Вы находитесь на странице: 1из 6

Special Article

Does nonmedical grade power cord compromise


the safety of medical equipment?

Address for correspondence: V Padmavathi, PS Vishnu Prasad1, Pankaj Kundra2


Er. V Padmavathi, Division of Biomedical Engineering, JIPMER, 1Department of Biomedical Engineering, MGMCRI,
Division of Biomedical 2
Department of Anaesthesiology and Critical Care, JIPMER, Puducherry, India
Engineering, JIPMER,
Puducherry, India.
Email:jipmerbme@gmail.com ABSTRACT

A tertiary care 1000 bedded hospital contains more than 10,000 pieces of equipment worth
approximately 41 million USD, while the power cords supplied along with the imported equipment
do not comply with countryspecific norms. Moreover, the local vendors procure power cords with
Access this article online typeD/M plug to complete installation and also onsite electrical safety test is not performed.
Website: www.ijaweb.org Hence, this project was undertaken to evaluate the electrical safety of all lifesaving equipment
purchased in the year 2013, referring to the guidelines of International Electrotechnical Commission
DOI: 10.4103/0019-5049.171556
62353, the Association for the Advancement of Medical Instrumentation(AAMI) and National Fire
Quick response code
Protection Association(NFPA)99 hospital standard for the analysis of protective earth resistance
and chassis leakage current. This study was done with a measuring device namely electrical
safety analyser 612 model from Fluke Biomedical.

Key words: Device under test, electrical safety test, equipment leakage current, protective earth
resistance

INTRODUCTION integrity and leakage current can be monitored for


necessary corrective actions. It is now clear that all the
The power source for all equipment is alternating lifesaving equipment must undergo electrical safety
current(AC) with frequency - 50Hz; unfortunately, test on recurrent intervals to ensure safe operation.
humans are most sensitive to this frequency.[1] Some of
the effects are tissue injury, burns, and fibrillation of The purpose of this project was to find out the root
the heart.[2] The main cause for these effects are leakage cause and influence of environmental factors for
current which occurs naturally in all the electrically equipment failures during the first year of purchase.
operated equipment due to stray capacitance between In addition, implementing electrical safety checks
two wires or wire and metal chassis. This can be
on recurrent intervals to guarantee safe usage of
eliminated by generating a low resistance path from
equipment on the patient are discussed.
equipment to ground, ideally at zero potential.
METHODS
Though the equipment is designed with the highest
degree of protection, safety is attained only when
We conducted electrical safety study on 200 lifesaving
there is a proper connection between the equipment
equipment purchased in the year 20122013 in
and hospital earth by a component called power cord.
If the protective earth resistance(PER) is not as per Jawaharlal Institute of Postgraduate Medical Education
the International Electrotechnical Commission(IEC)
This is an open access article distributed under the terms of the Creative
norms, the safety of equipment is violated. Therefore, Commons AttributionNonCommercialShareAlike 3.0 License, which allows
others to remix, tweak, and build upon the work noncommercially, as long as the
during installation of medical equipment, electrical author is credited and the new creations are licensed under the identical terms.
safety test is highly required to conform various safety
For reprints contact: reprints@medknow.com
parameters described by IEC 62353.[3] The initial tested
data serves as a reference guide for recurrent test How to cite this article: Padmavathi V, Vishnu Prasad PS, Kundra
throughout the working lifetime of each equipment. In P. Does non-medical grade power cord compromise the safety of
future, on the recurrent test, the deviations of ground medical equipment?. Indian J Anaesth 2015;59:769-73.

