Вы находитесь на странице: 1из 5

11/14/2017 Understanding Errors In Hand-Held Measuring Instruments : Modern Machine Shop

Understanding Errors In Hand-Held


Measuring Instruments

Different instruments (and different operators) are prone to different errors.

Article Post: 3/15/2003


RICHARD CLARK

Related Topics:

Inspection & Measuring Equipment

Measuring instruments have certain common errors built right into them, and the
instruments most commonly used in machine shops are no exception. Recognizing these
errors is a basic requirement for proper inspection. Its not necessary to comprehend the
precise physics or geometry underlying the error, but it is important to understand the
nature of the error and the extent to which it might affect or limit the instruments precision.

Take the micrometer. This is a very accurate and stable instrument, but even this
instrument has a certain capacity for error built in. Over-tightening the micrometers
spindle can cause the anvil portion of the gage to change shape. Lower-quality
micrometers are made of materials more prone to this error. While the amount of
deflection may be only 0.0001 or 0.0002 inch, that could be 50 percent of some tolerance
bands.

Calipers
A caliper is prone to error described by Abbs principle, which says that a source of
error is introduced anytime the reference line of a measuring system doesnt lie along the
same line as the dimension being measured. On a caliper, the scales or gears are not in
line with the measuring faces or contacts. As a result, the caliper shifts and wiggles (in
microscopic increments) in a way comparable to that of a table or chair when the legs
become loose. The error can be minimized by measuring as close to the rail as possible,
as shown in Figure 1.

Another error limits the instruments effectiveness at measuring an internal diameter. The
design of the standard caliper places the measuring contacts or jaws offset from one
another. That means the jaws will never find the maximum diameter of the workpiece.
Figure 2 illustrates this.

https://www.mmsonline.com/articles/understanding-errors-in-hand-held-measuring-instruments 1/5
11/14/2017 Understanding Errors In Hand-Held Measuring Instruments : Modern Machine Shop

One other matter to keep in mind with a caliper is the additive nature of the error amounts
that are permitted by the instruments calibration. To pass calibration, a dial caliper with
0.001-inch resolution must be accurate within 0.001 inch for length measurements and
allow no more than 0.001 inch for parallelism error. But the measurement of a large part
may be affected by both errors. In such a case, the possible error is 0.002 inch. According
to the 10:1 rule, that means the instrument should inspect a characteristic only if the total
tolerance is 0.020 inch or more.

Indicators
Indicators are prone to at least two common error sources resulting from the mechanics of
how they work. One is cosine error. This error is particularly likely with a test indicator,
which is the type of indicator that uses a tilting lever arm to measure a part moving
underneath it. In a correct measurement setup, the lever arm is adjusted so that the
movement of the tilt is as close as possible to being perpendicular to the measured
surface.

The test indicator setup shown in Figure 3 can check for the presence of cosine error. The
check involves a confirmation of the thickness difference between gage blocks that are
0.100 and 0.103 inch thick. The lever arm should be adjusted so that it will tilt along a line
very close to being straight up and down. If this is done successfully, then when we move
from one gage block to the other, the dial reading should show a change of 0.003 inch.

The same test can be used to illustrate cosine error by changing the tilt so that this error is
created, as shown in Figure 4. A subtle adjustment created an error of 0.0003 inch. If this
doesnt seem like much, consider that it represents a 10 percent difference. Imagine trying
to match a flatness, parallelism or roundness characteristic to the print tolerance when
your indicator is reading 110 percent of the actual value.

The other error common for indicators is hysteresis. This term refers to a lag between
action and reaction in a mechanical system. An example of hysteresis is the play that
may be detectable in a cars steering wheel. The steering wheel turns a tiny amount
before the cars wheels begin to change direction. Indicators have this play, too. As
precisely as the indicator may have been made, it still needs some clearance between
gear teeth. If you push the plunger of a dial indicator up from its rest position, there will be
a very short duration during which the plunger is moving but the gear that moves the
indicating hand on the dial face has not yet begun to move.

How often is a dial indicator moved along a surface to measure a taper or step? There will
be hysteresis error in any such measurement because of the small delay before the
indicators movement causes the indicator hand to move. The error is very small, but the
point to remember is that errors compound. An error allowed at calibration, combined with
cosine error, combined with hysteresis, combined with a part not being cleaned
sufficiently, can create more error than most would suspect.

https://www.mmsonline.com/articles/understanding-errors-in-hand-held-measuring-instruments 2/5
11/14/2017 Understanding Errors In Hand-Held Measuring Instruments : Modern Machine Shop

Measurement Variation
C. E. Johansonn, the first one to mass produce and market gage blocks, once observed
that No two things are exactly alike, and if they were, we would still get different values
when we measured them.

