Академический Документы
Профессиональный Документы
Культура Документы
Conductivity
Radioactivity
Pesticides Metals
analysis
POTENTIOMETRY FOR
IONS & GASES IN
AQUEOUS SOLUTION
Potentiometry analysis uses
Electrodes to measure specific ion in
solutions.
Type of electrodes use in water
analysis:
pH measurement pH electrode
Ions & hardness measurement Ions
Selective Electrode (ISE)
PH MEASUREMENT
pH Measurement using a Glass Membrane Electrode
To determine the activity of hydrogen ions in a solution
PH ELECTRODE
Combination electrode incorporates both glass and reference electrode in
one body
Ag(s)|AgCl(s)|Cl-(aq)||H+(aq,outside) H+(aq,inside),Cl-(aq)|AgCl(s)|Ag(s)
Glass membrane
Selectively binds H+
Junction potential
If ionic strengths differ between analyte and standard buffer, junction potential will differ resulting in
an error of 0.01
Sodium Error
At very low [H+], electrode responds to Na+ and the apparent pH is lower than the true pH
Acid Error
At high [H+], the measured pH is higher than actual pH, glass is saturated
ERRORS IN PH MEASUREMENTS
Equilibration Time
Takes ~30s to minutes for electrode to equilibrate with solution
Hydration of glass
A dry electrode will not respond to H+ correctly
Temperature
Calibration needs to be done at same temperature of
measurement
Cleaning
Contaminates on probe will cause reading to drift until properly
cleaned or equilibrated with analyte solution Larger errors occur at high
and low pH readings
IONS SELECTIVE ELECTRODE (ISE)
ISE is used for specific ion measurements such as fluoride, chlorine, chloride,
hardness (Ca2+ and Mg2+), dissolved oxygen, ammonia, etc.
AAS
Analysis
SAMPLE FILTRATION
To separate suspended metals from dissolved
metals.
To determine suspended metals on the filter, the
volume of water filtered must be recorded and the
filter must be stored in a clean container for further
analysis.
Polycarbonate or cellulose
acetate filters can be used
If the sample is very turbid, it is better to centrifuge
before filtering to avoid loading of fouling.
DIGESTION OF METAL SAMPLES
Metals associated with particulates need to brought into solution before
analysing with AAS/ICP.
Nitric acid is generally the most widely used with the addition of hydrochloric,
hydrofluoric or perchloric acid to extract more difficult metals.
Can be done in Erlenmeyer flask or beaker on a hotplate, where 50 to 100 ml
of the sample is mixed with acid.
The digestion is considered complete once the sample reach clear and light in
color.
Use of microwave digestion gives more reproducible results.
Rotor
The light source (hollow cathode lamp) emits a spectrum specific to the element of which it
is made, which is focused through the sample cell into the monochromator. The method of
heavy metal definition via Atomic-absorption spectrometer is based on the properties of
specific metal atoms to absorb specific wavelength of light. The electronics will measure
light attenuation and convert those readings to the actual metal concentration of the sample.
Putting lamps for various metals
The lamp holder with positions for 8 source lamps allows having 8 lamps
simultaneously that are prepared for work and can pass from one element to
another.
Process of measurements
Choose a desired method of atomization by means of the computer, with use of
software support WinLab32
1. When using the electrothermal atomization method the aliqoute part of a sample is
put in a graphite furnace of the spectometer using sampling device (autosampler)
According to the programme, heating of a sample is performed in the inert gas
(argon) stream, including drying, incineration, atomization of a sample and clean-out
of the furnaces.
2. The flame atomization method is also used. The injectoration of the analyzed
portion of the sample in the flame of a torch is performed by means of capillary. The
mix of air acetylene is used for the formation of a flame.
REACTANTS AND EQUIPMENT ARE NECESSARY FOR
MEASUREMENTS
A nitric acid of special cleanliness
Deionized water
Adjustable pipettes
Adjustable pipette
Measuring Flasks
Gaseous Argon of high cleanliness
Gaseous Acetylene of high cleanliness
Atomic Spectroscopy Single-Element Pure Standards.
Measuring Flasks
Standards
INDUCTIVELY COUPLED PLASMA METHODS
Well suited for water analysis because all elements desired can be analysed in one run.
Carbon Dioxide
and
Water
Under:
Specified Time 5 Days
Specified Temperature 200 C
Specified Conditions In the Dark
In the Presence of Bacteria
Measured Volume of Wastewater is
Added to BOD Bottle.
Contains:
Organics
Ammonia
Bacteria
NH3
NH3
CxHYOz
Dilution Water Added
Nutrients
O2
O2 Oxygen
O2 Measure
O2 D.O. Concentration
O2
CxHYOz
O2
Incubate 5 Days
Respiration
O2 Nitrification
O2
Measure
D.O. Concentration
O2
Measure Oxygen Loss
(Demand)
= Oxygen Demand
O2
O2
BOD
Distilled Water
Phosphate Buffer
Magnesium Sulfate
Calcium Chloride
Ferric Chloride
Provide Essential Nutrients
Buffer pH
DILUTION WATER
High Quality
Distilled Water No Toxics
No Organics
plus
BUFFER pH 7.2
plus
(Phosphorus
and Ammonia)
NUTRIENTS Magnesium
Calcium
Iron
DILUTION WATER PREPARATION
NEEDED VOLUME for Each Days Use
ADD NUTRIENTS (1 mL per Liter)
SATURATE WITH OXYGEN
Vacuum
Shake
(small volume)
or
Draw Vacuum
DILUTION WATER PREPARATION
NEEDED VOLUME
ADD NUTRIENTS
SATURATE WITH OXYGEN
STORE
In Incubator
ADD BUFFER
Day of Use
B.O.D. Calculation
D.O. In - D.O. Out
= DEPLETION
= Oxygen Demand of Diluted Sample
C1 X V1 = C2 X V2
Sample Sample BOD Diluted
BOD
X Volume = Diluted X Volume
Depletion
BOD Sample =
Sample Volume X 300 mL
B.O.D. mg/L =
= 40 mg/L
B.O.D. PRACTICE PROBLEM
Calculate the B.O.D. value to be reported for each of the samples below.
Be sure to consider the minimum depletion and residual requirements.
= 74 mg/L
B.O.D. PRACTICE PROBLEM
Sample 3 Dilution A Dilution B
mL Sample 15 30
Initial D.O., mg/L8.1 8.1
5-Day D.O., mg/L5.6 3.3
Depletion 2.5 4.8
50 + 48 = 49 mg/L
2
What is Chemical Oxygen Demand (COD)?
Definition
The quantity of oxygen needed to chemically oxidize
the organic compound in sample, converted to carbon
dioxide and water.
2 3 a
Cn Ha Ob Cr2 O 8H nCO2 2Cr 4 H2 O
7
2
Where: 2n a b
3 6 3
COD
The mixture is titrated with ferrous ammonium sulphate (FAS) to
measure the excess dichromate remaining in sample.
Condensor
O2
TOC - UV/PERSULFATE INSTRUMENT
Syringe
Sample
CO2 Detector Recorder
Inlet
Condensor
UV Reactor
O2
Persulfate
Solution
TOC - THE CO2 DETECTOR
Sensing Demodulator
Chopper Cell Amplifier
IR
Reference
Source
Sample
In Out
BOD mg/l
BOD mg/l
SS mg/l
SS mg/l
COD mg/l
COD mg/l
Oil / grease - not detectable
Oil / grease 10 mg/l
DISTRIBUTION OF
WATER SUPPLY
STREAM
UPSTREAM
DOWNSTREAM
SEWAGE TREATMENT PLANT
WATER INTAKE POINT