2015 Indian Journal of Anaesthesia | Published by Wolters Kluwer - Medknow 769


Padmavathi, etal.: Power cord and medical equipment

and Research, Pondicherry [Table1]. They belonged to Protective earth resistance/earth bond testing
ClassI category with detachable power cords of power Earth bond testing tests the integrity of the low
consumption<1.5 KVA(kilovolt ampere). ClassII resistance connection between the earth conductor
category, nondetachable power cord equipment and and any metal conductive parts, which may become
permanently installed equipment were excluded live in case of a fault on ClassI medical devices.
from the testing. Since all the tested equipment were Although many ClassI medical devices are supplied
newly purchased, the influence of film resistance with an earth reference point, most if not all medical
was excluded.[4] The electrical safety analyser(ESA) devices require multiple earth bond tests to validate
612 model was used which is capable of measuring low the connections of additional accessible metal parts of
resistance up to 2 with an accuracy of2% and leakage the enclosure. The test current is applied between the
current from 0 A to 1999 A with an accuracy of1%. earth pin of the mains supply plug and any accessible
The analyser incorporates test algorithm of Association metal part(including earth reference point) via a
for the Advancement of Medical Instrumentation(AAMI)/ dedicated earth bond test lead(clip/probe). The test
National Fire Protection Association (NFPA)-99 hospital method of ClassI equipment in reference to the 62353,
standard and IEC 62353.[5] All the measurements were IEC 2007(IEC 683/07) guideline is shown in Figure1.
manually obtained for better accuracy.
The MD ESA 612 applies a test current of 200mA
Test method to determine PER and the highest reading will
The standard test protocol involves two components; determine pass or fail criteria.[1] The allowable test
visual inspection and electrical testing.
limits are based on IEC 62353 and AAMI guideline:
100 mfor a detachable power cable up to 3m;
Visual inspection
300 mfor a Class I device including power cable
The equipment in working condition can only be
tested for electrical safety. This method identifies the (not exceeding 3m) and 500 mfor a medical
equipment type, model, power consumption, power system as per AAMI/NFPA99 hospital standard.
cord typeD/M and fuse ratings. In addition, it also An increase in resistance may indicate loosening/
checks site condition inclusive of input power and corroded connection or damaged or nonstandard
electrical socket where equipment is being used. cables, which need to be repaired before continuing
with the leakage test.
Electrical safety test
The electrical safety test involves PER and leakage Deviation
current measurements. The device under test(DUT) Class II equipment and doubleinsulated devices do
will be powered through the measuring device(MD) not have a protective earth. It is, therefore, neither
and testing probe performs the measurement. While necessary nor possible to measure earth continuity;
testing, all other external communication/data lines this poses no safety risk.[1]
such as video graphics array and Ethernet cables
must be removed from the DUT to eliminate parallel
earth.[1]

Table1: List of equipment tested


Equipment name Quantity
Anesthesia machine 7
Anesthesia ventilator 3
Cardiac output monitor 1
Defibrillator 23
Diathermy 18
Portable multichannel ECG machine 6
Humidifier 16
Multiparameter monitor 60
Portable ultrasound machine 1
Pulse oximeter 6
Syringe pumps 29
Figure 1: Test method for protective earth resistance for Class I
Ventilator 30 equipment

770 Indian Journal of Anaesthesia|Vol.59|Issue 12|Dec 2015


Padmavathi, etal.: Power cord and medical equipment

Leakage current Intensive Care Unit ventilators, humidifiers,


Research has shown that current, not voltage, is often electrosurgical unit syringe pumps, cardiac output
the source of injury or death.[2,6] It takes a small amount monitors and multiparameter monitors). Out of
of current to cause major consequences. For this 200 equipment, 122(61%) passed the testing of PER
reason, the International Electrotechnical Committee and leakage current as per the guideline IEC 62353 and
has set stringent rule on the design of medical AAMI/NFPA99 hospital standard test. Further, 62(31%)
equipment, so as to prevent any patient or operator equipment failed as per IEC 62353 and 49(24.5%)
being exposed to currents not part of the functional equipment failed as per AAMI/NFPA99 hospital
operation of the device. These currents are referred standard. Nevertheless, in both, 28 (14%) of equipment
to as leakage currents. The guideline for an inservice failed with safety standards. The test result comparison
test of medical electrical equipment IEC 62353 defines values for 200 equipment with IEC and AAMI/NFPA99
two different types of leakage current: equipment safe limit are shown in Figure2.
leakage current and applied part/patient leakage
current.[3] To measure these currents, the guideline Out of failed equipment, 20 numbers were
describes three methods: direct methodthe current again retested with medical grade power cord
flowing down the protective earth conductor of the (meeting IEC 60601 standards) and nonmedical
mains inlet to lead; differential methodthe result of grade power cords(supplied by the vendor). The
imbalance in current between the live conductor and comparative results of PER and leakage current are
neutral conductor and alternative methodcurrent presented in Table2. We applied MannWhitney
flowing through a person to earth from the applied Utest, assuming the hypothesis there is no difference
part or current flowing from a person to earth via the between medical grade power cord and nonmedical
applied part by applying unintended voltage from an grade power cord with a significance level of 0.99.
external source. The exact output significance of this test was 0.000,
hence the hypothesis was rejected. The results of PER
The direct method is identical to the method used in comparison are represented in Figure3. These values
the IEC 606011 standard, measuring the true leakage were also represented in boxplot shown in Figure4.
through a body model(MD) to earth(the measurement
of leakage current through body model circuit). This DISCUSSION
method can measure both AC and direct current
leakage with the highest accuracy compared to all other The conduct of electrical safety test has identified
methods. Therefore, in the analysis, we have selected a range of causes that circumvent safety aspects in
this method for measuring equipment leakage current. lifesaving equipment. The test results confirmed that
28 equipment failed the tests and unsafe to use on the
Equipment leakage current patient as well. These failed equipment on further
The total leakage current is derived from the applied investigation showed 80% failure due to deviant
enclosure and mains part. All applied(types B, BF power cord usage.
and CF) and nonearthed accessible conductive parts
are grounded together and connected to earth via 1
K MD(body model). The test is conducted with
the protective earth connection interrupted, to ensure
the measurements are done under worst conditions.
Measurements were carried out in both the polarities
and single fault condition. The allowable value for
leakage currents in single fault condition was 500 A,
and PER limit is 300 m referring to 62353 IEC:
2007(IEC 704/07).

