BVS CM
Efficient analyzer for battery discharge testing
failing cells
Reliable and easy to operate
Description
The Battery Voltage Supervisor Capacity Model performing a battery discharge test, providing
(BVS CM) is a battery monitoring system for detailed analysis of an individual cells condition.
real-time data gathering and viewing. The BVS CM system identifies a potential
The BVS CM main role is monitoring battery battery malfunction by monitoring the following
voltage values during a battery discharge cell parameters from an array of the cell voltage
testing. modules CVM-M:
The BVS CM is an integrated system consisting - Cell voltage
of: - Intercell connection voltage
- BVS Control Unit (BCM-CU) In addition, the BCM-CU provides ambient
- Cell Voltage Modules (CVM-C) temperature measurements on multiple
The BCM-CU performs monitoring and locations.
measurements data acquisition from the CVM-C The CVM-C modules detect cells that fail a
modules and their transfer to a PC. In addition, discharge test based on measured voltage
the BCM-CU provides power supply to all values, so these cells can be safely removed
connected CVM-C modules. from the battery string.
The cell voltage modules are installed on the Data acquisition and extensive analysis
individual batteries, performing battery voltage capabilities of data collected from the CVM-C
and intercell voltage measurements. are available in the DVB Win application
The BVS CM system acts as an add-on to the software suite, providing a user data viewing,
Battery Load Unit (BLU) series of devices while tracking and generation of comprehensive
B-B000CM-301-EN, 2016-08-12
reports.
1
Applications
In the BVS CM implementation, every CVM-C operating CVM-C modules and the BCM-CU.
module is installed directly on an individual The connection diagram of the BVS CM to a
battery. Failure of one of the CVM-C modules battery string is presented in the figure below.
will not affect communication between the rests
of
Connectivity diagram of BVS CM system components in combination with Battery Load Unit device
B-B000CM-301-EN, 2016-08-12
2
Features and Benefits Battery Internal Resistance Measurement
Using the BVS CM as a supplement to the The battery internal resistance measurement is
Battery Load Unit (BLU) devices provides an additional BVS CM feature available in
multiple advantages during a battery discharge combination with the BLU instruments, along
testing. with the intercell connection resistance
The CVM-C modules measure voltage values in measurement and the battery short circuit
real-time mode and forward recorded data to the current estimation. The test method used for
Control Unit in user defined intervals selected internal resistance measurement and a short
through DV-B Win application software UI. circuit current estimation is specifically designed
All out-of-tolerance measurement values are to comply with battery test methods
announced by a LED signalization on an recommended by the IEC standards.
individual CVM-M module, so the failing cells
can be detected before endangering the entire DV-B Win Software
battery system. This enables detecting and
All results transferred from the BVS CM system
safely bypassing the failing cell during a
can be viewed, analyzed and presented in
discharge process.
selectable report forms by using the DV-B Win
Up to 4 ambient temperature measurement
software application.
channels are provided on the BCM-CU for
Data can be transferred to a PC through a USB
measuring ambient temperature on multiple
or Bluetooth communication.
locations.
Both, the BVS CM and the BLU device provided
results can be viewed on the same report form.
A quick pass/fail test for a cell internal resistance
values may be performed based on a user
defined thresholds, adding additional insight in
an individual cells status.
B-B000CM-301-EN, 2016-08-12
3
DV-B Win Main Features
Both discharge results from BLU device and cell voltage values from CVM-C can be previewed
in one DV-B Win file
Test report for BLU and BVS CM are available in several formats (Excel spreadsheet, pdf, word
or RTF)
DV-B Win features: String View (BLU) & Cell View (BVS CM) functions
B-B000CM-301-EN, 2016-08-12
4
Technical Data
Mains Power Supply
Environment conditions
Temperature: Electromagnetic Compatibility:
All specifications herein are valid at cell temperature of + 25 C and recommended accessories.
Specifications are subject to change without notice.
Specifications are valid if the instrument is used with the recommended set of accessories.
5
Ordering Info:
Instrument Article No
Battery Voltage Supervisor Control Unit BVS-CUNN-000
Included accessories
DV-B Win software including USB cable