Service Manual DIG-610

SM01-02-01

SERIVEC MANUAL
MODEL: DIG-610
(PORTABLE X-RAY UNIT)

* ALL THE INFORMATION PRESCRIBED INSIDE THE MANUAL IS PRESENTED ONLY AS A GUIDE FOR THE OPERATION
OF OUR PRODUCTS. AND IT MAY BE CHANGED WITHOUT PRIOR NOTICE.
* DONGMUN CO., LTD.: 717-HO, UNITECHVIL B/D, 1141-2, BAEGSUK-DONG, ILSAN-GU, KOYANG CITY, SOUTH KOREA
TEL: 82-31-907-7222, FAX: 82-31-904-2231,E-MAIL: dongmun@kita.net, URL: dong-mun.co.kr
Service Manual DIG-610

CONTENTS

1 Notice ----------------------------------------------------------------------- 27
2 Operation Sheet ----------------------------------------------------------------------- 27
3 Block Diagram ----------------------------------------------------------------------- 28
4 Calibration Process ----------------------------------------------------------------------- 29
5 Calibration Code and Data ----------------------------------------------------------------------- 33
6 Error code & Countermeasure ----------------------------------------------------------------------- 33
7 Schematics ----------------------------------------------------------------------- 34
8 Maintenance ----------------------------------------------------------------------- 40

TABLES
1 mA Calibration Code ----------------------------------------------------------------------- 33
2 Filament Preheat Calibration ----------------------------------------------------------------------- 33
3 Technical guide ----------------------------------------------------------------------- 41

2
Service Manual DIG-610

1. NOTICE

1.1. If the output calibration is needed after periodical inspections or repairs, please observe following
rules and methods.

CAUTION
The ionized radiation is dangerous for the operator if the following
safety measures are not strictly observed.

1.2. This manual is designed to ensure correct use and operation of DIG-610. Please read all the lines
thoroughly before you use this equipment.
1.3. Incorrect use and operation exceeding described conditions in this manual may occur damage of
the machine and shorten its life. Particular attention must be paid to all the warnings, cautions and
notes incorporated herein.
1.4. This equipment should be used only by the legally qualified persons and practitioners.
1.5. Regarding the AC cable and software, only the one developed and supplied by Dongmun should
be used.
1.6. No responsibility is taken by Dongmun for any infringement of patients or other rights of the third
parties which many results from the use of this manual.
1.7. Always keep the manual at hand for your reference.

2. OPERATION SHEET

3
Service Manual DIG-610

3. BLOCK DIAGRAM

READY
1
PXP60HF-CTL-001
CON1 3V/35mA
2
X-RAY LASER POINTER
HAND SWITCH
3
35312-03
CONTROL BOARD CON2 1
2 * EXPORT : NORMAL
* DOMESTIC : NOT USED
OP PANEL 35312-02 LASER

1 CON6
2 & HT CONT. TANK
35312-02 T

2
CON5 CON4 35312-05
35312-11 CON3 THERMAL
35312-05 EP_FB(-)

10
11
EP(-)
1
2
3
4
5

1
2
3
4
5
6
7
8
9

1
2
3
4
5

1
AGND
GND

LAMP_CONT
FIL_CONT1
FIL_CONT2
EP_CONT1

EP_CONT2

/100V_SEL
IP_FB(+)
IP(+)

FIL_FB
AGND
DGND
+12V

-12V
+5V

10
11
1
2
3
4
5

1
2
3
4
5
6
7
8
9
35312-05 35312-11

CON2 CON3 L101


PXP-60HF-IGBT HT XFMR
TB1 TB2 TB4 T1
IGBT BOARD 3TURNS
T1
20 575
1 TB2 TB1 TB3 T2
2 CON5
3 CN1 CN2 CN3 CN4 COM
35312 35312 35312 35312
D-205
35312-03

1
2

1
2

1
2

1
2
35312
1
CON7 2

35312
PXP60HF-PWR CON10
1
2
FG
35312
1
CON8 2

35312
1
PWR & DRIVE CON9 2

CON4 T2
1 C1 3
FILAMENT DRIVE 2 2
3 +12.2
CB1 4 SHIELD GND SHIELD LINE 3
53258-04 C2
FIL XFMR
5
4 53258-05
3
2 CON1
1 COLLIMATOR DRIVECON6
250V/16A 1
250V/20A 2 COLLIMATOR
CN1 SHIELD LINE
PLUG AC MALE LAMP
53258-02 12V/50W

