Set Value Data Logger readings(average reading)

1 2 3 4 5 6 7 8
40.0C 40.1 39.9 40.1 40.1 40.1 39.9 40.0 40.0
75.0%RH 75.2 75.0 74.9 74.9 75.4 74.8 74.8 75.3

CALIBRATION CERTIFICATE
Set Value Chamber readings(average reading)
SCOPE
1 2 3 4 5 6 7 8
a 40.0C
) Company Name 40.0 : SRI40.1
KRISHNA PHARMACEUTICALS,
40.1 40.0 Unit-II,NACHARAM.
40.1 40.3 40.2 40.3
b75.0%RH
) Equipment 75.0 : Walk In
74.3 Humidity Chamber
74.6 74.9 74.3 75.7 74.7 75.4
c ) Model No. : NEC2280RSI
d ) Serial No. : WIHC/185/L

CALIBRATION DETAILS
a) Calibration Certificate No : ACPL/17-18/WIHC/HYD000
b) Calibration Performed On : 13-December-2017
c) Next Calibration Due On : 02- December -2018
d) Calibrated At : SRI KRISHNA PHARMACEUTICALS, Unit-II,NACHARAM.

SOURCE OF CALIBRATION
a) Equipment : Temperature / Humidity Sensor
b) Make And Model No : Newtronic,NW-70
c) Serial No. : NW10061722
d) Calibration No/Report No. : NLEPL/CAL/0717/10061722
e) Accuracy : Temperature : 0.5 C Humidity 2.0%RH
f) Due Date Of Calibration : 27-July-2018

RESULTS :
e) Tag No : QC/HCO/038

ACCURACY:
Temperature : 0.5 C With Respect To the Calibrator Value.
Humidity : 2.0%RH With Respect to the Calibrator Value.
CONCLUSION/REMARK
Temperature & humidity measurements are within the specified accuracy.
CALIBRATED BY : Mr. C.Obilesu

APPROVED BY :
_____________________________________
Page 1 of 1