Вы находитесь на странице: 1из 6

, .

An Introduction to Cyclic Voltammetry


Gary A. Mabbott
Department of Chemistry and Geology. Clemson University. Clemson. SC 29631

Cyclic voltammetry has become a popular tool in the last Epc


fifteen years for stu~ying electrochemical reactions. Organic
chemists have applied the technique to the study of biosyn­ (B)
thetic reaction pathways (1) and to studies of electrochemi­
cally generated free radicals (2). An increasing number of
inorganic chemists have been using cyclic voltammetry to
evaluate the effects of ligands on the. oxidation/reduction 120)lA
potential of the central metal ion in complexes and multinu­
clear clusters (3). This type of information plays an integral
part in many of the approaches directed toward solar energy EI
conversion (4) and in model studies of enzymatic catalysis (5).
Knowledge of the electrochemistry of a metal complex can be ~
useful in the selection of the proper oxidizing agent to put the t-0.5 - 0.5
metal complex in an intermediate oxidation state (6). Elec­ APPLIED VOLTS
trochemical methodology has also been exploited as a novel
meanS of introducing functional groups and removing blocking
agents (7).
There are several good texts that deal with the theory and
practice of modern voltammetry in depth (8-10). Also. many
instrumental analysis textbooks give a very brief overview of
the method. However, the needs of a researcher interested in Epa
(AI
applying the technique for the first time are somewhere be­
tween these two extremes. Instructors of instrumental analysis +
who want to teach modern electrochemistry are faced with a w
(!)
similar dilemma due to the lack of suitable background ma­
terial for the students to read. This article is intended to help '"o~
meet those needs. >
The CV Experiment
The voltage applied to the "workIng" electrode is scanned
linearly from an initial value. E i. to a predetermined limit. E ,\1,
(known as the switching potential) where the direction of the
scan is reversed (see Fig. la). The operator Can halt the scan FlgUte 1. a) Applied potential program. £),l and £),2 are switching potentials .
anywhere or let the instrument cycle between E;>,J and some b) Typical cyclic voltammogram of 1 mM K:'Fe(CN)e at a platinum electrode In
other preselected value, E >-2. The current response is plotted aqueous 0.1 M KCI solution. The scan rate was 100 mV!sec and the relerence
electrode was Ag!AgCl in 0.1 M KCI solullon.
as a function of the applied potential. Often there is very little
difference between the first cycle and successive scans.
However, the changes that do appear on repetitive eycles are conditions.) The term Ci is known as the transfer coefficient.
important keys to unlocking information about reaction It arises because only a fraction of the energy that is put into
mechanisms (as will be shown later in this paper). the system (in the form of the applied potential) lowers the
Figure 1b shows a current-voltage curve (or voltammogram) activation energy barrier. Its value varies from zero to unity
for Fe(CN)s3. As the potential is scanned in the negative di­ (often ~O.5) depending on the shape of the free energy sur­
rection the current rises to a peak and then decays in a regular faces for the reactants and products.!
manner. The current depends on two steps in the overall The exponential dependence of keon the applied potential
process, the movement of electroactive material to the surface accounts for the steep rise in the current. However, the elec­
and the electron transfer reaction. The electron transfer rate trolysis of the reactant depletes its concentration near the
constant for a reduction process is a function of potential and surface. Since the experiment is performed at a stationary
can be described ~heoretically. electrode in an unstirred solution, diffusion is the principr!l
means of moving the reactant to the surface. This relatively
-emF
kl = hO exp (Rr(E - EO')
) (1) slow mode of mass transport cannot maintain a steady-state
concentration profile in the region close to the electrode.
kO is the standard hetero~eneous electron-transfer rate con­ Therefore. the depletion zone grows. In a sense, the average
stant. (Its value is a property of the reaction between the distance that the reactant molecules must travel to reach the
particular compound and the electrode surface used.) The surface increases. Consequently, the rate of mass transport .
number of electrons transferred per molecule is n; F is the decreases. The dependence on mass transport, and the fact
Faraday; R is the universal gas constant; T is the Kelvin that a finite rate for the reverse electron transfer process is
temperature; and EO) is the formal reduction potential. (The possible, prevent the current from increasing exponentially
prime signifies that the effect on the free energy of the reac­
tants and products embodied in activity coefficients has been 1 this is a simplification. For a well-written and much more detailed
combined with the thermodynamic reduction potential to explanation of the significance of the transfer coefficient and symmetry
form a term that is directly measurable but subject to solution factors see pp . 917-929 of ref.(27).