RESULTS

Electrical safety test was performed on 200 critical


Figure 2: Protective earth resistance and leakage current values in
care equipment purchased in 2013 of 8 different single fault condition for all the tested 200 equipment (IEC 62353 and
categories(anaesthesia delivery systems, defibrillators, AAMl)

Indian Journal of Anaesthesia|Vol.59|Issue 12|Dec 2015 771


Padmavathi, etal.: Power cord and medical equipment

Table2: Electrical safety test results of selected 20 equipment tested with medical grade power cord and nonmedical
grade power cord
Medical equipment details Protective earth resistance(m) Leakage current in singe fault condition(A)
Name Make/model Nonmedical grade Medical grade Nonmedical grade Medical grade
Monitor GES5 Anaesthesia 332 97 110.00 115.00
Anaesthesia machine GEAestiva 7100 445 152 142.00 139.00
Monitor GES5 Anaesthesia 214 101 111.00 115.00
Anaesthesia machine GEAestiva 7100 356 83 153.00 156.00
Anaesthesia machine GEAestiva 7100 234 90 188.00 176
Diathermy ERBEVIO 300 S 543 84 14.00 23.1
Monitor GES5 Anaesthesia 230 76 111.00 116
Diathermy ERBEVIO 300 D 220 85 126.00 131
Pulse oximeter L and TComet 790 77 304.00 317
Ventilator HamiltonGalileo Gold 905 81 79.30 81.5
Ventilator HamiltonGalileo Gold 1230 83 80.50 78.5
Ventilator HamiltonGalileo Gold 980 82 84.80 83.2
Ventilator HamiltonGalileo Gold 1040 83 83.40 79.8
Ventilator HamiltonGalileo Gold 2000 76 84.80 82.9
Ventilator HamiltonGalileo Gold 1850 79 83.00 82.2
Monitor L and TStar Plus 393 182 201.00 202
Ventilator Care fusionVela 1800 90 175.00 169
ECG machine PhilipsPage writer trim I 1700 392 49.70 45
Defibrillator Nihon KodhenTEC 5521K 610 190 199.00 203
Defibrillator Nihon KodhenTEC 5521K 550 175 380.00 389

Figure 3: Describes the protective earth resistance values of selected


20 equipment, which was tested with medical and non-medical grade
power cords

Guideline IEC 606011 defines that a medical grade


power cord must have plug typeD and a mains supply
cable of 3m length with a minimum crosssectional Figure 4: Illustrates the distribution of equipment between medical
grade power cord and non-medical grade power cord. The boxplot
area of 0.75mm and resistance of the cord <100 m.[1] illustrates the difference between protective earth resistance and
If the wire thickness or purity of copper or materials leakage current values obtained between medical grade power cord
used in the cable do not meet the IEC medical and non-medical grade power cords for selected 20 equipment only.
Data shown as median (horizontal line), interquartile range (box), 25th
equipment guidelines, the ground resistance value to 75th percentile (vertical line). *P=0.000 Non medical grade versus
may deviate and results in higher PER. Hence, to find medical grade power cord
out significant differences between medical grade
and non-medical grade power cords, a comparative is a significant difference in the usage of power cord
study was performed. The hypothesis for this study types B and A.
assumed no significant differences between the
power cords. However, the acquired MannWhitney The study also confirmed that use of medical grade
exact output significance was 0.000. The median of power cord has maximized the flow of leakage current
PER value of medical grade power cord was 84.5 m to ground. Therefore, if PER increases, the fault/leakage
followed by a median of non-medical grade power current will find unintended paths that could include
cord 580 m. From this analysis, it is clear that there patient or operator. The other reasons of 18% failure

