Академический Документы
Профессиональный Документы
Культура Документы
18
Positron Emission
Tomography
The second major method for tomographic simultaneously, in 180-degree opposing direc-
imaging in nuclear medicine is positron emis- tions, usually within a few tenths of a mm to
sion tomography (PET). This mode can be a few mm of the location where the positron
used only with positron-emitting radionu- was emitted, depending on the energy and
clides (see Chapter 3, Section G). PET detec- range of the positrons. Near-simultaneous
tors detect the back-to-back annihilation detection of the two annihilation photons
photons that are produced when a positron allows PET to localize their origin along a line
interacts with an ordinary electron. Although between the two detectors, without the use
the annihilation photons could be detected of absorptive collimators. This mechanism
using single photon emission computed is called annihilation coincidence detection
tomography (SPECT) systems operating in (ACD). Detection of a pair of annihilation
conventional single-photon counting mode, photons in opposing detectors actually defines
these systems are not optimally designed for the volume from which they were emitted.
the relatively high energy of annihilation Most ACD detectors have square or rectangu-
photons (511keV). They have relatively low lar cross sections. Thus the volume is essen-
detection efficiencies at these energies and tially a box of square or rectangular cross
require relatively inefficient high-energy col- section, with dimensions equal to those of the
limators. As well, SPECT systems do not take detectors (Fig. 18-1).
advantage of the back-to-back directional Coincidence logic (Chapter 8, Section F and
characteristics of annihilation photons. This Fig. 8-15) is employed to analyze the signals
unique feature is exploited advantageously from the opposing detectors. For many PET
with special annihilation-coincidence detector scanners, this is accomplished by having the
systems for PET. electronics attach a digital time stamp to
PET has gained widespread clinical accep- the record for each detected event. Typically,
tance and now is firmly established alongside this is done with a precision of approximately
planar imaging and SPECT in clinical nuclear 1 or 2 nanoseconds (1 nsec = 109 sec). The
medicine. In this chapter, we describe the coincidence processor examines the time
basic features of annihilation coincidence stamp for each event in comparison with
detection, the design and performance char- events recorded in the opposing detectors. A
acteristics of PET detectors and scanners, coincidence event is assumed to have occurred
and some of the important clinical applica- when a pair of events are recorded within a
tions of PET. specified coincidence timing window, which
typically is 6 to 12 nanoseconds.
Although annihilation photons are emitted
A. BASIC PRINCIPLES OF
simultaneously, a small but finite coincidence
PET IMAGING window width is needed to allow for differ-
ences in signal transit times through the
1. Annihilation Coincidence Detection cables and electronics, as well as different
When a positron undergoes mutual annihila- distances of travel by the two photons from
tion with a negative electron, their rest the annihilation event to the detectors (see
masses are converted into a pair of annihila- Section A.2). In addition, the detectors in a
tion photons (see Fig. 3-7). The photons have PET scanner do not have perfect timing preci-
identical energies (511keV) and are emitted sion and therefore have a finite timing
307
308 Physics in Nuclear Medicine
Annihilation event
Detector Detector
Object containing
positron-emitting
radionuclide
Accepted by coincidence detection
Rejected by coincidence detection
FIGURE 18-1 Volume (green shaded area) from which a pair of simultaneously emitted annihilation photons can be
detected in coincidence by a pair of detectors. Not all decays in this volume will lead to recorded events, because it is
necessary that both photons strike the detectors. Outside the shaded volume, it is impossible to detect annihilation
photons in coincidence unless one or both undergo a Compton scatter in the tissue and change direction.
resolution. Uncertainties that govern the The ability of ACD to localize events on the
timing resolution can arise from the statisti- basis of coincidence timing, without the need
cal nature of the signal (which is produced by for absorptive collimation, is referred to as
the conversion of 511-keV photons into light, electronic collimation. As was discussed in
electrons, or electron-hole pairs in the detec- Chapter 14, Section C, the lead septa in stan-
tor) and from electronic noise in the detector dard parallel-hole collimators, which are nec-
and associated circuits. Uncertainties also essary to obtain adequate spatial localization,
can arise from the electronic method used to also are responsible for the relatively low sen-
determine the time at which the interaction sitivity of these collimators. Because ACD
occurred (see Chapter 8, Section F). For a pair does not require a collimator to define spatial
of similar detectors, the timing uncertainties location, its sensitivity (number of events
typically are well described by a gaussian dis- detected per unit of activity in the object) is
tribution, and the timing resolution is defined much higher than is obtainable with the
as the full width at half maximum (FWHM) absorptive collimators used for conventional
of this distribution. For scintillation detec- planar imaging and for SPECT. For compa-
tors, the timing uncertainty is reduced, and rable midplane resolution, the sensitivity of
the timing resolution improved, by using PET is many times higher than for SPECT.
brighter and faster scintillators that produce In addition, by incorporating multiple
a large number of light photons over a short opposing detectors in a complete ring or other
time interval immediately after an interac- geometric array around the patient, and oper-
tion occurs. Timing resolution is typically in ating each detector in the array in coincidence
the range of 0.5 to 5 nsec, depending on which with multiple detectors on the other side of
scintillator and photodetector is used. the array, data for multiple projection angles
The need for a finite window width permits can be acquired simultaneously (Fig. 18-2).
other types of events to occur in coincidence, Indeed, with a stationary ring or geometric
as discussed in Section A.9. Also, as discussed array that completely surrounds the patient,
in Section A.4, the annihilation photons are it is possible to acquire data for all projection
not always emitted in precise back-to-back angles simultaneously. This allows the perfor-
directions. The effects of these deviations mance of relatively fast dynamic studies and
from the ideal are discussed in the sections the reduction of artifacts caused by patient
indicated. motion.
18 Positron Emission Tomography 309
FIGURE 18-2 Array of detectors operating in electronic coincidence with detectors on the opposite side of the ring.
This allows simultaneous acquisition of projection views from many different angles. Solid and dotted lines illustrate
two simultaneously acquired projection views.
such as LSO and LaBr3, can be used (see a rectangular box of area d h at the face of
Tables 7-2 and 18-2). Several commercially either detector. Between these extremes, it is
built systems incorporate some level of time- the frustum of a pyramid with lower base size
of-flight information using these materials. equal to the size of the detectors and upper
Although there are scintillators with even base size increasing linearly from zero at mid-
faster decay components, such as BaF2, these plane to the size of the detectors at their face.
are not favored because the signal-to-noise Alternatively, consider an uncollimated
improvements that can be realized from time- pair of gamma camera detectors, also operat-
of-flight information is typically more than ing in coincidence mode (Fig. 18-4B). If their
offset by their lower density and therefore intrinsic spatial resolution (see Chapter 14,
lower efficiency for detecting 511-keV annihi- Section A.1) is a gaussian function with
lation photons. FWHM = Rint, then the spatial resolution of
the detector pair for ACD also is a gaussian
3. Spatial Resolution: Detectors function with FWHM = Rint / 2 at midplane.
The spatial resolution of ACD with discrete The ACD response profile becomes wider as
detector elements is determined primarily by the source moves toward either detector, with
the size of the individual detector elements. its FWHM eventually becoming equal to Rint
As shown in Figure 18-4A, for elements of at the face of either detector. Assuming that
width d, a one-dimensional (1-D) slice through the resolution of the imaging detector is the
the ACD point-source response profile at same in all directions, the 2-D ACD response
midplane between the detector pair is a profile is obtained by rotating the 1-D gauss-
triangle. The detector resolution, Rdet has a ian function around its center.
FWHM = d/2. As the source moves toward For both discrete or gamma cameratype
either detector, the response profile becomes detectors, the spatial resolution of ACD varies
trapezoidal, eventually becoming a box of by only approximately 30% in the central 60%
width d at the face of either detector. Consid- of the space between the detectors (Fig. 18-5).
ering a 2-D detector with width d and height By comparison, the resolution of a parallel-
h, the ACD response profile becomes a 3-D hole collimator can vary by several hundred
function, which is a pyramid at midplane and percent over a comparable range (see Fig.
