Pressure and Flow

Pressure Effects on Static Gases


Pressure and the Ideal Gas Law
Pressure is a measure of the force are compressible, and their molecules PV=nRT
exerted upon a surface. By the Ideal get closer to each other as pressure
Gas Law (at right), if the gas pressure increases. If the pressure decreases Double the
pressure:
V=nRT
(P) increases by a factor of 2, the by half, volume doubles. The molar
volume of the gas (V) decreases by mass (n, the number of particles in the 2 2P
a factor of 2, all other factors being volume) stays the same, regardless of Halve the
pressure:
2V=nRT
constant. This occurs because gases changes in static pressure.
P
Effects of Pressure on Mass and Volume P: pressure (static)
V: volume
Lets picture a flexible container filled If we halve the original pressure to 0.5
n: molar mass
with 500 cm3 of air at atmospheric atm, the volume expands to 1000 cm3. R: gas constant
pressure (1 atm, about 14.696 psia) and If we halve the pressure again to 0.25 T: temperature
standard ambient temperature (25C). atm, the volume grows to 2000 cm3.
If we double the pressure to 2 atm, the In each instance, we have neither
air molecules get closer together, and removed nor added any air, and so the
the volume is compressed to 250 cm3. molar mass of air inside the container Decreasing pressure increases
However, the number of air molecules does not change from its original 500 gas volume.
(molar mass) remains the same. scm3 of air.

Air at 1 atm Air at 2 atm Air at 0.5 atm Air at 0.25 atm
Mass: 500 scm3 Mass: 500 scm3 Mass: 500 scm3 Mass: 500 scm3
Volume: 500 cm3 Volume: 250 cm3 Volume: 1000 cm3 Volume: 2000 cm3

The Fastest Flow Controller Company in the World!

FLUID MECHANICS
alicat.com/u
FLUID MECHANICS

Pressure Effects on Gases in Motion


Effects of Line Pressure on Mass and Volumetric Flow Rates
When air is put into motion as a flow of line pressure halves the volumetric
gas, the actual space that the air takes flow rate, and vice versa. However, the Decreasing line pressure
up per unit of time (volumetric flow number of molecules of air that flow increases volumetric flow rate.
rate) varies with pressure in the same per unit of time (mass flow rate) does
manner as the static air. Doubling the not change.

Air flow at 0.25 atm


Mass: 500 sccm
Volume: 2000 ccm

Air flow at 0.5 atm


Mass: 500 sccm
Volume: 1000 ccm

FLOW

Air flow at 1 atm


Mass: 500 sccm
(500 scm3/min)
Volume: 500 ccm
(500 cm3/min)
Air flow at 2 atm
Mass: 500 sccm
Volume: 250 ccm

Sizing Flow Meters for Subatmospheric Pressures


Differential pressure-based flow meters meters built-in overrange of 28% Air flow at 0.25 atm (2000 ccm)
have internal flow channels that are allows it to read volumetric flows as
sized to accommodate the maximum high as 640 ccm (which it would see at 0.5 atm (1000 ccm)
expected volumetric flow rates passing 0.78 atm), but 1000 ccm at 0.5 atm is 1 atm (500 ccm)
through them. When these meters are too much and produces a VOV error.
used in vacuum applications, they may Because the flow meters are physically 2 atm (250 ccm)
need to be oversized in order to handle sized for volumetric flows, the solution
the increased volumetric flow rates at is to order an oversized flow meter at
subatmospheric pressures. twice the full-scale range (M-1SLPM-D). FLOW
Take for example an Alicat mass flow The ranges would be specified for the
meter built for 500 sccm (scm3) at full original mass flow range of 500 sccm M-500SCCM-D
scale (M-500SCCM-D). If we double and a doubled volumetric flow range Max volumetric flow: 500 ccm
the pressure to 2 atm and keep the of 1000 ccm. Similarly, at 0.25 atm the M-1SLPM-D
mass flow rate constant at 500 sccm, meter would be built for quadruple the Max volumetric flow: 1000 ccm
the volumetric flow rate decreases to flow (M-2SLPM-D) and ranged for 500 M-2SLPM-D
250 ccm. This decrease in the volume sccm and 2000 ccm. Max volumetric flow: 2000 ccm
of air flow poses no challenges to flow (Note that the maximum oversize is 5x,
measurement because the meter is which would nominally accommodate
designed to handle twice that amount. a line pressure of 0.2 atm, or 2.9 psia.
Now, if we reduce the line pressure to Mass flow ranges normally are not Oversize flow meters for use at
0.5 atm and keep the mass flow rate oversized in this scenario, in order to subatmospheric pressures.
constant at 500 sccm, the volumetric maintain measurement resolution and
flow rate doubles to 1000 ccm. The accuracy over the intended range.)
Pressure Drop under Low Pressures
Effects of Line Pressure on Instrument Pressure Drop
Pressure drop is the loss of line At subatmospheric pressures, pressure P Q
pressure caused by frictional resistance drops throughout the entire process
in the flow path. This is induced by will be higher than at atmospheric P: pressure drop
Q: volumetric flow
instruments on the line and even the pressures. Ignoring this can hinder the : viscosity
tube walls to a small degree. Pressure intended operation of the process.
drop tells you how much additional In the chart below, reducing the
differential pressure must be supplied pressure from 1 atm to 0.25 atm
to attain the desired flow rate. quadruples the volumetric flow rate A decrease in line pressure
Pressure drop increases proportionally to 2000 ccm, which quadruples the increases pressure drop.
to volumetric flow rate (flow velocity). pressure drop through the 500-ccm
Volumetric flow increases as pressure flow meter (red). This then requires 4
decreases. Thus, pressure drop also times the pressure differential (4 psi)
increases as line pressure decreases. to achieve a full-scale flow of 500 sccm.