APPROVED BY DESIGNED BY <OrgName>

DIG-610 Main Block


PXP-60HF-Main Block
CHECKED BY DATE
Size FSCM NO. DWG NO Rev
A3 A
<Cage Code>
Sheet
1 of 1

4
Service Manual DIG-610

-PREPARATION
Following tools and instruments are needed.
a. Tools: Philips screwdriver No.1 & 2, Straight slot screwdriver
b. Instruments: DVM, oscilloscope (Tektronics TDS 3032 and more than Equivalent)

4. CALIBRATION PROCESS

4.1. Line Power Supply


-Line switch should be at off position.
-Connect power line cord to the power outlet.
-Turn on the line switch.
-Then the LED on the operation panel illuminates.
4.2. Calibration procedures are composed of three steps.
-Tube voltage confirmation and calibration.
-Tube current confirmation and calibration.
-Filament preheat calibration
A. Calibration and Confirmation of Tube Voltage
-Turn off the line switch.
-Lift up the control panel from the unit, and then the PCB OP board appears.
-All DIP switches on the PCB OP board are turned off. (Note: DIP switch is usually located in this
location.)
-Connect the channel one probe of the oscilloscope to TP5 (IP F/B).
-Connect the channel another one probe of the oscilloscope to TP6 (EP F/B).

Channel 1 is mA, and 60mA = 9 V


Channel 2 is kV, and 40kV = 2V

-Connect the ground to TP11 (GND).


-Turn on the line switch and the display indicates as follows.

kV mAs

-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the
oscilloscope channel 1.
-If it is not about 9V, adjust mA by VR3 on the OP board. Then observe the kV waveform of the
oscilloscope channel 2.
-If it is not 2V, adjust kV by VR2 on the OP board. After adjustment, set the tube voltage kV to 80KV
by the kV up/down switch.
-Set the kV and the mAs at 80kV and 0.4mAs(40mA).Then observe the mAs, kV waveform of the
oscilloscope channel 1, 2.

5
Service Manual DIG-610

-If it is not about 6V, adjust mA by VR3 on the OP board.


-If it is not about 4V, adjust kV by VR2 on the OP board.

B. Calibration and Confirmation of Tube Current


-Turn off the line switch.
-Lift up the control panel from the unit, and then the PCB OP board appears.
-All DIP switches on the PCB OP board are turned off. (Note: DIP switch is usually in this position.)
-Connect the channel one probe of the oscilloscope to TP5 (IP F/B).
-Connect the channel another one probe of the oscilloscope to TP6 (EP F/B).

Channel 1 is mA, and 60 mA = 9V


Channel 2 is kV, and 40 kV = 2V

-Connect the ground to TP11 (GND).


-Turn on the line switch and the display indicates as follows.

kV mAs
-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the
oscilloscope channel 1.
-If it is not 7V, adjust mA by the calibration data on the OP board.
-Turn off the line switch, and wait until the display switch the light off.
-DIP switch 1 on the PCB OP board is turned on.
-Turn on the line switch and the display indicates as follows.
-Then push the mAs up switch (see operation sheet).
The display indicates as follows.

kV mAs
Calibration code Calibration data

C00 is calibration code of tube current (60mA at 40kV-50kV).


179 is calibration data of tube current (60mA at 40kV-50kV).

-Adjust mA by the mAs up/down switch (see operation sheet).


-After it is adjusted, push the store switch (see operation sheet).
-DIP switch 1 on the PCB OP board is turned off then exposure X-ray.
-If it is not 9V, set the calibration data again.
-DIP switch 1 on the PCB OP board is turned on.
-Then push the kV up switch (see operation sheet).

6
Service Manual DIG-610

The display indicates as follows.

kV mAs
Calibration code Calibration data

C01 is calibration code of tube current (40mA at 51kV-70kV).


153 is calibration data of tube current (40mA at 51kV-70kV).

-Adjust mA by the mAs up/down switch (see operation sheet).