18 Positron Emission Tomography 311
Intrinsic resolution
FWHM R int
FWHM d/2
FWHM R int / 2
Source
Intrinsic resolution
FWHM R int
Discrete Continuous
A detectors B detectors
FIGURE 18-4 Spatial resolution of detector pair (Rdet) for coincidence detection. A, For discrete detectors, spatial reso-
lution is determined by the width of the detector element, d. At midplane, the coincidence response function is a triangle
with full width at half maximum (FWHM) = d/2. As the source is moved closer to one of the detectors, the response
function becomes trapezoidal in shape, eventually becoming a rectangle of width, d. B, For continuous detectors, spatial
resolution is determined by the intrinsic resolution of the detector, Rint. At the midplane, the coincidence response
function is approximately gaussian, with FWHM = Rint / 2 . Near the face of a detector, it becomes FWHM = Rint.
312 Physics in Nuclear Medicine
12.9
12
11.7
8
)
m
10.5
t (c
8
ino
-p
id
m
m
Relative count rate
6
fro
e
nc
4
ta
is
4
D
9.8
0 0
20 0 20
Distance (mm)
FIGURE 18-5 Measured line-spread functions for a pair of 17-mm-wide coincidence detectors as a function of source
position between the two detectors. The detector separation was 42cm. The FWHM varies by only 30% within the
central 24cm (57%) of the space between the two detectors. (From Hoffman EJ, Huang S-C, Plummer D, Phelps ME:
Quantitation in positron emission computed tomography: VI. Effect of nonuniform resolution. J Comput Assist Tomogr
6:987-999, 1982.)
18 Positron Emission Tomography 313
TABLE 18-1
SOME POSITRON-EMITTING NUCLIDES USED FOR IN VIVO IMAGING
Maximum
Radionuclide Half-Life + fraction + Energy How Produced
11
C 20.4min 0.99 960keV Cyclotron
13
N 9.96min 1.00 1.19MeV Cyclotron
15
O 123sec 1.00 1.72MeV Cyclotron
18
F 110min 0.97 635keV Cyclotron
62 62
Cu 9.74min 0.98 2.94MeV Generator (from Zn)
64
Cu 12.7hr 0.19 580keV Cyclotron
68 68
Ga 68.3min 0.88 1.9MeV Generator (from Ge)
76
Br 16.1hr 0.54 3.7MeV Cyclotron
82 82
Rb 78sec 0.95 3.35MeV Generator (from Sr)
124
I 4.18 days 0.22 1.5MeV Cyclotron
The extrapolated range applies to the result is that the average distance measured
highest-energy positrons emitted by a radio- from the origin of the positrons to the end of
nuclide. However, positrons, like particles, their path is significantly smaller than their
are emitted with a spectrum of energies. Only extrapolated range.
a small fraction have the full amount of For purposes of defining the spatial resolu-
energy available from the decay (see Fig. 3-2). tion of ACD, the distance of interest is the
In addition, the extrapolated range is the effective positron range. This is the average
maximum distance that the electron would distance from the emitting nucleus to the end
travel if it were not significantly deflected of the positron range, measured perpendicu-
in any of its interactions and traveled in lar to a line defined by the direction of the
essentially a straight line to the end of its annihilation photons (Fig. 18-6). This dis-
range. In reality, most electrons (and posi- tance always is smaller than the extrapolated
trons) travel a tortuous path, often with mul- range for the positrons emitted by the
tiple large-angle deflections (see Fig. 6-4). The radionuclide.
Positron-emitting
radionuclide
Positron
Actual
positron Positron
range path
Effective
positron
range
511 keV
photon
511 keV
photon
Annihilation event
FIGURE 18-6 Blurring caused by positron range effects. The perpendicular distance from the decaying atom to the
line defined by the two 511-keV annihilation photons is referred to as the effective positron range.
314 Physics in Nuclear Medicine
Figure 18-7 shows the positron range dis- A second factor involving the physics of
tribution for point sources of 18F ( Emax = positrons is that the annihilation photons
0.635MeV) and 15O ( Emax = 1.72MeV). Accord- almost never are emitted at exactly 180-
ing to Figure 6-10, the extrapolated ranges degree directions from each other (Fig. 18-9).
for these positrons in water would be appro This effect, which is due to small residual
ximately 2mm and 8mm, respectively; momentum of the positron when it reaches
however, the FWHMs of their distribution the end of its range, is known as noncolinear-
profiles are only 0.1mm and 0.5mm. ity. The angular distribution is approximately
Note as well that the positron range distri- gaussian with FWHM approximately 0.5
butions shown in Figure 18-7 have long tails degree. The effect on spatial resolution,
and thus are not well described by gaussian expressed in terms of FWHM, is linearly
functions. Therefore the FWHM is not the best dependent on the separation of the ACD
indicator of the effect of positron range on ACD detectors, D, and is given by
spatial resolution. Instead, the root mean
square (rms) effective range often is used. R180 = 0.0022 D (18-2)
Figure 18-8 shows the general relationship
between rms effective range and maximum A typical value of D for a whole-body PET
positron energy. Typical rms effective ranges scanner is 80cm. Thus the FWHM for blur-
(and thus the blurring caused by positron ring caused by noncolinearity is approxi-
ranges) are on the order of 0.5 to 3mm. Note mately 2mm.
that positron range is inversely proportional The system resolution of an ACD or PET
to the density of the absorber. Thus rms ranges detector system is obtained by combining the
would be proportionately higher in lung tissue individual resolution components, in the same
and airways ( ~ 0.1-0.5g/cm3) and lower in manner as the component resolutions are
dense tissues such as bone ( ~ 1.3-2g/cm3). combined to determine the system resolution
10 2500
18F (Emax 635 keV) 100,000 Events 18F
2000
5
1500
Events
0
1000 0.102 mm FWHM
10,000 Events 0
10
10 5 0 5 10 2 1 0 1 2
End point coordinate (mm) End point coordinate (mm)
10 800
15O (Emax 1720 keV) 100,000 Events 15O
700
5 600
500
Events
400
0
300 0.501 mm FWHM
200 4.14 mm FWTM
5
100
10,000 Events 0
10
10 5 0 5 10 8 6 4 2 0 2 4 6 8
End point coordinate (mm) End point coordinate (mm)
FIGURE 18-7 Results of Monte Carlo simulations showing the distribution of annihilation sites for positron-emitting
point sources in water for 18F ( Emax = 0.635MeV) and 15O ( Emax = 1.72MeV). The profile of the distribution is broader
for 15O because of its higher average positron energy, which leads to a longer positron range prior to annihilation.
(From Levin CS, Hoffman EJ: Calculation of positron range and its effect on the fundamental limit of positron emission
tomography system spatial resolution. Phys Med Biol 44:781-799, 1999.)
18 Positron Emission Tomography 315
82
Rb
2.5
0
0 0.5 1 1.5 2 2.5 3 3.5
Maximum energy (MeV)
Positron
Non-colinearity
511 keV
photon 511 keV
photon
Annihilation event
Error due
to non-colinearity
FIGURE 18-9 Noncolinearity of annihilation photons resulting from residual momentum of the electron and positron
at annihilation. Noncolinearity leads to positioning errors. Angles are exaggerated in this example for purposes of
illustration. Actual range of angles is about 0.25 degree, centered at 180 degrees.
for a gamma camera system (see Chapter 14, the discrete detector elements or the intrinsic
Section C.4). Thus resolution of continuous detectors (see Section
A.3 and Fig. 18-4). For typical whole-body
Rsys 2
Rdet + Rrange
2
+ R180
2
(18-3) PET scanners, with either discrete detector
elements or gamma camera detectors, the
where Rdet is the spatial resolution of the effects of positron range and noncolinearity
detector system, as determined by the size of combine to add anywhere from a few tenths
316 Physics in Nuclear Medicine
that line. Thus a more appropriate measure hexagonal, or octagonal arrays. Each detector
for distributed sources is the average geomet- element is operated in coincidence with many
ric efficiency within the sensitive volume for detectors on the opposing side of the ring, as
ACD. Midway between the two detectors, this shown in Figure 18-12. This multicoincidence
is given by operation has useful and important conse-
quences for both the magnitude and unifor-
1 mity of geometric efficiency.
g ACD 2 [ Adet / D2 ] The simplest way to visualize its effects on
3
(18-7) geometric efficiency is to consider a complete
2 Adet ring of detectors on a diameter D, with detec-
3 D 2 tor height h in the axial dimension and detec-
tor width d << D in the plane of the ring.
where D is the distance between the detectors Assume that any interdetector gaps are very
and Adet is the area of the detector facing the small, so that the individual elements form a
source. virtually continuous ring of detector material.
