Вы находитесь на странице: 1из 7

Materials Characterization 49 (2002) 73 79

Characterization of different materials for corrosion


resistance under simulated body fluid conditions
I. Gurappa*
Surface Engineering Group, Defence Metallurgical Research Laboratory, Kanchanbagh PO, Hyderabad 500 058, India
Received 27 March 2002; received in revised form 14 September 2002; accepted 17 September 2002

Abstract

A systematic characterization study has been carried out on different materials such as commercial purity
titanium, Ti 6Al 4V, 316L stainless steel, and a cobalt-based alloy under simulated body fluid conditions at
37 C. Breakdown potential, corrosion rates, pitting/crevice corrosion resistance, and the ability to form
protective oxide scales were evaluated and compared. The advantages of the titanium alloy over other
materials were highlighted. An attempt has also been made to study the suitability of titanium alloy, IMI 834,
for biomedical applications. AC impedance measurements were also carried out in order to provide supportive
evidence for the above results.
D 2002 Elsevier Science Inc. All rights reserved.

Keywords: Titanium; Cobalt-based alloy; Simulated body fluid conditions; Corrosion resistance

1. Introduction human body is that it should be biocompatible and


not cause any adverse reaction in the body. The
Biomedical materials play an important and a material must withstand the body environment and
critical role in manufacturing a variety of prosthetic not degrade to a point that it cannot function in the
devices in a modern world. Prosthetic devices are body as intended.
artificial replacements that are used in a biological Corrosion of metal implants is critical because it
system, such as the human body in an effort to can adversely affect the biocompatibility and mech-
provide the function of the original part. These anical integrity. The materials used should not cause
devices are generally made of polymeric, metallic any adverse biological reaction in the body and
and ceramic materials, or combinations of these simultaneously, it must be stable and retain its func-
materials, depending on the intended use. Metals/ tional properties. Corrosion and surface oxide film
alloys are used as surgical implants in the human dissolution are the two mechanisms for introducing
body primarily for orthopedic purposes. The first additional ions into the body. Extensive release of
requirement for any material to be placed in the ions from prosthesis can result in adverse biological
reactions and can lead to mechanical failure of the
device.
Metals/alloys used in the human body must have a
high corrosion resistance and must not be treated or
* Tel.: +91-40-458-6515; fax: +91-40-434-1439. used in a configuration that would degrade the
E-mail address: igp1@rediffmail.com (I. Gurappa). corrosion behavior. Degradation of metals and alloys

1044-5803/02/$ see front matter D 2002 Elsevier Science Inc. All rights reserved.
PII: S 1 0 4 4 - 5 8 0 3 ( 0 2 ) 0 0 3 2 0 - 0
74 I. Gurappa / Materials Characterization 49 (2002) 7379