-After it is adjusted, push the store switch (see operation sheet).
-DIP switch 1 on the PCB OP board is turned off then exposure X-ray.
-If it is not 6V, set the calibration data again.
-Table .1 shows the calibration code and the calibration data against an each tube current.

C. Filament preheat calibration


-This is preheating filament for making tube current.
-The overshoot or the undershoot may occur from the tube current if the preheat calibration data or
the VR3(preheat gain volume) is set by mistake.
-There are two methods to make preheat calibration setting. The calibration data is possible to be
changed in accordance with the tube condition.
-The VR3(preheat gain volume) adjusts the whole range of tube current. Filament preheat
calibration data adjusts each tube current.
-Turn off the line switch.
-Lift up the control panel from the unit, and then the PCB OP Board appears.
-DIP switch 1 on the PCB OP board is turned on.
-Turn on the line switch.
-Push APR switch and then the display indicates as follows.

kV mAs
Preheat calibration code Preheat calibration data

P00 is Filament preheat calibration code (Preheat calibration code of 60mA).


17 is Filament preheat calibration data (Preheat calibration data of 60mA).
-Adjust data by the mAs up/down switch (see operation sheet).
-After it is adjusted, push the store switch (see operation sheet).
-DIP switch 1 on the PCB OP board is turned off then exposure X-ray.
If overshoot or undershoot occur in the waveform, set the calibration data again.
Then push the kV up switch (see operation sheet).
The display indicates as follows.

7
Service Manual DIG-610

kV mAs
Preheat calibration code Preheat calibration data

P01 is Filament preheat calibration code (Preheat calibration code of 40mA).


37 is Filament preheat calibration data (Preheat calibration data of 40mA).

-Adjust data by the mAs up/down switch (on the control panel).
-After it is adjusted, push the Store switch (on the control panel).
-DIP switch 1 on the PCB OP board is turned off then exposure X-ray.
-If overshoot or undershoot occur in the waveform, set the calibration data again.
-Table. 2 shows the calibration code and the calibration data against an each tube current.
The calibration data is possible to be changed in accordance with the tube condition.

8
Service Manual DIG-610

5. CALIBRATION CODE AND DATA

[Table 1. mA Calibration Code and Data]


AC voltage Code Data Tube voltage Tube current mAs
220V C00 235 40kV - 50kV 60mA 0.4 - 100
220V C01 156 51kV - 70kV 40mA 0.4 - 64
220V C02 155 51kV - 70kV 40mA 80 - 100
220V C03 155 71kV - 80kV 40mA 0.4 - 32
220V C04 136 71kV - 80kV 35mA 40 -80
220V C05 120 81kV - 90kV 30mA 0.4 - 40
220V C06 115 81kV - 90kV 30mA 50 - 80
220V C07 115 91kV - 100kV 30mA 0.4 - 50
220V C08 97 91kV - 100kV 25mA 64 - 80

[Table 2. Filament Preheat Calibration Code and Data]


AC voltage Code Data Tube current
220V P00 7 60mA
220V P01 60 40mA
220V P02 75 35mA
220V P03 85 30mA
220V P04 95 25mA

6. ERROR CODE & COUNTERMEASURE


[Table 3. Code & Measures]
Error Code Content Action
When the line switch is turned on, this is inspected. Change the hand switch
Err. 07 The ready contactor of the hand switch is closed even
it is not pushed.
When the line switch is turned on, this is inspected. Change the hand switch
Err. 08 The x-ray contactor of the hand switch is closed even
it is not pushed.
It is displayed when the FIL FB signal is not detected Replace operation board or
Err. 10
in the state of standby. Replace HV tank
It is displayed when the EP FB signal is not detected Replace operation board or
Err. 11
in the state of standby. Replace HV tank
It is displayed when the IP FB signal is not detected in Replace operation board or
Err. 12
the state of standby. Replace HV tank
It is displayed when the FIL FB signal is not detected Replace main board or
Err. 14
in the state of ready. Replace power board
It is displayed when the EP FB signal is low in the Replace operation board
Err. 15
state of exposure.
It is displayed when the IP FB signal is low in the Replace main board or
Err. 16
state of exposure. Replace power board
It is displayed when the IP FB and EP FB signal is Replace main board or
Err. 17
high in the state of exposure. replace power board
It is displayed when the mAs is over in the state of Check the IP FB level and adjust it
Err. 19
exposure. Replace operation board
It is displayed when the connector CON6 on the main Check CON3
Err. 21
board is not insert.

