Вы находитесь на странице: 1из 25

ARTICLE 10

T-1000 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189


T-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
T-1020 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
T-1021 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
T-1022 Referencing Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
T-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
T-1031 Gages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
T-1040 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1041 Cleanliness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1042 Openings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1043 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1044 Pressure /Vacuum (Pressure Limits) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1050 Procedure /Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1051 Preliminary Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1052 Specific Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1060 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1061 Pressure /Vacuum Gages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1062 Temperature Measuring Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1063 Calibration Leak Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
T-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
T-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
T-1081 Acceptance Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
T-1090 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
T-1091 Test Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
T-1092 Record Retention . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191

Mandatory Appendices
Appendix I Bubble Test Direct Pressure Technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1000 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1031 Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1032 Bubble Solution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1033 Immersion Bath. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1071 Soak Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1072 Surface Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1073 Application of Solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1074 Immersion in Bath . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
I-1075 Lighting and Visual Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193

185

ASME B&PVC sec5$$u165 05-24-99 12:29:59 pd: sec5 Rev 14.04


I-1076 Indication of Leakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
I-1077 Post-Test Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
I-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194

Appendix II Bubble Test Vacuum Box Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194


II-1000 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1031 Bubble Solution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1032 Vacuum Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1033 Vacuum Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1071 Surface Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1072 Application of Solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1073 Vacuum Box Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
II-1074 Pressure (Vacuum) Retention . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
II-1075 Vacuum Box Overlap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
II-1076 Lighting and Visual Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
II-1077 Indication of Leakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
II-1078 Post-Test Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
II-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195

Appendix III Halogen Diode Detector Probe Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195


III-1000 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
III-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
III-1011 Alkali-Ion Diode (Heated Anode) Halogen Leak Detectors . . . . . . . . . . . . . . . . . . . . . . 195
III-1012 Electron Capture Halogen Leak Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
III-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
III-1031 Tracer Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
III-1032 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
III-1033 Capillary Calibration Leak Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1060 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1061 Standard Leak Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1062 Warm Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1063 Scanning Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1064 Frequency and Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1071 Location of Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1072 Concentration of Tracer Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1073 Soak Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
III-1074 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
III-1075 Scanning Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
III-1076 Scanning Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
III-1077 Leakage Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
III-1078 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
III-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197

Appendix IV Helium Mass Spectrometer Test Detector Probe Technique . . . . . . . . . . . . . . . . . . . 197


IV-1000 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
IV-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
IV-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197

A99 186

ASME B&PVC sec5$$u165 05-24-99 12:29:59 pd: sec5 Rev 14.04


IV-1031 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
IV-1032 Auxiliary Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
IV-1033 Calibration Leak Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
IV-1060 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
IV-1061 Instrument Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
IV-1062 System Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
IV-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
IV-1071 Location of Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
IV-1072 Concentration of Tracer Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
IV-1073 Soak Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
IV-1074 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
IV-1075 Scanning Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
IV-1076 Scanning Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
IV-1077 Leakage Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
IV-1078 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
IV-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199

Appendix V Helium Mass Spectrometer Test Tracer Probe and Hood Techniques . . . . . . . . . . 199
V-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
V-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199.1
V-1031 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199.1
V-1032 Auxiliary Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
V-1033 Calibration Leak Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
V-1060 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
V-1061 Instrument Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
V-1062 System Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
V-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
V-1071 Tracer Probe Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
V-1072 Hood Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
V-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
V-1081 Tracer Probe Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
V-1082 Hood Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202

Appendix VI Pressure Change Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202


VI-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
VI-1020 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
VI-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
VI-1031 Pressure Measuring Instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
VI-1032 Temperature Measuring Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
VI-1060 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
VI-1061 Pressure Measuring Instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
VI-1062 Temperature Measuring Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1071 Pressure Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1072 Vacuum Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1073 Test Duration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1074 Small Pressurized Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1075 Large Pressurized Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1076 Start of Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1077 Essential Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203

187 A99

ASME B&PVC sec5$$u165 05-24-99 12:29:59 pd: sec5 Rev 14.04


VI-1081 Acceptable Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VI-1082 Rejectable Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203

Appendix VII Glossary of Terms for Leak Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203


VII-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
VII-1020 General Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
VII-1030 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204

Appendix VIII Thermal Conductivity Detector Probe Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204


VIII-1000 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
VIII-1010 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1011 Thermal Conductivity Leak Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1030 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1031 Tracer Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1032 Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1033 Capillary Calibration Leak Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1060 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1061 Standard Leak Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1062 Warm-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1063 Scanning Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1064 Frequency and Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1
VIII-1070 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1071 Location of Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1072 Concentration of Tracer Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1073 Soak Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1074 Scanning Distance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1075 Scanning Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1076 Scanning Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1077 Leakage Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1078 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2
VIII-1080 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.2

Table
III-1031 Tracer Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
VIII-1031 Tracer Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204.1

Nonmandatory Appendix
Appendix A Supplementary Leak Testing Formula Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
A-10 Applicability of the Formulas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205

A99 188

ASME B&PVC sec5$$u165 05-24-99 12:29:59 pd: sec5 Rev 14.04


ARTICLE 10
LEAK TESTING

T-1000 INTRODUCTION (a) extent of the examination;


A99 T-1010 SCOPE (b) type of equipment to be used for detecting leaks
This Article describes methods and requirements for or measuring leakage rates;
the performance of leak testing. (c) surface cleanliness preparation and type of equip-
(a) When a leak testing method or technique of ment used;
Article 10 is specified by a referencing Code Section, (d) method or technique of the test that will be
the leak test method or technique shall be used together performed;
with Article 1, General Requirements. (e) temperature, pressure, gas, and percent concentra-
(b) Definition of terms used in this Article are in tion to be used.
Mandatory Appendix VII of this Article.
(c) The test methods or techniques of these methods T-1022 Referencing Code
can be used for the location of leaks or the measurement
of leakage rates. For the leak testing method(s) or technique(s) speci-
The Nonmandatory Standards of Article 27 may be fied by the referencing Code, the referencing Code
used in the preparation of leak test specifications. The Section shall then be consulted for the following:
specific test method(s) or technique(s) and Glossary of (a) personnel qualification /certification
Terms of the methods in this Article are described in (b) technique(s) /calibration standards
Mandatory Appendices I through VIII and Nonmanda- (c) extent of examination
tory Appendix A as follows: (d) acceptable test sensitivity or leakage rate
Appendix I Bubble Test Direct Pressure Tech- (e) report requirements
nique (f) retention of records
Appendix II Bubble Test Vacuum Box Tech-
nique
Appendix III Halogen Diode Detector Probe Test T-1030 EQUIPMENT
Appendix IV Helium Mass Spectrometer Test T-1031 Gages
Detector Probe Technique
Appendix V Helium Mass Spectrometer Test (a) Gage Range. When dial indicating and recording
Tracer Probe and Hood Techniques pressure gage(s) are used in leak testing, they should
Appendix VI Pressure Change Test preferably have the dial(s) graduated over a range of
Appendix VII Glossary of Terms about double the intended maximum pressure, but in
Appendix VIII Thermal Conductivity Detector no case shall the range be less than 112 nor more than
Probe Test four times that pressure. These range limits do not
Appendix A Supplementary Leak Testing Formula apply to dial indicating and recording vacuum gages.
Symbols Range requirements for other types of gages given in
an applicable Mandatory Appendix shall be as required
by that Appendix.
T-1020 GENERAL (b) Gage Location. When components are to be
T-1021 Procedure pressure /vacuum leak tested, the dial indicating gage(s)
When written procedures are required by the referenc- shall be connected to the component or to the component
ing Code Section, they shall include, as a minimum, from a remote location, with the gage(s) readily visible
the following information: to the operator controlling the pressure /vacuum through-