The term in brackets in Equation 18-7 is For a point source located precisely at the
the geometric efficiency for a single detector center of the ring, the geometric efficiency
for a point source located at the midpoint of would be equal to the solid angle subtended
the centerline between the detectors. (As dis- by the ring, because if either annihilation
cussed in Chapter 11, Section A.2, this expres- photon is intercepted by the ring, it is virtually
sion is valid when the detector dimensions assured that the second photon is traveling
are small in comparison with the source-to- the proper direction to be intercepted as well.
detector distance.) The factor of 2 accounts for From simple geometric considerations, if h <<
the fact that two detectors are used and that D, it can be shown that the solid angle, and
if one photon is emitted in a proper direction thus the geometric efficiency, for a point source
toward one detector, the other photon is virtu- precisely at the center of the ring is given by
ally assured of being emitted in the proper
direction toward the other. The factor of 1/3 gACD,RING h/D (18-8)
is the average geometric efficiency across the
sensitive volume at midplane, that is, the Under the conditions described, geometric
average height of a pyramid. efficiency is relatively constant as the source
Actual PET systems typically employ many is moved away from the center of the ring but
small detector elements arranged in circular, still in its center plane; however, as the source
Useful field-
of-view
is moved axially toward the ends of the ring, photons onto the detectors and by gaps
the geometric efficiency still has a triangular between detector elements. Corrections for
shape. Thus the average geometric efficiency this and other image nonuniformities are
for a source distributed within the sensitive described in Section D.1.
volume for ACD across the width of the It is noteworthy that, by segmenting large
ring is half the value given by Equation 18-8, detectors into smaller elements and operating
that is, them in coincidence with multiple elements
h in the opposing array, it is possible to improve
g ACD, RING (18-9) the spatial resolution in PET with only a
2D modest loss of geometric efficiency. This effect
Equations 18-8 and 18-9 also are valid for is seen in Equation 18-9, in which geometric
polygonal arrays, with D representing the efficiency depends on the diameter of the ring,
diameter of a circle drawn tangential to the D, but not on the width d of the individual
surface of the array. As long as h << D, they detector elements. Most of the loss of sensitiv-
also apply to continuous detectors that use ity that does occur is due to the requirement
gamma camera electronics, rather than dis- for interelement spacing and shielding, which
crete detector elements, to determine event is only approximately 0.2 to 0.3mm in practi-
locations. cal systems. For comparison, from fundamen-
In addition to increasing geometric effi- tal principles, the geometric efficiency of
ciency and improving its uniformity, multico- absorptive collimators is degraded approxi-
incidence detection with a ring or polygonal mately as the square of spatial resolution
array of detectors also allows simultaneous (Equation 14-8). This presents a formidable
acquisition of multiple projection views challenge for improving spatial resolution in
without moving the detectors. Suppose the imaging applications based on single-photon
ring consists of N individual detector ele- counting, including SPECT.
ments. When each detector in the ring or The benefits of multicoincidence operation
array is operated in coincidence with a bank extend as well to the third (axial) dimension
of detectors on the opposite side, as illustrated in multi-ring PET systems. This is discussed
in Figure 18-12, a total of N/2 fan-beam further in Section C.
projections are acquired. These fan-beam pro- As is the case for any imaging system, the
jections typically are arranged to form parallel- sensitivity of a PET system also depends
beam profiles, as illustrated in Figures 16-20 critically on the detection efficiency of the
and 18-2. However, as illustrated in Figure detector, which enters as a squared term
18-11, data from adjacent pairs of detector in Equation 18-6. As was discussed in Chapter
usually are assigned to the same projection 11, Section A.3, detection efficiency is given
profile, thereby decreasing the number of by
views to N/4.
The number of detectors that are enabled = 1 e x
l (18-10)
for multicoincidence detection determines the
width of the fan-beam projections and thus where l is the linear attenuation coefficient
the diameter of the useful FOV. Sources of the detector material and x is the detec
located within the circle illustrated in Figure tor thickness. Values of l for several detector
18-12 are seen in all projections. Once a materials of interest for PET are given in
source is included within the useful FOV, a Table 18-2. Also indicated are values of for
further increase in fan-beam width does not 2-cm-thick detectors of each material, without
increase the counts recorded from that source. a low-energy threshold and with an energy
Sources outside the circle are not seen in threshold that eliminates 50% of the detected
some views, which could be a cause of image pulses. Values for 2 in this table are useful
artifacts (see Chapter 16, Section C.2). Typical for calculating scanner sensitivity and ACD
PET systems operate with each detector in counting rates (Equation 18-6). These values
coincidence with approximately two-thirds illustrate why materials such as BGO, LSO,
the total number of detectors in the ring. and LYSO (Chapter 7, Section C.4) are pre-
Geometric efficiency varies somewhat ferred over NaI(Tl) for PET imaging.
across the useful FOV of the detector ring. In Overall sensitivities for PET systems for a
part, this is because the solid-angle for coin- small-volume source of activity located near
cidence detection changes with source posi- the center of the scanner range from 0.2%
tion. There also are geometric effects caused to 0.5% (0.002-0.005cps/Bq) for single-ring sys
by differences in angle of incidence of the tems or for multi-ring systems operated in
322 Physics in Nuclear Medicine
TABLE 18-2
LINEAR ATTENUATION COEFFICIENTS AND DETECTION EFFICIENCIES FOR SOME
SCINTILLATORS AT 511keV*
Detection efficiency (see Equation 18-10), assuming no low-energy threshold (all pulses counted).
Detection efficiency, assuming low-energy threshold is used, with 50% of pulses counted (f = 0.5 in Equation 11-4).
2-D acquisition mode (one slice per ring; events are called prompt coincidences. The
Section C.1). For multi-ring systems in which discussion and analysis presented thus far
coincidences between rings are allowed for 3-D assumes that all prompt coincidences arise
data acquisition (see Section C.2), the sensitiv- from a pair of photons produced from the
ity typically is 2% to 10% (0.02-0.10cps/Bq). same annihilation event and that the anni-
For comparison, the sensitivities for SPECT hilation event occurs somewhere within the
systems with a general-purpose parallel-hole coincidence volume between the detectors
collimator are in the range of 0.01% to 0.03% (see Fig. 18-1). These events are called true
(0.0001-0.0003cps/Bq), depending on the number coincidences. Equation 18-6 describes the
of detector heads (see Chapter 14, Section E.7). sensitivity of the system for these events.
The substantially greater sensitivity of PET However, other prompt coincidence events
versus SPECT systems is due primarily to also can occur within the resolving time of
their ability to achieve a high degree of spatial the detector system.
resolution without the use of absorptive Two examples are shown in Figure 18-13.
collimators. Random coincidences (also called accidental
coincidences) occur when annihilation photons
from two unrelated positron annihilation
9. Event Types in Annihilation events are detected in two different detectors,
Coincidence Detection within the coincidence timing window, and
ACD produces an output whenever two recorded as a single coincidence event. This
events are recorded within a specified coinci- can happen if one photon from each annihila-
dence timing window. Generically, any such tion event is detected in each detector element.