9
Service Manual DIG-610

7. SCHEMATICS

(2F-5)
SEG_A (2G-5)
SEG_B (2H-5)
SEG_C (2J-5)
SEG_D (2K-5)
SEG_E (2L-5)
SEG_F (2N-5)
SEG_G (2O-5)
SEG_DP

R21
15 1/4W
1 2

R22
15 1/4W
1 2

10

10

10

10

10

10
7

5
U8 R23

DP

DP

DP

DP

DP

DP
G

G
A

E
C

D
F

F
(1B-13) 1 16 15 1/4W
SEG01_OUT I1 Q1
(1B-13) 2 15 1 2
SEG02_OUT (1B-12) 3 I2 Q2 14
SEG03_OUT I3 Q3
(1B-12) 4 13
SEG04_OUT I4 Q4
(1B-12) 5 12 R24
SEG05_OUT I5 Q5
CC
CC

CC
CC

CC
CC

CC
CC

CC
CC

CC
CC
(1B-12) 6 11 15 1/4W
SEG06_OUT (1B-12) 7 I6 Q6 10 1 2
SEG07_OUT 8 I7 Q7 9
GND VCC
3
8

3
8

3
8

3
8

3
8

3
8
SEG1 uD_SR105K SEG2 uD_SR105K SEG3 uD_SR105K SEG4 uD_SR105K SEG5 uD_SR105K SEG6 uD_SR105K
TD62304 R25
15 1/4W
1 2

+5V R26
15 1/4W
1 2

LED2 READY LED3 X-RAY LED4 WAIT LED5 M1 LED6 M2 LED7 M3 LED8 M4 LED9 M5
R27 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1
15 1/4W GREEN YELLOW RED GREEN GREEN GREEN GREEN GREEN
1 2

LED10 M6 LED11 M7 LED12 M8 LED13 ERROR LED14 REVERSE


R28 2 1 2 1 2 1 2 1 2 1
15 1/4W GREEN GREEN GREEN RED GREEN
(1M-6) 1 2
SEG08_OUT1

APPROVED BY DESIGNED BY
HTP-24 PORTABLE X-RAY

CONTROL BOARD OP PANEL & HT CONT.


CHECKED BY DATE
Size FSCM NO. DWG NO Rev
A3
<Cage Code> HTP24-CTL-001 0
Sheet
3 of 7

10
Service Manual DIG-610

+5V

1
RA7
4.7K

2
3
4
5
CON6
ISO_PWR (1B-10),(7G-13)
1 /FAULT
(1B-10),(5N-13)
IP_FB_IN
2 R29
ISO_GND 270
3 1 2

2
35312-0310
D33
1N4148 PC1
1 8 (1B-10)
HS_READY

1
CON1
2 7
1
2 3 6 (1B-9)
3 HS_X-RAY
R30
4 270
35312-0410 1 2

2
4 5
D34
1N4148 TLP620-2

+5V
D35 D36
CON2
2 1 2 1
1
1N4148 1N4148

(1M-7)
2 /LASER
35312-0210

APPROVED BY DESIGNED BY
HTP-24 PORTABLE X-RAY

CONTROL BOARD OP PANEL & HT CONT.


CHECKED BY DATE
Size FSCM NO. DWG NO Rev
A3
<Cage Code> HTP24-CTL-001 0
Sheet
4 of 7