189

ASME B&PVC sec5$$u165 05-24-99 12:29:59 pd: sec5 Rev 14.04


T-1031 1998 SECTION V T-1063.1

out the duration of pressurizing, evacuating, testing, T-1050 PROCEDURE /TECHNIQUE


and depressurizing or venting of the component. For
T-1051 Preliminary Leak Test
large vessels or systems where one or more gages
are specified or required, a recording type gage is Prior to employing a sensitive leak testing method,
recommended, and it may be substituted for one of it may be expedient to perform a preliminary test to
the two or more indicating type gages. detect and eliminate gross leaks. This shall be done
(c) When other types of gage(s) are required by an in a manner that will not seal or mask leaks during
applicable Mandatory Appendix, they may be used in the specified test.
conjunction with or in place of dial indicating or
recording type gages.
T-1052 Specific Requirements
It is recommended that leak testing be performed
before hydrostatic or hydropneumatic testing. For the
leak testing method(s) or technique(s) specified by the
T-1040 REQUIREMENTS referencing Code, the referencing Code Section shall
T-1041 Cleanliness then be consulted for the following:
(a) personnel qualification /certification
The surface areas to be tested shall be free of oil, (b) technique(s) /calibration standards
grease, paint, or other contaminants that might mask (c) extent of examination
a leak. If liquids are used to clean the component or (d) acceptable test sensitivity or leakage rate
if a hydrostatic or hydropneumatic test is performed (e) report requirements
before leak testing, the component shall be dry before (f) retention of records
leak testing.

T-1060 CALIBRATION

A99 T-1042 Openings T-1061 Pressure /Vacuum Gages

All openings shall be sealed using plugs, covers, (a) All dial indicating and recording type gages used
sealing wax, cement, or other suitable material that can shall be calibrated against a standard deadweight tester,
be readily and completely removed after completion a calibrated master gage, or a mercury column, and
of the test. Sealing materials shall be tracer gas free. recalibrated at least once a year, when in use, unless
specified differently by the referencing Code Section
or Mandatory Appendix. All gages used shall provide
results accurate to within the Manufacturers listed
T-1043 Temperature accuracy and shall be recalibrated at any time that
there is reason to believe they are in error.
The minimum metal temperature for all components (b) When other than dial indicating or recording
during a test shall be as specified in the applicable type gages are required by an applicable Mandatory
Mandatory Appendix of this Article or in the referencing Appendix, they shall be calibrated as required by that
Code Section for the hydrostatic, hydropneumatic, or Mandatory Appendix or referencing Code Section.
pneumatic test of the pressure component or parts. The
minimum or maximum temperature during the test shall
not exceed that temperature compatible with the leak T-1062 Temperature Measuring Devices
testing method or technique used. When temperature measurement is required by the
referencing Code Section or Mandatory Appendix, the
device(s) shall be calibrated in accordance with the
requirements of that Code Section or Mandatory Ap-
T-1044 Pressure /Vacuum (Pressure Limits)
pendix.
Unless specified in the applicable Mandatory Appen-
dix of this Article or by the referencing Code Section,
T-1063 Calibration Leak Standards
components that are to be pressure-leak tested shall
not be tested at a pressure exceeding 25% of the Design T-1063.1 Permeation Type Leak Standard. This
Pressure. standard shall be a calibrated permeation type leak

190

ASME B&PVC sec5$$u165 05-24-99 12:29:59 pd: sec5 Rev 14.04


T-1063.1 ARTICLE 10 LEAK TESTING T-1092

through fused glass or quartz. The standard shall have T-1090 DOCUMENTATION
a helium leakage rate in the range of 1 106 to 1
T-1091 Test Report A99
1010 std cm3 /sec.
The test report shall contain, as a minimum, the
A99 T-1063.2 Capillary Type Leak Standard. This stan- following information as applicable to the method or
dard shall be a calibrated capillary type leak through technique:
a tube. The standard shall have a leakage rate equal (a) date of test;
to or smaller than the required test sensitivity times (b) certification level and name of operator;
the actual percent test concentration of the selected (c) test procedure (number) and revision number;
tracer gas. (d) test method or technique;
(e) test results;
(f) component identification;
T-1070 TEST (g) test instrument, standard leak, and material identi-
fication;
See applicable Mandatory Appendix of this Article. (h) test conditions, test pressure, tracer gas, and gas
concentration;
(i) gage(s) manufacturer, model, range, and identi-
fication number;
T-1080 EVALUATION
(j) temperature measuring device(s) and identifica-
T-1081 Acceptance Standards tion number(s);
(k) sketch showing method or technique setup.
Unless otherwise specified in the referencing Code
Section, the acceptance criteria given for each method
or technique of that method shall apply. The supplemen-
T-1092 Record Retention
tal leak testing formulas for calculating leakage rates
for the method or technique used are stated in the The test report shall be maintained in accordance
Mandatory Appendices of this Article. with the requirements of the referencing Code Section.

191

ASME B&PVC sec5$$u165 05-24-99 12:29:59 pd: sec5 Rev 14.04


ARTICLE 10
MANDATORY APPENDICES

APPENDIX I BUBBLE TEST I-1070 TEST


DIRECT PRESSURE TECHNIQUE I-1071 Soak Time
I-1000 INTRODUCTION Prior to examination the test pressure shall be held
I-1010 SCOPE for a minimum of 15 min.

The objective of the direct pressure technique of


bubble leak testing is to locate leaks in a pressurized
component by the application of a solution or by I-1072 Surface Temperature
immersion in liquid that will form bubbles as leakage
As a standard technique, the temperature of the
gas passes through it.
surface of the part to be examined shall not be below
40F (4C) nor above 125F (52C) throughout the
examination. Local heating or cooling is permitted
provided temperatures remain within the range of 40F
I-1030 EQUIPMENT
(4C) to 125F (52C) during examination. Where it
I-1031 Gases is impractical to comply with the foregoing temperature
limitations, other temperatures may be used provided
Unless otherwise specified, the test gas will normally
that the procedure is demonstrated.
be air; however, inert gases may be used.

NOTE: When inert gas is used, safety aspects of oxygen deficient


atmosphere should be considered.
I-1073 Application of Solution
The bubble forming solution shall be applied to the
I-1032 Bubble Solution surface to be tested by flowing, spraying, or brushing
the solution over the examination area. The number of
(a) The bubble forming solution shall produce a film bubbles produced in the solution by application should
that does not break away from the area to be tested, be minimized to reduce the problem of masking bubbles
and the bubbles formed shall not break rapidly due to caused by leakage.
air drying or low surface tension. Household soap or
detergents are not permitted as substitutes for bubble
testing solutions.
(b) The bubble forming solution shall be compatible I-1074 Immersion in Bath
with the temperature of the test conditions.
The area of interest shall be placed below the surface
of the bath in an easily observable position.

I-1033 Immersion Bath


(a) Water or another compatible solution shall be
I-1075 Lighting and Visual Aids
used for the bath.
(b) The immersion solution shall be compatible with When performing the test, the requirements of Article
the temperature of the test conditions. 9, T-952 and T-953 shall apply.

