Академический Документы
Профессиональный Документы
Культура Документы
MOT Lab
Guide
ANDREW M. C. DAWES
C HAPTER 1
Getting
Started
2
S ECTION 2 scan appears on the scope as shown to the right. Use the fre-
quency (C) and span (D) knobs to shift and zoom respectively.
Trap Laser When the scan matches the photo here, use the error offset
knob (E) to bring the blue line into the scan as shown. The
blue line should cross the zero-voltage axis just to the left of
the resonance blip. At this point, the laser is ready to lock, so
flip switch (F) from scan to lock.
The laser power is enabled by switch (A) and the current ad-
justed by knob (B). Usually the current is close to the correct
value from day to day. Adjust the current until the resonance
B C D
F
A E
3
S ECTION 3 The repump laser has a
dedicated box for each
Repump Laser control (temperature,
current, and scan). The
temperature control
should not need adjust-
ment. To tune this laser,
adjust the current until a
shallow dip is visible on
the scope.
If the dip is not centered, use the offset knob on the scan con-
Even in the cycling transition, not all atoms excited by the trol to shift the scan onto resonance. Finally, turn the ampli-
trap laser decay back to the F = 2 ground state. Those that re- tude fully CCW to park the laser on the resonance.
lax to the F = 1 state must be pumped back to the F = 2 state
to interact with the trap laser again. This repumping action is
provided by the repump laser. To operate correctly, it must be
tuned to the F = 1 F resonances.
Probe Laser
5
6
S ECTION 5 This is the hardest part of the lab, but you can do it.
Making a MOT Turn the Rb Getter current on (switch located on the mini-
MOT). The current should be between 3.15 and 3.3 A. This pro-
vides Rb atoms for the trap, and it takes a few minutes to load
the cell with atoms. Instead of waiting, make sure the trap and
repump lasers are on resonance, and practice locking the trap
laser.
When the lasers are all tuned to resonance, and the trap laser
is locked, turn on the magnet power supply (near the MOT).
Watch the camera for a mot cloud as you lock the trap laser
and gently adjust the ERROR OFFSET. This is what makes it
hard, the knob is sensitive so you dont need to turn it much
(small, slow turns by about as little as you can manage).
The ERROR OFFSET knob shifts the lock point for the laser
and you should see the MOT cloud start to form. As you tune
the lock point, wait whenever you see the mot form as it takes
a few second for the cloud to grow to a reasonable size.
O BJECTIVES
7
C HAPTER 2
Lab activities
I/Is
R=
2 1 + I/Is + 4(2 / 2)
9
The most important part of this measurement is an accurate Analysis
assessment of the background light (i.e. light that does not
The detector calibration has been performed already:
come from the trap cloud). There are several ways to do this,
but keep in mind that blocking the laser beams eliminates the 0.92 0.01 V/W
trap fluorescence and a source of background light.
You can use this calibrated output to calculate the detected
Using the magnets is a more reliable way to turn off the trap power based on the measured voltage. Compare the observed
without affecting the amount of measured light. The figure be- power to the expression for the expected power scattered by
low shows a data trace where the detector sees the MOT and the atom cloud and solve for the number of atoms in the
the background light, then the magnetic field is switched off cloud. Be sure to take into account the fact that we only detect
and on and the trap begins to re-form. You are interested in a small portion of the light scattered from the MOT.
the difference in optical power from with-the-trap to without-
the-trap.
Scatter Detector (V)
0.06
0.05
5 10 15 20
Time (s)
10
S ECTION 2 The trap fill rate is the rate at which atoms accumulate in the
trap. This rate indicates a variety of things but the parameter
Lab II: Fill Rate it depends on most strongly is the background vapor pressure
in the trap cell. If there is a lot of Rubidium in the vapor cell
(i.e. a high background pressure) the trap fills quickly as at-
oms are readily caught. However, this also means atoms
leave the trap quickly and it is not always desirable. On the
other hand if the background pressure is low, atoms accumu-
late slowly and tend to remain in the trap for a longer period
of time. For these reasons, the fill rate is a reasonable measure
of the background pressure in the vapor cell.
With the MOT running and the trap full, set the scope to col-
lect at least 25 s of data (it will be in untriggered mode). With
a bright MOT, turn the magnet power supply off and back on
O BJECTIVES
while the scope is near the beginning of a full trace. The detec-
1. Measure the trap fill rate tor signal should drop several mV (hopefully at least 20 mV)
and begin to increase along an exponential growth curve. This
2. Calculate the background vapor pressure growth is the brightness of the MOT increasing as atoms fill
the trap.
11
t=?s
Analysis
In your graphical data program of choice, fit or model an expo-
nential growth function to the data. You only need to recover
the time constant from the data, but it is important to allow
other parameters in your fit or model in order for it to be accu-
rate. Expect a result on the order of 10 seconds.
12
S ECTION 3 One question that often comes up is why do we want to trap
atoms? The main reason is that if we dont trap them, they
Lab III: Cold atom absorp- tend to fly all over the place. Room-temperature atoms move
tion spectrum at several hundred meters per second. This means they inter-
act with each other and with beams of light very briefly. If we
want to have a sample of atoms to interact with repeatedly, or
for a long time, we need to hold them still.
13
10 ms per div., adjust the vertical scales of each channel to see
the full signal. The screenshot below, should be your goal.
Once the MOT cloud is well formed, three distinct dips should
appear in your transmission scan. These correspond to the
three resonances in Rubidium vapor. Note they also corre-
spond to three of the features in the saturated absorption
scan.
0.5
0.7
Probe signal (V)
Sat. Abs. (V)
0.4 0.6
0.5
0.05 0 0.05
Time (s)
14
C HAPTER 3
193.7407(46) MHz
gF o=o2/3
(0.93 MHz/G)
F=3
25
266.6500(90) MHz
52P3/2
72.9112(32) MHz
gF o=o2/3 F=2
229.8518(56) MHz (0.93 MHz/G)
302.0738(88) MHz 156.9470(70) MHz
gF o=o2/3 F=1
72.2180(40) MHz (0.93 MHz/G)
F=0
gF o=o1/2 F=2
(0.70 MHz/G)
2.563 005 979 089 109(34) GHz
52S1/2
6.834 682 610 904 290(90) GHz
gF o=o-1/2 F =1
(-o0.70 MHz/G)
Figure 2: Rubidium 87 D2 transition hyperfine structure, with frequency splittings between the hyperfine energy
levels. The excited-state values are taken from [9], and the ground-state values are from [29]. The relative hyperfine
shifts are shown to scale within each hyperfine manifold (but visual spacings should not be compared between
manifolds or to the optical splitting). The approximate Lande gF -factors for each level are also given, with the
corresponding Zeeman splittings between adjacent magnetic sublevels.