used as surgical implant orthopedic devices is usually Table 1


a combination of electrochemical and mechanical Different metallic materials tested for biomedical
effects. Modern-day life has resulted in the increased applications
use of implants at a younger age. The average life Material Type Chemical
span has also been increased. Both these factors composition (wt.%)
necessitated the use of materials with enhanced life Stainless AISI 316L 18 Cr, 12 Ni,
and/or innovation of newer implant materials with steel 2.5 Mo, < 0.03 C,
prolonged life. and balance iron
Corrosion is the first consideration for a material Cobalt Co Ni Cr Mo 35 Ni, 35 Co,
of any type that is to be used in the body because alloys 20 Cr, and 10 Mo
metal ion release takes place mainly due to corrosion Titanium Commercial 99.6%
of surgical implants [1]. Therefore, various in vitro purity (0.10 C and 0.3 Fe)
Titanium Ti 6Al 4V 6 Al, 4 V,
and in vivo tests have to be carried out in order to
alloys and balance titanium
identify appropriate materials for use as surgical IMI 834 5.8 Al, 4 Sn,
implants. It is desirable to keep the metal ion release 3.6 Zr, 0.7 Nb,
to a minimum by the use of corrosion-resistant 0.54 Mo, 0.32 Si,
materials. Some effects of incompatible materials and balance titanium
include interference with normal tissue growth near
the implant, interference with systematic reactions of
the body, and transport and deposition of metal ions
at selective sites or organs may occur [2 5]. There is terials were tested by using various electrochemical
always a concern about the carcinogenic effects of techniques in de-aerated Hanks solution of pH 7.4
foreign materials in the body both short- and long- (Table 2), known to simulate body fluid envir-
term periods exceeding 20 years. Some individuals onment [6,8,9], at 37 C (98.4 F), which is equi-
are sensitive to metals and some develop metal valent to human body temperature [6 9]. The
sensitivity later after receiving an implant. accuracy of temperature maintained in the present
The environment of the body is extremely well investigation is 1 C. Disc specimens of 14 mm
buffered so that the pH is maintained at around 7.4 diameter and 4 mm height were machined from the
and it is, of course, held at 37 C. Two features corresponding alloy ingots, polished up to 800 grit
control the severity of this environment. Firstly, the surface, washed with distilled water followed by
saline solution is an excellent electrolyte and facili- acetone before testing the materials. A potentiostat,
tates the electrochemical mechanisms of corrosion model 273A, in association with corrosion analysis
and hydrolysis. Secondly, there are many molecular software supplied by EG&G PARC, USA, was used
and cellular species in the tissues that have the to test different materials. A standard three-electrode
ability to catalyze certain chemical reactions or cell in which graphite is the counter-electrode and
rapidly destroy certain components identified as saturated calomel electrode is the reference electrode
foreign. Corrosion behavior of materials can be was used. The total amount of solution taken in the
studied either in vivo or in vitro using artificial cell is about 600 ml. Argon gas was used for de-
physiological fluids, in which the oxygen content aeration of the electrolyte. For recording impedance
is controlled at a suitably low value at 37 C. Hanks plots, the above potentiostat in association with
solution is an example of an artificial solution, which impedance software was used.
has been used over a period for corrosion testing in Potentiodynamic tests were carried out to deter-
the laboratory [6]. mine the breakdown potentials as well as corrosion
In the present investigation, a systematic study rates of different materials. The isolated electrode
was carried out in order to understand the stability of potentials as a function of time plots were used to
various materials both for short- and long-term bio- assess the ability of various materials to form a
medical applications and thereby suggest an appro- protective oxide scale on their surfaces upon immer-
priate material for surgical implants. sion in the simulated body fluid environment. A
cyclic polarization technique was used to evaluate
pitting and crevice corrosion resistance of various
2. Experimental materials. AC impedance plots were used to provide
supportive evidence to the above studies. Finally,
Different materials tested in the present investiga- scanning electron microscopy (SEM) was used to
tion are provided in Table 1. understand the surface morphology of the titanium
The corrosion resistance of different materials alloys, Ti 6Al 4V and IMI 834, after corrosion
that are being used and intended to use as bioma- studies. The corrosion morphology of titanium
I. Gurappa / Materials Characterization 49 (2002) 7379 75

Table 2 Table 3
The composition of Hanks solution Breakdown potentials and corrosion rates of different
Compound Composition (g/l) materials in de-aerated Hanks solution at 37 C
NaCl 8.00 Material Breakdown Corrosion rate
KCl 0.40 potential (V) (mils/year)
CaCl2 0.14 316L stainless steel 0.28 0.17
NaHCO3 0.35 Cobalt alloy 0.6 0.056
Na2HPO42H2O 0.06 Ti 6Al 4V 1.9 0.007
KH2PO4 0.60
MgSO47H2O 0.06
MgCl26H2O 0.10 requirement is to determine the breakdown potential,
Glucose 1.00 i.e., at which potential the passive film breaks down.
For selecting any material for biomedical application,
alloys, Ti 6Al 4V and IMI 834, was studied by the breakdown potential plays a major role.
SEM. Table 3 illustrates breakdown potentials measured
in de-aerated Hanks solution at 37 C for various
materials. As can be seen, the maximum breakdown
3. Results potential is observed for the titanium alloy, Ti 6Al
4V, and followed by the cobalt alloy. The minimum
Determination of the chemical interaction of breakdown potential is observed for 316L stainless
metallic materials/bone prosthesis with the body fluid steel. The corrosion rates of different materials were
environment is essential in order to understand their evaluated and are presented in Table 3. The maximum
stability in the human body. One simple way to study corrosion rate was observed for the 316L stainless
the film formation and passivation of implants/alloys steel followed by the cobalt alloy. The minimum
in a solution is to monitor the open-circuit electrode corrosion rate is observed for the titanium alloy, Ti
potential as a function of time. A rise of potential in 6Al 4V. In essence, the Ti 6Al 4V alloy exhibits
the positive direction indicates the formation of a the minimum corrosion rate and maximum break-
passive film, and a steady potential indicates that the down potential when compared to other materials.
film remains intact and protective. A drop of potential Another most important requirement is the pitting
in the negative direction indicates breaks in the film, and crevice corrosion resistance. Any metal/alloy
dissolution of the film, or no film formation. An intended for use as a biomaterial should exhibit
open-circuit potential as a function of time for differ- excellent pitting and crevice corrosion resistance.
ent materials tested in the present investigation is This can be determined by carrying out cyclic polar-
provided in Fig. 1. As can be seen, the potential ization experiments in Hanks solution at 37 C. The
shifted in the positive direction for all the tested cyclic polarization curves recorded for various mate-
materials, which clearly indicates the formation of a rials in de-aerated Hanks solution at 37 C are
passive film on their surfaces. Then, the second presented in Fig. 2. No hystersis loop is observed