193
I-1076 1998 SECTION V II-1073

I-1076 Indication of Leakage (c) The bubble forming solution shall be compatible
with the temperature conditions of the test.
The presence of continuous bubble growth on the
surface of the material indicates leakage through an
orifice passage(s) in the region under examination. II-1032 Vacuum Box
The vacuum box used shall be of convenient size
I-1077 Post-Test Cleaning [e.g., 6 in. (152 mm) wide by 30 in. (762 mm) long]
After testing, surface cleaning may be required for and contain a window in the side opposite the open
product serviceability. bottom. The open bottom edge shall be equipped with
a suitable gasket to form a seal against the test surface.
Suitable connections, valves, lighting, and gage shall
I-1080 EVALUATION be provided. The gage shall have a range of 0 psi
(0 kPa) to 15 psi (103 kPa), or equivalent pressure
(a) Unless otherwise specified by the referencing units such as 0 in. Hg to 30 in. Hg. The gage range
Code Section, the area under test is acceptable when limit requirements of T-1031(a) do not apply.
no continuous bubble formation is observed.
(b) When leakage is observed, the position of the
leak(s) shall be marked. The component will then be II-1033 Vacuum Source
depressurized, and the leak(s) repaired as required by
the referencing Code Section. After repairs have been The required vacuum can be developed in the box
made, the repaired area or areas shall be retested in by any convenient method (e.g., air ejector, vacuum
accordance with the requirements of this Appendix. pump, or motor intake manifold). The gage shall register
a partial vacuum of at least 2 psi (4.1 in. Hg) (13.8 kPa)
below atmospheric pressure or the partial vacuum re-
quired by the referencing Code Section.
APPENDIX II BUBBLE TEST
VACUUM BOX TECHNIQUE
II-1000 INTRODUCTION II-1070 TEST
II-1010 SCOPE II-1071 Surface Temperature
The objective of the vacuum box technique of bubble As a standard technique, the temperature of the
leak testing is to locate leaks in a pressure boundary surface of the part to be examined shall not be below
that cannot be directly pressurized. This is accomplished 40F (4C) nor above 125F (52C) throughout the
by applying a solution to a local area of the pressure examination. Local heating or cooling is permitted
boundary surface and creating a differential pressure provided temperatures remain in the range of 40F
across that local area of the boundary causing the (4C) to 125F (52C) during the examination. Where it
formation of bubbles as leakage gas passes through is impractical to comply with the foregoing temperature
the solution. limitations, other temperatures may be used provided
that the procedure is demonstrated.

II-1030 EQUIPMENT
II-1031 Bubble Solution II-1072 Application of Solution

(a) The bubble forming solution shall produce a film The bubble forming solution shall be applied to the
that does not break away from the area to be tested, surface to be tested by flowing, spraying, or brushing
and the bubbles formed shall not break rapidly due to the solution over the examination area before placement
air drying or low surface tension. The number of of the vacuum box.
bubbles contained in the solution should be minimized
to reduce the problem of discriminating between existing
II-1073 Vacuum Box Placement
bubbles and those caused by leakage.
(b) Soaps or detergents designed specifically for The vacuum box shall be placed over the solution
cleaning shall not be used for the bubble forming coated section of the test surface and the box evacuated
solution. to the required partial vacuum.

194
II-1074 ARTICLE 10 MANDATORY APPENDICES III-1032

II-1074 Pressure (Vacuum) Retention III-1010 SCOPE


The required partial vacuum (differential pressure) The halogen detector probe test method is a semiquan-
shall be maintained for at least 10 sec examination time. titative method used to detect and locate leaks, and
shall not be considered quantitative.

II-1075 Vacuum Box Overlap


An overlap of 2 in. minimum for adjacent placement
III-1011 Alkali-Ion Diode (Heated Anode)
of the vacuum box shall be used for each subsequent
Halogen Leak Detectors
examination.
The alkali-ion diode halogen detector probe instru-
ment uses the principle of a heated platinum element
II-1076 Lighting and Visual Aids (anode) and an ion collector plate (cathode), where
When performing the test, the requirements of Article halogen vapor is ionized by the anode, and the ions
9, T-952 and T-953 shall apply. are collected by the cathode. A current proportional to
the rate of ion formation is indicated on a meter.

II-1077 Indication of Leakage


The presence of continuous bubble growth on the III-1012 Electron Capture Halogen Leak
surface of the material or weld seam indicates leakage Detectors
through an orifice passage(s) in the region under exami-
nation. The electron capture halogen detector probe instru-
ment uses the principle of the affinity of certain molecu-
lar compounds for low energy free electrons usually
II-1078 Post-Test Cleaning produced by ionization of gas flow through an element
with a weak radioactive tritium source. When the gas
After testing, cleaning may be required for product
flow contains halides, electron capture occurs causing
serviceability.
a reduction in the concentration of halogen ions present
as indicated on a meter. Non-electron capturing nitrogen
or argon is used as background gas.
II-1080 EVALUATION
(a) Unless otherwise specified by the referencing
Code Section, the area under test is acceptable when
no continuous bubble formation is observed.
III-1030 EQUIPMENT
(b) When leakage is observed, the position of the
leak(s) shall be marked. The vacuum box shall then III-1031 Tracer Gas
be vented and the leak(s) repaired as required by the
Gases that may be used are shown in Table III-1031.
referencing Code Section. After repairs have been made,
the repaired area or areas shall be retested in accordance III-1031.1 For Alkali-Ion Diode. Halogen leak de-
with the requirements of this Appendix. tectors, select a tracer gas from Table III-1031 that
will produce the necessary test sensitivity.

APPENDIX III HALOGEN DIODE III-1031.2 For Electron Capture. Halogen leak
DETECTOR PROBE TEST detectors, sulfur hexafluoride, SF6, is the recommended
tracer gas.
III-1000 INTRODUCTION
The more sophisticated electronic halogen leak detec-
tors have very high sensitivity. These instruments make
III-1032 Instrument A99
possible the detection of halogen gas flow from the
lower pressure side of a very small opening in an An electronic leak detector as described in III-1011
envelope or barrier separating two regions at different or III-1012 shall be used. Leakage shall be indicated
pressures. by one or more of the following signaling devices.

195

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


III-1032 1998 SECTION V III-1073

TABLE III-1031 III-1063 Scanning Rate A99


TRACER GASES
The instrument shall be calibrated by passing the
Commercial Chemical Designation Chemical probe tip across the orifice of the capillary leak standard
Designation Symbol in III-1061. The probe tip shall be kept within 18 in.
Refrigerant-11 Trichloromonofluoromethane CCl3 F (3.2 mm) of the orifice of the capillary leak standard.
Refrigerant-12 Dichlorodifluoromethane CCl2 F2 The scanning rate shall not exceed that which can
Refrigerant-21 Dichloromonofluoromethane CHCl2 F detect leakage rate Q from the capillary leak standard.
Refrigerant-22 Chlorodifluoromethane CHCIF2 The meter deflection shall be noted or the audible
Refrigerant-114 Dichlorotetrafluoroethane C2 Cl2 F4 alarm or indicator light set for this scanning rate.
Refrigerant-134a Tetrafluoroethane C2 H2 F4
Methylene Chloride Dichloromethane CH2 Cl2
Sulfur Hexafluoride Sulfur Hexafluoride SF6
III-1064 Frequency and Sensitivity A99

Unless otherwise specified by the referencing Code


Section, the sensitivity of the detector shall be deter-
mined before and after testing and at intervals of not
(a) Meter a meter on the test instrument, or a more than 2 hr during testing. During any calibration
probe, or both. check, if the meter deflection, audible alarm, or indicator
(b) Audio Devices a speaker or set of headphones light indicates that the detector cannot detect leakage
that emits audible indications. from the calibration standard of III-1061, the instrument
(c) Indicator Light a visible indicator light. shall be recalibrated and areas tested after the last
satisfactory calibration check shall be retested.