Pressure Drop and Oversized Flow Meters


We increase flow channel size to accommodate the higher
volumetric flows at subatmospheric pressures. Doing this also
provides the benefit of reduced pressure drop. Line Pressure
atm psia
For example, the flow meters M-500SCCM-D, M-1SLPM-D and
M-2SLPM-D each have pressure drops of 1 psi at their full-scale 30
2
P: 0.5 (2%)
volumetric flow rates. Flowing 500 ccm in the 1000-ccm meter
(50% full scale at 1 atm) halves the pressure drop to 0.5 psi.
Likewise, flowing 500 ccm in the 2000-ccm meter (25% full
25
Pressure Drop (psi)
scale at 1 atm) yields 0.25 psi of pressure drop. as % of line pressure
When 500 sccm of mass flow becomes 1000 ccm of volumetric 1.5 M-500SCCM-D flow meter
flow at 0.5 atm, using the larger 1000-ccm meter results in the Oversized flow meter
full-scale drop of 1 psi. This is double the drop of 0.5 psi at 500 20
Oversized Whisper flow meter
sccm and 1 atm, because the volumetric flow rate has doubled.
Similarly, flowing 500 sccm at 0.25 atm in the 2000-ccm meter
also becomes a full-scale flow with a pressure drop of 1 psi.
1 15
P: 0.07 (0.4%)
P: 1.0 (7%)

Use low pressure drop meters for


subatmospheric apps. 10

0.5 P: 0.07 (1%)


Low Pressure Drop Meters P: 1.0 (14%)
5 P: 2.0 (27%)
Using larger meters to handle subatmospheric volumetric 0.25 P: 0.07 (2%)
flows also helps to reduce pressure drop, but sometimes this P: 1.0 (27%)
is not enough. In the 0.25 atm example, the lower drop of 1
psi nevertheless represents 27% of the available pressure (3.7 P: 4.0 (108%)
psia), which may be too much to lose. In these cases, it is best to 250 500 1000 1500 2000
choose a meter with lower pressure drop. An Alicat Whisper
series meter, sized for 2000 ccm (MW-2000SCCM-D) yields a
Volumetric Flow Rate (ccm)
full-scale pressure drop of only 0.07 psi at 0.25 atm, which is
only 2% of the available pressure.
FLUID MECHANICS

Pressure Effects and Flow Controllers


Sizing Flow Controllers for Vacuum and Subatmospheric Conditions
Unlike meters, differential pressure- below, the measurement head inside
based mass flow controllers do the flow controller always sees flow
not need to be oversized for use in rates at ambient air pressure (1 atm).
subatmospheric conditions. For these Only after the gas passes through the
Request downstream valves on
applications, the proportional control valve does it expand to four times its
controllers for vacuum apps.
valve is located at the downstream volume under 0.25 atm of pressure.
side of the controller, forming a sonic As in the previous examples, the mass
barrier that shields the sensor from flow rate remains constant as the gas
the volumetric expansion of the transitions from ambient pressure to
gas under vacuum. In the example subatmospheric pressure.

MC-500SCCM-D

Air flow at 1 atm SCCM


Air flow at 0.25 atm
Mass: 500 sccm Air Mass: 500 sccm
valve
Volume: 500 ccm Volume: 2000 ccm

FLOW FLOW

Key Concepts
Decreasing pressure increases gas volume.
Decreasing line pressure increases volumetric flow rate.
Oversize flow meters for use at subatmospheric pressures.
A decrease in line pressure increases pressure drop.
Use low pressure drop meters for subatmospheric apps.
Request downstream valves on controllers for vacuum apps.

Learn more!

A Whisper series meter is a


great choice for monitoring flows
at subatmospheric pressures.
alicat.com/u
Alicat Scientific, Inc 888-290-6060