FIGURE 18-13 True coincidence event (left), scatter coincidence event (center), and random or accidental coincidence
(right). Scatter and accidental coincidences yield incorrect positional information and contribute a relatively uniform
background to the image that results in a loss of contrast. (Courtesy Dr. Magnus Dahlbom, University of California
Los Angeles.)
18 Positron Emission Tomography 323
Random coincidences are not rare events, In actual PET scanners, the ratio of
because the volume of tissue from which the random-to-true coincidence counting rates
photons for a random coincidence event could typically ranges from approximately 0.1 to 0.2
arise generally is much larger than the poten- for brain imaging to greater than 1 for appli-
tial volume for true coincidence events. cations where large amounts of activity may
The random coincidence counting rate in a be nearby, but outside the true coincidence
detector pair is given by volume of the scanner. The latter could apply,
for example, to some types of abdominal
Rrandom = T Rsingle,1 Rsingle,2 (18-11) imaging when large amounts of activity are
excreted into the bladder. Random coinci-
where T is the coincidence timing window dences occur more or less uniformly across
used by the system* and Rsingle,1 and Rsingle,2 the FOV of the scanner, causing a loss of
are the single-channel counting rates in the image contrast as well as inaccuracies in
two detectors of the pair. Unlike true coinci- quantification of activity within the patient.
dence events, which can occur only when the Methods for correcting for random coinci-
source is located within the volume that is dences are discussed in Section D.2.
geometrically defined by the sides of the A second category of nonvalid prompt
detector pair (see Fig. 18-1), random coinci- coincidences are scatter coincidences. These
dences can arise from activity anywhere in occur when one (or both) of the photons from
the region between the detectors, including an annihilation event outside the sensitive
activity outside the useful FOV for a ring or volume for true coincidence events under-
array of detectors. Thus the single and goes scattering and is detected in a detector
random coincidence counting rates depend in other than the one that would be appropri-
a complicated way on both the source and ate for a true coincidence event. The scatter-
detector geometry. A detailed analysis of these ing event shown in Figure 18-13 occurs
factors is beyond the scope of this text. Refer- within the patient, but it also can occur
ence 1 provides a more comprehensive analy- within components of the scanner. Because
sis. Nonetheless, some general observations the two annihilation photons were emitted
can be made. simultaneously, they reach the detectors vir-
In general, the greater the total amount of tually simultaneously, apart from small
activity used in a study, the higher the ratio time-of-flight differences (see Section A.2).
of random-to-true coincidence rates. This is Because these differences are very small, the
because the random coincidence rate increases detector system and its associated coinci-
as the square of the amount of activity present dence logic cannot typically distinguish them
(product of single-channel counting rates in from valid events.
Equation 18-11), whereas the true coincidence As is the case for the random-to-true coin-
rate increases only linearly with the amount cidence ratio, the ratio of scatter-to-true
of activity administered (Equation 18-6). A coincidence counting rates depends in a com-
second general observation is that the ratio of plicated way on the source distribution and
random-to-true coincidence rates decreases in detector geometry. Placement of lead shield-
proportion to the width of the coincidence ing on either side of the detector ring, or of
timing window. However, as noted in Sections thin tungsten septa between detector rings in
A.1 and A.2, there are lower limits for this a multi-ring PET system, reduces the likeli-
value, because of electronic and time-of-flight hood of accepting scattered photons. However,
considerations. Finally, a general way to unlike the random-to-true ratio, the scatter-
reduce the single-channel counting rate from to-true ratio does not depend on the amount
activity outside the true coincidence volume is of activity administered, because both the
to use tungsten septa to restrict the FOV of scatter and true coincidence rates increase
individual detectors (see Section B.3). In turn, linearly with this parameter. It also does not
this reduces the random coincidence rate. depend on the width of the coincidence timing
window because scatter coincidences arise
from the same positron annihilation event,
*The value of T used here refers to the time separation and the two photons actually do arrive almost
between any two events that is determined by the elec- simultaneously at the two detectors. In clini-
tronics to indicate a prompt coincidence event (see cal studies, the scatter-to-true coincidence
Chapter 8, Section F). This differs from the definitions
used in some articles and texts and leads to an additional
ratio ranges from 0.2 to 0.5 for brain imaging
factor of 2 in the version of Equation 18-11 in those and from 0.4 to 2 for abdominal imaging. The
publications. higher end of these ranges applies for 3-D
324 Physics in Nuclear Medicine
acquisitions, which do not use interplane The block detector, designed in the mid-
septa (see Section C.2). 1980s by Casey and Nutt,3 allows small detec-
Scatter coincidences provide incorrect tor elements to be used (improving spatial
localization of the positron annihilation event. resolution) while reducing the number of
The degree of position error depends on the PMTs required to read them out (controlling
scattering angle and location of the scatter cost). Figure 18-14 shows a typical block
event. Scatter coincidences lead to a broad detector. A large piece of scintillator (most
distribution of mispositioned events, gener- commonly BGO, LSO, or LYSO), is segmented
ally peaked toward the center of the object. into an array of many elements by making
Methods for minimizing the acceptance of partial cuts through the crystal with a fine
scattered photons and for correcting for resid- saw. The cuts between the elements are filled
ual scatter coincidences are discussed in with a reflective material that serves to reduce
Section D.3. and control optical cross-talk between scintil-
lator elements. The array of crystals is read
out by four individual PMTs. The depth of the
B. PET DETECTOR AND SCANNER saw cuts is determined empirically to control
DESIGNS the light distribution to the four PMTs in a
fairly linear fashion.
As discussed in Section A.6, detection effi- To determine the segment of the crystal in
ciency (Equation 18-10) is an important which an annihilation photon is detected, the
parameter in PET scanner sensitivity and per- signals from a four-PMT array are combined
formance. Sodium iodide detectors, which are as follows:
the workhorse for many nuclear medicine
applications, also have been used for PET ( PMTA + PMTB ) ( PMTC + PMTD )
X=
scanners. Indeed, as discussed in Section B.3, PMTA + PMTB + PMTC + PMTD
it is possible to use appropriately modified
dual-headed SPECT systems for PET imaging. ( PMTA + PMTC ) ( PMTB + PMTD )
Y=
However, because of the relatively high energy PMTA + PMTB + PMTC + PMTD
of the 511-keV annihilation photons, sodium
iodide generally is not the detector material of (18-12)
choice for PET imaging. For these reasons,
most PET scanners use denser higher-Z scin- where PMTA, PMTB, and so forth are the
tillation detectors arranged in rings or banks signals from different PMTs. It will be recog-
of discrete elements around the scanned object. nized that these are essentially identical to
These systems not only provide a high detec- Equations 13-1 and 13-2 for position localiza-
tion efficiency but they allow the simultaneous tion for an Anger camera, except that only
collection of data for all projection angles with four PMTs are used here. The X and Y signals
a completely stationary set of detectors. In this then are used to determine the subelement of
section, we discuss the design of modern PET the array in which the annihilation photon
detector systems and scanners. Reference 2 is was detected.
a useful review describing emerging detector Figure 18-15 shows the image obtained
technologies for PET. from uniform irradiation of a block detector.
