Академический Документы
Профессиональный Документы
Культура Документы
Procedure for autoclave validation including steam penetration, heat distribution and penetration, biochallenge study, estimation of F0 value and acceptance
criteria of steam sterilizer validation in pharmaceutical industry.
Sterilization process in pharmaceutical industry in an autoclave will be considered qualified for consistent and reliable performance validated on successful
completion of the following tests
Bowie Dick Test for steam penetration 3 trails.
Empty Chamber Heat distribution studies 3 trails with temperature mapping probe at different locations
of the sterilizer chamber.
Loaded chamber heat Distribution & penetration studies 3 trails for each sterilization load of fixed
loading pattern
1 Sterile area garments 20 number Garments packs, Each pack contains 01 Nos. Boiler suit, 01 Nos.
Headgear, 02 Nos booties, 01 pairs gloves
2 Glassware S.S Mannifold holder 06 holder 02 Nos, Sampling unit of Compressed air 02 Nos, 500 ml
sampling bottles 10 Nos, 250 ml sampling bottles 25 Nos, 04 Nos S.S Bin.
3 Media SCDA Medium 500 ml 09 Nos. Conical flask, SCDM 100 ml 20 Nos. tubes, FTM 100 ml 10
Nos. tubes, MSA 250 ml 01 Nos Conical flask, CA 250 ml 01 Nos Conical flask, BGA 250 ml 01 Nos
Conical flask, BSA 250 ml 01 Nos Conical flask, MCA 250 ml 01 Nos Conical flask, Peptone Water 500
ml 06 Nos. Conical flask.
With temperature mapping probes along with Biological Indicator Geobacillus stearothermophillus spore vials containing 10^6 or more spores per vials
inside the innermost possible layer of the load subjected for sterilization.
Estimation of the F0 Value achieved during the sterilization hold period at each temperature mapping probe.
To qualify these tests the equipment should fulfill the acceptance criteria described in the individual test procedures. After completion of the qualification
tests all the data generated will be compiled together to evaluate ability of the steam sterilizer to sterilize different components at the set parameters and
set loading pattern.
A Bowie Dick test for Steam Penetration
Objective
Objective of this test is to ensure that the vacuum pulses applied the sterilization hold period are sufficient to remove the entrapped air so as to facilitate
rapid and even steam penetration into all parts of the load and maintaining these conditions for the specified temperature holding time 17 minutes at 121
deg.C
If air is present in the chamber, it will collect within the Bowie Dick test pack as a bubble. The indicator in the region of the bubble will be of
different color as compared to the color on the remaining part of the test paper, because of a lower temperature, lower moisture level or both. In this
condition the cycle parameters to be reviewed and the normal sterilization cycles to be modified accordingly.
Bowie Dick cycle should be normally preceded by a warm up cycle, as the effectiveness of air removal may depend on all parts of the sterilizer being at
working temperature.
Procedure
1. Record the set parameters for the Bowie Dick test cycle in The Annexure.
2. Place one Bowie Dick test pack near the drain of the sterilization chamber.
3. Select cycle Bowie Dick on the control panel & operate the steam sterilizer.
4. The print out taken during the Bowie Dick test cycle & the Bowie Dick test indicator should be preserved.
5. Compile the observation made during the qualification test for complete evaluation of the system.
Acceptance Criteria
The Bowie Dick Test indicator should show a uniform color change, non uniform change and/or air entrapment bubble spot on the pattern indicates
inadequate air removal from the sterilization chamber.
Observation & Results
Record the observation and results in format.
F0 = dt S10T121/z
Where
dt = Time interval between two following temperature measurements 1 minutes.
T = The observed Temperature at that particular time.
Z = The change in the heat resistance of Geobacillus stearothermophillus spores as temperature is changed 10 deg.C.
Acceptance Criteria
The calculated minimum F0 value by equation should be more than biological F0 value for the biological indicator vial exposed for the biochallenge
studies.
The biological F0 value for the specific biological indicator spore vial is calculated as per the following equation
F0 = D121 Log A Log B
Where
D121 = D value of the of the biological indicator at 121 deg.C.
A = Biological indicator concentration or spore population.
B = Desired level of non sterility. 10 deg.C
Click here for further details
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of Pharmaceutical Guidelines, a widelyread pharmaceutical
blog since 2008. Signup for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question
Offline
20082017 All rights reserved to Ankur Choudhary and data may not be reproduced on other websites. Designed by Pharmaceutical Guidelines.
Privacy Policy Disclaimer