Volume 60 Number 9 September 1983 697


...

with potential. Eventually the mass transport step becomes not proportional to u m except. when the peaks are so widely
rate determining and the current reaches a maximum. Since separated that the system is more appropriately described as
the concentration gradient continues to decrease, the rate of totally irreversible . <

mass transport continues to decrease causing the current to Although the average of the peak potentials can be a good
decay. Beyond the peak the current is actually dependent on estimate of the EO ' for a reversible redox reaction, one must
time and independent of the applied potential. [n this diffu­ be careful not to attribute high accuracy to the determination.
sion-limited region the current is proportional to t - II2.2 In Nicholson and Shain's classic paper on the theory of vol­
An advantage of the cyclic voltammetry experiment is the tammetry at stationary electrodes (J2), they point. out that
fact that a significant concentration of product (in this case, the position of the return peak, even for a reversible system
the reduced form) has been generated near the electrode on can shift as much as 5 m V depending em the choice of the
the forward scan. When the scan direction is reversed, the switching potential, E~ . (The farther Bh is from Ep, the more
r.educed form is oxidized back to the original starting material nearly symmetric about the reduction potential the peaks will
and the current for the reverse process is recorded (see Fig. be.) Also their work indicates that the midpoint. between the
Ib). The electron transfer rate constant for the reverse (or peaks is really an estimate of E 112." (E J/2 is a term from po­
oxidation) process is similarly controlled by the applied po­ larography that was given to the potential where the current
tentiaL is half the value of that on the current plateau.) Fortunately,
,
11, =
kO
exp
((1-RT(\')n.t)-) l E-' - EO ' ) (4)
the dilfusion coefficients have a small effect. Even if DolDH
= 2, the error introduced by assuming E 112 = EO' would be
only ~9m V for a single electron exchange process.
Some workers have found that the broad point at the top
Deierminatlon of the Formal Reduction Potential of the current peak makes it difficult to determine the true
It is common practice to report the average of the forward position of the peak precisely. They prefer to read the po­
and return peak potentials as the formal reduction potential tential where the current is half the value of the peak current.
for the redox couple. This is an approximation that is most This half-peak potential (for a reversible redox couple at
accurate when the electron transfer procesS is reversible and 25°C) is re.lated to the polarographic E liZ value by the fol­
t he diffusion coefficients for the oxidized and reduced forms lowing equation:
are the same. 1f the reaction is reversible, then the separation , 28.0
in the peak potentials, M p , will be close t.o 58/11. m V (at 25°C). Epl2 = E J!~ ±- mV (7)
II
(This relationship can be used to evaluate 11..)
By reversible, electrochemists mean that the reaction is fast (The sign is positive for a reduction process.)
enough to maintain the concentrations of the oxidized and Estimating formal reduction potentials from quasi·re­
reduced forms in equilibrium with each other at the electrode versible voltammograms is less reliable the farther the peak
surface. The proper equilibrium ratio at a .given potential is separation deviates from t he reversible case. Theoretical work
determined by the Nernst Equation: shows that as the scan rate is increased, slow electron transfer
kinetics can make the peak potentials shift in such a manner
E = EO' - RT In (lHI) (i) that they are no longer symmetric about t.he E l/2 for the redox