The image is not uniform. Rather, the calcu-
lated locations for recorded events are clus-
1. Block Detectors tered in small localized areas corresponding
Early PET systems used individual detector to the individual detector elements. There is
units consisting of a piece of scintillator a small amount of overlap, but the individual
coupled to a photomultiplier tube (PMT). The elements are clearly resolved. Although the
individual detectors were arranged in a ring array pattern is nonlinear, the separation is
or in multiple rings around the subject. As sufficiently clear to allow each (x,y) location
illustrated in Figure 18-4, the response profile in the image to be assigned to a specific detec-
at midplane of a pair of coincidence detectors tor element in the array, for example, by using
is a triangle with FWHM equal to one half the a look-up table.
width of the detector. Thus to improve The major advantage of the block detector
the intrinsic resolution of a PET scanner, the is that it enables many detector elements
detectors must be made smaller. However, (e.g., 8 8 = 64) to be decoded using only four
the cost increases rapidly if each detector PMTs. This dramatically lowers the cost per
element requires its own PMT. detector element while providing high spatial
18 Positron Emission Tomography 325
Reflector material
packed in saw cuts
Four single-channel
photomultiplier tubes
PMTC
Y PMTA
PMTD PMTB
X
FIGURE 18-14 Block detector commonly used in clinical PET scanners. A piece of BGO or LSO scintillator is cut into
an array of smaller elements that are read out using four single-channel photomultiplier tubes (PMTs). The cuts in
the material are filled with an opaque reflective material that, along with the depths of the cuts, helps control the
distribution of scintillation light reaching the PMTs.
(Chapter 13, Section B), with the continuous the overall detector performance is slightly
scintillator plate replaced by an array of dis- degraded as compared with a detector of the
crete scintillator elements. Although this same dimensions made purely from LSO. The
approach reduces the cost per detector phoswich design also can be combined with
element of a PET system (by reducing the the quadrant-sharing approach.
number of PMTs and electronic channels
required compared with a block detector), it 3. Whole-Body PET Systems
has the disadvantage of higher dead time Figure 18-18 illustrates several different
losses, because each PMT views signals from whole-body PET scanner designs that have
a larger volume of scintillator. been developed, some using block detectors
The second modification of the basic block comprising discrete scintillator elements as
detector design is to use layers of two differ- introduced previously (A, B, C) and others
ent scintillator materials, creating what is that use continuous large-area gamma camera
known as a phoswich (Fig. 18-17). This approach detectors (D, E, F). Systems that use a sta-
makes use of the difference in decay times of tionary ring or polygonal array of detectors (A,
the two scintillators. By monitoring the decay C, E, and F), with the detectors operating in
time of the pulse, the event can be localized multicoincidence mode, can acquire data for
into either the upper or lower layer. For all projection angles simultaneously and these
example, combinations of LSO (decay time designs have been the basis for most commer-
~40nsec) and GSO (decay time ~60nsec) cial systems. Others (B, D) use only a few
scintillators can be used. Because the location opposing banks of detectors, which must be
of photon interaction can be determined to rotated to get full tomographic information.
within half the total scintillator thickness, Most PET systems use a ring diameter of 80
this reduces the DOI effects (described in to 90cm. After inserting scatter shielding and
Section A.5) by approximately a factor of 2. a shroud to cover the detectors and other com-
The disadvantage of this approach is that ponents, the clear bore of the scanner typically
manufacturing of the detectors is more is 55 to 60cm, which is sufficient to comfort-
involved and that the light output and stop- ably accommodate most patients. The FOV in
ping power of GSO are worse than LSO. Thus the axial direction is determined by the axial
Scintillator 1
Scintillator 2
Signals
Four single-channel
photomultiplier tubes
FIGURE 18-17 Phoswich detector constructed from two scintillator materials with different decay times. By analyzing
pulse decay time, an event can be assigned to either the upper or lower layer, reducing the effective thickness of the
detector by one half and providing some depth of interaction information.
A B C
D E F
FIGURE 18-18 PET scanner geometries based on discrete scintillator elements (top row) or continuous scintillator
plates (bottom row). A, Full ring of modular block detectors. B, Partial ring of modular block detectors. C, Hexagonal
array of quadrant-sharing panel detectors. D, Dual-headed gamma camera with coincidence circuitry. E, Hexagonal
array of gamma camera detectors. F, Continuous detectors using curved plates of NaI(Tl). A complete set of profiles
can be acquired without motion with systems shown in A. C, E, and F, whereas detector motion is required with systems
shown in B and D.
328 Physics in Nuclear Medicine
extent of the detectors and typically is in the (thickness). The inside diameter of the detec-
range of 15 to 40cm. tor ring is 92.7cm and the clear bore of the
Figure 18-19 shows schematically the design of scanner is 59cm. The 18 crystals (three rings
a representative whole-body PET scanner based of three blocks) in the axial direction cover an
on block detectors. This scanner employs 336 axial FOV of 15.2cm. The gantry can be tilted
BGO block detectors, arranged in three rings 20 degrees from the vertical, which can
of 112 blocks per ring.4 Each block is cut into be useful for aligning the scan planes with the
a 6 6 array of elements, with element sizes optimal viewing angle for an organ of interest.
of 4mm (transaxial) 8.1mm (axial) 30mm The system contains a set of tungsten interplane
Septa
68Ge rod
source
Lead end
shielding
Septa
15.2 cm
68Ge rod
source
BGO block
detector
59 cm
92.7 cm
~120 cm
FIGURE 18-19 Drawings showing transaxial (top) and axial (bottom) cross-sections through a representative whole-
body PET scanner.
18 Positron Emission Tomography 329
septa of 1-mm thickness and 12-cm length direction and uses the scintillator LYSO. Its
between the crystal rings. The septa can design and performance capabilities are
be extended or retracted to provide varying broadly similar to those of the scanner des
levels of scatter rejection, as described in more cribed previously. The trend in recent years
detail in Section C.2. They also provide shield- has been to improve spatial resolution by
ing from potential high concentrations of further reducing the dimensions of the detec-
activity outside the scanning volume of inter- tor elements (4mm 4mm is current state-
est, which helps control the random and of-the-art), and to increase the number of
scatter coincidence rates, as described in detectors along the axial direction to improve
Section A.9. axial coverage of the body, increase sensitivity
The scanner also incorporates a rod source and thus reduce imaging times for studies
made from 68Ge (T1/2 = 273 days) to perform that cover the whole body. A second trend has
transmission scans for attenuation correc- been to improve the detectors and electronics
tions. The source is permanently mounted in to provide sufficient timing resolution that
the system and is retracted into a lead shield time-of-flight information (see Section A.2)
when not in use. This is discussed further in can be extracted. Commercial systems with a
Section D.4. Typically, the system uses a coin- coincidence timing resolution of better than
cidence timing window of 12.5 nanoseconds 600 psecs (corresponding to a spatial localiza-
and an energy window of 300 to 650keV. The tion of ~9cm) are now available.
intrinsic spatial resolution of the detectors is Gamma camera technology similar to that
approximately 3mm, whereas the system used for conventional planar imaging and
resolution is approximately 4.5mm near the SPECT (see Chapters 13, 14, and 17) also has
center of the FOV and approximately 6.2mm been employed for PET imaging. In one
near the periphery of the scanner bore, the approach, coincidence timing circuitry has
difference being due primarily to DOI effects been installed between the heads of dual-
(see Section A.5). The axial (slice thickness) headed scanners and the collimators removed
resolution is approximately 4.2mm at the for PET imaging. The spatial localization pro-
center of the scanner, whereas near the periph- vided by the detector heads allows many coin-
ery of the FOV, approximately 20-cm distance cidence lines to be acquired simultaneously.
axially from the center, it is approximately The basic concept is illustrated in Figure
6.9mm. The scanner simultaneously acquires 18-18D. These systems can still be used for
data for 35 slices, separated center-to-center planar or SPECT imaging, by replacing the
by 4.25mm, in 2-D acquisition mode. In 3-D collimators.
acquisition mode (see Section C.2), the number The performance of standard gamma
of slices and slice thickness in the axial direc- cameras for PET suffers from a number of
tion can be chosen arbitrarily. limitations. Chief among these is the rela-
Figure 18-20 shows photographs for tively low detection efficiency of the camera
another whole-body PET scanner, which is detectors for 511-keV annihilation photons
described in reference 5. This system has (see Figs. 11-4 and 11-5). As well, although
smaller detector dimensions in the axial removing the collimator allows simultaneous
A B
FIGURE 18-20 A, Modular cassette from a PET scanner containing eight block detectors. These cassettes are mounted
on the PET scanner gantry to form complete rings of detector blocks that surround the patient. B, Clinical PET scanner
based on rings of these block detectors. (Courtesy GE Healthcare, Waukesha, WI.)
