DIO 562
CONTENTS
1
2
3
4
5
6
5 6
MAIN MENU (described below in 1.4) allows setting of special modes. Access to the Main
Menu is possible only during switching the instrument ON by simultaneous depressing
and ON.
2.2. Switching ON
Press the ON key to switch the instrument ON. will appear on the
display. Then press any arrow (cursor) key to continue. The instrument is ready and PARAM menu is
displayed.
While the instrument is switched OFF, press simultaneously and ON keys. Main menu will
appear on the display. Use and keys to move the cursor and or key to select the
appropriate item (command).
Since these functions are not needed for routine operation of the flaw detector, their description
is given in the section Special functions. The use of particular items should be avoided if operator
is not sure or does not have appropriate skills. Otherwise, data on the memory chip may be lost
and/or principal changes of internal parameters of the instrument might occur!
2.7. Maintenance
DIO562 does not require any special maintenance, except timely charging the accumulator.
Setup number
1 2 3 4 5 6 7 8 9
within group
This allows the direct access to the nine setups within a group. The group means choice from ten
setups (1-9, 11-19, 21-29 91-99). The desired group of setups shall be set in the TOOLS SETUP
sub-menu. For example, if the memory number #35 had been set, then setup number 31 will be
recalled by pressing F1 key (not setup number 1) see section Sub-menu SETUP.
selection of the gate triggering logic: (triggers above positive, or below negative;
switch the sound signalization ON and OFF
MEASURE selection of the measuring mode within the GATE 1 (also THC. FF, or
THC. PP in the GATE 2, which is switched ON automatically);
OFF
THC. FF = measuring echo-echo (flank to flank)
THC. PP = measuring echo-echo (peak to peak)
THC. 0F = measuring zero-echo (time base zero to echo flank)
THC. PP = measuring zero-echo (time base zero to echo peak)
mm symbols displayed under the time base, which allow zoom and time base settings,
the same as in the PARAM menu
ESC (F1) leave menu; press F1 to move the cursor from any item to the ESC position, then
press or to leave the menu
REF-ECHO (BW, DSR or SDH); setting the reference reflector type for AVG/DGS or DAC
curve: BW back wall; DSR disc shaped reflector (flat bottomed hole); SDH
side drilled hole
REF-SIZE reference reflector size for AVG/DGS or DAC curve; diameter in mm for DSR or
SDH. For BW the INFIN. (infinite surface) value will be set automatically
ATT-TEST (0.0 dB/m 99.9 dB/m) attenuation in the inspected material. May be
adjusted even after creating the curve in the PARAM menu
AMPL-CORR (-20.0 dB - +20.0 dB) gain correction, in dB, for surface roughness
ESC (F1) leave menu; press F1 to move the cursor from any item to the ESC position, then
press or to leave the menu
REF-ECHO (BW, DSR or SDH); setting the reference reflector type for AVG/DGS or DAC
curve: BW back wall; DSR disc shaped reflector (flat bottomed hole); SDH
side drilled hole
REF-SIZE reference reflector size for AVG/DGS or DAC curve; diameter in mm for DSR or
SDH. For BW the INFIN. (infinite surface) value will be set automatically
R1: R5: (-99.9 0 dB) amplitude value, in dB, of the particular points of the DAC
curve