n.F rO],= () couple. The problem is worse the farther the electron transfer
where 0 is the oxidized form and R is the reduced form. coefficient, l¥, is from 0.5. A rough indication of the magnitude
How fast is fast enough? Many systems look reversible when of the error that may be introduced by averaging the pair of
the voltage is scanned slowly but at higher scan rates Mp cathod ic and anodic peak potentials for a quasi-reversible
appears greater than 58/71. m V. Reversibility is, then, a matter voltammogram can be derived from graphical data prepared
of degree and depends on the stress that is applied to the by Mat.suda and Ayabe U J). For an a·value of 0.3 and peak
system. Matsuda and Ayabe indicated that for scanning vol­ separations of 108, 312, and 592 mY, the midpoint between
tammetry any deviations from reversible behavior will be the peaks would be on the negative side ofthe actual E 112 value
imperceptible if t.he value of hO (in em/sec) is greater than the by 2, 44, and 104 m V, respectively. (For an (i-value of 0.7 the
numerical value ofO.3v 1./2 (where v is the scan rate in V/sec)3 error would have the same magnitude with the opposite
(J 1). Voltammograms recorded at scan rates up to 10 V/sec sign.)
are common. Some instruments are capable of scanning up
to 1000 V/sec. Tberefore. electron transfer reactions with rate
constants greater than 10 em/sec will be reversible even in the 2 The mathematics that describe 'a scanning experiment are quite
very fastest experiments. complicated. However, an application of Fick's First law of Diffusion
Redox couple$ whose peaks shift farther apart with in­ tells us that the current at any time is proportional 10 the concentration
gradient for the reactant :
creasing scan rate are categorized as quasi-reversible. (Some
authors merely say irreversible.) There are some cases in which (2)
the peaks are so widely separated (k ° ~ 2 X IO-5 v 1/ 2 cm/sec)
A is the electrode area; Do is the diffUSion coefficient of ,the OXidized
that no parts of the two peaks overlap on the potential axis ,a t
species; t is time: and x is the distance from the eleotrode surface. An
all. These are generally known as " to tally irreversible" sys­ experiment in which the voltage 15 stepped instead of scanned out to
tems. A subset of this class are those reactions that yield a pOtential where the current 15 diffusion controlled is much eaSier to
products that cannot be recycled electrochemically to give model mathematically. (This 15 chronoamperometry .) In that case
back the original reactants (for example, those that involve
extensive bond breaking and/or loss of substituents to soh.l ­ (aColaX).=O., = C~/V 7rDot (3)
tion). These are "chemically irreversible" reactions, and many where C~ is the concentration of the oxidized material in the bulk so­
yield no return peak at all. lution (In moleslcm 3 ). II seems reasonable, then that scanning to the
Another characteristic of reversible systems is the depen­ same voltage would also cause the current to fall off as t- 1I2 .
dence of the peak height on tbe square root of the scan rate. 4 3 Rates of heterogeneous reactions are referenced to unit surface
At 25°C the peak current is area and have units of mol sec- i cm- 2. Since the rate 15 expressed
as the product of kC, the rate constant, k, must have units of cm sec- 1
ip = (2.69 X 10r,)fl :\12AD6/2uJ /'2C ~ (6 ) when concentration is given in mol cm- 3 .
• Equation (6) assumes a planar electrode. For a spherical electrode
The current will be ill amperes when A is in cm 2, Do is in +
(such as a hanging mercury drop): ip = ip(pJane) (0.752) nFADoC~/ro
cm 2/sec, v is in volts/sec, and CI~ is the bulk concentration in where '0 is the radius 01 the electrode.
moles/cm:). The peak (~urrent for a quasi-reversible system is 5E1I2 = E'" - RTlnFln [Do/~J1/2