Measurement variation is present every time a part is measured. In some cases, when
using a caliper or a 0.001-inch dial indicator, the step of the gage increment is large
enough to hide most of the variation. In other situations, the 10:1 rule will require that an
instrument of 0.0001- or 0.00005-inch resolution be used to inspect the part feature. When
anything is measured to this level of discrimination, the measurement becomes a shade of
gray instead of black or white.

Here are some common types of variation:

Instrument variation. Not all instruments repeat as well as others. A vernier


micrometer of 0.0001-inch resolution may be more stable than a digital micrometer
of 0.00005-inch (50 millionths) resolution. And differences in measurement variation
between a micrometer and a caliper can be large indeed.
Part variation. The true form of a supposedly flat surface or a supposedly cylindrical
bore can depart considerably from the expected ideal. For there to be confidence in
any precision measurement, there needs to be a specific, agreed-upon definition as
to what each measurement means and entails.

For example, measuring a particular bore may entail measuring the diameters in the X
and Y directions at each end of the bore as well as in the middle. The diameter of the bore
may be defined as the grand average of all of these values.

Operator variation. Everyone measures a little differently from everyone else. This
seldom means anyone is wrong. However, if subtle differences from one operator
to the next arent addressed, then we could all be wrong as a group. Reference
standardssample parts of known measurement that can be used to make a quick
confirmation of a gagecan be used to identify individual measuring tendencies. For
example, one operator may consistently measure a tenth on the plus side compared
to the rest of the shop. Further variation comes from the simple fact that we all have
good days and bad days. Given these circumstances, reference standards can help
the overall team. Making a quick measurement of the reference standard just before
the critical feature is measured can ensure that a particular combination of operator
and gage is measuring accurately at that moment. Think of this as a variation on the
maxim, Measure twice, machine once.

The R Words

https://www.mmsonline.com/articles/understanding-errors-in-hand-held-measuring-instruments 3/5
11/14/2017 Understanding Errors In Hand-Held Measuring Instruments : Modern Machine Shop

The most common terms used to describe measurement variation are repeatability and
reproducibility. Repeatability describes the range of values obtained when a single
operator measures the same part several times. For example, Operator 1 may measure a
plug five times with the following results: 1.4948, 1.4948, 1.4949, 1.4951 and 1.4950. We
could say the repeatability of the measurement was 0.0003, which is the difference
between the highest and lowest values obtained during the test.

Reproducibility describes the difference between the average measurements obtained


by two or more operators. Operator 2, for example, may measure the same plug and
obtain the results 1.4950, 1.4952, 1.4951, 1.4952 and 1.4952. While Operator 1s average
was 1.4949, Operator 2s average was 1.4951. The reproducibility of the measurement is
the difference between the averages, 0.0002. (And Operator 2s repeatability is also
0.0002.)

What these numbers mean is that, on any given day, the two operators might measure the
same feature 4 or 5 tenths differently. Which operator is right? They both are, because
that level of variation falls within the range of the combined value of reproducibility and
repeatability.

So what is the diameter of the plug? It may be as small as 1.4948 or as large as 1.4952.
Using this particular measurement system, it may not be possible to know any more
accurately than this.

Know the uncertainty is not an oxymoron. Instead, understanding sources of error and
understanding the scope of variation are important requirements for getting meaningful
results from even the most familiar measuring device.

About the author: Richard Clark is a metrologist working in Portland, Indiana. He can be
contacted at rcmetrology@yahoo.com

You might also like:

Same Problem, Different Resol Calipers: Ideal For Measurement On The


Article In shopfloor measurements, dont Column Although it has been around for a long time,
count partial increments, but do keep a the caliper is still an extremely versatile and useful tool

https://www.mmsonline.com/articles/understanding-errors-in-hand-held-measuring-instruments 4/5
11/14/2017 Understanding Errors In Hand-Held Measuring Instruments : Modern Machine Shop

https://www.mmsonline.com/articles/understanding-errors-in-hand-held-measuring-instruments 5/5

Вам также может понравиться