Thermo Scientific Niton XL2 GOLDD Series Mining Analyzers
Elemental Limits of Detection in SiO2 and SRM Matrices Using Mining Analysis
The Niton XL2 GOLDD Series x-ray fluorescence (XRF) analyzer
is the performance choice for your toughest testing applications. In addition to the offices listed below, The chart below details the sensitivity, or LODs1, of the Niton XL2 Thermo Scientific GOLDD Series using mining analysis for an SiO2 matrix, a typical soil Niton Anaylzers matrix (SiO2 with Ca/Fe), and SRM matrix. maintains a network of sales and service organizations Limits of Detection in ppm (mg/kg) throughout the world. Time 60s per filter Limits of detection (LODs) are Matrix SiO2 SiO2+Fe+Ca SRM dependent on the following factors: Americas Billerica, MA USA Ba 75 100 100 Testing time Interferences/matrix US Toll Free: 800 875-1578 Sb 30 30 30 Level of statistical confidence +1 978 670 7460 niton@thermofisher.com Sn 30 35 35 LODs are calculated as three standard Cd 25 25 25 Europe, Middle East, deviations (99.7% confidence interval) Africa & South Asia Pd 15 15 15 for each element for a 120-second Munich, Germany Ag A/S A/S A/S total analysis time. +49 89 3681 380 Mo A/S A/S A/S Please Note: niton.eur@ thermofisher.com Nb 35 35 35 Ongoing research and advancements in our Niton XL2 Series analyzers with Zr 15 18 20 Asia Pacific geometrically optimized large area Central, Hong Kong Sr 6 8 15 drift detector (GOLDD) technology will +852 2869 6669 Rb 5 8 10 lead to continual improvement in many niton.asia@ of the values detailed in this chart. Bi 10 15 15 Contact a Thermo Fisher Scientific thermofisher.com As 6 15 15 office or your local representative for the latest performance specifications. www.thermoscientific.com/niton Se 6 10 10 Pb 10 20 20 Analysis time is based on your requirements, and, in most cases, W 75 100 100 Elements
shorter times will give you the
Zn 15 25 25 detection limits you require. For Cu 25 35 40 example, if analysis time was reduced from 60 seconds/filter to Ni 45 50 60 15 seconds/filter, then the detection Co 35 200 200 limits obtained would be twice the values shown in the chart. Similarly, Fe 60 N/A N/A increasing the analysis time will Mn 75 125 125 reduce the detection limits by the Cr 150 150 150 square root of the increased time.
V 450 450 450 A/S = Application-specific
Ti 700 700 700 N/A = Not applicable 2010 Thermo Fisher Scientific Inc. Ca 65 N/A N/A 1. Definition and Procedure for the Determination All rights reserved. K 125 N/A N/A of the Method of Detection Limit, 40 CFR, Part 136, Appendix B. Revision 1.11. U.S. All trademarks are the property of Thermo Fisher Cl 75 75 100 Environmental Protection Agency. U.S. Government Printing Office: Washington, Scientific Inc. and its S 120 150 200 DC, 1995. subsidiaries. Specifications, terms and P 350 500 500 pricing are subject to change. Not all products are available Si N/A N/A N/A in all countries. Please consult your local sales representative Al 750 1500 3000 for details. Mg 7500 8000 8000 Element list shown is not exhaustive. For limits of detection for elements not shown, please AN44822 07/2010 contact a Thermo Fisher Scientific office or your local representative.