698 Journal of Chemical Edtlcation


The point to remember is that one should use conditions tionship is plotted in Figure 3. Experimentally we can use the
that give the most reversible behavior possible. For quasi­ graph to obtain an estimate of k for the chemical step. (For
re'-.rersible systems the scan rate should be slowed down to experimental values of ipa/ipc, find the corresponding theo­
minimize !lE p . There is a lower limit for the scan rate, how­ retical value of log (kT). Solve for kT and divide by the ex­
ever. This limit is set by our ability to maintain a convec· perimental value of T.)
tion·free solution. Convection often sets in after times as short Other electrochemical mechanisms can be identified. A list
as 30 seconds, unless special precautions against stray vibra­ of some of the more common ones is given in the table. Figure
tions and other sources of convection are taken.
The type and condition of the working electrode surface can 1.0
make a difference in the electron transfer kinetics. Often solid ;;"
electrodes need to be polished and/or cleaned by exposing ~
0 0.8
them to oxidizing agents. In some cases workers advocate ;;

"activating" the electrode surface (particularly carbon elec­


trodes) by application of a sequence of extreme positive and
" .
....
0 .6

negative potentials. (For more discussion about surface effects ic


0
0.4 ,
and pretreatment procedures see references (13-15). 0

_0. 0.2
"
Peak Currents and Coupled Chemical Reactions
o.
The height of the current peak can be used to determine the -2.0 -1.0 0.0
concentration of the reactant in the bulk solution. However,
IO~ ( k1:)
other electrochemical methods (such as differential pulse
voltammetry) are better suited for this task. The real forte of Figure 3. Ratio of anodic to cathodic (or reverse to forward) peak current as a
this technique is the analysis of homogeneous chemical re­ function of k for a reversible electron transfer followed by a chemical step.
actions that are coupled to the electron transfer process. Di· (Adapted with permission from Anal. Chern.• 38, 706-723 (1964).)
agnosis of coupled chemical reactions is often based on the Electrochemical Mechanisms Involving Coupled Chemical

relative heights of the anodic and cathodic peaks. For example, Reactions

consider a reversible charge transfer followed by an irrevers­


ible chemical reaction (entry III in the table). If the conversion Reversible electron transfer. no chemical complications
of R to Z is fast and the potential is scanned slowly, some of
R will be lost to Z. Less R will be available for the oxidation
back to 0 on the return scan and, therefore, the anodic peak II. Reversible eleC1ron transfer followed by a reversible chemical

will be smaller. In the extreme case, k may be so fast relative reactlon-ErGr mechanism

to the scan rate, that all of R will be converted to Z and the O+ne=R
return peak will be absent entirely. Figure 2 illustrates how
the voltammograms change depending on the ratio of the R~Z"
chemical rate constant to the scan rate. (Note that the y-axis
is the normalized current as giveri by the function: III. Reversible electron transfer followed by an Irreversible chemical

i/(n 3/ 2FAD 61 2C~V 1/2).) reaction-ErG, mechanism

Nicholson and Shain have demonstrated that the ratio of O+ne=R


the anodic to cathodic peak currents can be predicted from
the rate constant for the chemical step, k, and the time, T,
spent between E 112 and the switching potential, E>. (15). (E 1/ 2
will be the midpoint between the peaks at very high scan rates IV. ReverSible ch8mical reaction preceding a reversible electron

where the following chemical reaction does not have time to


consume a significant amount of R.) The theoretical rela­
transfer-C,E, mechanism

Z~O
.,

<-,
k O+ne-='R
O+ne~R ~z.
V. Reversible chemical reaclion preceding an irreversible eleC1ron

transfer-C,E, mechanism

k/a =
004
z O+ne-.R
Q
I­ VI. Reversible electron transfer followed by an Irreversible regeneration
~ O.l of starting material-catalytic mechanism
=>
LL.
O+ne=R

VII. Irreversible electron transfer followed by an irreversible regeneration


of starting- material
-0.2

180 120 60 o -60


(E -E~n, mV VIII. Multiple electron transfer with Intervening chemical reaction-ECE
mechanism
Figure 2. Cyclic voltammograms for a reversible electron transfer followed by
an irreversible chemical step for various ratios of the chemical rate constant 0+ n,e = R R=Y
10 scan rate. k , / a where a = nFv/ RTand v is volts/second. (Adapted with per·
mission from Anal. Chern.• 36, 706-723 (1964).)