11
Service Manual DIG-610

(7B-13)
IP_FB

R31 10K 1%
(1J-5)
IP_FB_ADC

R32
+12V
10K 1%

2
D37 U9A

4
LM358 U10 C21 BOX332

2
1N4148 2 - R33 39K 14 5 1 2
40mA(0 ~ 8V) R34 4.7k 1 1 3 1 +IN 1SC 10
-IN CPI

1
3 + 11
VR1 +12V COM

2
10K 12 7 (1B-10),(4J-11)
IP_FB_IN

2
+VS FOUT

8
+ D38 4 13
R35 C62 -VS VOUT
1/2W 200 1N4148 -12V VFC32

1
C22 MN104

1
C24 BOX103
6.8uF/16V tantal 1 2

CON3
IP_FB(+)
1

2
EP_FB(+) C60 39P
3 2 1
4
R75 100K 1%
5
35312-0510
(1A-10) (6B-14),(7B-11)
/HT_CHECK
EP_FB
+12V +12V C28 MN104
2 1

2
D39

8
100KV(0 ~ -5V) 1N4148 5 R36 1K
+
R37 100K 1% 7 (1J-5)
EP_FB_ADC

1
6
-
2

2
C25 C26 C27 LM358 R84

2
R

4
D40 U9B
BOX223 BOX223 BOX223 C58 MN104
1N4148
1

R76 2 1
1K

2
1

1
-12V C63
-12V

1
MN104
2

VR2 2K
1 3 R38
10K 1%
1 2
C29 MN101
2 1
TP3 C30 MN101
2 1 TP4
+12V
C31 MN104
1 2 EP_REF
1

IP_REF

1
R39
8

10K 1% R40 U11B

8
(1F-7) 1 2 3 10K 1%
EP_CONT +
1 (6B-14) (1F-7) 1 2 5
EP_REF IP_CONT
1

+
2 7 (6C-5)
IP_REF
2

1
-
C32 U11A 6
1

1
-
R41 LM358

1
4

10M BOX224 R43 C59 MN104 R42 C33 LM358


1

4
10K 1% 1 2 10M MN101 R44
2

2
2 10K 1%
-12V
2

2
APPROVED BY DESIGNED BY
HTP-24 PORTABLE X-RAY

CONTROL BOARD OP PANEL & HT CONT.


CHECKED BY DATE
Size FSCM NO. DWG NO Rev
A3 0
<Cage Code> HTP24-CTL-001
Sheet
5 of 7

12
Service Manual DIG-610

+12V

+12V
R45
5.6K U12
(5J-9),(7B-11) 1 2 1 12 R47 910 CON4
EP_FB +IN1 VCC
(5F-4) 1 2 1 2
EP_REF 1
R46 1 2 2 8 EP_CONT1
5.6K 2 1 1 2 -IN1 C1 9 2

1
C36 3 E1 R50 910
C34 R48 R49 F.B. 1 2
BOX102 BOX682 470K 680 EP_CONT2 3
4

2
2 1 5 11
C37 BOX102 CT C2 10
1 2 6 E2 FIL_CONT1
R51 RT FIL_CONT2 5
6.2K 1% 4 6
DT
7
15 GND 14 LAMP_CONT 7
16 -IN2 VR 13 8
+IN2 OC
(6B-7)
TL494 FIL_FB 9
10

1
1
C38 + R52 /100V_SEL 11
47K
U13A 1uF/25V TANTAL

2
14066 35312-1110

2
1 2

1
R53
13
2K

2
(1M-6)
/HV_ON

R54 VR3 R55 R56


750 2K 15K 10K 1%
1 2 1 3 1 2 1 2 (1J-5)
FIL_FB_ADC

2
+5V
2

R57

-12V 10K 1%
1

U13B
U14 +12V 14066
4
8

OP07 R59 11 10
R60 2 - 5.6K U15
4.7K 6 1 2 1 12

1
(6K-11) 1 2 3 + 1 2 +IN1 VCC
2

FIL_FB 1 2 2 8

12
R80 R61 2 1 1 2 -IN1 C1 9 R58
2

E1
7
1

C40 C41 5.6K 3 1K


BOX104 MN104 1K C39 R62 R63 F.B.

2
2 1 BOX473 470K 10K
1

2 1 5 11
C42 BOX102 CT C2 10
+12V 3 1 2 6 E2
1 2 2 U17 R64 RT

1
PN2222 DRV1 1PN2222 5K 1% 4
2

R82 10K DT R65 R66


R81 7 1K 1K
1 2 1 2 15 GND 14
3.3K 16 -IN2 VR 13 (1M-6)
+IN2 OC /LAMP_ON

2
R79 5.6K R70 470K
1

1 2 2 1
TL494
(5L-4) R78 30K C61 BOX104
IP_REF

1 2
IP_FB
R77 5.6K
1
1

C43 R67
BOX473 47K
U13C
2

14066
2

4 3
1

R68
5

2K
2

APPROVED BY DESIGNED BY
(1M-6)
/PREP HTP-24 PORTABLE X-RAY

CONTROL BOARD OP PANEL & HT CONT.