Fig. 1. Open-circuit electrode potential as a function of time for various materials in de-aerated Hanks solution at 37 C.
76 I. Gurappa / Materials Characterization 49 (2002) 7379

Fig. 2. Cyclic polarization plots of different materials in de-aerated Hanks solution at 37 C.

for cobalt and titanium alloys. In the case of 316L the surfaces of both the alloys under simulated body
stainless steel, the reverse scan takes an entirely fluid environmental conditions. This indicates that the
different path when compared to the forward scan. alloy is suitable for biomedical applications. How-
This clearly indicates that the titanium and cobalt ever, further studies are essential before using it for
alloys have excellent pitting and crevice corrosion biomedical applications.
resistance, while 316L stainless steel is highly sus- Impedance measurements are one of the most
ceptible to pitting and crevice corrosion. useful and informative methods of corrosion assess-
With a view to understanding the suitability of a ment [10 15]. The response of a corroding specimen
recently developed titanium alloy, IMI 834, for bio- to an applied small amplitude signal will depend on
medical applications, studies were carried out on IMI the frequency of the signal. The magnitude and phase
834. It was observed that the alloy forms a protective shift at each frequency is noted, and the corrosion cell
oxide scale on its surface upon immersion in a response is expressed as a combination of resistive,
simulated body fluid environment and exhibits a very capacitive, and inductive components. The imped-
low corrosion rate and good breakdown potential. ance behavior of a specimen can be expressed either
The corrosion morphology observed under the SEM in Bode plots of impedance modulus (|Z|) as a
was found to be similar to that of the normally used function of frequency or in Nyquist plots of Z00(w)
titanium alloy, Ti 6Al 4V (Fig. 3). In both cases, as a function of Z0(w), where w = 2pf. The Nyquist
grooves and ridges typical for titanium alloy surfaces plot shows semicircle, with increasing frequency in a
were observed. Further, no corrosion pits are seen on counterclockwise direction. At very high frequency,

Fig. 3. SEM photographs of titanium alloys, Ti 6Al 4V and IMI 834, showing similar corrosion morphology after corrosion
studies in de-aerated Hanks solution at 37 C.
I. Gurappa / Materials Characterization 49 (2002) 7379 77

Fig. 4. Nyquist plots for various biomaterials in de-aerated Hanks solution at 37 C.