Volume 60 Number 9 September 1983 699


..I

1.8 (A)
IV
U
'.x:&
;; 1.4

D.

.Il
1.0
."
0
C
.s
._D. 0.8

0.01 0.1 1.0 10 100

SCAN RATE

Figure 4. Ratio of anOdic to cathodic (or reverse 10 forward) peak currents as


a function of rate 01 voltage scan for various electrochemical mechanisms.
(Adapted with permission lromAnal. Chem.. 36, 706-723 (1964).)

~ 30/n -------------------------------

-301ft

0.01 0.1 1.0 10 )00


Figure 6. MethOds lor determining the proper baseline for measurement of the
SCAN RATE
peak current for the reverse scan. (a) Recording the signal versus time and
stopping the first scan at E~. (b) Using an X- Y recorder and stopping the scan
F\gIIe 5. Rate of shift of potential as a function of scan rate lor varlo\JS electrode at E~ until i approaches a steady state before scanning back. (c) Parameters
mechanisms. (Adapted with permission from Anal. Chem.. 36, 706-723 used for calculation of the current ratio using Nicholson's methOd, eqn. (10).
(1964).)

There are occasions when neither of these two methods is


4 shows the general trend that the peak current ratio follows convenient. Nicholson (19) has indicated that the proper
as a function of scan rate for each of the cases involving re­ current ratio can also be calculated using the following
versible electron transfer listed in the table. Also of some di­ equation.
agnostic value is the rate at which Ep/2 shifts as a function of
scan rate. These trends are presented in Figure 5. (Reference i". (ipa)o 0.485(i~)o 0086
-:;--+ + . (8)
(I2) gives a detailed development of each of these cases. Pol­ ipc ipo ipo
cyn and Shain also described multistep charge-transfer re­ In eqn. (8), ipc is the peak current for the forward process; (i>.)o
actions where an intermediate oxidation state is formed (16). is the absolute current at the switching potential; and (ipa)o
Treatment of a multi-electron reaction with an intervening is the uncorrected return peak current measured from the
chemical step has also been given (17, 18). Data from other current axis. (See Fig. 6c.)
types of electrochemical experiments is often needed to
evaluate rate constants for reactions with coupled chemical
a
There are some systems in which coupled chemical reac­
tion yields electroactive by-products. In these cases multiple
steps.) scans can be beneficial. For example, the product of electron
Measurement of the peak current is fairly simple for the transfer in the oxidation of aniline is thought to be a free
forward scan. The proper baseline can be obtained by re­ radical that very rapidly dimerizes (13) .
cording the background current for a scan without the analyte
under the same conditions (same electrolyte, surface pre­
treatment, etc.). The reverse scan is more complicated since ONH. OH+
~ g~. +.
the electrolysis for the forward process still contributes to the lJ
total current until the scan has passed the foot of the (forward)
wave again. The generally accepted approach is to assume that
the contribution of the forward process to the total current oa~ + .V~+H o-~-o-NH. + 2H'
continues to decrease with the square root of the time during III
the reverse scan. The baseline curve for the return scan can
be obtained b:v stopping the forward scan at the switching The p-aminodiphenylamine (III) that is formed is much
potential with the recorder sweeping along the x-axis as a more easily oxidized at the applied potential and a further two
function of time. This is shown in Figure 6a. electron oxidation occurs.
A second approach is to stop the scan at a convenient spot
(at least 35 m V past Ep for the forward scan) and hold the
potential until the current is relatively constant. The appro­
o-~-o-NH' .-. Q-NVNH + 2tr + 2.
priate baseline is shown in Figure 6b (J 3). IV

