Calibration Cert
Indicated Force (F) True Force(F) Correction (F)
0 0 0
9.99 9.88 0.11
50.02 49.44 0.58
100.4 100 0.4
200.14 199.59 0.55
401.48 400.42 1.06
600.68 599.83 0.85
800.17 802.11 -1.94
851.49 853.33 -1.84
Page 1
Round specimens TT1
Reference(F) S1 = 2.74 kN
K1 = 2
U1 = S1/K1 kN
1.37
correction(F) S2 = 1.94 kN
U2 = S2/(2x3)
0.561
UF = (u1 + u2)
1.481
Reference(D) S3 = 0.01 mm
K3 = 2
U3 = S1/K1 mm
0.005
Correction(D) S4 = 0.02 mm
U4 = S4/(2x3)
0.006
Repeatability(D) S5 = 0.068
U5 = S2n mm
= 0.04 mm
Uc = (Cfuf + C(D)uD )
Cf = 4/PID2
0.009
C(D) = -8F/PID^3
-0.083
UC = 0.004
U = Uc x k K= 2
= 0.008 kN/mm2
0.008 kN/mm2
Page 2
Rectangular specimens TT1
Calibration Cert
Indicated Force (F) True Force(F) Correction (F)
9.8 9.7 0.1 0.1
49.1 48.9 0.2 0.2
98 98.2 -0.2 0.2
196 196.1 -0.1 0.1
392 393 -1 1
490 491.8 -1.8 1.8
588 590.8 -2.8 2.8
784.1 789.2 -5.1 5.1
837.1 843 -5.9 5.9
max 5.9
Force As per test report 140.1 kN
Tensile Strength = Indication + Reference (F)
Correction (F)
Reference (W)
Correction (W)
Repeatability (W)
Reference (T)
Correction (T)
Repeatability (T)
Page 3
Rectangular specimens TT1
Reference(F) S1 = 2.95 kN
K1 = 2
U1 = S1/K1 kN
1.475
correction(F) S2 = 5.9 kN
U2 = S2/(2x3)
1.704
Cf=1/A = 0.047
CW=-F/TW2 = 0.042 kN
CT=-F/T2W = 0.025
Uc = 0.445 kN/mm2
0.445 kN/mm2
U = Uc x k K= 2
= 0.89 kN/mm2
0.89 kN/mm2
Page 4