CHECKED BY DATE
Size FSCM NO. DWG NO Rev
A3 0
<Cage Code> HTP24-CTL-001
Sheet
6 of 7

13
Service Manual DIG-610

+12V +12V

TP5
+12V

2
C44
R69 R85 1 2
100k IP_FB
64K MN104

1
U16A

8
1

3
+
1
(5I-15) 1 2 2
IP_FB -
R71 LM393

4
100K
R86 C45
100k BOX473
(1B-10),(4J-11)

2
TP6 /FAULT

1 EP_FB

U16B +5V

8
5
+
7
1 2 6
EP_FB -
(5J-9),(6B-14)
R72 LM393 74HC573 27512 TD62304AP TD62304AP 14066

1
4
100K
2

C46 C47 C48 C49 C50 C51


R74 BOX103 MN104 MN104 MN104 MN104 MN104
+ C23 R73 + C35

2
10u/16V 56K 10u/16V 100K
1

TP7
+5V

+5V
1

TP8 H1 H2 H3 H4 H5 H6 H7 H8 H9 H10 H11 H12


1

C52 + C53
MN104 100uF/50V DGND HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE
2

1
2

CON5 TP9

1
2 +12V +12V
3
1

4
5 TP11
TP10
1

35312-0510
1

C54 +
100uF/50V C55
MN104 AGND AGND
2

TP12
1

C56 + C57
100uF/50V MN104
-12V
2
2

-12V APPROVED BY DESIGNED BY


HTP-24 PORTABLE X-RAY

CONTROL BOARD OP PANEL & HT CONT.


CHECKED BY DATE
Size FSCM NO. DWG NO Rev
A3
<Cage Code> HTP24-CTL-001 0
Sheet
7 of 7

14
Service Manual DIG-610

CAPACITOR BANK BOARD


53258-05
CON1 224/AC250V 224/AC250V TB1 TB3
SC-10-D050 330/400V 68K 1W
5 DC+ OUT 1 1
4 R7 C53 C54 C55
3 1/2W470K L2 C8
2 R8 C13 102/AC250V C9 C10 C15 C16 C11 C14
1 C12 - + + + +
102/AC250V + + + +
C17
R9 FET GATE DRIVE CIRCUIT
R12 NTC10D22 BD1
D25XB60 TB2 TB4
DC- OUT S1
68K 1W 1 1 +24V
330u/400V/30mm*40mm GND 1
BTA26-600 SIG1(+) 2
Q11 R13 F1 50T2A 3
10K 4
5
L3 6
SQE20 7
8
CON7
DRV2A G
1
S 2
S1
+24V 35312
C18 100/400V GND 1
2
D14 SIG1(-) 3
1N4937 T1 PQ32/30 4
1 5
C19 6
M104/100V + D6 1N4937 7
R14 1W100 C20 47/25V 8 CON10
2 DRV2A G
4 1
S 2
C21 S1 35312
C47 R15 222/1KV TP1 +5V 0.5A TP3 +24V
R16 2W24K GND 1
1/400V 2W 100K +5V 2
D8 S2L60 U2 LM7805CT GND 3
D7 7 1 3 4

1
TP2 1N4937 IN OUT +5V 5
GND 1 6

GND
3 C24 +12.2V 2 7
Q21 C2235 R29 C45 GND 3 8
+ + CON8
27K C22 HS3 C46 100/16V -12V 4 G
DRV2A
1

NP 2200u/25V 104 104 C23 5 1


+

2
100/16V CON2 +24V S 2
S1
C48 HS1 35312-5 +24V 35312
1
2

22/25V U7 UC3844 TP4 8 CON3


GND
Q12 2
7

R17 4.7 IXIF 16N90Q C25 C26 1 SIG2(-) 3


2 1
+
100/16V TT22/16V TP5 2 4
TC7662A1
6 3 5
+
1

1 R18 220 PG_S C28 1 8 4 6


C27 TT22/16V 2 7 5 FIL_CONT 7
3

8 3 D15 S2L60 3 6 6 8
+ CON9

1
4 5 7
4

BOX104 C29 C30 9 -12V 20mA LAMP_CONT DRV2A G


Z5 R21 8 1
U3
R19 BOX102 BOX472 5W 0.2 C51 9 FIL_FB S 2
R22 3.9K R20 1N4747 TP6 +24V TP4 +24V 0.2A 10
100V_SEL 35312
20K 470/25V 1 11
+