700 Journal of Chemical Education


trodes see references (25) and (26).) A three-electrode system
also permits one to minimize voltage errors due to ohmic (or
iR) loss through solution by placing the reference electrode
close to the working electrode surface. The voltage represented
by the product of the current and uncompensated resistance
(mainly the solution resistance between the reference and
working electrodes) is wasted and does not appear across the
electrode/solution interface. Since the peak current increases
with the square root of the scan rate, the voltage error in­
creases at the same rate. In cases where the error is of the order
of millivolts, the peaks from even a totally reversible redox
couple will appear to separate as though the system were
quasi-reversible. Therefore the experimenter should be cau­
tious about drawing conclusions concerning electron transfer
kinetics when either the current or uncompensated resistance
is large. Some manufacturers incorporate a positive feedback
circuit in their equipment to compensate for iR loss.
Another source of error in the applied voltage arises from
the variation in the rates at which ions diffuse across the ion
bridge separating the reference electrode from the sample
solution. The net difference in t.he movement of cations
compared to anions results in a charge separation and,
therefore an electrochemical potential difference. These
junction potentials can be minimized by choosing supporting
1- Ie-+- II -m+z.H~
electrolyte ions with similar diffusion coefficients (such as K+
m- Z.e -+nl +2H+
and Cl-).
Mercury working electrodes are limited t{) the negative
Figure 7. CV of aniline at a glassy carbon electrode In 0.1 M KCI and 0.05 M potential range. Platinum and various carbon electrodes are
potassium hydrogen phthalate solution (pH4). Scan rate Is 50 mV Isec; Agi AgCl
popular for performing oxidations. However, solid electrodes
in 0. 1 M KCI reference. Done<! line Indicates second scan.
are susceptible to adsorption, or surface fouling, and surface
oxidation. For example, platinum forms an oxide film that
This last reaction is a reversible process and the reduction step shows a large reduction peak near +0.8 V in 0.5 M H 2 S0 4
is observed (Fig. 7) at 0.2 Von the return scan. The oxidation versus the normal hydrogen electrode. Fortunately, a surface
of the p-aminodiphenylamine appears as a new peak on the current can be identified easily as such since the peak height
second scan (dotted line). The appearance of the new peak was is directly proportional to the scan rate. Often, judiciou.., choice
an important clue to the identity of the product of the oxi­ of solution conditions and electrode pretreatment can mini­
dation reaction. mize this problem.
In summary, cyclic voltammetry is a convenient tool for
The Calculation of Formation Constants ohaining qualitative information about electron transfer
The formation constants for complexes in both the oxidized processes. It is also a rapid method for obtaining good esti­
and reduced forms are useful in evaluating the u-donor and mates of formal reduction potentials, formation constants and,
1t'-acceptor properties of the ligands (20,21). The shift in the sometimes, the number of electrons transferred per reactant
formal reduction potential for the complex from the reduction molecule and rate constants, if the user is aware of its limita­
potential for the aquo-complex is dependent on the ratio of tions.
the formation constants for the reduced and oxidized forms.
