Digital Multimeter Measurement Errors Series

Resistance; DC Current;
AC Current; and Frequency
and Period Measurement
Errors in Digital Multimeters
Application Note AN 1389-2

Introduction
When you make measurements with a digital
multimeter (DMM), common errors will crop up.
The following discussion will help you eliminate
potential measurement errors and achieve the
greatest accuracy with a DMM. This paper covers
resistance; dc current; ac current; and frequency
and period measurement errors. For a discussion of
system cabling errors and dc voltage measurement
errors, see Application Note AN 1389-1. For an
overview of ac voltage measurement errors, see
Application Note AN 1389-3. (NOTE: The Agilent
34401A, a 6-1/2-digit, high-performance DMM with
both benchtop and system features, will be used as
an example throughout this article).
Resistance Measurement Errors

The Agilent 34401A offers two Range Test Current DUT Power at Full Scale
methods for measuring resistance:
2-wire and 4-wire ohms. For both 100 1 mA 100 W
1 k 1 mA 1 W
methods, the test current flows
10 k 100 A 100 W
from the input HI terminal and
100 k 10 A 10 W
then through the resistor being
1 M 5 A 30 W
measured. For 2-wire ohms, the
10 M 500 nA 3 W
voltage drop across the resistor
being measured is sensed internal Figure 1.

to the multimeter. Therefore, test


lead resistance is also measured.
For 4-wire ohms, separate sense Settling Time Effects The Agilent High-Resistance Measurement Errors
connections are required. Since 34401A has the ability to insert When you measure large
no current flows in the sense automatic measurement settling resistances, significant errors can
leads, the resistance in these leads delays. These delays are adequate occur due to insulation resistance
does not give a measurement for resistance measurements with and surface cleanliness. You
error. less than 200 pF of combined should take the necessary
cable and device capacitance, pre-cautions to maintain a "clean"
Errors for dc voltage measurements high-resistance system. Test leads
which is particularly important
(see Application Note 1389-1) also and fixtures are susceptible to
when measuring resistances above
apply to resistance measurements. leakage due to moisture absorption
100 k. Settling due to RC time
However, there are additional in insulating materials and "dirty"
constant effects can be quite long.
error sources that are unique surface films. Nylon and PVC
Some precision resistors and
toresistance measurements. arerelatively poor insulators
multi-function calibrators use
Power Dissipation Effects When large parallel capacitors (1000 pF (109ohms) compared to PTFE
measuring resistors designed for to 0.1 F) with high resistor values insulators (1013 ohms). Leakage
temperature measurements (or to filter out noise currents injected from nylon or PVC insulators can
other resistive devices with large by their internal circuitry. easily contribute a0.1% error
temperature coefficients), be Non-ideal capacitances in cables when measuring a 1M
aware that the multimeter will and other devices may have much resistance in humid conditions.
dissipate some power in the DUT. longer settling times than expected
If power dissipation is a problem, just by RC time constants, due
select the multimeters next todielectric absorption (soak)
higher measurement range to effects. Errors will be measured
reduce the errors to acceptable when settling after the initial
levels. Figure 1 shows several connection and after a range
examples. change.

2
DC Current Measurement Errors

When you connect the multimeter


in series with a test circuit to
measure current, a measurement
error is introduced, caused by
the multimeters series burden
voltage. A voltage is developed
across the wiring resistance and
current shunt resistance of the
multimeter, as shown in Figure 2.

V s = Source Voltage
R s = DUT Source Resistance 100 x Vb
V b = Multimeter Burden Voltage Error (%) =
Vs
R = Multimeter Current Shunt

Figure 2.

AC Current Measurement Errors

Burden voltage errors, which The burden voltage increases as


apply to dc current, also apply the input frequency increases.
to ac current measurements. Some circuits may oscillate when
However, the burden voltage for performing current measurements
ac current is larger due to the due to the multimeters series
multimeters series inductance inductance and the specific
and the measurement connections. measurement connections.

Frequency and Period Measurement Errors

The Agilent 34401A uses a counters are susceptible to errors will also occur when you attempt
reciprocal counting technique to when measuring low-voltage, low- to measure the frequency (or
measure frequency and period to frequency signals. The effects of period) of an input following a
generate constant measurement both internal noise and external dc offset voltage change. To avoid
resolution for any input frequency. noise pickup are critical when this, allow the multimeters input
The 34401As ac voltage measure- measuring "slow" signals. The dc blocking capacitor to fully
ment section performs input error is inversely proportional to settle before making frequency
signal conditioning. All frequency frequency. Measurement errors measurements.

3
Conclusion

When measuring small resistance Agilent Technologies' Test and Measurement By internet, phone, or fax, get assistance
or low current, it is important to Support, Services, and Assistance with all your test & measurement needs
Agilent Technologies aims to maximize the value Online assistance:
take extra precautions to make you receive, while minimizing your risk and www.agilent.com/find/assist
quality measurements. Once the problems. We strive to ensure that you get the
system settling times have been test and measurement capabilities you paid for Phone or Fax
determined, the reading delay and obtain the support you need. Our extensive United States:
support resources and services can help you (tel) 1 800 452 4844
times can be set to optimize choose the right Agilent products for your Canada:
measurement time and increase applications and apply them successfully. Every (tel) 1 877 894 4414
system throughput. instrument and system we sell has a global (fax) (905) 282 6495
warranty. Support is available for at least five China:
years beyond the production life of the product. (tel) 800 810 0189
Two concepts underlie Agilent's overall support (fax) 800 820 2816
For more information about the policy: "Our Promise" and "Your Advantage." Europe:
Agilent 34401A DMM, go to Our Promise (tel) (31 20) 547 2323
www.agilent.com/find/34401a Our Promise means your Agilent test and (fax) (31 20) 547 2390
measurement equipment will meet its advertised Japan:
performance and functionality. When you are (tel) (81) 426 56 7832
choosing new equipment, we will help you with (fax) (81) 426 56 7840
product information, including realistic Korea:
performance specifications and practical (tel) (82 2) 2004 5004
recommendations from experienced test (fax) (82 2) 2004 5115
engineers. When you use Agilent equipment, we
Latin America:
can verify that it works properly, help with
(tel) (305) 269 7500
product operation, and provide basic (fax) (305) 269 7599
measurement assistance for the use of specified
Taiwan:
capabilities, at no extra cost upon request. Many
(tel) 080 004 7866
self-help tools are available.
(fax) 080 028 6331
Your Advantage
Your Advantage means that Agilent offers Other Asia Pacific Countries:
(tel) 1 800 375 8100
a wide range of additional expert test and
(fax) (65) 836 0252
measurement services, which you can purchase
(e-mail) tm_asia@agilent.com
according to your unique technical and business
needs. Solve problems efficiently and gain a Product specifications and descriptions in this
competitive edge by contracting with us for document subject to change without notice.
calibration, extra-cost upgrades, out-of-warranty Agilent Technologies, Inc. 2012
2002
repairs, and on-site education and training, as Printed in the USA May 7, 2012
January 31, 2002
well as design, system integration, project 5988-5512EN
management, and other professional engineering
services. Experienced Agilent engineers and
technicians worldwide can help you maximize
your productivity, optimize the return on
investment of your Agilent instruments and
systems, and obtain dependable measurement
accuracy for the life of those products.