Вы находитесь на странице: 1из 4

/ APPLICATION NOTE

8 Steps to Validating/Mapping a Chamber

Validation in 8 Steps

Step 1
Write a plan
Step 2
Check equipment &
documentation
Step 3
Confirm device operation
Step 4
Set up the data loggers
Step 5
Place the data loggers
Periodic mapping of conditions within environmental chambers is Step 6
critical for compliance in FDA-regulated applications, where validation Periodically check the progress
of conditions such as temperature, and sometimes humidity, is Step 7
mandated. (1, 2, 3, 4, 5) This application note recommends some options Retrieve and store the data
to help meet cGMP requirements for validation projects. Note: we refer Step 8
to probes, sensors, data loggers interchangeably within this application Report the findings
note. These are broad guidelines only, and most recommendations are
based upon using data loggers as the sensing device.

Step 1 many regulatory bodies mandate within a chamber (with the exception
Write a Validation Plan temperature mapping of controlled of very small bench-top applications).
spaces, they do not specify any This comprises two layers of four
First, define the objective of the
particular method so the onus is loggers, one near each corner and one
validation in writing, creating an
therefore on us to document and logger in the center. Or,
outline of the methods to be used,
provide justification for our mapping
and list any anticipated barriers. In
procedures. Fifteen (15): three layers of four
most cases, these three items will form
loggers one near each corner and one
the bulk of the validation protocol.
How many data points do you need logger in the center.
Below there are several questions,
to monitor/map?
the answers to which will ideally be in
This number will vary based on Or, four or five loggers per shelf.
writing as part of the Validation plan.
a number of factors, including
environment, temperature/RH range, Where will each data logger
What regulations and
application, but a typical number of be placed?
requirements must you comply with?
units used to map small chambers can Spacing the loggers in an evenly
Start with a review of those listed in
include: distributed grid is recommended,
the quality guidelines of the facility
however, its also important to monitor
(i.e. CFRs 210, 211, etc) and scan for
Nine (9): In most cases, this is the the worst-case locations within a
recent changes or updates. Although
bare minimum of sample locations chamber for heat-loss and/or air
movement. Monitoring the corners of How long will the mapping study be? mapping study, and a measurement
the chamber and near any openings/ Again, it varies and justification may interval of 15-30 seconds. With the
pass-throughs will cover most worst- be required. Typical study lengths chamber temperature stabilized and
case locations; however shelving/ are 24, 48, or 72 hours. Occasionally data loggers recording data, the
racking within the chamber may facilities map for one week, one month, door to the chamber is opened for
require the identification of additional or even longer. For laboratory scale a period of time that is typical for
worst-case locations. Place a sensor chambers and walk-in chambers, 24- normal operation of the chamber
at or near the control sensor for the 72 hours can be sufficient. However (1 minute for laboratory scale
temperature control unit, and any for larger scale projects (i.e. a chambers, up to 5 minutes for large
alarm sensors within the chamber. temperature controlled warehouse) walk-in chambers used in shipping/
where the temperature can be receiving areas) and then closed.
What is the chamber load? influenced by the temperature outside Data collection continues until the
Will the mapping take place with the the facility, longer studies may be chamber returns to specified
chamber empty, as for an Operation required. Larger scale chambers may operating range. This test shows
Qualification (OQ), or full of product, also require studies during summer that product temperatures are not
as for a Performance Qualification and winter months depending on the adversely affected during normal
(PQ)? For most pharmaceutical and seasonal temperature variations at the operation of the chamber.
biotechnical applications, both tests facilitys location.
are important. Consider how OQ Insulation Thermal Conductivity/
and PQ might affect the process. What will the duration be? Temperature Change Study
As well, some regulatory bodies(2) Pre-determining a test schedule will This study also uses the same setup
mandate the use of maximum and allow the validation technician to set as the mapping study, with the same
minimum loads during the validation all data loggers to simultaneously start measurement interval of the
process. An empty chamber can be and stop collecting data at the same mapping study. With the chamber
considered the minimum load and is time. temperature stabilized and data
usually the worst-case for temperature loggers recording data, power is
fluctuations within the chamber. How frequently will testing removed from the temperature
be repeated? control unit for the chamber. Data is
Tracking air temperature inside While some industries only require an usually collected for a 12-hour
the chamber Vs. interior product initial qualification, some require that period. This study provides
temperatures: temperature mapping be repeated on a information on how long the
Tracking interior product regular basis. Once again, review any chamber remains within the
temperatures, such as the temperature applicable regulations and be sure to specified operating range in the
inside a solution bottle, is sometimes document the intended test frequency event of a power failure. Data
considered more meaningful as it in the validation plan/protocol. from this study can be used to help
insulates the data against minor Generally, while the initial qualification determine if product is adversely
disturbances such as periodic door will usually include mapping of the affected following a power failure,
openings and closings. empty chamber and the loaded as well as for developing procedures
chamber, subsequent re-qualifications for transferring product to a
What parameters will you measure? only require mapping of the loaded different chamber/facility in the
Should the chamber be mapped for chamber. event of failure.
temperature only? Or temperature
and relative humidity? If there is any What additional tests may be useful? Many of these questions may already
plan to store products with humidity In most cases regulatory bodies be answered by past practices,
sensitivities, then the chamber should only require validation of an empty protocols, or SOPs.
be mapped for relative humidity as chamber and a loaded chamber.
well as temperature. However, other types of tests may