330 Physics in Nuclear Medicine
C B
FIGURE 18-21 A, Photograph of a high-resolution brain imaging system. B, Interior of the scanner, showing the eight
panels of detectors arranged in an octagonal geometry that are made up of phoswich detectors read out by photo
multiplier tubes in a quadrant-sharing configuration. C, Phantom images from this system (right) compared with those
obtained from a typical whole-body PET scanner (left). The improvement in spatial resolution arising from the smaller
detector dimensions is apparent. (Courtesy Dr. Adriaan Lammertsma VU Medical Center, Amsterdam, Netherlands,
and Siemens Medical Solutions, Knoxville, TN.)
18 Positron Emission Tomography 331
A
2-D direct and cross planes /1
D
FIGURE 18-24 Two-dimensional (2-D) and 3-D data acquisition schemes for PET. Axial cross-sections through a multi-
ring scanner are shown on the left and corresponding axial sensitivity profiles on the right. A, 2-D direct plane data
acquisition. B, 2-D direct and cross-plane data acquisition. C, 2-D high-sensitivity data acquisition. D, Full 3-D data
acquisition. For clarity, lines of response are shown only for selected axial positions in C and D.
334 Physics in Nuclear Medicine
structures of interest as close to the center of compared with Equation 18-9, this adds
the axial FOV as possible. Multi-ring PET another factor of 2 to the sensitivity at the
systems have relatively high overall sensitiv- center of the FOV. The final result is
ity, as shown by the following example.
EXAMPLE 18-2 Sensitivity 3D 32 2 0.18 0.00409
Consider a 32-ring PET scanner with BGO 0.047 = 4.7%
detector elements that are 6-mm-wide 6-mm
axial height 2-cm-thick. The crystals are For an extended source, for example, a line
tightly packed on a 73.3-cm-diameter ring, source that has length comparable to the total
such that each ring contains 384 crystals. In thickness of the 32-ring array, the average
single-slice mode, each ring is operated in sensitivity of the scanner across the axial
coincidence only with other detector elements FOV is given by Equation 18-9, and the 3-D
within the same ring. In 3-D acquisition result given previously would be reduced by
mode, each ring is operated in coincidence a factor of 2.
with all other rings. Estimate the sensitivity
for a source located at the center of scanner The estimated sensitivity in Example 18-2
bore for single-slice and 3-D modes. Assume for 3-D operation is 2-3 orders of magnitude
that the source is comparable in size to the greater than the sensitivity achieved with a
axial length of the detector crystals, and that gamma camera for single-photon imaging
a low-energy threshold that passes 50% of with absorptive collimation (see Chapter 14,
detected events is used. Section E.7). Note as well that the additional
rings extend the volume of coverage, so that
Answer a volume of tissue can be imaged in less time
For a small-volume source, comparable in size with a multi-ring scanner, as compared with
to the axial thickness of a single detector ring, a single-ring device.
Equation 18-9 applies. For such a source, the Reconstruction of 3-D PET data also is
average geometric efficiency at the center of more complex, because the projection data
a single-ring is arises not only from transverse slices used
for 2-D reconstruction, but also from many
1 6 mm oblique angles through the subject. Thus the
g RING 0.00409 = 0.409% full 3-D image volume must be considered
2 733 mm during the reconstruction process. Fully 3-D
Fourier-based and iterative reconstruction
From Table 18-2, the intrinsic efficiency algorithms are both available; however, com-
(squared) for coincidence detection with putation times are roughly an order of mag-
2-cm-thick BGO crystals and an energy nitude longer than for 2-D reconstructions,
threshold that passes 50% of detected pulses because they involve backprojections and
is 0.18. When combined with geometric effi- computations in three dimensions rather
ciency (Equation 18-6), this gives a total than two. Approximate 3-D reconstruction
detection efficiency (sensitivity) for single- algorithms have been developed in which the
slice acquisition of 3-D dataset is reduced to a 2-D dataset using
rebinning methods (see Chapter 16, Section
E.3). In many situations, any loss in accuracy
Sensitivity RING 0.18 0.00409
resulting from the approximations made in
0.00074 = 0.074% these algorithms is small when compared
with the benefit of enabling 3-D PET data
In 3-D mode, all 32 rings are operated in to be reconstructed in clinically acceptable
coincidence with each other (Fig. 18-24D). timeframes.
Because of the increased solid-angle of cover- Despite the increased computational and
age, this immediately increases the geometric data storage requirements for 3-D PET, the
efficiency for a source located at the center of large increases in sensitivity that it produces
the axial FOV by approximately a factor of 32 has resulted in it being offered as an option on
(ignoring small geometric effects). As well, a all commercial whole-body PET systems. In
small-volume source at the center of the axial some systems, interplane septa have been
FOV more closely approximates a true point completely eliminated and only 3-D acquisi-
source, for which Equation 18-8 applies. As tion is possible. All small-animal and breast
18 Positron Emission Tomography 335
normalization factor for a specific pair of the end of each clock cycle (typically 256
detectors is computed from nanoseconds), the computer checks to see if
any events have occurred, and if so, whether
N i, j they occurred with arrival times within T
Normi, j = (18-13)
<N> nanoseconds of each other, where T is width
of the coincidence timing window (typically 4
where <N> is the average value of N(i,j) for to 12 nanoseconds). If two photons arrive
all of the coincidence detector pairs in the within this time interval, they are recorded
scanner. Hence, the average normalization as a valid event and the appropriate memory
factor is equal to 1. location corresponding to that particular
The normalization factor then is used to detector pair is incremented by +1.
correct the counts recorded for each detector An estimate of the random coincidence rate
pair in a scan of a patient, C(i,j), as follows: can be obtained by delaying the coincidence
timing window by a time that is much greater
Ci, j than its width. For example, the coincidence
CNormi, j = (18-14) timing window might be delayed by 64 nano-
Normi, j
seconds, for example, from 64 to 76 nano
where CNormi,j are the corrected counts. This seconds for a 12-nanosecond window. With this
correction is applied to the projection (sino- amount of time delay, only events that have
gram) data prior to image reconstruction. arrival times separated by between 64 and 76
Statistical errors caused by the finite nanoseconds are accepted. No true (or scat-
number of counts in the normalization scan tered) prompt coincidences will be detected in
will increase the noise levels in the corrected the delayed window, because photons from the
data, which is undesirable. In 3-D mode, PET same decay will always arrive at the detectors
scanners can have on the order of 108 lines of within a few nanoseconds of each other.
response. To achieve a statistical uncertainty However, the rate of random coincidences will
of ~3% in the normalization factor would be the same in the delayed and undelayed
therefore require a normalization scan with a windows because the rate at which uncorre-
total of approximately 1000 108 = 1011 counts. lated photons strike the detector is the same
Even at relatively high total counting rates of for both windows. Thus the delayed window
approximately 500,000cps, this would require count provides an estimate of the number of
approximately 55 hours of scanning time. random coincidence events. This number is
Thus the straightforward approach outlined subtracted from the total number of coinci-
previously must be modified to reduce the dence events for the detector pair. The correc-
number of counts required without increasing tion occurs on-line in most PET systems and
statistical noise. Most of the modified methods usually is transparent to the user.