For example, consider the reactions below. Acknowledgment
The author gratefully acknowledges the helpful comments
and encouragement of Dr. James C. Fanning and Dr. John D.
Petersen in the preparation of this article.
Literature Cited
(I) Bobbitt, J. M.. and Willo. John P., J. (Jrp . CM"", 45, 1978 (1900).
(2) Neisen, Stephen f'f K688el , Carl R.. Brien, I)flvid .J., and Weinhold, Frank, J . OrJl.
Chem .• 45. 2lt6 (1900).
(3) Poowe,.. Michael .l.. and Meyer, Thorna. J .. J . Arner. Chern . Soc .. 102.1289 (980).
(4.) Kalyan8SUnd8J'8m, K., Kiwi, J., Grabel, M., Hp.lv. Chim. Acta. 61 , 2720 (197A) .
(5) Ric<. C. A.. and Spence../. T .. ln org. Chern., 19.2845 (1980).
(6) HeadriciKe, ,J. B., "Electrochemica.l Techniques ror Inorganic Chemist.s,'· Academic
RT Pres.tJ, London and New York, 1969.
+ -In (KIlIK III ) (7) Henton. D. R., McCreer),. R. L., Swenlon, J. S., J .Org. t:hem., 45, 369-378 (1980) .
nF (8) Bard. Allen ,I., and ~·.ulkner. J.arry R.. "Electrochemical Method., Fundamentals .nd
. Often the formal potential can be evaluated by cyclic vol­ ApplicatiOns," John Wiley and Sons. New York, 19~0 .
(9) MacDonald. Digby D.• "Transient Techniques in glectrochemisuy," Plenum Pres.'.
tammetry when one of the complexes is too unstable to permit New York. 1977.
a direct determination of EO! by potentiometric measurement (10) Galus, Z., "Flllldamenlals of Elect.rocbemic.a l Analy';s," John Wiley & Son>!. New York,
1976.
of an equimolar solution of the two forms. If either Kill or Kn (II) Mauuda, H., and Ayabe. Y.. Z. Ekclrochem. . 59, 494 (1955) .
can be evaluated independently (22), then the other can be (12) Nicholson, R. S., and Shain, l.. AIUII. C/",m .. 36. 706 (1964).
(13) Adams. Ralph N., "Electrochemistry at Solid Electrod ..... MaruI Dokker. New York.
calculated from EO' data. 1969.
(14) Gilman. S .. Ekctfoo""l. Chem .. 2. III (1967).
Practical Consider allons (t,) Pa"",ns. R.. J. Electronnni. Ch~m. , 21, 35 (1969).
(16) Polcyn, D ., and Shain, I., Anal. Ch.rn .. 38. ;no, 376 (966).
Most modem equipment uses a three-electrode cell in which (t 7) Nichol",", R. S .. and Shain. I., Annl. Chern .. 37, 178 (I()65).
a counter or auxiliary electrode provides the current that is (18) Saveanl. J. M.. t;lectrochim . Acta, 12,7,,3 (1967).
(19) Nicholson. R. S., Anal. Chern .. 38, 1406 (966).
needed at the working electrode (23,24). Therefore, virtually (20) Kuehn. C. G .• and Tauhe. C. G.. J . Amer. Chern. Soc.• 98. 689 (1976).
no current nows through the reference electrode and its po­ (21) Tome. Henrique E.• and Cruetz, Cllml,/nufg. Cht'm., IG. 54;' (1977).
(22) For exrunple, see the ItIWctmphotomctric mcth"d of continuQu8 variAtion in Skoog,
tential remains constant. (For discussion of reference elec­ D. A., and We.~ D. M .. " Fundamentals of Annlylienl Chemi,try," 2nd &t.• Holt.
Rinehart, Bnd Win.lilon, New York . 1969.

Volume 60 Number 9 September 1983 701


(23) BionnalyticaJ Syste:l1lll, Int.• WeetLaf.yetu!.IN.: E. G. &G. Princeton Applied Research (2M Caler, Jr.. Roy D .• J. CH.EM. EDUe.. 51. A7 (1974).
Cotp., PTinceton. NJ j IBM Instrwnenta, lnc., White Plain., NY; Atka Scientific (26) Iv.... D.vid J .••nd Jan%, George J .. "Reference Electrodes Theory .nd Practi""," Ac­
International. S.nta Clara. CA; EGO Instruments. Cambridge. MA. .d.micPr.... NewYo,k. 1961.
(24) Li. Chi.· Yu. and Gillikin. Jr.. Jesse E.. J, CHF.M. Enuc., 54. A217 (1977). (27) Hockri.. John O. M•• and Reddy. Amulya K. N.• "Modern Electrochemistry;" Vo'.. 2.
Plenum Preae. New York, 1970.

702 Journal of Chemical Education

Вам также может понравиться