772 Indian Journal of Anaesthesia|Vol.59|Issue 12|Dec 2015


Padmavathi, etal.: Power cord and medical equipment

were due to the instrumentation errors, harmonic CONCLUSION


distortion issues and possible intermittent dilemmas.
On consideration of patient safety, this review provides
It is wellknown that medical equipment is one of evidence that safety is attained only when there is a
the largest capital investments of every healthcare proper connection between equipment and hospital
organisation, often involved in patient incidents that earth by a major component called power cord. Hence,
resulted in serious injuries or deaths.[7] At present, we recommend performing onsite electrical safety
our system guarantees preventive maintenance for test using a suitable electrical safety analyser during
equipment, but safety is not ensured. International installation and annual maintenance to validate and
Standard Guideline IEC 623532007 medical electrical confirm that medical equipment along with power
equipmentrecurrent test and test after repair of medical cord are safe and reliable to use in patient care.
electrical equipment, document applies to testing of
medical equipment and medical electrical systems, which Financial support and sponsorship
comply with IEC 606011 before putting into service, Nil.
during maintenance, inspection, servicing and after repair
or on occasion of recurrent test to assess the safety of such Conflicts of interest
medical equipment. Therefore, this study proposes to There are no conflicts of interest.
implement an electrical safety test protocol in healthcare
REFERENCES
organisations to get benefits in 3 ways from electrical
hazards. First, protect patient, user and equipment from 1. ClarkJT, LaneM, RafuseL. Medical equipment quality
possible abuse to which the device may be exposed. assurance: Inspection program development and procedures.
Electrical Safety. 2ndRevision. Burlington: University of
Second, it will save equipment from harmful influence
Vermont, Fluke Biomedical; 2009. p.2436.
of harmonics caused by neighbouring equipment. Third, 2. JacobsonB, MurrayA. Medical Devices Use and Safety:
equipments leakage current can be conformed to the safe Electricity and Safety. 1sted., Vol.3. Tolstoy Marg, NewDelhi:
Reed Elsevier India Pvt. Ltd.; 2011. p.3354.
limit during any catastrophic failure. 3. Bureau of Indian Standards (Medical electrical equipment
- recurrent test and test after repair of medical electrical
Moreover, the conduct of electrical safety test will also equipment IS/IEC 62353:2007) Act, 1986. New Delhi:
eliminate the subsequent incidents reported in our Published by BIS; 2012.
4. BackesJ, editor. Electrical Safety: Current Thinking on
operation theatre which was one of the reasons that Testing Protective Earthing. UK: Rigel MedicalSeaward
encouraged us to conduct this study. We experienced Group; 2015. Available from: http://www.ebme.co.uk/
articles/electricalsafety/339currentthinkingon
a fire incident, where power cords plug top burnt off
testingprotectiveearthing. [Last accessed on 2014 Oct 01].
completely. On root cause analysis, we found that the 5. Electrical Safety Analyzer 612 User Manual [Internet]. 2nd Rev.
plug type was nonIngress Protection(IP) rated; the USA: Fluke Biomedical Corporation; 2009. Available from:
http://www.flukebiomedical.com/biomedical/usen/Support/
saline solution had entered into the socket and created Manuals/default.htm. [Last cited on 2013 Apr 17].
a short circuit path. This was because the supplier had 6. Seaward Group. A practical guide to IEC 62353: IEC 62353
cutoff the plug typeF from the cord and replaced leakage measurements [Internet]. 1st ed. USA: Seaward
group; 2009. Available from: http://www.rigelmedical.com/
with plug typeD(Indian standard). Both root downloads/Guide-to-IEC-62353-2009-UK.pdf. [Last cited on
cause analysis and our study concur with the result. 2012 Aug 14].
Hence, we recommend using medical gradeIP 44 or 7. Wang B. Core functions of medical equipment maintenance and
management. In: Enderle John D, editor. Medical equipment
above rated power cord with typeD plugs; if not, it maintenance. UK: Morgan and Claypool Publishers; 2012.
compromises the safety of medical equipment. p.17-24.

Indian Journal of Anaesthesia|Vol.59|Issue 12|Dec 2015 773


Copyright of Indian Journal of Anaesthesia is the property of Medknow Publications & Media
Pvt. Ltd. and its content may not be copied or emailed to multiple sites or posted to a listserv
without the copyright holder's express written permission. However, users may print,
download, or email articles for individual use.

Вам также может понравиться