the imaginary component (Z00) disappears, leaving decreases with increase in the cold work and the
only the solution resistance, RV. At very low fre- maximum breakdown potential is observed for the
quency, Z00 again disappears, leaving a sum of RV and steel without any cold work [7]. The results on pitting
the faradic reaction resistance or polarization resist- corrosion of 316L stainless steel showed that it is
ance, Rp. Because of this behavior, measurements susceptible to both pitting and crevice corrosion (Fig.
taken at high frequencies will generally reveal the 2). Pitting corrosion is evidenced by observing a
solution resistance, while measurements taken at low hystersis loop, while crevice corrosion is seen
frequencies will measure both solution resistance and through the difference in potential between break-
charge transfer resistance. Thus, the difference down and protective potentials. From these studies, it
between low and high frequencies gives the charge is also very clear that the ability of 316L stainless
transfer resistance, which is inversely proportional to steel to repassivate is considerably less compared to
the corrosion rate. AC impedance plots recorded for other materials. Stress corrosion cracking may also
different biomaterials in Hanks solution at 37 1 C occur in vivo in the case of possible film breakdown
are provided in Fig. 4. The charge transfer resistance and extremely high loads. Therefore, they are better
measured with the help of Nyquist plots clearly suited only for temporary components in orthopedic
confirms the superior performance of the titanium treatments like bone screws, bone plates, intramodul-
alloy followed by cobalt and 316L stainless steel by lary rods, and other temporary fixation devices.
the increasing charge transfer resistance in the same Cobalt chromium alloys are important biomate-
order: titanium alloy < cobalt alloy < 316L stainless rials because of their corrosion resistance (an order of
steel. magnitude greater than stainless steels) and high
mechanical properties [8,9]. Although the cobalt
alloys do not passivate as is the case for titanium
4. Discussion alloys, good resistance against pitting corrosion and
crevice corrosion has been confirmed for cobalt
Stainless steels predominate as materials for pros- alloys containing molybdenum as evidenced from
thetic devices because they are relatively inexpensive the present results. Therefore, cobalt-based alloys
and formable by common techniques. Further, their are widely used safely for the fabrication of various
mechanical properties are controllable over a wide devices that are surgically implanted in the body.
range, providing optimum strength and ductility. These devices must be highly reliable because the
However, stainless steels are the least corrosion failure of certain implants (e.g., heart valves) can be
resistant over the long term and cause rashes or pain fatal and the failure of other implants can necessitate
due to release of nickel ions as evidenced from the further surgery. They are also used in total knee
above results. The present results clearly reveal that replacements and, to a lesser extent, as implants that
the breakdown potential is relatively low over Co and fix bone fractures like bone screws, staples, and
Ti alloys (Table 3). Further, the breakdown potential plates. The support structures for heart valves are
78 I. Gurappa / Materials Characterization 49 (2002) 7379

also often fabricated from cobalt-based alloys. A of Al ions results in phosphorus deficiency in blood
variety of dental implants have also been produced and bone due to the combination of Al with inorganic
from cobalt alloys. phosphorus while vanadium causes cytotoxicity.

5. Advantages of titanium and its alloys 6. Future research

Titanium possesses three outstanding character- The corrosion behavior of several existing materi-
istics for biomedical applications. They are als used today is not ideal in all respects. Although a
small number of biomaterials have been in use, post-
(a) corrosion resistance in tissue, healing surgery is essential in order to remove the
(b) absence of tissue toxicity and allergic reactions, implanted material. Further, the implant is obviously
and an item for which complete resistance of the metal
(c) good strength, a factor for the safety of the device should be aimed at throughout the period of use.
and low elastic modulus, which appears desirable Therefore, it is essential to develop a material that can
for reducing stress-shielding effects of bone due satisfy all the required properties over the stipulated
to the presence of the implant. period and then start degrading in the body itself
without further surgery. In fact, it is a real challenge
The corrosion resistance of titanium and its alloys to scientists and surgeons and has tremendous poten-
is obtained due to the formation of a stable oxide tial for future research. This will direct the attention
scale on their surfaces. It can reform if damaged in to the development of an excellent biocompatible and
physiological fluids at body temperature. A variety biodegradable polymeric material, which will cer-
of titanium alloys have been developed by the tainly be a smart material and thereby eliminate costly
addition of different alloying elements. However, post-healing operations.
only Ti 6Al 4V alloy and four grades of commer-
cially available pure titanium are customarily used in
the body. Pure titanium is used in cases where there Acknowledgements
is low mechanical load on the implant, while Ti
6Al 4V alloy is used where good mechanical load is The Defence Research and Development Organ-
there. isation is gratefully acknowledged for financial
Ti 6Al 4V alloy offers the best combination of assistance.
properties of all structural implant materials today as
evidenced from its maximum breakdown potential,
minimum corrosion rate, and high resistance to pit- References
ting and crevice corrosion. This alloy has a much
greater yield and ultimate strength than pure titanium. [1] Donachie M. Biomedical alloys. Adv Mater Process
It is generally produced by forging, but can be cast. 1998;7:63 5.
However, release of titanium into adjacent tissue has [2] Steinemann SG. In: Winter GD, Leray JL, De Goot
been reported upon prolonged exposure. It is believed K, editors. Corrosion of surgical implants in vivo and
in vitro tests. Evaluation of biomaterials: Wiley; 1980.
that a proper pretreatment before surgery can prevent
p. 1 34.
this phenomenon [16].
[3] Therin M, Meurier A, Chrisled P. A histomorphometric
A titanium-based alloy, Ti 5Al 2.5Fe, was dev- comparison of the muscular tissue reaction to stainless
eloped as an implant material, and it was reported that steel, pure titanium and titanium alloy implant materi-
its mechanical properties are comparable to the exist- als. J Mater Sci Mater Med 1991;2:1 8.
ing alloy, Ti 6Al 4V, and safe for human use. [4] Leitao E, Burbosa MA, De Goot K. In vitro testing of
Various titanium alloys with different alloying ele- surface modified biomaterials. J Mater Sci Mater Med
ments were studied for their corrosion resistance [17]. 1998;9:543 8.
The modified titanium-based alloys, Ti 15%Zr [5] Granchi D, Cenni E, Ciapetti G, Savarino L, Stea S,
4%Nb 2%Ta and Ti 15%Sn 4%Nb 2%Ta Gamberini S, Gori A, Rizzoferrato A. Cell death in-
duced by metal ions: necrosis or apoptosis. J Mater Sci
0.2%Pd, with improved biological safety, strength,
Mater Med 1998;9:31 7.
and ductility for medical implants were developed
[6] Gurrappa I, Kiran Kumar M. Evaluation of corrosion
[18,19]. These alloys were reported to exhibit super- resistance of plasma sprayed alumina, magnesia stabi-
ior properties over that of Ti 6Al 4V extra low lized zirconia and yttria stabilized zirconia coatings on
interstitial (ELI) alloy since Ti 6Al 4V ELI has 316 stainless steel for biomedical applications. Proceed-
created recent concerns due to the suspected effects ings of Global 2000 Corrosion Meet, Mumbai, India,
of aluminium (Al) and vanadium (V). The presence vol. 1. Quest Publications, Mumbai; 2000. p. 561 72.
I. Gurappa / Materials Characterization 49 (2002) 7379 79