are based on computing the efficiencies of the The events recorded in the delayed window
individual detector elements (rather than all are not the same ones as are recorded in
possible detector pairs) and then combining the undelayed window. Rather, the delayed
them to estimate the efficiency of the detector window provides a separate and independent
pairs. Details of these methods are beyond measure of the random event rate. Subtract-
the scope of this text but are discussed in ing the number of random events recorded in
reference 12. the delayed window results in an increase in
the statistical noise level for the measure-
2. Correction for Random Coincidences ment (see Chapter 9, Section C.1). Specifi-
As discussed in Section A.9, random coinci- cally, if Ntrue is the number of true coincidence
dences add a relatively uniform background events recorded, Nscatter the number of scatter
across the reconstructed image, suppressing coincidences, and Nrandom is the number of
contrast and distorting the relationship random coincidences subtracted from the
between image intensity and the actual total, the uncertainty in the remaining (true
amount of activity in the image. There are two plus scatter) coincidences is
approaches to estimating the random coinci-
dences so that they can be subtracted from the ( Ntrue + Nscatter ) = ( Ntrue + Nscatter ) + (2 Nrandom)
measured projection data: the delayed window
method and the singles method. (18-15)
In most PET scanners, the arrival time of
each photon is recorded and tagged with an Thus even if accurate corrections can be
accuracy of approximately 2 nanoseconds. At made, the random coincidence rate should be
18 Positron Emission Tomography 337
minimized to avoid unduly increasing the sta- It is not possible to distinguish between
tistical noise level of the image. scatter events in the body versus scatter
The second method for estimation of events in the detector crystal on the basis of
random coincidences is based on Equation pulse amplitude. Therefore simple correction
18-11. If the rate at which single (not coinci- schemes based on dual-energy windows are
dence) events occur in each detector is mea- far less successful in PET imaging than in
sured, and the coincidence timing window T SPECT imaging. Two main approaches cur-
is known, then the rate of random coinci- rently are used for scatter correction in PET.
dences for any pair of detectors can be com- The first approach uses information from the
puted. Because the rate of single events is original scatter-contaminated image and
typically at least an order of magnitude higher transmission image (see Section D.4) to derive
than the rate of coincidence events, the sta- the correction. The emission image shows the
tistical noise level in the estimate of the distribution of the activity in the subject. The
number of random events is small in compari- transmission image reflects the attenuation
son with that in the measurement of the coefficient of the tissue. At 511keV, virtually
number of prompt coincidences, and the all attenuation is due to Compton scatter.
uncertainty in the remaining coincidences Using these two images and computer model-
after random coincidences have been sub- ing of photon interaction physics (see Chapter
tracted is given by 6, Section C.3) with some simplifying assump-
tions, it is possible to derive an estimate of
( N true + Nscatter ) N true + Nscatter + N random the underlying distribution of scattered
events and their contribution to individual
(18-16) profiles. The estimated contribution of scat-
tered radiation then is subtracted from the
This method requires that each detector projection profiles and the reconstruction is
module continuously monitors the rate at repeated with the scatter-corrected data.
which it is detecting single events. As described in Chapter 19, Section E, with
the advent of hybrid PET/CT scanners, the
3. Correction for Scattered Radiation scatter distribution also can be computed
Scattered radiation in PET imaging leads to from the registered CT images, in which the
a hazy background in the reconstructed CT image is used in place of a PET transmis-
images, generally more concentrated toward sion image. This method works very well
the center of the image. As with random coin- when all the sources of radioactivity that
cidences, this leads to a decrease in image could lead to detected scatter events are con-
contrast and to errors in the quantitative tained within the FOV of the scanner. When
relationship between image intensity and the large amounts of activity are just outside the
amount of activity in the object. The fraction FOV of the scanner, problems can arise.
of scattered events in PET can be very high, Another drawback of this approach is that it
especially in 3-D imaging of the abdomen, is computationally intensive.
where it may be as high as 60% to 70%. This A second method for scatter correction is
large value has three major causes. First, based on an examination of projection profiles
only one of the two annihilation photons immediately outside the object. After correct-
needs to be scattered for a scatter coincidence ing for random coincidences, the only events
to occur. Second, the energy resolution of PET that should fall into these projection elements
detectors using dense scintillators such as are those that are mispositioned because of
BGO and LSO is inferior to NaI(Tl) detectors scatter. Based on the premise that scatter is
because of their lower light output. This a low-frequency phenomenon with little struc-
requires the use of a wider pulse-height ana- ture, data from the tails of the projection
lyzer window to capture the photopeak events. profiles can be extrapolated using simple
Finally, the predominant mode of interaction smoothly varying functions across the entire
in scintillators at 511keV is Compton scatter- projection. Both gaussian and cosine func-
ing, and many unscattered annihilation tions have been used for this purpose. The
photons deposit less than 511keV of energy extrapolated scatter distribution then is sub-
in the detector. Thus to increase the detection tracted from the projections prior to image
efficiency for photons that undergo Compton reconstruction. This method is rapid and,
scattering in the crystal, the analyzer window because it involves a direct measurement of
is widened even further to capture these scatter levels, it accounts for scatter from
events. radioactivity outside the FOV. However, it can
338 Physics in Nuclear Medicine
only approximate the true scatter distribu- measurement, called the blank scan, is made
tion and, in situations in which the scatter without the subject in the scanner. The subject
distribution is complex, or when the object then is placed in the scanner and the mea-
fills the whole FOV with no portion of the surement is repeated. This is known as
profile to examine outside the object, the the transmission scan. The attenuation
technique may result in significant errors. correction factor A for a detector pair (i,j) is
These can range from a few percent for brain given by
imaging to tens of percent at the heart-lung
interface. Blanki, j
Ai, j = (18-18)
Transi, j
4. Attenuation Correction
Attenuation correction is by far the largest where Blanki,j and Transi,j are the counts in
single correction in PET. Fortunately, the cor- the blank and transmission scans for the
rection is relatively easy to derive. Consider detector pair.
a source located at a depth x inside an object To obtain transmission data for all coinci-
of thickness T as shown in Figure 18-25. Both dence detector pairs, it is necessary to scan
of the photons from an annihilation event in the transmission source around the scanning
the source must be detected to record a valid volume for both the blank and transmission
event. Assuming that they are emitted in scans. Typically, a rod source, with its length
the appropriate directions, the probability extending along the axis of the scanner, is
that both photons will reach the detector is placed in a holder near the surface of the
given by the product of their individual scanner bore, and the holder rotates around
probabilities the central axis so that data are acquired for
all pairs. The most commonly used source
Pdet = e x e (T x ) material is 68Ge (parent of 68Ga, T1/2 = 273
(18-17) days). The blank scan needs to be performed
= e T
only once a day because it remains constant
where is the linear attenuation coefficient over a period that is short compared with
of tissue at 511keV and is approximately the half-life of the radionuclide in the
0.095cm1 for soft tissue, 0.12 to 0.14cm1 for rod source. The transmission scan is per-
bone, and 0.03 to 0.04cm1 for lung. Note that formed prior to injecting the patient with
the probability that both photons will reach the radiotracer. It is important that the
the detector is independent of the source loca- patient not move between the transmission
tion along the line joining the two detectors. and emission scan. Otherwise serious arti-
Equation 18-17 is similar to Equation 17-5 facts can occur, including the appearance of
for the geometric mean in SPECT, except that areas of abnormally high or low radiotracer
it applies for all source distributions, whereas uptake.
the geometric mean equation applies only for Conceptually, the simplest approach is to
point or plane sources at a fixed depth, x. As obtain the transmission scan before injecting
was the case for SPECT, transmission mea- the radiotracer to be imaged. This eliminates
surements can be used to correct for attenu- any possible interference between the activity
ation in PET. In PET, two measurements are that is present for the two scans.
taken with a source located on a line joining A second approach, called postinjection
each pair of coincidence detectors. The first transmission scanning, is to perform the
x
Detector 1 Detector 2
Point source
of activity
FIGURE 18-25 Parameters involved in the derivation of attenuation correction for PET (equation 18-17 in the text).