A99 III-1033 Capillary Calibration Leak Standard


A capillary type leak standard per T-1063.2 using III-1070 TEST
100% tracer gas as selected per III-1031.
III-1071 Location of Test
(a) The test area shall be free of contaminants that
could interfere with the test or give erroneous results.
III-1060 CALIBRATION (b) The component to be tested shall, if possible,
A99 III-1061 Standard Leak Size be protected from drafts or located in an area where
drafts will not reduce the required sensitivity of the test.
The maximum leakage rate Q for the capillary leak
standard described in III-1033 containing 100% tracer
concentration for use in III-1062 shall be calculated
as follows: III-1072 Concentration of Tracer Gas A99

The concentration of the tracer gas shall be at least


actual percent test concentration 10% by volume at the test pressure, unless otherwise
Q p 1 104 specified by the referencing Code Section.
100

where the actual percent test concentration is the concen- III-1073 Soak Time
tration of the tracer gas that is to be used for the test.
See III-1072. Prior to examination, the test pressure shall be held a
minimum of 30 min. When demonstrated, the minimum
allowable soak time may be less than that specified
above due to the immediate dispersion of the halogen
A99 III-1062 Warm Up
gas when:
The detector shall be turned on and allowed to warm (a) a special temporary device (such as a leech box)
up for the minimum time specified by the instrument is used on open components to test short segments;
manufacturer prior to calibrating with the capillary leak (b) components are partially evacuated prior to initial
standard of III-1061. pressurization with halogen gas.

196

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


III-1074 ARTICLE 10 MANDATORY APPENDICES IV-1032

A99 III-1074 Scanning Distance leakage is detected that exceeds the allowable rate of
1 104 std cm3 /sec.
After the required soak time per III-1073, the detector
(b) When unacceptable leakage is detected, the posi-
probe tip shall be passed over the test surface. The
tion of the leak(s) shall be marked. The component
probe tip shall be kept within 18 in. (3.2 mm) of the
will then be depressurized, and the leak(s) repaired as
test surface during scanning. If a shorter distance is
required by the referencing Code Section. After repairs
used during calibration, then that distance shall not be
have been made, the repaired area or areas shall be
exceeded during the examination scanning.
retested in accordance with the requirements of this
Appendix.
III-1075 Scanning Rate
The maximum scanning rate shall be as determined
in III-1063. APPENDIX IV HELIUM MASS
SPECTROMETER TEST
DETECTOR PROBE TECHNIQUE
A99 III-1076 Scanning Direction
IV-1000 INTRODUCTION
The examination scan should commence in the upper-
most portion of the system being leak tested while IV-1010 SCOPE
progressively scanning downward. This technique describes the use of the helium mass
spectrometer to detect minute traces of helium gas in
III-1077 Leakage Detection pressurized components. The high sensitivity of this
leak detector makes possible the detection of helium
Leakage will be indicated and detected according to gas flow from the lower pressure side of a very small
either III-1032(a), III-1032(b), or III-1032(c). opening in an envelope or barrier separating two regions
at different pressures, or the determination of the pres-
III-1078 Application ence of helium in any gaseous mixture. The detector
probe is a semiquantitative technique used to detect and
The following are two examples of applications that locate leaks, and shall not be considered quantitative.
may be used (note that other types of applications may
be used).
III-1078.1 Tube Examination. To detect leakage IV-1030 EQUIPMENT
through the tube walls when testing a tubular heat
IV-1031 Instrument A99
exchanger, the detector probe tip should be inserted into
each tube end and held for the time period established by A helium mass spectrometer leak detector capable
demonstration. The examination scan should commence of sensing and measuring minute traces of helium shall
in the uppermost portion of the tubesheet tube rows be used. Leakage shall be indicated by one or more
while progressively scanning downward. of the following signaling devices.
(a) Meter a meter on, or attached to, the test
III-1078.2 Tube-to-Tubesheet Joint Examination.
instrument.
Tube-to-tubesheet joints may be tested by the encapsula-
(b) Audio Devices a speaker or set of headphones
tor method. The encapsulator may be a funnel type
that emits audible indications.
with the small end attached to the probe tip end and
(c) Indicator Light a visible indicator light.
the large end placed over the tube-to-tubesheet joint.
If the encapsulator is used, the calibration response
time is determined by placing the encapsulator over IV-1032 Auxiliary Equipment
the orifice on the capillary calibration leak standard
(a) Transformer. A constant voltage transformer shall
and noting the time required for an indicated instrument
be used in conjunction with the instrument when line
response.
voltage is subject to variations.
(b) Detector Probe. All areas to be examined shall
be scanned for leaks using a detector probe (sniffer)
III-1080 EVALUATION
connected to the instrument through flexible tubing or
(a) Unless otherwise specified by the referencing a hose. To reduce instrument response and clean up
Code Section, the area tested is acceptable when no time, the tubing or hose length shall be less than 15

197

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


IV-1032 1998 SECTION V IV-1071

ft (4.6 m), unless the test setup is specifically designed IV-1033, containing 100% helium concentration for use
to attain the reduced response and clean up time for in IV-1062.2, shall be calculated as follows:
longer tubing or hose lengths.
actual percent test concentration
Q p 1 104
100
IV-1033 Calibration Leak Standards
Calibration leak standards may be either a permeation where the actual percent test concentration is the concen-
or capillary type standard per T-1063.1 and T-1063.2. tration of helium that is to be used for the test. See
The type of leak standard used shall be established by IV-1072.
the instrument or system sensitivity requirement, or as
IV-1062.2 Scanning Rate. After connecting the de- A99
specified by the referencing Code Section.
tector probe of IV-1032(b) to the instrument, the system
shall be calibrated by passing the detector probe tip
across the orifice of the capillary leak standard in
IV-1060 CALIBRATION IV-1062.1. The probe tip shall be kept within 18 in.
of the orifice of the capillary leak standard. The scanning
IV-1061 Instrument Calibration rate shall not exceed that which can detect leakage
IV-1061.1 Warm Up. The instrument shall be turned rate Q from the capillary leak standard. If the capil-
on and allowed to warm up for the minimum time lary leak standard helium source simulates the actual
specified by the instrument manufacturer prior to cali- test concentration, the scanning rate shall not exceed
brating with the calibrated leak standard. that which can detect actual allowable leakage of 1
104 std cm3 /sec.
IV-1061.2 Calibration. The instrument shall be cali-
brated as specified by the instrument manufacturer using IV-1062.3 Response Time. The time required for
a permeation type standard as stated in T-1063.1. an indication to appear and stabilize on the instrument
output is the response time and it should be observed
98 IV-1061.3 Sensitivity Acceptance during system calibration. It is usually desirable to
(a) For helium mass spectrometers with leak indicator keep this time as short as possible to reduce the time
meter range selector switch settings marked in actual required to pinpoint detected leakage.
leakage rate units and with the instrument tuned to
direct read when calibrated, the sensitivity of the instru- IV-1062.4 Clean Up Time. The time that is required
ment shall be a minimum of 1 109 std cm3/sec. If for the instrument to reduce its detected output signal
the sensitivity of the instrument at any calibration is to 37% of the signal indicated at the time tracer gas
less than 1 109 std cm3/sec, the instrument shall be ceases to be applied to the test system. It is usually
retuned, cleaned or repaired, and recalibrated until this desirable to keep this time as short as possible.
sensitivity can be attained. IV-1062.5 Frequency and Acceptance. Unless other-
(b) For helium mass spectrometers with leak indicator wise specified by the referencing Code Section, the
meter range selector switch settings marked in multiplier system sensitivity shall be determined before and after
units such as 1, 3, 10, 30, 100, etc., the sensitivity of testing and at intervals of not more than two hours
the instrument shall be a minimum of 1 109 std during the test. During any calibration check, if the
cm3/sec/div (a division is the unit on the instrument meter deflection, audible alarm, or visible light indicates
meter selected as the smallest readable signal indicating that the system cannot detect leakage per IV-1062.2,
real leakage. The meter unit selected as a division shall the instrument shall be recalibrated and all areas tested
be the same for both instrument and system calibration). after the last satisfactory calibration check shall be
If the sensitivity of the instrument at any calibration retested.
is less than 1 109 std cm3/sec/div, the instrument
shall be retuned, cleaned or repaired, and recalibrated
until this sensitivity can be obtained.
IV-1070 TEST
IV-1071 Location of Test
IV-1062 System Calibration
The component to be tested shall, if possible, be
A99 IV-1062.1 Standard Leak Size. The maximum leak- protected from drafts or located in an area where drafts
age rate Q for the capillary leak standard described in will not reduce the required sensitivity of the test.