[7] Gurrappa I. Effect of cold work on corrosion character- perimental approach and results. Corrosion 1982;38:
istics of cobalt alloy used in biomedical applications. J 570 9.
Mater Sci Lett Med [submitted]. [14] MacDonald D, Mckubre M. Electrochemical impe-
[8] Kumar P, Hicki AJ, Asphahani AI, Lawley A. Properties dance techniques in corrosion science. Electrochemical
and characteristics of cast, wrought and powder metal- corrosion testing. ASTM Spec Tech Publ 1981;727:
lurgy processed cobalt chromium molybdenum im- 110 49.
plant materials. ASTM Spec Tech Publ 1985;859: [15] Mansfeld M, Kendig M, Tsai S. Evaluation of corro-
30 41. sion behavior of coated metals with AC impedance
[9] Gurrappa I. Corrosion and its importance in selection measurements. Corrosion 1982;38:478 85.
of materials for biomedical applications. Corros Prev [16] Solar RJ, Pollack SR, Korostoff E. Titanium Release
Control 2001;48:23 37. from implants: a proposed mechanism. J Biomed Ma-
[10] Gurrappa I. The influence of Nb and Ga addition on ter Res 1974;5:219 26.
the corrosion resistance of permanent magnets in ma- [17] Haruyuki K. Effect of different alloying elements on
rine environments: Part I. Corros Prev Control 2001; corrosion resistance of titanium alloys for biomedical
48:116 21. applications. Bull Jpn Inst Metals 1992;31:1033 9.
[11] Gurrappa I. Suitability of Nd Fe B permanent mag- [18] Okazaki Y, Ito A, Tateishi T, Ito Y. Effect of alloying
nets for biomedical applicationsa corrosion study. elements on anodic polarization properties of titanium
J Alloys Compd 2002;339:241 7. alloys in acid solutions. Mater Trans JIM 1994;35:
[12] Mansfeld F. Recording and analysis of AC impedance 58 66.
data for corrosion studies: I. Background and methods [19] Okazaki Y, Ito A, Tateishi T. Effect of alloying el-
of analysis. Corrosion 1981;36:301 7. ements on mechanical properties of titanium alloys
[13] Mansfeld F, Kendig M, Tsai S. Recording and analysis for medical implants. Mater Trans JIM 1993;34:
of AC impedance data for corrosion studies: II. Ex- 1217 22.

Вам также может понравиться