in the light beam
handle, and reliable and has been of the spectrometer and alignment
proven in tens of thousands of is provided by the instrument’s
laboratories worldwide. burner adjustment controls. Its

w w w. p e r k i n e l m e r. c o m
quartz cell can be easily tilted out of By manually actuating a pneumatic cury into its metallic state. Excess
the flame using a swivel mount. valve, an argon stream transports the hydrogen and the argon carrier gas
reducing agent (NaBH4) from a reser- transport the hydride (or metallic
Due to the reaction chemistry in the
voir into the sample. The special mercury) vapor into the quartz cell
mercury/hydride process, the ana-
conical form of the immersion tube where the absorption of the metal is
lyte is completely separated from its
and reaction flask provides excellent measured. During mercury determi-
matrix during analysis, eliminating
mixing of the sample and NaBH4, nation, the cell remains at ambient
interferences.
guaranteeing spontaneous reaction. temperatures. Mercury determina-
No stirring device is required for tion can also be performed using
Operating procedure this procedure. stannous chloride (SnCl2) as the
During analysis, the sample contain- reducing agent. For the determina-
Hydrogen, liberated during the reac-
ing the analyte in ionic form is placed tion of hydride-forming elements,
tion, reduces the metal ions to gaseous
in the reaction flask. The flask is then the quartz cell is heated by an
hydrides. In the case of mercury
installed into the analyzer assembly. air/acetylene flame to thermally
determination, NaBH4 reduces mer-
decompose the metal hydrides.

Detection limits using Technical data


the MHS-15 with the
Absorption cell Quartz cell, 165 mm long, 12 mm diameter
AAnalyst™ 700
Cell heating Up to approx. 1000 °C with air/acetylene flame
(EDL lamps were used for all elements)
Reaction flask Polypropylene flask with special, conically formed inner side.
Element Absolute (ng) Relative (µg/L) Connection to the analyzer assembly with special joint
(in 10-mL (in 50-mL with O-ring gasket.
volume) volume) Sample volume Max. 50-mL volume
As 1.0 0.02 Reducing agent NaBH4 or SnCl2
Se 1.0 0.02 Inert gas supply Argon, inlet pressure 2.5-3.5 bar (36-51 psig)
Bi 2.0 0.10 approx. consumption 1.1 L/min

Sb 5.0 0.10 Power supply No electrical supply is required.

Te 2.0 0.05 Dimensions 70 mm wide x 160 mm deep x 375 mm high

Hg 1.0 0.02 Weight Approximately 4.3 kg

Ordering information
Part Number Description
B3140760 MHS-15 Mercury/Hydride System – Manually operated system for the high-sensitivity determination of Hg
and metallic hydride-forming elements such as As, Bi, Sb, Sn, Se and Te. Includes analyzer with sample flask
and reagent reservoir. Manual selection of NaBH4 or SnCl2 reductant solution. Also includes a quartz cell, which
may be heated with an air/acetylene flame. No electrical connections are required. For use with all PerkinElmer
flame AA instruments.
B3000350 Cell holder – To mount an MHS-15 quartz cell on the burner system.
B0094321 Reaction flask for MHS-15 – Specially formed polypropylene reaction vessel with cover.
B0087354 Reductant reservoir for MHS-15
B0094415 Quartz cell for MHS-15 – Complete with graphite cooling rings and silicone rubber sleeves.

PerkinElmer Life and


Analytical Sciences
710 Bridgeport Avenue
Shelton, CT 06484-4794 USA
Phone: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com

For a complete listing of our global offices, visit www.perkinelmer.com/lasoffices

©2004 PerkinElmer, Inc. All rights reserved. The PerkinElmer logo and design are registered trademarks of PerkinElmer, Inc. AAnalyst is a trademark and PerkinElmer is a registered trademark
of PerkinElmer, Inc. or its subsidiaries, in the United States and other countries. All other trademarks not owned by PerkinElmer, Inc. or its subsidiaries that are depicted herein are the
property of their respective owners. PerkinElmer reserves the right to change this document at any time without notice and disclaims liability for editorial, pictorial or typographical errors.

006385C_01 ELEC110400 Printed in USA