Moisture determination (chronological sequence of the measurement cycle and status display User Guide)
1 2 3 4 5 6 7
Swtich on Standby mode Load pan Tare Weigh in Start Drying Read result
Enter drying temperature Heating module adjustment Heating module adjustment Balance adjustments
(standard and rapid drying)
95 Menu W.CAL: NO Menu W.CAL: NO
W.CAL: YES
or:
125 T.CAL: NO
7 8 9
T.CAL: YES
130 4 5 6 W.CAL: 50.000g
0 .
125
3:30
Alphanumeric entries
4:00 A...Z A
wait
Start W.CAL: 50.000g
Select method
M 1. Read off 1st temperature on tempera-
Select method: ture adjustment set and enter:
Method 01 T.CAL: 102
A...Z MA
wait
12
2. Read off 2nd temperature on tempe-
MARGARINE
rature adjustment set and enter:
T.CAL: 158
Function keys
Selection of the switch-off criterion Selection of the display mode Selection of the drying program
%
Automatic switch-off Timed Moisture ATRO moisture content Standard drying Step drying
1: Quick drying sample switch-off content
5: Slow drying sample
Manual switch-off with Test Dry content ATRO dry content Rapid drying
the Stop key measurement
with record
Free switch-off one free printout Current sample Gentle drying
switch-off criterion for each weight in grams (with temperature
method ramp)
PW.SET: NO
Password protection F.SOC: NO
YES PW-1 OFF ABORT? Free switch-off criterion
ON CLEAR? Activate/deactivate 7 SYMBOL: ON
(= switch-off criterion F) YES WT.MG: 1 (1-10 mg)
TEST-1 (var. Temp.) Activate/deactivate 11 F.PRT.IN: OFF 40
free print interval ON
W.SET: NO
Reset to factory High resolution RES: STD
YES ABORT?
settings (0.0001 g) HIGH
SET?
Stat. and Journal 12 ST/JOU: ON
Function OFF ABORT?
M.CLR: NO
Clear method memory CLEAR? Standby STD.BY: NO
YES ABORT?
clear completely (Temperature) YES S.TEMP: 40 (40 100 C)
CLEAR?
C.CLR: 1-4
Selective clearing 13
Settings of code lines
2-4 S.TIME: PER
VAR 30
Enter current 1 TIME: 13:04 3-4 (30 480 Min.)
time
4
OFF S.SWOF: NO
YES 18:00
(0:00 23:59)
Enter current 2 DAT: 20.02.08 Define test weight 14 TEST.W: 50.000 g (0.1 81.0 g)
date (Balance-Test)
Selecting the setting within a menu item:
Set vibration 3 VIBRA: MED:
0.002 g (0.001 0.010 g) Quits the menu without saving the last menu
HIGH Define 15 TEST.T: Reset
adapter
tolerance (Balance-Test)
item you changed
LOW
Saves the menu item and skips to the next
Automatic/manual 4 START: AUTO.
menu item
sample champer MAN.
Select dialog DEUTSCH FRANCAIS
language ITALIANO Calls up the next menu item
ESPANOL
Set audio signal 5 BEEP: LOW RUSSIAN Menu
HIGH ENGLISH EU
OFF ENGLISH US
Select settings
KEY.PROT: OFF
Protect settings 6
ON