provide valuable information. Two
How frequently will take readings? common tests that many companies
Typical sample rates are once every find useful are:
minute or once every five minutes,
however, as with most other aspects Temperature Recovery Study This
of validation, be prepared to justify study uses the same number of
sample rate and include that rationale probes in the same locations as the
in the plan and/or protocol.
Step 2 Step 3 the start/stop times will eliminate the
Check Equipment & Confirm Device Operation collection of extraneous data;
Documentation Check for damage before use. The facility Set a sample interval that is
In a GMP environment it is essential to may have an SOP that covers this type appropriate to the application. Again,
ensure that the equipment being used to of check. If not, there are several ways to a typical safe sample interval might
perform any chamber qualification is not ensure proper operation. For example, be once per minute;
only appropriate for the job but in good if the data loggers have been stored Set a meaningful data logger
working order. together and set to wrap (take readings description for each of the data
Two considerations: on a continuous sliding window basis), loggers. For example, each data logger
simply transfer the information collected might have an ID that describes its
Have the loggers been validated for this
by the data loggers to a PC and compare location in the chamber.
type of application? Ideally the
the readings from the various data
equipment manufacturer will provide
an IQ/OQ validation protocol expressly
loggers. If the loggers have not been Step 5
for this purpose.
stored together, another way to confirm Place the Data Loggers
their function is to place the loggers side
Are the sensors currently in calibration? Place the sensors in the chamber
by side for a short time and compare the
Check the Calibration Certificates according to the validation protocol,
readings. To minimize the risks of using
(also provided by the manufacturer). or according to a diagram or grid.
any equipment that might have drifted
The validation software should also The interior of the chamber can be
out of calibration, there are several
show calibration date information, often photographed to show both logger
possible ways to check calibration.
stored the memory of each device. If and product placement. If using probe
There may be obvious signs of problems sensors (rather than just loggers with
using thermocouples, a pre- and post-
such as measurement readings that vary internal sensors), it is important to
calibration is required. For NIST-
widely from the norm. If the environment ensure that they are not in contact any
traceable calibrated data loggers, no
has been stable, look for chemical surfaces within the chamber.
pre- and post calibration is required.
contamination that may have corrupted
Be sure to also place a data logger
the sensor measurements. Or if the logger
outside the chamber to monitor
has been dropped, physical damage
ambient temperature, or both ambient
in the form of a broken circuit may be
temperature and relative humidity.
the cause. For less obvious causes of
Failure to do so may invalidate the
out-of-cal performance, a quick accuracy
results as exterior temperatures can
check can be performed without sending
sometimes significantly influence
the loggers to a calibration lab. Simply
interior performance. Place this outside
compare the logger(s) to other calibrated
data logger away from areas of undue
(NIST-traceable) data loggers or a NIST-
temperature influence such as an
traceable instrument for a reference.
outside wall, a heat-generating piece of
Finally, comparing any suspect loggers
equipment, near a window, or in a high
to an accepted reference standard (such
traffic area.
as an ice bath or saturated salt solution
environment) is another sanity check.
Step 6
Step 4 Periodically Check the
Set Up the Data Loggers Progress
Set up the test start and stop times If possible, its a good idea to check the
within the mapping software as well study results periodically to ensure that
as the sample intervals and the logger the mapping is progressing as expected.
descriptions; As an example, a typical validation
Use a delayed start time to ensure protocol may call for the validation
sufficient time to set up and place technician to establish when the
all the data loggers, and to allow temperature in the chamber stabilizes or
the chamber to stabilize before data to verify operation throughout the course
collection begins; of the study. Using a flat cable can allow
for periodic checks of the data for one
Configure the same start and stop time
logger without disturbing the chamber.
for all the data loggers. Synchronizing
It is important in the report to identify
the maximum, minimum and average References
temperatures recorded at each data
logger location. In some cases it is also 1 U.S. Pharmacopeia: Good
valuable to include the Mean Kinetic Storage and Shipping Practices
Temperature at each location. This <1079> USP 32-NF 27
information is valuable in identifying any
hot or cold spots within the chamber, as 2 MHRA: Rules and Guidance for
Step 7 these are the locations where sensors Pharmaceutical Manufacturers
Retrieve and Store the for continuous monitoring and alarms and Distributors, 2007.
Data should be located (3). 3 Health Canada (Guide-0069)
Once the study is complete, to transfer Guidelines for Temperature
the data from each data logger to a PC. Control of Drug Products During
Customize the filename format to make it Storage and Transportation
easy to track files and relate them to the 4 EU European Commission Annex
deployment. It is recommended that the 15 Guide to Good Manufacturing
software be set to automatically include Practices
the following information in each file
5 FDA CFR 21 211.142
name: serial number of the data logger,
time and date of the data download,
and user description of the data logger.
Most validation software will provide
files that are tamper-proof and secure in
order to meet the FDAs electronic record
requirements of 21 CFR Part 11. Each file
must be uniquely identified, allowing
users to relate each printout to the
original data logger files.

Step 8
Report the Findings
The results of the mapping study can be
reported in several different ways:
All data logger files can be plotted
on one graph to show the variances
in different parts of the chamber.
Or, a raw data printout of all readings
and associated times and dates can be
printed to support the above graph.
Or, data can be exported to Excel to
facilitate further reports.

B211103EN-A Vaisala 2011


This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications technical included are subject
to change without notice.

Вам также может понравиться