18 Positron Emission Tomography 339
transmission scan immediately after the Although these techniques work well, the
emission scan but while there is still radio- widespread use of hybrid PET/CT scanners
tracer activity in the patient. This can save has significantly reduced the use of transmis-
a significant amount of time, because the sion scans using external radionuclide sources
patient does not have to be on the table for attenuation correction. Instead, informa-
waiting for uptake after the transmission tion from the CT scan is used to perform
scan before the emission scan is performed. attenuation correction. The methods for
Another advantage of this approach is that CT-based attenuation correction are discussed
it reduces the chances of patient motion and in Chapter 19, Section E.
misalignment between the emission and
transmission scans. However, it requires the 5. Dead Time Correction
ability to distinguish transmission events Like all radiation detectors, PET detectors
from emission events caused by residual exhibit dead time and pile-up effects at high
radiotracer in the body. This is possible counting rates. The mispositioning of events
because the emission radiations generally caused by pile-up and possible approaches for
are emitted from locations spread through- minimizing pile-up described in the context of
out the body, whereas the transmission radi- the gamma camera (see Chapter 14, Section
ations are emitted from a very small volume. A.4) apply as well to block detectors used for
Thus the transmission source irradiates only PET. Dead time corrections must be applied.
a small subset of detector pairs at any one Otherwise, the amount or concentration of
time, and its counting rate in those irradi- radioactivity will be underestimated at high
ated pairs generally is much higher than the counting rates. Most PET scanners use empir-
counting rate in the same detectors caused ical dead time models in which the observed
by emission radiations. To implement this counting rate as a function of radioactivity
method, the scanner must have the means concentration is measured for a range of object
to track the location of the transmission sizes and at different energy thresholds. The
source and identify which detectors are resulting data are then fit with paralyzable or
being irradiated by it. A disadvantage of the nonparalyzable dead time models (see Chapter
postinjection approach is that the count-rate 11, Section C.2). Some systems apply a global
performance of the detectors must be suffi- dead time correction factor for the system,
cient to handle the emission and transmis- whereas others apply corrections to individual
sion activities simultaneously. pairs of detector modules.
Finally, it is possible to acquire transmis- Dead time losses are dominated by the
sion and emission data at the same time. single-channel counting rate, which are much
This approach, known as simultaneous higher than the coincidence counting rate.
emission/transmission scanning, is the most Corrections can be as large as a factor of 2,
efficient way to use scanner time in many although generally it is desirable to keep
situations. As with postinjection transmis- them below this level. Situations in which the
sion scanning, it is necessary to track the corrections can be large include first-pass
location of the rod source. Because the rod cardiac studies, imaging studies near the
source irradiates only a small and known bladder when there are high levels of excreted
subset of detector pairs at any one time, radioactivity, and studies with very short-
emission data can be acquired simultane- lived radiotracers, such as 15O, which require
ously from the remaining nonirradiated high starting levels of activity to maintain
pairs. For irradiated detector pairs, the adequate counting statistics over the course
counting rate from the transmission source of a study.
is much higher than the emission counting
rate, so the emission counts do not seriously
affect the accuracy of the transmission data. 6. Absolute Quantification of PET
A disadvantage of simultaneous emission/ Images
transmission scanning is that the relatively All of the corrections described earlier are
hot transmission source can contribute applied to the projection or sinogram data
random and scatter coincidence events to the prior to reconstruction of the image. If accu-
emission data. Even with corrections, these rately applied, after reconstruction, the voxel
events contribute to statistical noise and intensity in the image will be directly propor-
some degradation of image quality. For this tional to the amount of radioactivity in that
reason, postinjection transmission scanning voxel. Calibration to absolute concentrations
generally is the preferred approach. of radioactivity usually is accomplished by
340 Physics in Nuclear Medicine
103
600
400 Rtrue
Count/sec
Rrandom
Rscatter
NECR
200
0
0 2 4 6 8 10
Activity (Ci/cc)
FIGURE 18-26 Example of various coincidence counting rates and noise equivalent counting rate (NECR) for a clinical
whole-body scanner. These data predict that for the phantom used in this study, the best signal-to-noise (corresponding
to the peak of the NECR curve) in the reconstructed image would be achieved with an activity concentration of ~4 Ci/
cc (~150kBq/cc).
13
N-ammonia or 82RbCl to evaluate myocar- Control Alzheimers disease
dial viability and stratify patients with coro-
nary artery disease with regard to bypass
surgery. In the brain, PET is used diagnosti-
cally in a range of neurodegenerative diseases
(Alzheimers disease, Parkinsons disease)
and dementia, for epilepsy, neurodevelopmen-
tal disorders and in psychiatric disorders.
Metabolic FDG images of a patient with
Alzheimers disease compared with those of a
normal control are shown in Figure 18-28.
Many other PET radiotracers are used for
research studies and are being developed for
future clinical use. These include radioligands
that bind to specific receptors systems in the
brain, and radiotracers that target cell-surface
molecules specific to certain types of tumors.
Positron-emitting radionuclides also have
been used to radiolabel cells (e.g., stem cells)
and drug delivery vehicles such as nanopar-
ticles. Many of these radiotracers are first
FIGURE 18-28 Transaxial image slices showing 18F-
evaluated in animal models using a dedicated fluorodeoxyglucose uptake at two different levels of the
animal PET scanner. Figure 18-29 shows an brain in a normal volunteer (control) and in a patient at
image of the binding of the radioligand an early stage of Alzheimers disease. Data acquisition
11
C-raclopride to dopamine receptors in the times were ~30 minutes and injected doses were 370MBq
rat brain. (10mCi). Arrows indicate metabolic deficits in the
patients images. This distinct pattern of reduced metab-
olism is seen in all patients with Alzheimers disease and
increases in severity and extent are seen as the disease
progresses.
FIGURE 18-29 Coronal image of rat brain acquired on a small-animal PET scanner following injection of 11C-raclopride.
The location of the brain is indicated by the ellipse. This radiotracer binds to the dopamine receptors in the brain,
which are located primarily in the striatum (arrows). Images were acquired before (left) and after (right) a pharmaco-
logic intervention that damages the dopaminergic neurons on one side of the brain. A clear reduction in the binding
of the radiotracer is observed on that side.
3-dimensional lutetium oxyorthosilicate-based whole- 12. Badawi RD, Marsden PK: Developments in component-
body PET/CT scanner with the national electrical based normalization for 3D PET. Phys Med Biol 44:
manufacturers association NU 2-2001 standard. 571-594, 1999.
J Nucl Med 46:2083-2091, 2005. 13. Performance Measurements of Positron Emission
6. De Jong HWAM, van Velden FHP, Kloet RW, et al: Tomographs: National Electrical Manufacturers
Performance evaluation of the ECAT HRRT: an LSO- Association (NEMA) Standards Publication NU2-
LYSO double layer high resolution, high sensitivity 2007, Rosslyn, VA, 2007, NEMA.
scanner. Phys Med Biol 52:1505-1524, 2007. 14. Performance Measurements of Small Animal Posi-
7. MacDonald L, Edwards J, Lewellen T, et al: Clinical tron Emission Tomographs: National Electrical Man-
imaging characteristics of the positron emission ufacturers Association (NEMA) Standards Publication
mammography camera: PEM Flex Solo II. J Nucl NU4-2008, Rosslyn, VA, 2007, NEMA.
Med 50:1666-1675, 2009.
8. Constantinescu CC, Mukherjee J: Performance eval-
uation of an Inveon PET preclinical scanner. Phys BIBLIOGRAPHY
Med Biol 54: 2885-2899, 2009.
Additional general references are the texts indi-
9. Rowland DJ, Cherry SR: Small-animal preclinical
cated in references 1 and 11. Also pertinent are the
nuclear medicine instrumentation and methodology.
following:
Sem Nucl Med 38:209-222, 2008.
10. Larobina M, Brunetti A, Salvatore M: Small animal Phelps ME, editor: Molecular Imaging and its Biological
PET: A review of commercially available systems. Applications, New York, 2004, Springer-Verlag.
Curr Med Imag Rev 2:187-192, 2006. Zanzonico P: Positron emission tomography: A review of
11. Bendriem B, Townsend DW: The Theory and Practice basic principles, scanner design and performance, and
of 3D PET, Netherlands, Kluwer, 1998, Dordrecht. current systems. Sem Nucl Med 34:87-111, 2004.