120 C31 C32 35312-11


PG D9 D10LC20U +15V TP7 +15V 6A +15V
BOX103 BOX102 10 1 +5V
1

D10 +12.2V
5 C35 Q14 M472/100V C37
+ + + PN2222 R27 10 Q15 +15V
+ C36 C33 C34 R23 R41 1N4746 IRF640 D12
1N4937 2200/16V(13pie) 2200/16V 1W680 47K R28 1
47/25V 2200/16V R24 +15V Q17 R30 Z6 2
U4 6 1/2W470 PN2907 1N4937 1W1K 3
D11 1N4937 78L05 12 4
+ C40 20K
TG 3 1 R25 Q18 22/25V CON4
IN OUT R26 100 5.1K PN2222 R32 10 1N4746 Q19 53258-4
GND

IRF640
1

U5 PC817 Q16 Z7 R33


+ + C39 C1815 Q20 C41
3

C38 10uF/25V PN2907 R34 D13 1W1K M472/100V


2

100uF/25V R31 20K R35


1%9.3K 5W/0.1
2

1N4937
4

TH
1

C42 C43 R36


3

104 5D11 HS4


101/3KV R37 4.7K R38 10K Q13
U6
TL431
1 t
50N60 APPROVED BY DESIGNED BY
PXP-60HF PWR & DRV
2

ISO_5V
DIG-610 PWR &DRV
2

ISO_GND R39
1%1.8K
C44 222/AC250V
3
2
1

2
1

HAND SW PWR COLIMETER CHECKED BY DATE


Size FSCM NO. DWG NO Rev
CON5 CON6 A3 PXP60HF-PWR A
<Cage Code>
35312-3 53258-2 Sheet
1 of 1

15
Service Manual DIG-610

8. Maintenance

8-1 Maintenance and periodic operator

Maintenance has checked periodically with manual by the user.


User checks items as following.
- Cleaning of product is clean on the outside surface when external form is dirty and you clean
by cleaning material etc with soft cloth.
But do not clean surface of product with water.
-. Check If the bolt of Hand switch housing under hand switch is loosed when you used the product.

- When main power of product is on, and display will show values (kV and mAs) as following .
And this product is normal, please use safely.

kV

mAs

- If you have a problem, show error code on display and see page 28 and check it.
After the error code on display check it, if this error code is not appear
Please just let me know.
Tel +82-31-907-7222
Fax +82-31-904-2231
Homepage: www.dong-mun.co.kr
E-mail: dongmun@kita.net

16
Service Manual DIG-610

8. Technical
Technical date

TECHNICAL GUIDE

LOCATION DISTANCE GRID KVP mA TIME SIZE

CHEST 100-110 + 65 30 0.2-0.3 10 12

CHEST(LAT) 100-110 65 30 0.3-0.4 10 12

CHEST 100-110 75 25 0.5-0.8 14 14

SKULL(AP) 100 75 25 0.8-1 8 10

SKULL(LAT) 100 75 25 0.3 8 10

ABDOMEN 100 75(80) 25(20) 1.5-2 14 17

FOREARM 100 65 30 0.2 - 0.3 10 12

WATER'S VIEW 80 75 25 0.8 - 1 8 10

ANKIE 80-100 65 30 0.2-0.3 8 10

KNEE 80-100 65 30 0.2 - 0.3 10 12

WRIST 80-100 65 30 0.14-0.18 8 10

RIB LOWER 100 80 20 1.7 - 2 14 17

C-SPINE(AP) 100 + 75 25 0.8-1 8 10

C-SPINE(LAT) 100 75 25 0.3 8 10

T-SPINE(AP) 100 + 75 25 2 14 14

T-SPINE(LAT) 100 + 80 20 2 - 2.5 7 17

L-SPINE(AP) 100 + 90 15 2.5 - 3 14 14

L-SPINE(LAT) 100 + 90 15 3-5 7 17

PEVRIS 100 + 80 20 2.5 - 3 14 17

HIP JOINT 100 90 15 2.5 - 3 10 12

17