198

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


IV-1072 ARTICLE 10 MANDATORY APPENDICES V-1010

A99 IV-1072 Concentration of Tracer Gas in the lowermost portion of the tubesheet tube rows
while progressively scanning upward.
The concentration of the helium tracer gas shall be
at least 10% by volume at the test pressure, unless IV-1078.2 Tube-to-Tubesheet Joint Examination.
otherwise specified by the referencing Code Section. Tube-to-tubesheet joints may be tested by the encapsula-
tor method. The encapsulator may be a funnel type
with the small end attached to the probe tip end and
IV-1073 Soak Time
the large end placed over the tube-to-tubesheet joint.
Prior to examination, the test pressure shall be held If the encapsulator is used, the calibration response
a minimum of 30 min. The minimum allowable soak time is determined by placing the encapsulator over
time may be less than that specified above due to the the orifice on the capillary calibration leak standard
immediate dispersion of the helium gas when: and noting the time required for an indicated instrument
(a) a special temporary device (such as a leech box) response.
is used on open components to test short segments;
(b) components are partially evacuated prior to initial
pressurization with helium gas.
IV-1080 EVALUATION

A99 IV-1074 Scanning Distance (a) Unless otherwise specified by the referencing
Code Section, the area tested is acceptable when no
After the required soak time per IV-1073, the detector leakage is detected that exceeds the allowable rate of
probe tip shall be passed over the test surface. The 1 104 std cm3 /sec.
probe tip shall be kept within 18 in. (3.2 mm) of the (b) When unacceptable leakage is detected, the posi-
test surface during scanning. If a shorter distance is tion of the leak(s) shall be marked. The component
used during system calibration, then that distance shall will then be depressurized, and the leak(s) repaired as
not be exceeded during test scanning. required by the referencing Code Section. After repairs
have been made, the repaired area or areas shall be
retested in accordance with the requirements of this
IV-1075 Scanning Rate
Appendix.
The maximum scanning rate shall be as determined
in IV-1062.2.

APPENDIX V HELIUM MASS


IV-1076 Scanning Direction SPECTROMETER TEST TRACER
The examination scan should commence in the lower- PROBE AND HOOD TECHNIQUES
most portion of the system being tested while progres- V-1010 SCOPE
sively scanning upward.
These techniques describe the use of the helium mass
spectrometer to respectively detect and measure minute
IV-1077 Leakage Detection traces of helium gas in evacuated components.
Leakage will be indicated and detected according to (a) The high sensitivity of this leak detector, when
either IV-1031(a), IV-1031(b), or IV-1031(c). tracer probe testing, makes possible the detection and
location of helium gas flow from the higher pressure
side of very small openings through the evacuated
IV-1078 Application envelope or barrier separating the two regions at differ-
ent pressures. This is a semiquantitative technique and
The following are two examples of applications that
shall not be considered quantitative.
may be used (note that other types of applications may
(b) The high sensitivity of this leak detector, when
be used).
hood testing, makes possible the detection and measure-
IV-1078.1 Tube Examination. To detect leakage ment of total helium gas flow from the higher pressure
through the tube walls when testing a tubular heat side of all hooded, very small openings through the
exchanger, the detector probe tip should be inserted into evacuated envelope or barrier that separates the two
each tube end and held for the time period established by regions at different pressures. This is a quantitative
demonstration. The examination scan should commence measurement technique.

199

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


V-1030 1998 SECTION V V-1031

V-1030 EQUIPMENT
A99 V-1031 Instrument
A helium mass spectrometer leak detector capable
of sensing and measuring minute traces of helium shall
be used. Leakage shall be indicated by one or more
of the following signaling devices.
(a) Meter (Tracer Probe and Hood Technique)
a meter on or attached to the test instrument.
(b) Audio Devices (Tracer Probe Technique) a
speaker or set of headphones that emits audible indica-
tions.
(c) Indicator Light (Tracer Probe Technique) a
visible indicator light.

199.1

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


V-1032 1998 SECTION V V-1062.1

V-1032 Auxiliary Equipment retuned, cleaned or repaired, and recalibrated until this
sensitivity can be attained.
(a) Transformer. A constant voltage transformer shall
(b) For helium mass spectrometers with leak indicator
be used in conjunction with the instrument when line
meter range selector switch settings marked in multiplier
voltage is subject to variations.
units such as 1, 3, 10, 30, 100, etc., the sensitivity of
(b) Auxiliary Pump System. When the size of the
the instrument shall be a minimum of 1 109 std
test system necessitates the use of an auxiliary vacuum
cm3/sec/div (a division is the unit on the instrument
pump system, the ultimate absolute pressure and pump
meter selected as the smallest readable signal indicating
speed capability of that system shall be sufficient to
the real leakage. The meter unit selected as a division
attain required test sensitivity and response time.
shall be the same for both instrument and system
(c) Manifold. A system of pipes and valves with
calibration). If the sensitivity of the instrument at any
proper connections for the instrument gages, auxiliary
calibration is less than 1 109 std cm3/sec/div, the
pump, calibration leak standard, and test component.
instrument shall be retuned, cleaned or repaired, and
(d) Tracer Probe. Tubing connected to a source of
recalibrated until this sensitivity can be obtained.
100% helium with a valved fine opening at the other
end for directing a fine stream of helium gas.
(e) Hood. Any suitable envelope or container, such
as a plastic bag, with a through aperture for the V-1062 System Calibration
manifold.
V-1062.1 Tracer Probe Technique. The capillary
(f) Vacuum Gage(s). The range of vacuum gage(s)
type calibrated leak standard, as stated in T-1063.2,
capable of measuring the absolute pressure at which
shall be attached to the component as far as possible
the evacuated system is being tested. The gage(s) for
from the instrument connection to the component. The
large systems shall be located on the system as far as
calibrated leak standard shall remain open during system
possible from the inlet to the pump system.
calibration.
(a) Calibration. With the component evacuated to
V-1033 Calibration Leak Standards an absolute pressure sufficient for connection of the
helium mass spectrometer to the system, the system
The calibration leak standard may be either a perme- shall be calibrated for the test by passing the tracer
ation or a capillary type standard as per T-1063.1 and probe tip across the orifice of the capillary leak standard
T-1063.2. The type of standard leak used shall be in V-1033. The probe tip shall be kept within 14 in.
established by the instrument or test sensitivity require- (6 mm) of the orifice of the capillary leak standard.
ments or as specified by the referencing Code Section. For a known flow rate from the tracer probe of 100%
helium, the scanning rate shall not exceed that which
can detect leakage (0) of 1 105 std cm3 /sec of
V-1060 CALIBRATION helium through the calibration leak standard into the
V-1061 Instrument Calibration test system. If the flow rate from the tracer probe is
reduced, the system calibration must be repeated to
V-1061.1 Warm Up. The instrument shall be turned determine a new scanning rate.
on and allowed to warm up for the minimum time (b) Normal Scanning. With the component evacuated
specified by the instrument manufacturer prior to cali- to an absolute pressure sufficient for connection of the
brating with the calibration leak standard. helium mass spectrometer to the system, the scanning
rate shall be determined by passing the tracer probe
V-1061.2 Calibration. The instrument shall be cali-
across the orifice of the capillary leak standard. The
brated as specified by the instrument manufacturer using
scanning rate shall not exceed that which can detect
a permeation type standard as stated in T-1063.1.
the leakage rate as stated in (a) above.
98 V-1061.3 Sensitivity Acceptance (c) Response Time. The time required for an indica-
(a) For helium mass spectrometers with leak indicator tion to appear and stabilize on the instrument output
meter range selector switch settings marked in actual should be observed during system calibration. It is
leakage rate units and with the instrument tuned to desirable to keep this time as short as possible to
direct read when calibrated, the sensitivity of the instru- reduce the time required to pinpoint detected leakage.
ment shall be a minimum of 1 109 std cm3/sec. If (d) Clean Up Time. The time that is required for
the sensitivity of the instrument at any calibration is the instrument to reduce its detected output signal to
less than 1 109 std cm3/sec, the instrument shall be 37% of the signal indicated at the time tracer gas

