Вы находитесь на странице: 1из 1

Procedures for Using Steam Autoclaves

QUESTIONS: Contact DRS, 217-333-2755, for additional information.


MAINTENANCE & CARE: CHOOSE PROPER OPERATING CONDITIONS: DECONTAMINATING
Check the Pressure: Sterilization will occur only when the conditions of time, temperature, pressure, and humidity have been met. BIOHAZARDOUS WASTE BAGS:
Check all pressure gauges. Incorrect selection of time or exhaust cycle may damage the autoclave, cause liquid to boil over, or bottles to Use autoclavable, high
Jacket pressure gauge break. strength polymer bags
should be a minimum of 15 Proper use of autoclave will minimize the chance of serious injury. imprinted with chemical
psi (refer to indicator (if possible).
manufacturers Biological Waste Liquids Dry Items Glassware Add approximately 250 mL
Items
instructions for maximum (Gravity Cycle) (Liquid Cycle) (Gravity Cycle) (Gravity Cycle) water to bag before closing
psi). and transporting to autoclave
Loosen caps or use
Use Sterilization Indicators Open the bag >2", Dirty: Place in area.
a vented closure, Fabrics Wrap;
Place in tray, middle of the pan;
Preparation Fill containers no Instruments: Clean, Place bag in a polypropylene
Visual indicators, e.g., Place indicator if Clean: wash, rinse,
more than 75% dry, lay in pan or stainless tray before
chemical/tape indicators, needed wrap
capacity
measure one or more autoclaving.
Dirty: In detergent
Fabrics: Separated, Open the bag at least 2 to
physical conditions of the Placement in and pan;
In the center Upright in pan on edge;
autoclave cycle. Autoclave
Instruments: Flat
Clean: On side or allow steam to enter.
inverted When complete, the
Mechanical indicators
record the time- sterilization indicators must
Temperature 121oC 121oC 121oC 121oC show adequate
temperature profile
attained during a cycle. decontamination before
Biological indicators, i.e., 60-120 min. 22 min. for volumes disposal as municipal solid
Treatment Time in depending on load <100mL; waste.
spore vials, are the only 30-60 min. 30-60 min.
Minutes size and packing 40 min. for volumes
approved way to prove density >100mL
Repeat cycle if sterilization
sterility. indicators do not show
NOTIFY FACILITIES SUPPORT Dirty: Slow exhaust; evidence of sterilization.
Exhaust Cycle Slow exhaust Slow exhaust Fast exhaust and dry
Clean: Fast/dry Overbag biohazardous waste
WHEN A PROBLEM OCCURS:
bag with an opaque trash
No steam. Avoid puncturing Hot bottles may Check reference for Glassware with bag before placing in the
Valves leaking. Notes: bags. Overbag and explode. Let cool proper packaging cracks or deep
dispose of properly. before moving. methods scratches may crack regular trash.
Door gasket deterioration.
Jacket pressure below 15 psi. CAUTION: Reminders
Temperature not reaching NEVER AUTOCLAVE FLAMMABLE, REACTIVE, CORROSIVE, TOXIC or RADIOACTIVE MATERIALS, e.g., Store biohazardous waste in
121C (250F). bleach. a closed leak-proof
Erratic temperature or Materials that melt (plastic lab wear) at 121C will block chamber exhaust drain if not placed in a container.
pressure during cycle. shallow autoclave pan able to withstand that temperature. Never allow waste to
Steam escapes from around o
Use caution when increasing autoclave temperature to 135 C because plastics (including some plastic accumulate in the lab.
door during cycle. pans) melt at this temperature, causing difficult clean-up and damage to temperature sensors. Never leave waste
Water on floor or in chamber Always wear safety glasses, goggles, or face shield, lab coat or apron, and heat-protective non-asbestos unattended.
after a cycle. gloves when opening door or removing item(s) from autoclave. Clean up leaks and spills with
Reference:
a suitableUsing the Gravity
disinfectant.
Articles very wet after a fast Do not mix loads that require different exposure times and exhaust.
Displacement Steam Autoclave in
exhaust or dry cycle. Open door only after chamber pressure returns to zero. Leave door open for several minutes to allow
the Biomedical Laboratory
Excessive steam in area pressure to equalize and for materials to cool. DHHS/PHS/HIH/DS
during cycle or after opening Open door slowly. Beware of rush of steam or water.
door.

Вам также может понравиться