200

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


V-1062.1 ARTICLE 10 MANDATORY APPENDICES V-1071.4

ceases to be applied to the test system. It is usually system upon completing the preliminary system sensitiv-
desirable to keep this time as short as possible. ity calibration.
(e) Calibration Frequency and Sensitivity. Unless (e) Final Calibration. Upon completing the test of
otherwise specified by the referencing Code Section, the system, and with the component still under the
the system sensitivity shall be determined before and hood, the instrument output reading M3 shall be deter-
after testing and at intervals of not more than 2 hrs mined with the calibrated leak closed. Again, the cali-
during testing. During any calibration check, if the brated leak shall be opened into the system being
meter deflection, audible alarm, or visible light indicates tested. The increase in instrument output M4 shall
that the system cannot detect leakage of 1 105 std be used in calculating the final system sensitivity as
cm3 /sec, the instrument shall be recalibrated and all follows:
areas tested after the last satisfactory calibration check
has been retested. CL
S2 p std cm3 / sec / div
V-1062.2 Hood Technique. A calibrated leak CL M 4 M3
standard as per T-1063.1 with 100% helium shall be
attached, where feasible, to the component as far as
If the final system sensitivity S2 has decreased below
possible from the instrument connection to the compo-
the initial system sensitivity S1 by more than 35%, the
nent. The calibrated leak standard shall remain open
instrument shall be cleaned and /or repaired, recalibrated
during system calibration until the response time has
and the component or system retested.
been determined.
(f) Measured Leakage Rate. The measured leakage
(a) Evacuation. With the component evacuated to
rate of the component shall be determined as follows:
an absolute pressure sufficient for connection of the
helium mass spectrometer to the system, the calibrated
leak standard shall be opened to the system. The Q1 p S2 (M3 M 2 ) std cm3 / sec
calibrated leak standard shall remain open until the
instrument signal becomes stable, and the response time (g) Actual Leakage Rate. Calculation of actual leak-
has been determined. age rate (corrected for tracer gas concentration used):
(b) Response Time. The time is recorded when the
calibrated leak standard is opened to the component
and when the increase in output signal becomes stable. Q1 100
Q2 p std cm3 / sec
The elapsed time between the two readings is the % He
response time. The stable instrument reading shall be
noted and recorded as M1 in divisions.
(c) Background Reading.1 Background M2 in divi-
sions is established after determining response time.
The calibration leak standard is closed to the system V-1070 TEST
and the instrument reading shall be recorded when it
becomes stable. V-1071 Tracer Probe Technique
(d) Preliminary Calibration. The preliminary system V-1071.1 Scanning Rate. The maximum scanning
sensitivity shall be calculated as follows: rate shall be as determined in V-1062.1(a).

V-1071.2 Scanning Direction. The examination scan


CL
S1 p p std cm3 / sec / div should commence in the uppermost portion of the system
M1 M 2
being tested while progressively scanning downward.

V-1071.3 Scanning Distance. The tracer probe tip


The calibration shall be repeated when there is any
shall be passed over the test surface. The probe tip
change in the leak detector setup (e.g., a change in
shall be kept within 14 in. (6 mm) of the test surface
the portion of helium bypassed to the auxiliary pump,
during scanning. If a shorter distance is used during
if used) or any change in the calibrated leak. The
system calibration, then that distance shall not be
calibrated leak standard shall be isolated from the
exceeded during the examination scanning.

1 Systembackground noise. For definition of symbols, see Nonmanda- V-1071.4 Leakage Detection. Leakage will be indi-
tory Appendix A. cated and detected according to V-1031.

201
V-1072 1998 SECTION V VI-1061

V-1072 Hood Technique vacuum retention are examples of techniques that may
be used whenever pressure change testing is specified
V-1072.1 Hood. For a single wall component or
as a means of determining leakage rates. The tests
part, the hood (envelope) container may be made of
specify a maximum allowable change in either pressure
a material such as plastic.
per unit of time, percentage volume, or mass change
V-1072.2 Filling of Hood with Tracer Gas. After
per unit of time.
completing preliminary calibration per V-1062.2(d), the
space between the component outer surface and the
hood shall be filled with helium after the component
has been evacuated. VI-1020 GENERAL
V-1072.3 Estimating or Determining Hood Tracer Each of the tests shall be performed by either pneu-
Gas Concentration. The tracer gas concentration shall matic pressurization or evacuation of a closed compo-
be determined or estimated in the hood enclosure. nent or system to a specific pressure or vacuum. Temper-
ature, pressure, or vacuum are systematically recorded
for a specified period of time. Analysis of data deter-
V-1080 EVALUATION mines the component or system acceptability with re-
spect to leakage rate or pressure change per unit of time.
V-1081 Tracer Probe Technique
Unless otherwise specified by the referencing Code
Section, the area tested is acceptable when no leakage
VI-1030 EQUIPMENT
is detected that exceeds the allowable rate of 1 105
std cm3 /sec. VI-1031 Pressure Measuring Instruments
(a) Gage Range. Dial indicating and recording type
V-1082 Hood Technique gages shall meet the requirements of T-1031(a). Liquid
manometers or quartz Bourdon tube gages may be used
Unless otherwise specified by the referencing Code over their entire range.
Section, the component tested is acceptable when the (b) Gage Location. The location of the gage(s) shall
actual leakage rate Q2 is equal to or less than 1 be that stated in T-1031(b).
106 std cm3 /sec of helium. (c) Types of Gages. Regular or absolute gages may
(a) When the actual leakage rate exceeds the permis- be used in pressure change testing. When greater accu-
sible value, all welds or other suspected areas shall be racy is required, quartz Bourdon tube gages or liquid
retested using a tracer probe technique. All leaks shall manometers may be used. The gage(s) used shall have
be marked and temporarily sealed to permit completion an accuracy, resolution, and repeatability compatible
of the tracer probe retest. The temporary seals shall with the acceptance criteria.
be of a type which can be readily and completely
removed after testing has been completed.
(b) The component will then be vented and the VI-1032 Temperature Measuring Instruments
leak(s) repaired as required by the referencing Code
Section. After repairs have been made, the repaired Dry bulb or dew point temperature measuring instru-
area or areas shall be retested in accordance with the ments, when used, shall have accuracy, repeatability, and
requirements of this Appendix. resolution compatible with the leakage rate acceptance
criteria.

APPENDIX VI PRESSURE
CHANGE TEST VI-1060 CALIBRATION
VI-1010 SCOPE VI-1061 Pressure Measuring Instruments
This test method describes the techniques for de- All dial indicating, recording, and quartz Bourdon
termining the leakage rate of the boundaries of a closed tube gages shall be calibrated per T-1061(b). The
component or system at a specific pressure or vacuum. scale of liquid manometers shall be calibrated against
Pressure hold, absolute pressure, maintenance of pres- standards that have known relationships to national
sure, pressure loss, pressure decay, pressure rise, and standards, where such standards exist.

202
VI-1062 ARTICLE 10 MANDATORY APPENDICES VII-1010

VI-1062 Temperature Measuring Instruments VI-1076 Start of Test


Calibration for dry bulb and dew point temperature At the start of the test, initial temperature and pressure
measuring instruments shall be against standards that (or vacuum) readings shall be taken and thereafter at
have known relationships to national standards, where regular intervals, not to exceed 60 min, until the end
such standards exist. of the specified test duration.

VI-1077 Essential Variables


VI-1070 TEST (a) When it is required to compensate for barometric
VI-1071 Pressure Application pressure variations, measurement of the test pressure
shall be made with either an absolute pressure gage
Components that are to be tested above atmospheric or a regular pressure gage and a barometer.
pressure shall be pressurized per T-1044. (b) When it is required by the referencing Code
Section, or when the water vapor pressure variation
can significantly affect the test results, the internal dew
VI-1072 Vacuum Application point temperature or relative humidity shall be measured.
Components that are to be tested under vacuum shall
be evacuated to at least 2 psi (4.1 in. Hg) (13.8 kPa)
below atmospheric pressure or as required by the refer- VI-1080 EVALUATION
encing Code Section.
VI-1081 Acceptable Test
When the pressure change or leakage rate is equal
VI-1073 Test Duration to or less than that specified by the referencing Code
Section, the test is acceptable.
The test pressure (or vacuum) shall be held for the
duration specified by the referencing Code Section or,
if not specified, it shall be sufficient to establish the VI-1082 Rejectable Test
leakage rate of the component system within the accu-
racy or confidence limits required by the referencing When the pressure change or leakage rate exceeds
Code Section. For very small components or systems, that specified by the referencing Code Section, the
a test duration in terms of minutes may be sufficient. results of the test are unsatisfactory. Leak(s) may be
For large components or systems, where temperature located by other methods described in the Mandatory
and water vapor corrections are necessary, a test duration Appendices. After the cause of the excessive pressure
in terms of many hours may be required. change or leakage rate has been determined and repaired
in accordance with the referencing Code Section, the
original test shall be repeated.
VI-1074 Small Pressurized Systems NOTE: For more information regarding this method of testing refer
to the following:
For temperature stabilization of very small pressurized (a) 10 CFR 50, Appendix J, Primary Containment Leakage Testing
systems, such as gasket interspaces, where only system for Water Cooled Power Reactors.
(b) ANSI /ANS 56.8-1981, American National Standard Contain-
(metal) temperature can be measured, at least 15 min ment System Leakage Testing Requirements, published by the Ameri-
shall elapse after completion of pressurization and before can Nuclear Society.
starting the test.

APPENDIX VII GLOSSARY OF


VI-1075 Large Pressurized Systems
TERMS FOR LEAK TESTING
For temperature stabilization of large pressurized
VII-1010 SCOPE
systems where the internal gas temperature is measured
after completion of pressurization, it shall be determined This Mandatory Appendix is used for the purpose
that the temperature of the internal gas has stabilized of establishing standard terms and definitions of terms
before starting the test. which appear in Article 10, Leak Testing.

203

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


VII-1020 1998 SECTION V VIII-1000

VII-1020 GENERAL REQUIREMENTS temperature that the leak detector is capable of indicating
for a particular division on the leak indicator scale
(a) This standard terminology for nondestructive ex-
leakage the fluid, either liquid or gas, flowing
aminations ASTM E 1316 has been adopted by the
through a leak and expressed in units of mass flow;
Committee as SE-1316.
i.e., pressure and volume per time
(b) SE-1316 Section E provides the definitions of
leak standard see standard leak in VII-1030(a)
terms listed in VII-1030(a).
quartz Bourdon tube gage this high accuracy
(c) For general terms such as Discontinuity, Evalua-
gage is a servonulling differential pressure measuring
tion, Flaw, Indication, Inspection, etc., refer to Article
electronic instrument. The pressure transducing element
1, Mandatory Appendix I.
is a one piece fused quartz Bourdon element.
(d) Paragraph VII-1030(b) provides a list of terms
regular pressure see gage pressure in VII-1030(a)
and definitions which are in addition to SE-1316 and
sensitivity the size of the smallest leakage rate
are Code specific.
that can be unambiguously detected by the leak testing
instrument, method, or technique being used
soak time the elapsed time between when the
desired differential pressure is attained on a system
and the time when the test technique is performed to
detect leakage or measure leakage rate
A99 VII-1030 REQUIREMENTS
standard dead weight tester a device for hydrau-
(a) The following SE-1316 terms are used in conjunc- lically balancing the pressure on a known high accuracy
tion with this Article: absolute pressure; background weight against the reading on a pressure gage for the
signal; gage pressure; gas; halogen; halogen leak purpose of calibrating the gage
detector; hood test; leak; leakage rate; leak testing; system calibration introduction of a known size
mass spectrometer; mass spectrometer leak detector; standard leak into a test system with a leak detector
sampling probe; standard leak; tracer gas; vacuum. for the purpose of determining the smallest size leakage
(b) The following Code terms are used in conjunction rate of a particular gas at a specific pressure and
with this Article. temperature that the leak detector as part of the test
background reading see background signal in system is capable of indicating for a particular division
VII-1030(a) on the leak indicator scale
calibration leak standard see standard leak in thermal conductivity detector a leak detector that
VII-1030(a) responds to differences in the thermal conductivity of
detector probe see sampling probe in VII-1030(a) a sampled gas and the gas used to zero it (i.e.,
dew point temperature that temperature at which background atmosphere)
the gas in a system would be capable of holding no vacuum box a device used to obtain a pressure
more water vapor and condensation in the form of dew differential across a weld that cannot be directly pressur-
would occur ized. It contains a large viewing window, special easy
dry bulb temperature the ambient temperature of seating and sealing gasket, gage, and a valved connec-
the gas in a system tion for an air ejector, vacuum pump, or intake manifold
halogen diode detector see halogen leak detector water vapor gaseous form of water in a system
in VII-1030(a)
helium mass spectrometer see mass spectrometer
and mass spectrometer leak detector in VII-1030(a)
hood technique see hood test in VII-1030(a)
immersion bath a low surface tension liquid into
which a gas containing enclosure is submerged to detect APPENDIX VIII THERMAL A99
leakage which forms at the site or sites of a leak or leaks CONDUCTIVITY DETECTOR
immersion solution see immersion bath PROBE TEST
inert gas a gas that resists combining with other
VIII-1000 INTRODUCTION
substances. Examples are helium, neon, and argon.
instrument calibration introduction of a known These instruments make possible the detection of a
size standard leak into an isolated leak detector for tracer gas flow from the lower pressure side of a very
the purpose of determining the smallest size leakage small opening in an envelope or barrier separating two
rate of a particular gas at a specific pressure and regions at different pressures.

204

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


VIII-1010 ARTICLE 10 MANDATORY APPENDICES VIII-1064

TABLE VIII-1031 (b) Audio Devices. A speaker or sets of headphones


TRACER GASES that emit(s) audible indications.
(c) Indicator Light. A visible indicator light.
Chemical
Designation Chemical Designation Symbol
... Helium He VIII-1033 Capillary Calibration Leak Standard
... Argon Ar
... Carbon Dioxide CO2
A capillary type leak standard per T-1063.2 using
Refrigerant-11 Trichloromonofluoromethane CCl3F 100% tracer gas as selected per VIII-1031.
Refrigerant-12 Dichlorodifluoromethane CCl2F2
Refrigerant-21 Dichloromonofluoromethane CHCl2F
Refrigerant-22 Chlorodifluoromethane CHClF2
Refrigerant-114 Dichlorotetrafluoroethane C2Cl2F4
VIII-1060 CALIBRATION
Refrigerant-134a Tetrafluoroethane C2H2F4 VIII-1061 Standard Leak Size
Methylene Chloride Dichloromethane CH2Cl2
Sulfur Hexafluoride Sulfur Hexafluoride SF6 The maximum leakage rate Q for the capillary leak
standard described in VIII-1033 containing 100% tracer
concentration for use in VIII-1062 shall be calculated
as follows:

VIII-1010 SCOPE actual percent test concentration


Q p Qs
100
The thermal conductivity detector probe test method
is a semiquantitative method used to detect and locate
leaks, and shall not be considered quantitative. where Qs (in std cm3/sec) is the required test sensitivity
and the actual percent test concentration is the concentra-
tion of the tracer gas that is to be used for the test.
VIII-1011 Thermal Conductivity Leak Detectors See VIII-1072.
The thermal conductivity detector probe instrument
uses the principle that the thermal conductivity of a VIII-1062 Warm-up
gas or gas mixture changes with any change in the
concentration(s) of the gas or gas mixture (i.e., the The detector shall be turned on and allowed to warm
introduction of a tracer gas in the area of a leak). up for the minimum time specified by the instrument
manufacturer prior to calibrating with the capillary leak
standard of VIII-1061.
VIII-1030 EQUIPMENT
VIII-1031 Tracer Gas VIII-1063 Scanning Rate
In principle, any gas having a thermal conductivity The detector shall be calibrated by passing the probe
different from air can be used as a tracer gas. The tip across the orifice of the capillary leak standard in
sensitivity achievable depends on the relative differences VIII-1061. The probe tip shall be kept within 12 in.
of the thermal conductivity of the gases [i.e., background (13 mm) of the orifice capillary leak standard. The
air (air used to zero the instrument) and the sampled scanning rate shall not exceed that which can detect
air (air containing the tracer gas) in the area of a leak]. leakage rate Q from the capillary leak standard. The
Table VIII-1031 lists some of the typical tracer gases meter deflection shall be noted or the audible alarm
used. The tracer gas to be used shall be selected based or indicator light set for this scanning rate.
on the required test sensitivity.
VIII-1064 Frequency and Sensitivity
VIII-1032 Instrument
Unless otherwise specified by the referencing Code
An electronic leak detector as described in VIII- Section, the sensitivity of the detector shall be deter-
1011 shall be used. Leakage shall be indicated by one mined before and after testing and at intervals of not
or more of the following signaling devices: more than 2 hr during testing. During any calibration
(a) Meter. A meter on the test instrument, or a probe, check, if the meter deflection, audible alarm, or indicator
or both. light indicate that the detector cannot detect leakage

204.1

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


VIII-1064 1998 SECTION V VIII-1080

from the calibration standard of VIII-1061, the instru- VIII-1076 Scanning Direction
ment shall be recalibrated and areas tested after the
For tracer gases that are lighter than air, the examina-
last satisfactory calibration check shall be retested.
tion scan should commence in the lowermost portion
of the system being tested while progressively scanning
upward. For tracer gases that are heavier than air, the
examination scan should commence in the uppermost
VIII-1070 TEST
portion of the system being tested while progressively
VIII-1071 Location of Test scanning downward.
(a) The test area shall be free of contaminants that
could interfere with the test or give erroneous results. VIII-1077 Leakage Detection
(b) The component to be tested shall, if possible, Leakage shall be indicated and detected according
be protected from drafts or located in an area where to either VIII-1032(a), VIII-1032(b), or VIII-1032(c).
drafts will not reduce the required sensitivity of the test.
VIII-1078 Application
VIII-1072 Concentration of Tracer Gas The following are two examples of applications that
may be used (note that other types of applications may
The concentration of the tracer gas shall be at least be used).
10% by volume at the test pressure, unless otherwise
specified by the referencing Code Section. VIII-1078.1 Tube Examination. To detect leakage
through the tube walls when testing a tubular heat
exchanger, the detector probe tip should be inserted
VIII-1073 Soak Times into each tube and held for the time period established
by demonstration.
Prior to examination, the test pressure shall be held a
VIII-1078.2 Tube-to-Tubesheet Joint Examination.
minimum of 30 min. When demonstrated, the minimum
Tube-to-tubesheet joints may be tested by the encapsula-
allowable soak time may be less than that specified
tor method. The encapsulator may be a funnel type
above due to the immediate dispersion of the tracer
with the small end attached to the probe tip end and
gas when: the large end placed over the tube-to-tubesheet joint.
(a) a special temporary device (such as a leech box) If the encapsulator is used, the calibration response
is used on open components to test short segments; time is determined by placing the encapsulator over
(b) components are partially evacuated prior to initial the orifice on the capillary calibration leak standard
pressurization with tracer gas. and noting the time required for an indicated instrument
response.

VIII-1074 Scanning Distance


VIII-1080 EVALUATION
After the required soak time per VIII-1073, the
detector probe tip shall be passed over the test surface. (a) Unless otherwise specified by the referencing
The probe tip shall be kept within 12 in. (13 mm) of Code Section, the area tested is acceptable when no
the test surface during scanning. If a shorter distance leakage is detected that exceeds the maximum leakage
is used during calibration, then that distance shall not rate Q, determined per VIII-1061.
be exceeded during the examination scanning. (b) When unacceptable leakage is detected, the posi-
tion of the leak(s) shall be marked. The component
will then be depressurized, and the leak(s) repaired as
required by the referencing Code Section. After repairs
VIII-1075 Scanning Rate
have been made, the repaired area or areas shall be
The maximum scanning rate shall be as determined retested in accordance with the requirements of this
in VIII-1063. Appendix.

204.2

ASME B&PVC Sec5$$u170 06-07-99 05:17:52 pd: sec5 Rev 14.04


ARTICLE 10
NONMANDATORY APPENDIX

APPENDIX A SUPPLEMENTARY determined in S2 less background), std


LEAK TESTING FORMULA cm3 /sec
SYMBOLS (3) Actual leakage rate (corrected to 100% He):
Q2pactual leakage rate (corrected for tracer gas
A-10 APPLICABILITY OF THE concentration), std cm3 /sec
FORMULAS (4) Helium or Halogen concentration:
percent /100
(a) The formulas in this Article provide for the
(5) Calibrated standard:
calculated leak rate(s) for the technique used.
CLpcalibrated leak leakage rate, std cm3 /sec
(b) The symbols defined below are used in the
(6) Instrument reading sequence:
formulas of the appropriate Appendix. M1pmeter reading before test with calibrated
(1) System sensitivity calculation: leak open to the component (divisions)
S1ppreliminary sensitivity (calculation of sensi- M2pmeter reading before test with calibrated
tivity), std cm3 /sec /div leak closed to component (divisions) (system
S2pfinal sensitivity (calculation of sensitivity), background noise reading)
std cm3 /sec /div M3pmeter reading (registering component leak-
(2) Measured leakage rate calculations: age) with calibrated leak closed (divisions)
Q1pmeasured leakage rate of the component M4pmeter reading (registering component leak-
(calculation for the actual leakage rate as age) with calibrated leak open (divisions)

205

Вам также может понравиться